Omron FQ-CR1 User Manual

Page 1
Fixed Mount Multi Code Reader
FQ-CR1
User's Manual
Cat. No. Z315-E1-01
Page 2
Introduction
Thank you for purchasing the FQ-CR1. This manual provides information regarding functions, performance and operating methods that are required for using the FQ-CR1. When using the FQ-CR1, be sure to observe the following:
• The FQ-CR1 must be operated by personnel knowledgeable in electrical engineering.
• To ensure correct use, please read this manual thoroughly to deepen your understanding of the product.
• Please keep this manual in a safe place so that it can be referred to whenever necessary.
Page 3
APPLICATION CONSIDERATIONS (Please Read)
1
User's Manual
Introduction
Installation and Connections
Taking Images
Setting Up Inspections
Testing and Saving Settings
Operation
1
2
3
4
5
6
Fixed Mount Multi Code Reader
FQ-CR1
Convenient Functions
Communications with External Devices
Troubleshooting
Appendices
7
8
9
10
Page 4
READ AND UNDERSTAND THIS DOCUMENT
Please read and understand this document before using the products. Please consult your OMRON representative if you have any questions or comments.
WARRANTY
OMRON’s exclusive warranty is that the products are free from defects in materials and workmanship for a period of one year (or other period if specified) from date of sale by OMRON.
OMRON MAKES NO WARRANTY OR REPRESENTATION, EXPRESS OR IMPLIED, REGARDING NON­INFRINGEMENT, MERCHANTABILITY, OR FITNESS FOR PARTICULAR PURPOSE OF THE PRODUCTS. ANY BUYER OR USER ACKNOWLEDGES THAT THE BUYER OR USER ALONE HAS DETERMINED THAT THE PRODUCTS WILL SUITABLY MEET THE REQUIREMENTS OF THEIR INTENDED USE. OMRON DISCLAIMS ALL OTHER WARRANTIES, EXPRESS OR IMPLIED.
LIMITATIONS OF LIABILITY
OMRON SHALL NOT BE RESPONSIBLE FOR SPECIAL, INDIRECT, OR CONSEQUENTIAL DAMAGES, LOSS OF PROFITS OR COMMERCIAL LOSS IN ANY WAY CONNECTED WITH THE PRODUCTS, WHETHER SUCH CLAIM IS BASED ON CONTRACT, WARRANTY, NEGLIGENCE, OR STRICT LIABILITY.
In no event shall responsibility of OMRON for any act exceed the individual price of the product on which liability is asserted.
IN NO EVENT SHALL OMRON BE RESPONSIBLE FOR WARRANTY, REPAIR, OR OTHER CLAIMS REGARDING THE PRODUCTS UNLESS OMRON’S ANALYSIS CONFIRMS THAT THE PRODUCTS WERE PROPERLY HANDLED, STORED, INSTALLED, AND MAINTAINED AND NOT SUBJECT TO CONTAMINATION, ABUSE, MISUSE, OR INAPPROPRIATE MODIFICATION OR REPAIR.
SUITABILITY FOR USE
THE PRODUCTS CONTAINED IN THIS DOCUMENT ARE NOT SAFETY RATED. THEY ARE NOT DESIGNED OR RATED FOR ENSURING SAFETY OF PERSONS, AND SHOULD NOT BE RELIED UPON AS A SAFETY COMPO­NENT OR PROTECTIVE DEVICE FOR SUCH PURPOSES. Please refer to separate catalogs for OMRON’s safety rated products.
OMRON shall not be responsible for conformity with any standards, codes, or regulations that apply to the combination of products in the customer’s application or use of the product.
At the customer’s request, OMRON will provide applicable third party certification documents identifying ratings and limitations of use that apply to the products. This information by itself is not sufficient for a complete determination of the suitability of the products in combination with the end product, machine, system, or other application or use.
The following are some examples of applications for which particular attention must be given. This is not intended to be an exhaustive list of all possible uses of the products, nor is it intended to imply that the uses listed may be suitable for the products:
• Outdoor use, uses involving potential chemical contamination or electrical interference, or conditions or
uses not described in this document.
2
FQ User’s Manual
Page 5
• Nuclear energy control systems, combustion systems, railroad systems, aviation systems, medical
equipment, amusement machines, vehicles, safety equipment, and installations subject to separate industry
or government regulations.
• Systems, machines, and equipment that could present a risk to life or property.
Please know and observe all prohibitions of use applicable to the products.
NEVER USE THE PRODUCTS FOR AN APPLICATION INVOLVING SERIOUS RISK TO LIFE OR PROPERTY WITHOUT ENSURING THAT THE SYSTEM AS A WHOLE HAS BEEN DESIGNED TO ADDRESS THE RISKS, AND THAT THE OMRON PRODUCT IS PROPERLY RATED AND INSTALLED FOR THE INTENDED USE WITHIN THE OVERALL EQUIPMENT OR SYSTEM.
PERFORMANCE DATA
Performance data given in this document is provided as a guide for the user in determining suitability and does not constitute a warranty. It may represent the result of OMRON’s test conditions, and the users must correlate it to actual application requirements. Actual performance is subject to the OMRON Warranty and Limitations of Liability.
CHANGE IN SPECIFICATIONS
Product specifications and accessories may be changed at any time based on improvements and other reasons.
It is our practice to change model numbers when published ratings or features are changed, or when significant construction changes are made. However, some specifications of the product may be changed without any notice. When in doubt, special model numbers may be assigned to fix or establish key specifications for your application on your request. Please consult with your OMRON representative at any time to confirm actual specifications of purchased products.
DIMENSIONS AND WEIGHTS
Dimensions and weights are nominal and are not to be used for manufacturing purposes, even when tolerances are shown.
ERRORS AND OMISSIONS
The information in this document has been carefully checked and is believed to be accurate; however, no responsibility is assumed for clerical, typographical, or proofreading errors, or omissions.
PROGRAMMABLE PRODUCTS
OMRON shall not be responsible for the user’s programming of a programmable product, or any consequence thereof.
COPYRIGHT AND COPY PERMISSION
This document shall not be copied for sales or promotions without permission.
This document is protected by copyright and is intended solely for use in conjunction with the product. Please notify us before copying or reproducing this document in any manner, for any other purpose. If copying or transmitting this document to another, please copy or transmit it in its entirety.
FQ User’s Manual
3
Page 6

