Red-Green Display Allows Easy Recognition
of Judgment Results
• The measurement value display can be set to switch between red and green in accordance with the status of comparative outputs. This means that the status can be
ascertained at a distance.
Position Meter Enables Easy Monitoring of
Operating Status Trends
• The present value with respect to the measurement or display range (full scale) can be viewed on a bar display. The
operating status can be grasped intuitively, allowing easy
judgement of levels and threshold values.
HH
Max
Min
H
P
L
LL
MAX/MIN
B
L
TG HH H
T-ZR
CMW
Zero
TLLL
Hold
UPSHIFTMODELEVEL
Position mete
Many Measurement and Discrimination
Functions Using External Event Input
• Offers a wide variety of application possibilities, such detection and judgement while synchronizing on an external signal.
Input
Measurement
OFF
TIMING
input
(b1−a1)
'a2
Sampling hold value
value
ON
Peak-to-peak value
'b2
(b2−a2)
Time
Time
Sampling
measurement
Peak hold value
Bottom hold value
Peak-to-peak hold
measurement
Time
Measurement value
TIMING
input
Peak/bottom hold
measurement
Input
Measurement value
TIMING
input
OFF
Input
OFF
ON
ON
b1
a1
High-speed Sampling at 50 Times per Second
(20 ms)
• Provides an input sampling cycle at least three times faster
than earlier models (12.5 times faster for temperature input
models) at 50 times/second (see note). In addition to improved response times for judgement output and transfer
output, average processing can be used to increase the stability of measurements.
Note: The K3HB-S Linear Sensor Indicator features high-speed re-
sponse of 2,000 times/second.
Sampling at 15 times/secondSampling at 50 times/second
Short Body with Depth of Only 95 mm (from
Behind the Front Panel)
• A short body of only 95 mm (see note) contributes to the development of slimmer and smaller control panels and installations.
27%
shorter
than earlier
Depth: 95 mm (See note.)
(The depth is 100 mm when mounted to the terminal cover.)
Note: Depth of DeviceNet models is 97 mm.
models
Select a Comparative Output Pattern to Suit
the Discrimination or Control Application
• The output pattern for comparative outputs can be selected.
In addition to high/low comparison with set values, output
based on level changes is also possible. (Use the type of output pattern appropriate for the application.)
Comparative
set value HH
Comparative
set value H
Comparative
set value L
Comparative
set value LL
Output HH
Output H
Output PASS
Output L
Output LL
Comparative
set value HH
Comparative
set value H
Comparative
set value L
Comparative
set value LL
Output HH
Output H
Output PASS
Output L
Output LL
Standard Output
Zone Output
Measurement
value
Higher
Lower
ON
OFF
Measurement
value
Higher
Lower
ON
OFF
Comparative
set value HH
Comparative
set value H
Comparative
set value L
Comparative
set value LL
Output HH
Output H
Output PASS
Output L
Output LL
Level Output
Note: The HH, H, L, or LL outputs must be set in that order for the
zone outputs to output correctly.
(This is because the comparative set values and outputs for
standard and level outputs are in a 1-to-1 relationship, whereas
the meaning of zone outputs depends on the settings of all the
comparative set values.)
Measurement
value
Higher
Lower
ON
OFF
Lineup Includes DeviceNet Models Enabling
High-speed Data Communications with PLCs
without Special Programming
• DeviceNet compliance enables high-speed data transmission by allocating setting and monitoring parameters in the
I/O memory of the PLC. This capability greatly reduces labor
spent in developing communications programs.
Ethernet
DeviceNet
Master
Peak value
Time
100
Timing of
internal
sampling
0
Time
P
Zero
P
Zero
96
Timing of
internal
sampling
2Digital IndicatorsK3HB Series (Analog Input Series)
H
K3HB
Digital Indicators
DeviceNet
Configurator
Note: The applications provided in this catalog are intended as reference only. Do not attempt to use any of them in real systems without first con-
firming machine and device functions and safety. For applications that require safety, ensure that there is sufficient leeway in ratings and
performances, install fail-safe measures, and take any other safety measures required by the application. In addition, contact your nearest
OMRON representative and confirm specifications.
K3HB-series Product Lineup
■ K3HB-X Process Indicator (page 4)
Indicates Measurements for Voltage/Current Signals
• DC Voltage Input Type: K3HB-XVD
(±199.99 V, ±19.999 V, ±1.9999 V, 1.0000 to 5.0000 V)
• DC Current Input Type: K3HB-XAD
(±199.99 mA, ±19.999 mA, ±1.9999 mA, 4.000 to 20.000 mA)
• AC Voltage Input Type: K3HB-XVA
(0.0 to 400.0 V, 0.00 to 199.99 V, 0.000 to 19.999 V, 0.0000 to 1.9999 V)
• AC Current Input Type: K3HB-XAA
(0.000 to 10.000 A, 0.0000 to 1.9999 A, 0.00 to 199.99 mA, 0.000 to
19.999 mA)
■ K3HB-V Weighing Indicator (page 8)
Digital Indicators K3HB Series (Analog Input Series)3
Process Indicator
K3HB-X
A Process Indicator Ideal for Discriminating and
Displaying Measurements for Voltage/Current
Signals
• Easy recognition of judgement results using color display that can be
switched between red and green.
• Equipped with a position meter for monitoring operating status trends.
• External event input allows use in various measurement and discrimination applications.
• Series expanded to include DeviceNet models.
• Shor t body with depth of only 95 mm (from behind the front panel), or
97 mm for DeviceNet models.
• UL certification approval (Certification Mark License).
• CE Marking conformance by third party assessment body.
• Water-resistant enclosure conforms to NEMA 4X (equivalent to IP66).
• Capable of high-speed sampling at 50 times per second (20 ms)
• Easy-to-set two-point scaling allows conversion and display of any userset values.
Refer to Precautions on page 30.
Model Number Structure
■ Model Number Legend
Base Units and Optional Boards can be ordered individually or as sets.
Base Units
K3HB-X@
1. Input Sensor Codes
VD: DC voltage input
AD: DC current input
VA: AC voltage input
AA: AC current input
5. Supply Voltage
100-240 VAC: 100 to 240 VAC
24 VAC/VDC: 24 VAC/VDC
Optional Board
Sensor Power Supply/Output Boards
K33-@
2
Relay/Transistor Output Boards
K34-@
3
Event Input Boards
K35-@
4
Note: 1. CPA can be combined with relay outputs only.
2. Only one of the following can be used by each Digital Indicator:
5
1
RS-232C/RS-485 communications, a linear output, or DeviceNet communications.
Linear outputLinear output 0 to 20 mA DC, 4 to 20 mA:
Note: 1. DC power supply models require a control power supply capacity of approximately 1 A per Unit when power is turned ON. Particular
attention is required when using two or more DC power supply models. The OMRON S8VS-series DC Power Supply Unit is recommended.
2. PNP input types are also available.
3. For K3HB-series DeviceNet models, use only the DeviceNet Connector included with the product. The crimp terminals provided are for
Thin Cables.
100 to 240 V: 18 VA max. (max. load)
24 VAC/DC: 11 VA/7 W max. (max. load)
ON residual voltage: 3 V max.
ON current at 0
Max. applied voltage: 30 VDC max.
OFF leakage current: 1.5 mA max.
NPN open collector or no-voltage contact signal
ON residual voltage: 2 V max.
ON current at 0
Max. applied voltage: 30 VDC max.
OFF leakage current: 0.1 mA max.
Mechanical life expectancy: 5,000,000 operations, Electrical life expectancy: 100,000 operations
Load: 500
Linear output 0 to 5 VDC, 1 to 5 VDC, 0 to 10 VDC:
Load: 5 k
(1 V or less: ±0.15 V; not output for 0 V or less)
7-segment digital display (Character height: PV: 14.2 mm (green/red); SV: 4.9 mm (green)
ro, zero-limit, output hysteresis, output OFF delay, output test, teaching, display value selection, display color
selection, key protection, bank selection, display refresh period, maximum/minimum hold, reset
−10 to 55°C (with no icing or condensation)
−25 to 65°C (with no icing or condensation)
a DeviceNet connector (Hirose HR31-5.08P-5SC(01)) and crimp terminals (Hirose HR31-SC-121) (See note
(100 m/s2 for relay outputs) 3 times each in 3 axes, 6 directions
CISPR 11 Group 1, Class A: CISPRL16-1/-2
CISPR 11 Group 1, Class A: CISPRL16-1/-2
EN61000-4-2: 4 kV (contact), 8 kV (in air)
EN61000-4-3: 10 V/m 1 kHz sine wave amplitude modulation (80 MHz to 1 GHz)
EN61000-4-4: 2 kV (power line), 1 kV (I/O signal line)
EN61000-4-5: 1 kV with line (power line), 2 kV with ground (power line)
EN61000-4-6: 3 V (0.15 to 80 MHz)
EN61000-4-11: 0.5 cycle, 0
°/180°, 100% (rated voltage)
2
, 10 sweeps of 5 min each in X, Y, and Z directions
µs/100 ns)
µs/100 ns)
6Process IndicatorK3HB-X
■ Input Range (Measurement Range and Accuracy) CAT II
Input typeRangeSet valueMeasurement rangeInput
impedance
AccuracyAllowable
instantaneous
overload (30 s)
K3HB-XVD
DC voltage
Aa Ud±199.99 V10 M
Bb Ud±19.999 V1 M
Ω min.±0.1%rdg ± 1
Ω min.±200 V
digit max.
±400 V
Cc Ud±1.9999 V
Dd Ud1.0000 to 5.0000 V
K3HB-XAD
DC current
K3HB-XVA
AC voltage
(See note 4.)
K3HB-XAA
AC current
Aa ad±199.99 mA1
Bb ad±19.999 mA10
Cc ad±1.9999 mA33
Dd ad4.000 to 20.000 mA10
Aa Ua0.0 to 400.0 V1 M
Bb Ua0.00 to 199.99 V
Ω max.±0.1%rdg ± 1
Ω max.±200 mA
digit max.
Ω max.
Ω max.
Ω min.±0.3%rdg ± 5
digits max.
Cc Ua0.000 to 19.999 V±0.5%rdg ± 10
Dd Ua0.0000 to 1.9999 V
Aa aa0.000 to 10.000 A(0.5 VA CT)
(See note 3.)
digits max.
±0.5%rdg ± 20
digits max.
±400 mA
700 V
400 V
20 A
Bb aa0.0000 to 1.9999 A(0.5 VA CT)
(See note 3.)
Cc aa0.00 to 199.99 mA1
Dd aa0.000 to 19.999 mA10
Ω max.±0.5%rdg ± 10
Ω max.
digits max.
2 A
Note: 1. The accuracy is for an input frequency range of 40 Hz to 1 kHz (except for AD current input A and B ranges) and an ambient temperature
of 23 ±5
°C. The error, however, increases below 10% of the maximum input value.
DC voltage input (all ranges): 10% or less of max. input = ±0.15% FS
DC current input (all ranges): 10% or less of max. input = ±0.1% FS
AC voltage input (A: 0.0 to 400.0 V): 10% or less of max. input = ±0.15% FS
AC voltage input (B: 0.00 to 199.99 V): 10% or less of max. input = ±0.2% FS
AC voltage input (C: 0.000 to 19.999 V; D: 0.0000 to 1.9999 V): 10% or less of max. input = ±1.0% FS
AC current input (A: 0.000 to 10.000 A): 10% or less of max. input = ±0.25% FS
AC current input (B: 0.0000 to 1.9999 A): 10% or less of max. input = ±0.5% FS
AC current input, (C: 0.00 to 199.99 mA; D: 0.000 to 19.999 A): 10% or less of max. input = ±0.15% FS
When DC voltage input models are used with a ±1.9999 V range, make sure that the connections between input terminals are not open.
If the input terminals are open, the display will show large variations. Connect resistance of approximately 1 M
Ω between the input ter-
minals if they are open.
2. The letters “rdg” mean “reading” and refer to the input error.
3. The value (0.5 VA CT) is the VA consumption of the internal CT (current transformer).
E2
+
−
1 MΩ
E3
E4
E5
E6
COM
±1.9999 V
4. The K3HB-XVA@@ complies with UL standards when the applied input voltage is within the range 0 to 150 VAC.
If the input voltage is higher than 150 VAC, install an external transformer or take other measures to drop the voltage to 150 VAC or lower.
Process Indicator K3HB-X7
Weighing Indicator
K3HB-V
An Ideal Indicator for OK/NG Judgements in
Automated and Picking Machines, Measuring
Factors such as Pressure, Load, Torque, and
Weight Using Load Cell Signal Input.
• Easy recognition of judgement results using color display that can be
switched between red and green.
• Equipped with a position meter for monitoring operating status trends.
• External event input allows use in various measurement and discrimination applications.
• Series expanded to include DeviceNet models.
• Shor t body with depth of only 95 mm (from behind the front panel), or
97 mm for DeviceNet models.
• UL certification approval (Certification Mark License).
• CE Marking conformance by third party assessment body.
• Water-resistant enclosure conforms to NEMA 4X (equivalent to IP66).
• Capable of high-speed sampling at 50 times per second (20 ms)
• Easy-to-set two-point scaling allows conversion and display of any userset values.
Refer to Precautions on page 30.
Model Number Structure
■ Model Number Legend
Base Units and Optional Boards can be ordered individually or as sets.
Base Units
K3HB-V@
1. Input Sensor Codes
LC: Load cell input (DC low-voltage input)
5. Supply Voltage
100-240 VAC: 100 to 240 VAC
24 VAC/VDC: 24 VAC/VDC
Optional Board
Sensor Power Supply/Output Boards
K33-@
Relay/Transistor Output Boards
K34-@
Event Input Boards
K35-@
Note: 1. CPB can be combined with relay outputs only.
Accessories (Sold Separately)
K32-DICN: Special Cable (for event inputs, with 8-pin connector)
K32-BCD: Special BCD Output Cable
1
5
2
3
4
2. Only one of the following can be used by each Digital Indicator:
RS-232C/RS-485 communications, a linear output, or DeviceNet communications.
Linear outputLinear output 0 to 20 mA DC, 4 to 20 mA:
Note: 1. DC power supply models require a control power supply capacity of approximately 1 A per Unit when power is turned ON. Particular
attention is required when using two or more DC power supply models. The OMRON S8VS-series DC Power Supply Unit is recommended.
2. PNP input types are also available.
3. For K3HB-series DeviceNet models, use only the DeviceNet Connector included with the product. The crimp terminals provided are for
Thin Cables.
100 to 240 V: 18 VA max. (max. load)
24 VAC/DC: 11 VA/7 W max. (max. load)
ON residual voltage: 3 V max.
ON current at 0 Ω: 17 mA max.
Max. applied voltage: 30 VDC max.
OFF leakage current: 1.5 mA max.
NPN open collector or no-voltage contact signal
ON residual voltage: 2 V max.
ON current at 0
Max. applied voltage: 30 VDC max.
OFF leakage current: 0.1 mA max.
Mechanical life expectancy: 5,000,000 operations, Electrical life expectancy: 100,000 operations
Load: 500
Linear output 0 to 5 VDC, 1 to 5 VDC, 0 to 10 VDC:
Load: 5 k
(1 V or less: ±0.15 V; not output for 0 V or less)
7-segment digital display (Character height: PV: 14.2 mm (green/red); SV: 4.9 mm (green)
zero, zero-limit, output hysteresis, output OFF delay, output test, teaching, display value selection, display
color selection, key protection, bank selection, display refresh period, maximum/minimum hold, reset
−10 to 55°C (with no icing or condensation)
a DeviceNet connector (Hirose HR31-5.08P-5SC(01)) and crimp terminals (Hirose HR31-SC-121) (See note
Note: 1. The accuracy is for an ambient temperature of 23±5°C. For all ranges,10% or less of max. input ±0.1% FS.
2. The letters “rdg” mean “reading.”
Input typea lcb lcc lcd lc
(mV)
Connected
terminals
200.000
150.000
100.000
50.000
0.00
−50.00
−100.00
−150.00
−200.00
E2E6E3E6E4E6E5E6
199.99199.99
100.00
19.999
0.000.000
−100.00
−199.99
The area shown in dark shading indicates the factory setting.
instantaneous
overload (30 s)
■ Load Cell Wiring Example■ Scaling Example Using Range A
Indicated on the K3HB-V as 0 to 49N in the load cell specifications
(rated load 49N, recommended applied voltage 10 V, rated output 2
mV/V) (See note.)
Display value
49 N
20 mV
Scaling values:
inp.a1=000.00
dsp.a1=00000
inp.a2=020.00
dsp.a2=49000
dp decimal point position: 00.000
Input value
Note: 2 mV/V indicates a load cell output of 2 mV for 1 V applied volt-
age for the rated load (when using a load of 1 N). When the applied voltage is 10 V, the load cell output is 20 mV (2 mV × 10).
B5
B6
B
E
E2
E3
E4
E5
E6
+10 V
+
−
0 V
+OUT
+IN
Load cell
−IN
−OUT
Weighing Indicator K3HB-V11
Temperature Indicator
K3HB-H
New High-speed, High-precision Temperature
Indicator
• Easy recognition of judgement results using color display that can be
switched between red and green.
• Equipped with a position meter for monitoring operating status trends.
• External event input allows use in various measurement and discrimination applications.
• Series expanded to include DeviceNet models.
• Shor t body with depth of only 95 mm (from behind the front panel), or
97 mm for DeviceNet models.
• UL certification approval (Certification Mark License).
• CE Marking conformance by third party assessment body.
• Water-resistant enclosure conforms to NEMA 4X (equivalent to IP66).
• Capable of high-speed sampling at 50 times per second (20 ms).
• High-resolution of 0.01
input. Thermocouple sensor inputs also support a resolution of 0.1
for all ranges.
• Temperature input shift is easily set using two points.
Refer to Precautions on page 30.
Model Number Structure
°C with platinum-resistance thermometer Pt100
°C
■ Model Number Legend
Base Units and Optional Boards can be ordered individually or as sets.
Base Units
K3HB-H@
1. Input Sensor Codes
TA: Temperature input
5. Supply Voltage
100-240 VAC: 100 to 240 VAC
24 VAC/VDC: 24 VAC/VDC
Optional Board
Sensor Power Supply/Output Boards
K33-@
Relay/Transistor Output Boards
K34-@
Event Input Boards
K35-@
Note: 1. CPA can be combined with relay outputs only.
Linear outputLinear output 0 to 20 mA DC, 4 to 20 mA:
100 to 240 V: 18 VA max. (max. load)
24 VAC/DC: 11 VA/7 W max. (max. load)
Thermocouple: K, J, T, E, L, U, N, R, S, B, W
ON residual voltage: 3 V max.
ON current at 0
Max. applied voltage: 30 VDC max.
OFF leakage current: 1.5 mA max.
NPN open collector or no-voltage contact signal
ON residual voltage: 2 V max.
ON current at 0
Max. applied voltage: 30 VDC max.
OFF leakage current: 0.1 mA max.
Mechanical life expectancy: 5,000,000 operations, Electrical life expectancy: 100,000 operations
Load: 500
Linear output 0 to 5 VDC, 1 to 5 VDC, 0 to 10 VDC:
Load: 5 k
(1 V or less: ±0.15 V; not output for 0 V or less)
7-segment digital display (Character height: PV: 14.2 mm (green/red); SV: 4.9 mm (green)
output hysteresis, output OFF delay, output test, display value selection, display color selection, key protection,
bank selection, display refresh period, maximum/minimum hold, reset
Linear outputLinear output 0 to 20 mA DC, 4 to 20 mA:
100 to 240 VAC (50/60 Hz),
85% to 110% of the rated power supply voltage,
100 to 240 V: 18 VA max. (max. load)
24 VAC/DC: 11 VA/7 W max. (max. load)
ON residual voltage: 3 V max.
ON current at 0
Max. applied voltage: 30 VDC max.
OFF leakage current: 1.5 mA max.
NPN open collector or no-voltage contact signal
ON residual voltage: 2 V max.
ON current at 0
Max. applied voltage: 30 VDC max.
OFF leakage current: 0.1 mA max.
Mechanical life expectancy: 5,000,000 operations, Electrical life expectancy: 100,000 operations
Load: 500
Linear output 0 to 5 VDC, 1 to 5 VDC, 0 to 10 VDC:
Load: 5 k
(1 V or less: ±0.15 V; not output for 0 V or less)
7-segment digital display (Character height: PV: 14.2 mm (green/red); SV: 4.9 mm (green)
comparison, forced-zero, zero-limit, output hysteresis, output OFF delay, output test, teaching, display value selection,
display color selection, key protection, bank selection, display refresh period, maximum/minimum hold, reset
The range shown in dark shading indicates the factory setting.
Sampling and Comparative Output Response Times
The K3HB-S sampling and comparative output response times depend on the calculation methods, timing hold type, and, for simple averaging, the
averaging times. Refer to the following description for details.
■ Output Refresh Period
The K3HB-S repeats input reads, calculation, and judgement output
processing. The output refresh period differs depending on whether
there are one or two inputs, as outlined below.
One Input
0.5 ms
Input
↓
Input A
or
Input B
Input read
Output
refresh
Two inputs
0.5 ms
Input
↓
Input A
or
Input B
Input read
Output
refresh
0.5 ms
Input
↓
↓
Output
Every 0.5 ms
Every 0.5 ms
0.5 ms
Input
Output
Input A: Every 1 ms
Input B: Every 1 ms
Every 0.5 ms
Input
↓
↓
Output
Input
↓
↓
Output
0.5 ms
0.5 ms
Input
↓
↓
Output
Input
Output
Input
↓
↓
Output
Input
↓
↓
Output
Input
↓
↓
Output
Input
Output
Input
↓
↓
Output
■ Output Response Time
The comparative output response time is the sum of the data processing time and the output (relay or transistor) response time.
One Input
0.5 ms0.5 ms
R
Two Inputs
0.5 ms0.5 ms0.5 ms0.5 ms
RRRR
Note: For transistor outputs:
For relay outputs:
0.5 ms
Data processing time
RR
R Output response time
(See note.)
Data processing time
R Output response time
(See note.)
For one input: OFF to ON 1 ms and ON to OFF 1.5 ms
For two inputs: OFF to ON 2 ms and ON to OFF 2.5 ms
The relay operation time of 15 ms is added to the transistor
output response times.
Linear Sensor Indicator K3HB-S19
Common Specifications
■ Event Input Ratings
Input typeS-TMR, HOLD, RESET, ZERO, BANK1, BANK2,
ContactON: 1 k
No-contactON residual voltage: 2 V max.
Ω max., OFF: 100 kΩ min.---
OFF leakage current: 0.1 mA max.
Load current: 4 mA max.
Maximum applied voltage: 30 VDC max.
BANK4
ON residual voltage: 3 V max.
OFF leakage current: 1.5 mA max.
Load current: 17 mA max.
Maximum applied voltage: 30 VDC max.
■ Output Ratings
Contact OutputTransistor Output
ItemResistive loads
Rated load5 A at 250 VAC
Rated through
current
Mechanical life
expectancy
Electrical life
expectancy
(250 VAC, cos
30 VDC, L/R=0 ms)
5 A at 30 VDC
5A
5,000,000 operations
100,000 operations
φ=1;
Linear Output
Item0 to 20 mA4 to 20 mA0 to 5 V1 to 5 V0 to 10 V
Allowable load impedance500
ResolutionApprox. 10,000
Output error±0.5%FS
Ω max.5 kΩ min.
Inductive loads
(250 VAC, closed
circuit, cos
30 VDC, L/R=7 ms)
1 A at 250 VAC
1 A at 30 VDC
φ=0.4;
Maximum load voltage24 VDC
Maximum load current50 mA
Leakage current100
TIMING
µA max.
Serial Communications Output
ItemRS-232C, RS-485
Communications method Half duplex
Synchronization method Start-stop synchronization
Baud rate9,600, 19,200, or 38,400 bps
Transmission codeASCII
Data length7 bits or 8 bits
Stop bit length2 bits or 1 bit
Error detectionVertical parity and FCS
Parity checkOdd, even
Note: For details on serial and DeviceNet communications, refer to
the Digital Indicator K3HB Communications User's Manual
(Cat.No. N129).
BCD Output I/O Ratings
(Input Signal Logic: Negative)
I/O signal nameItemRating
InputsREQUEST
HOLD
MAX
MIN
RESET
Outputs DATA
POLARITY
OVER
DATA VALI D
RUN
HH
H
PA SS
L
LL
Note: For details on serial and DeviceNet communications, refer to
the Digital Indicator K3HB Communications User's Manual
(Cat.No. N129).
Input signalNo-voltage
Input current for
no-voltage input
Signal
ON voltage 1.5 V max.
level
OFF voltage 3 V min.
Maximum load
voltage
Maximum load
current
Leakage current100
Maximum load
voltage
Maximum load
current
Leakage current100
contact input
10 mA
24 VDC
10 mA
µA max.
24 VDC
50 mA
µA max.
20Digital Indicators K3HB-X/-V/-H/-S
DeviceNet Communications
Communications protocolConforms to DeviceNet
Supported communications
Connection methodsCombination of multi-drop and T-branch connections (for trunk and drop lines)
Baud rateDeviceNet: 500, 250, or 125 Kbps (automatic follow-up)
Communications mediaSpecial 5-wire cable (2 signal lines, 2 power supply lines, 1 shield line)
Communications distance
Communications power supply24-VDC DeviceNet power supply
Allowable voltage fluctuation range11 to 25-VDC DeviceNet power supply
Current consumption50 mA max. (24 VDC)
Maximum number of nodes64 (DeviceNet Configurator is counted as one node when connected)
Maximum number of slaves63
Error control checksCRC errors
DeviceNet power supplySupplied from DeviceNet communications connector
I/O allocationsAllocate any I/O data using the Configurator.
Allocate any data, such as DeviceNet-specific parameters and variable area for Digital
Indicators.
Input area: 2 blocks, 60 words max.
Output area: 1 block, 29 words max.
(The first word in the area is always allocated for the Output Execution Enabled Flags.)
CompoWay/F communications commands can be executed (using explicit message
communications)
Baud rateNetwork length
(max.)
Drop line length
(max.)
500 Kbps100 m (100 m)6 m39 m
250 Kbps100 m
6 m78 m
(250 m)
125 Kbps100 m
6 m156 m
(500 m)
The values in parentheses are for Thick Cable.
Total drop line
length (max.)
■ Internal Block Diagram
Analog input
BCD I/O
Event input
Linear output
Sensor power
supply
Input
circuit
BCD
• Input circuit
• Output circuit
• Transistor
output circuit
Digital input
circuit
Linear output
circuit
Filter
V
O
ADC
EEP-ROM
Drive
circuit
Micro-
Waveform
shaping circuit
Drive
circuit
Power supply circuit (isolated)
Power supply
computer
Keys
Indications
COM
Transistor
output
Relay output
DeviceNet
Communications
Drive
circuit
Drive
X
circuit
Drive
circuit
Drive
circuit
DD
V
DeviceNet
circuit
Communi-
cations
driver
V
V
O
DC-DC
Converter (isolated)
Digital Indicators K3HB-X/-V/-H/-S21
■ Power Supply Derating Curve for Sensor (Reference Value)
With 12 V
With 10 V
Max. current (mA)
140
120
100
80
60
40
20
0
−20 −100 102030405060
Ambient temperature (°C)
1
Max. current (mA)
140
120
100
80
60
40
20
0
−20 −100 102030405060
Ambient temperature (°C)
1
Note: 1. The above values are for standard mounting. The derating curve differs depending on the mounting conditions.
2. Do not use the Sensor outside of the derating area (i.e., do not use it in the area labeled A in the above graphics). Doing so may occa-
sionally cause deterioration or damage to internal components.
■ Component Names and Functions
Max/Min status indicator
Turns ON when the maximum value or minimum value is
displayed in the RUN level.
PV display
Displays PVs, maximum values,
minimum values, parameter
names, and error names.
Level/bank display
In RUN level, displays the bank if the bank function is
ON. (Turns OFF if the bank function is OFF.)
In other levels, displays the current level.
Comparative output
status indicators
Display the status of comparative outputs.
Status indicators
Display
Turns ON when the tare zero function is
executed. Turns OFF if it is not executed or
T-ZR
is cleared.
Turns ON when the forced-zero function is
Zero
executed. Turns OFF if it is not executed or
is cleared. (Excluding the K3HB-H.)
Turns ON/OFF when hold input turns
Hold
ON/OFF.
Function
HH
H
P
L
LL
MAX/MIN
Max
Min
L
T-ZR
Zero
TG HH H
CMW
TLLL
Hold
UPMODELEVEL
SHIFT
Position meter
Displays the position of the PV with
respect to a desired scale.
SV display
Displays SV and monitor values.
SV display status indicators
DisplayFunction
Turns ON when the timing signal turns
ON. Otherwise OFF.
TG
Turns ON when parameters for which
teaching can be performed are
T
displayed.
In RUN level, turn ON when the
HH, H,
comparative set values HH, H, L, and
L, LL
LL are displayed.
MAX/MIN KeyLEVEL KeyMODE KeySHIFT KeyUP Key
Used to switch the
display between the
PV, maximum value,
and minimum value and
to reset the maximum
and minimum values.
Used to switch level.
Used to switch the
parameters displayed.
Used to change
parameter settings.
When changing a set
value, this key is used
to move along the
digits.
When changing a set value,
this key is used to change
the actual value.
When a measurement value
is displayed, this key is
used to execute or clear the
forced-zero function or to
execute teaching.
22Digital Indicators K3HB-X/-V/-H/-S
■ BCD Output Timing Chart
A REQUEST signal from a Programmable Controller or other external device is required to read BCD data.
Single Sampling Data Output
20-ms pulse min. (50 ms max.)
REQ.
MAX.MIN.
DATA
DATA
VALID
The data is set in approximately 30 ms from the rising edge of the
REQUEST signal and the DATA VALID signal is output.
When reading the data from a Programmable Controller, start
reading the data when the DATA VALID signal turns ON.
The DATA VALID signal will turn OFF 40 ms later, and the data will
turn OFF 16 ms after that.
All data "High"All data "High"
Approx.
30 ms
K3HB
(1)
K3HB
(2)
K3HB
(3)
DATA (including POL and
OVER) and DATA VALID
can be used in a wired OR.
RUN, HH, H, PASS, L, and
LL are always output, with or
without a REQUEST signal.
Do not used a wired OR
connect for these signals.
Data
40 ms
Programmable
Controller
16 ms
REQ. (1)
REQ. (2)
REQ. (3)
DATA
DATA
VALID
Note: Leave 20 ms min. between DATA VALID turning OFF and the REQUEST signal.
Continuous Data Output
REQ.
MAX.MIN.
DATA
All data "High"
DATA
VALID
Approx.
30 ms
Measurement data is output every 64 ms while the REQUEST signal
remains ON.
Note: If HOLD is executed when switching between data 1 and data
2, either data 1 or data 2 is output depending on the timing of
the hold signal. The data will not go LOW.
(1)(2)(3)
(See note.)(See note.)
Data 1Data 2
40 ms40 ms
(See note.)
24 ms24 ms
64 ms64 ms
Programmable Controller Connection Example
Digital Indicator
+5 V
240 Ω
240 Ω
Connector pin No. (See note.)
1.COMMON
2.1
3.2
10°
4.4
5.8
23.DATA VALID
24.RUN
25.COMMON
26.REQUEST
240 Ω
30.RESET
240 Ω
31.POLARITY
(+/− polarity)
SYSMAC Programmable Controller
DC Input Unit
COM
IN
IN
IN
IN
IN
IN
Transistor Output Unit
OUT
OUT
COM
(0 V)
24 VDC
+24 V
DC power
supply
0 V
Internal circuit
Internal circuit
Internal circuit
Note: The BCD output connector pin number is the D-sub connector pin number when
the BCD Output Cable (sold separately) is connected. This number differs from
the pin number for the Digital Indicator narrow pitch connector (manufactured by
Honda Tsushin Kogyo Co., Ltd.).
Display Unit Connection Example
Digital Indicator
Connector pin No. (See note.)
1.COMMON
2.1
3.2
10°
4.4
5.8
23.DATA VALID
24.RUN
1
To 10
2
To 10
+5 V
240 Ω
240 Ω
240 Ω
240 Ω
25.COMMON
26.REQUEST
30.RESET
31.POLARITY
(+/− polarity)
Shortcircuit
V
V
B
A
D
C
O
O
LE
DP
8
M7E-01D@N2, 01H@N2
<M7E Digital Display Unit>
D
8
V
V
B
B
C
A
A
D
C
O
O
LE
LE
DP
DP
8
SEC
Refer to the following User's Manual for application precautions and other information required when using the Digital Indicator:
K3HB-S/-X/-V/-H Digital Indicator User's Manual (Cat. No. N128)
The manual can be downloaded from the following site in PDF format: OMRON Industrial Web http://www.fa.omron.co.jp
• Special Cable (Sold separately)
K32-DICN (OMRON)
(XG4M-1030 with 3 m cable)
2: S-TMR
2
1
4: RESET
6: COM
8: BANK2
10
9
10: COM
Weighing Indicator
K3HB-V
E1
N/C
E2
A
E3
B
E4
C
E5
D
E6
COM
Temperature Indicator
K3HB-H
E1
E2
N/C
E3
A
E4
Pt
E5
B
TC
E6
COM
+
B'
Linear Sensor Indicator
Current input
Voltage input
• Use terminal pin D6 as the common terminal.
• Use NPN open collector or no-voltage contacts for event input.
PNP types are also available.
4.7 kΩ
3.9 kΩ
12 V
TIMING
COM
D1
D6
S-TMR: D2
HOLD: D3
RESET: D4
ZERO: D5
COM
D6
K3HB-S
N/C
560 Ω
750 Ω
E1
E2
E3
E4
E5
E6
12 V
Input B
Input A
COM
Input A
Input B
BCD Output Cable
ModelShapePin arrangement
K32-BCD
K3HB end
38 mm46.5 mm
Cover: HDR-E50LPA5
(manufactured by Honda Tsushin Co., Ltd)
Connector: HDR-E50MAG1
(manufactured by Honda Tsushin Co., Ltd)
300 mm
D-sub connector (37-pin female)
Cover: 17JE-37H-1A (manufactured by DDK)
Connector: Equivalent to 17JE-13370-02
(manufactured by DDK)
Stand: 17L-002A
Connected device end
(PLC, display device, etc.)
manufactured by DDK
COMMON
1
20
4
4
2
10
1
21
8
3
2
0
22
10
4
4
5
8
6
1
7
7
2
1
10
8
4
9
8
10
1
11
2
2
10
12
4
13
8
14
1
15
2
3
10
16
4
17
8
18
1
4
10
19
2
OVER
23
DATA VALI
24
RUN
25
COMMON
26
REQUEST
27
MAX REQ.
28
MIN REQ.
29
HOLD
30
RESET
31
POLARITY
32
HH
33
H
34
PASS
35
L
36
LL
37
COMMON
Note: The BCD Output Cable has a D-sub plug. Cover: 17JE-37H-1A (manufactured by DDK); Connector: equivalent to 17JE-23370-02 (D1) (manufactured by DDK)
Special Cable (for Event Inputs with 8-pin Connector)
ModelAppearanceWiring
K32-DICN
910
12
Cable marking
3,000 mm
(3 m)
Pin No.
1
2
3
4
5
6
7
8
9
10
Signal name
N/C
S-TMR
HOLD
RESET
N/C
COM
BANK4
BANK2
BANK1
COM
Digital Indicators K3HB-X/-V/-H/-S25
■ Main Functions
e
e
Measurement
Input Calculation
S
• Two input circuits are provided. The input ranges for these circuits
can be set independently. For example, one can be set to 4 to 20
mA and the other can be set to 1 to 5 V.
• In addition to calculations such as K (constant)–A (input for one circuit), it is possible to perform calculations based on the inputs for
both circuits, such as A+B and A–B, making it possible to perform
thickness measurement and level-difference measurement using
displacement and length-measuring sensors.
Timing Hold
X V H S
Normal
• Continuously performs measurement and always outputs based on
comparative results.
Input
H comparative set value
Measurement value
Time
H output
Peak Hold/Bottom Hold
• Measures the maximum (or minimum) value in a specified period.
Input
Measurement value
ON
TIMING
input
OFF
Peak hold value
Bottom hold value
Time
A
K
K−A
Sampling Hold
• Holds the measurement at the rising edge of the TIMING signal.
Input
Measurement value
ON
TIMING
input
OFF
Sampling hold valu
Time
Peak-to-peak Hold
• Measures the difference between the maximum and minimum val-
ues in a specified period.
Input
Measurement value
ON
TIMING
input
OFF
(b1−a1)
b1
a1
'a2
Peak-to-peak value
(b2−a2)
'b2
Time
ScalingAverage Processing
Scaling converts input signals in any way required before displaying
them. The values can be manipulated by shifting, inverting, or +/–
reversing.
Display value
Display value 2
(
dsp.a2
Display value 1
(
dsp.a1
)
)
Input value 1
Input value 2
(inp.a1)
(inp.a2)
(Scaling)(Reverse scaling)
Display value
Display value 2
Display value 1
Input value
Teaching
Settings for scaling can be made using the present measurement
values instead of inputting values with the SHIFT and UP Keys. This
is a convenient function for making the settings while monitoring the
operating status.
Standby Sequence
X V SX V H S
Turns the comparative output OFF until the measurement value
enters the PASS range.
Average processing of input signals with extreme changes or noise
smooths out the display and makes control stable.
(
dsp.a2
)
(
dsp.a1
)
Input value 2
X V S
X V H S
(
inp.a2
)
Input value 1
(
inp.a1
Input valu
)
Previous Average Value Comparison
Slight changes can be removed from input signals to detect only
extreme changes.
Temperature Input Shift
Shifts the temperature input value.
Supported Models
The models that support the functions shown here are indicated by symbols as follows:
K3HB-X
X
K3HB-V
V
K3HB-H
H
K3HB-S
S
X V H S
H
26Digital Indicators K3HB-X/-V/-H/-S
■ Input Compensation/Display
Forced-zero
Forces the present value to 0. (Convenient for setting reference values or deducting tares for weight measurement.)
Tare Z e ro
Shifts the current value measured with a forced zero to 0 again. It is
possible to measure two or more compounds separately and then, by
releasing the tare zero and forced-zero, measure the combined total.
Zero-trimming
Compensates for mild fluctuations in input signals due to factors
such as sensor temperature drift, based on OK (PASS) data at measurement. (This function can be used with sampling hold, peak hold,
or bottom hold.)
Zero-limit
Changes the display value to 0 for input values less than the set
value. It is enabled in normal mode only. (This function can be used,
for example, to stop negative values being displayed or to eliminate
flickering and minor inconsistencies near 0.)
Display
X V S
V S
X V H S
X V H S
The current display value can be selected from the present value, the
maximum value, and the minimum value.
It is possible to specify (i.e., restrict) the values that the smallest displayed digit can change by. For example, if the setting is 2, the smallest digit will only take the values 0, 2, 4, 6, or 8 and if the setting is 5,
it will only take the values 0 or 5. If the setting is 10, it will only take
the value of 0.
Zero-limit
setting
Input
Display Refresh Period
The display refresh period can be lengthened to reduce flickering
and thereby make the display easier to read.
Display Color Selection
Values can be displayed in either red or green. With comparative output models, the display color can also be set to change according to
the status of comparative outputs (e.g., green to red or red to green).
X
Digital Indicators K3HB-X/-V/-H/-S27
■ Output
e
Comparative Output Pattern
X V H S
The output pattern for comparative outputs can be selected. In addition to high/low comparison with set values, output based on level
changes is also possible. (Use the type of output pattern appropriate
for the application.)
Comparative
set value HH
Comparative
set value H
Comparative
set value L
Comparative
set value LL
Output HH
Output H
Output PASS
Output L
Output LL
Comparative
set value HH
Comparative
set value H
Comparative
set value L
Comparative
set value LL
Output HH
Output H
Output PASS
Output L
Output LL
Output Logic
Standard Output
Zone Output
Measurement
value
Higher
Lower
ON
OFF
Measurement
value
Higher
Lower
ON
OFF
Comparative
set value HH
Comparative
set value H
Comparative
set value L
Comparative
set value LL
Output HH
Output H
Output PASS
Output L
Output LL
Level Output
X V H S
Measurement
Reverses the output operation of comparative outputs for comparative results.
value
ON
OFF
Higher
Lower
Hysteresis
X V H S
Prevents comparative output chattering when the measurement
value fluctuates slightly near the set value.
Measurement can be stopped for a set time using external input.
Input
H comparative set value
Tim
H output
H input
PASS Output Change
Startup compensation time
set time
X V H S
Comparative results other than PASS and error signals can be output
from the PASS output terminal.
■ Dimensions
101.2
91
96
Terminal cover (included)
Character Size for Main Display (mm)
PV displaySV display
14.2
100
(112)
12
1.3
48
7.6
95*
2
*DeviceNet models: 97 mm
Terminal: M3, Terminal Cover: Accessory
3.5
4.9
44.8
Panel Cutout Dimensions
75 min.
+0.8
92
0
120 min.
45
+0.6
0
28Digital Indicators K3HB-X/-V/-H/-S
■ Wiring Precautions
• For terminal blocks, use the crimp terminals suitable for M3 screws.
• Tighten the terminal screws to the recommended tightening torque
of approx. 0.5 N
• To prevent inductive noise, separate the wiring for signal lines from
that for power lines.
Wiring
• Use the crimp terminals suitable for M3 screws shown below.
⋅m.
5.8 mm max.
■ Mounting Method
1. Inser t the K3HB into the mounting cutout in the panel.
2. Inser t watertight packing around the Unit to make the mounting
watertight.
Watertight
packing
5.8 mm max.
Unit Stickers
• Select the appropriate units from the unit sticker sheets provided
and attach the sticker to the Indicator.
Note: When using for meters, such as weighing meters, use the units
specified by regulations on weights and measures.
3. Inser t the adapter into the grooves on the left and right sides of
the rear case and push until it reaches the panel and is fixed in
place.
Adapter
■ LCD Field of Vision
The K3HB is designed to have the best visibility at the angles shown
in the following diagram.
10°
30°
■ Waterproof Packing
The waterproof packing ensures a level of waterproofing that conforms to NEMA 4X. Depending on the operating environment, deterioration, contraction, or hardening may occur and replacement may
be necessary. In this case, consult your OMRON representative.
Digital Indicators K3HB-X/-V/-H/-S29
■ Precautions
!WARNING
Do not touch the terminals while power is being supplied.
Doing so may possibly result in electric shock. Make sure
that the terminal cover is installed before using the
product.
Always provide protective circuits in the network. Without
protective circuits, malfunctions may possibly result in
accidents that cause serious injury or significant property
damage.
Provide double or triple safety measures in external
control circuits, such as emergency stop circuits, interlock
circuits, or limit circuits, to ensure safety in the system if
an abnormality occurs due to malfunction of the product
or another external factor affecting the product's
operation.
!CAUTION
Do not allow pieces of metal, wire clippings, or fine
metallic shavings or filings from installation to enter the
product. Doing so may occasionally result in minor
electric shock, fire, or malfunction.
Do not use the product in locations where flammable or
explosive gases are present. Doing so may occasionally
result in explosion, causing minor or moderate injury, or
property damage.
Do not attempt to disassemble, repair, or modify the
product. Doing so may occasionally result in minor or
moderate injury due to electric shock.
Do not use the equipment for measurements within
Measurement Categories III and IV for K3HB-X and II, III,
and IV for K3HB-S, K3HB-V, and K3HB-H (according to
IEC61010-1). Doing so may occasionally cause
unexpected operation, resulting in minor or moderate
injury, or damage to the equipment. Use the equipment
for measurements only within the Measurement Category
for which the product is designed.
Perform correct setting of the product according to the
application. Failure to do so may occasionally cause
unexpected operation, resulting in minor or moderate
injury, or damage to the equipment.
Ensure safety in the event of product failure by taking
safety measures, such as installing a separate monitoring
system. Product failure may occasionally prevent
operation of comparative outputs, resulting in damage to
the connected facilities and equipment.
Tighten the screws on the terminal block and the
connector locking screws securely using a tightening
torque within the following ranges. Loose screws may
occasionally cause fire, resulting in minor or moderate
injury, or damage to the equipment.
Terminal block screws: 0.43 to 0.58 N·m
Connector locking screws:0.18 to 0.22 N·m
Make sure that the product will not be adversely affected
if the DeviceNet cycle time is lengthened as a result of
changing the program with online editing. Extending the
cycle time may cause unexpected operation, occasionally
resulting in minor or moderate injury, or damage to the
equipment.
Before transferring programs to other nodes or changing
I/O memory of other nodes, check the nodes to confirm
safety. Changing the program or I/O memory of other
nodes may occasionally cause unexpected operation,
resulting in minor or moderate injury, or damage to the
equipment.
Precautions for Safe Use
1. Do not use the product in the following locations.
• Locations subject to direct radiant heat from heating equipment
• Locations where the product may come into contact with water or
oil
• Locations subject to direct sunlight
• Locations where dust or corrosive gases (in particular, sulfuric or
ammonia gas) are present
• Locations subject to extreme temperature changes
• Locations where icing or condensation may occur
• Locations subject to excessive shocks or vibration
2. Do not use the product in locations subject to temperatures or
humidity levels outside the specified ranges or in locations prone
to condensation. If the product is installed in a panel, ensure that
the temperature around the product (not the temperature around
the panel) does not go outside the specified range.
3. Provide sufficient space around the product for heat dissipation.
4. Use and store the product within the specified temperature and
humidity ranges. If several products are mounted side-by-side or
arranged in a vertical line, the heat dissipation will cause the internal temperature of the products to rise, shortening the service life.
If necessary, cool the products using a fan or other cooling
method.
5. The service life of the output relays depends on the switching
capacity and switching conditions. Consider the actual application
conditions and use the product within the rated load and electrical
service life. Using the product beyond its service life may result in
contact welding or burning.
6. Install the product horizontally.
7. Mount to a panel between 1 and 8-mm thick.
8. Use the specified size of crimp terminals (M3, width: 5.8 mm
max.) for wiring. To connect bare wires, use AWG22 (cross section: 0.326 mm
power supply terminals and AWG28 (cross section: 0.081 mm
to AWG16 (cross section: 1.309 mm
of exposed wire: 6 to 8 mm)
9. In order to prevent inductive noise, wire the lines connected to the
product separately from power lines carrying high voltages or currents. Do not wire in parallel with or in the same cable as power
lines. Other measures for reducing noise include running lines
along separate ducts and using shield lines.
10.Ensure that the rated voltage is achieved no longer than 2 s after
turning the power ON.
11.Allow the product to operate without load for at least 15 minutes
after the power is turned ON.
12.Do not install the product near devices generating strong high-frequency waves or surges. When using a noise filter, check the voltage and current and install it as close to the product as possible.
13.Do not use thinner to clean the product. Use commercially available alcohol.
14.Be sure to confirm the name and polarity for each terminal before
wiring the terminal block and connectors.
15.Use the product within the noted supply voltage and rated load.
16.Do not connect anything to unused terminals.
17.Output turns OFF when the mode is changed or settings are ini-
tialized. Take this into consideration when setting up the control
system.
18.Install an external switch or circuit breaker that complies with
applicable IEC60947-1 and IEC60947-3 requirements and label
them clearly so that the operator can quickly turn OFF the power.
19.Use the specified cables for the communications lines and stay
within the specified DeviceNet communications distances. Refer
to the User's Manual (Cat. No. N129) for details on communications distance specifications and cables.
20.Do not pull the DeviceNet communications cables with excessive
force or bend them past their natural bending radius.
2
) to AWG14 (cross section: 2.081 mm2) to wire the
2
) for other terminals. (Length
2
)
30Digital Indicators K3HB-X/-V/-H/-S
21.Do not connect or remove connectors while the DeviceNet power
is being supplied. Doing so will cause product failure or malfunction.
22.Use cables with a heat resistance of 70ºC min.
■ Noise Countermeasures
1. Do not install the product near devices generating strong high-fre-
quency waves or surges, such as high-frequency welding and
sewing machines.
2. Mount a surge suppressor or noise filter to peripheral devices
generating noise, in particular, motors, transformers, solenoids,
and magnet coils.
Line filter
+
Digital
Indicator
Power supply input
Indicator
Signal input
-
Surge
suppressor
Digital
Power supply input
3. In order to prevent inductive noise, wire the lines connected to the
terminal block separately from power lines carrying high voltages
or currents. Do not wire in parallel with or in the same cable as
power lines. Other measures for reducing noise include running
lines along separate ducts and using shield lines.
Example of Countermeasures for Inductive Noise on
Input Lines
+
Signal
input
2 conductors
with shield
4. If a noise filter is used for the power supply, check the voltage and
current, and install the noise filter as close to the product as possible.
5. Reception interference may occur if the product is used close to a
radio, television, or wireless.
Digital
Indicator
-
Digital Indicators K3HB-X/-V/-H/-S31
Warranty and Limitations of Liability
■ WARRANTY
OMRON's exclusive warranty is that the products are free from defects in materials and workmanship for a period of one year (or other period if
specified) from date of sale by OMRON.
OMRON MAKES NO WARRANTY OR REPRESENTATION, EXPRESS OR IMPLIED, REGARDING NON-INFRINGEMENT, MERCHANTABILITY,
OR FITNESS FOR PARTICULAR PURPOSE OF THE PRODUCTS. ANY BUYER OR USER ACKNOWLEDGES THAT THE BUYER OR USER
ALONE HAS DETERMINED THAT THE PRODUCTS WILL SUITABLY MEET THE REQUIREMENTS OF THEIR INTENDED USE. OMRON DISCLAIMS ALL OTHER WARRANTIES, EXPRESS OR IMPLIED.
■ LIMITATIONS OF LIABILITY
OMRON SHALL NOT BE RESPONSIBLE FOR SPECIAL, INDIRECT, OR CONSEQUENTIAL DAMAGES, LOSS OF PROFITS, OR COMMERCIAL LOSS IN ANY WAY CONNECTED WITH THE PRODUCTS, WHETHER SUCH CLAIM IS BASED ON CONTRACT, WARRANTY, NEGLIGENCE, OR STRICT LIABILITY.
In no event shall the responsibility of OMRON for any act exceed the individual price of the product on which liability is asserted.
IN NO EVENT SHALL OMRON BE RESPONSIBLE FOR WARRANTY, REPAIR, OR OTHER CLAIMS REGARDING THE PRODUCTS UNLESS
OMRON'S ANALYSIS CONFIRMS THAT THE PRODUCTS WERE PROPERLY HANDLED, STORED, INSTALLED, AND MAINTAINED AND NOT
SUBJECT TO CONTAMINATION, ABUSE, MISUSE, OR INAPPROPRIATE MODIFICATION OR REPAIR.
Application Considerations
■ SUITABILITY FOR USE
OMRON shall not be responsible for conformity with any standards, codes, or regulations that apply to the combination of products in the customer's application or use of the products.
At the customer's request, OMRON will provide applicable third party cer tification documents identifying ratings and limitations of use that apply
to the products. This information by itself is not sufficient for a complete determination of the suitability of the products in combination with the end
product, machine, system, or other application or use.
The following are some examples of applications for which particular attention must be given. This is not intended to be an exhaustive list of all
possible uses of the products, nor is it intended to imply that the uses listed may be suitable for the products.
• Outdoor use, uses involving potential chemical contamination or electrical interference, or conditions or uses not described in this catalog.
• Nuclear energy control systems, combustion systems, railroad systems, aviation systems, medical equipment, amusement machines, vehicles,
safety equipment, and installations subject to separate industry or government regulations.
• Systems, machines, and equipment that could present a risk to life or property.
Please know and observe all prohibitions of use applicable to the products.
NEVER USE THE PRODUCTS FOR AN APPLICATION INVOLVING SERIOUS RISK TO LIFE OR PROPERTY WITHOUT ENSURING THAT THE
SYSTEM AS A WHOLE HAS BEEN DESIGNED TO ADDRESS THE RISKS, AND THAT THE OMRON PRODUCTS ARE PROPERLY RATED
AND INSTALLED FOR THE INTENDED USE WITHIN THE OVERALL EQUIPMENT OR SYSTEM.
ALL DIMENSIONS SHOWN ARE IN MILLIMETERS.
To convert millimeters into inches, multiply by 0.03937. To convert grams into ounces, multiply by 0.03527.
Cat. No. N131-E1-04
In the interest of product improvement, specifications are subject to change without notice.
OMRON Corporation
Industrial Automation Company
Control Devices Division H.Q.
Analog Controller Division
Shiokoji Horikawa, Shimogyo-ku,
Kyoto, 600-8530 Japan
Tel: (81)75-344-7080/Fax: (81)75-344-7189
OMRON EUROPE B.V.
Wegalaan 67-69, NL-2132 JD Hoofddorp
The Netherlands
Tel: (31)2356-81-300/Fax: (31)2356-81-388
OMRON ELECTRONICS LLC
1 East Commerce Drive, Schaumburg, IL
60173
U.S.A.
Tel: (1)847-843-7900/Fax: (1)847-843-8568