Meanings of Signal Words

The following signal words are used in this manual.
Indicates a potentially hazardous situation which, if not avoided, will result in minor or moderate injury, or may result in serious injury or death. Additionally there may be significant property damage.

Meanings of Alert Symbols

The following alert symbols are used in this manual
Indicates general prohibitions for which there is no specific symbol.
Indicates the possibility of laser radiation.
Indicates the possibility of explosion under specific conditions.
Indicates prohibition when there is a risk of minor injury from electrical shock or other
source if the product is disassembled.
This product is not designed or rated for ensuring safety of persons.
Do not use it for such purposes.
The Sensor emits visible light, which may adversely affect the eyes in rare instances.
Do not look directly into the light emitted from the Sensor. When the subject is a specular
reflective object, protect your eyes from reflected light.
A lithium ion battery is built into the Touch Finder and may occasionally combust, explode, or
burn if not treated properly.
Dispose of the Touch Finder as industrial waste, and never disassemble, apply pressure that
would deform, heat to 100 °C or higher, or incinerate the Touch Finder.
High-voltage parts inside; danger of electrical shock. Do not open the product cover.
4
FQ User’s Manual
Page 7

Precautions for Safe Use

The following points are important to ensure safety, so make sure that they are strictly observed.
1. Installation Environment
• Do not use the product in environments where it can be exposed to inflammable/explosive gas.
• To secure the safety of operation and maintenance, do not install the product close to high-voltage devices
and power devices.
• Install the product in such a way that its ventilation holes are not blocked.
• Tighten mounting screws at the torque specified in this manual.
2. Power Supply and Wiring
• The power supply voltage must be within the rated range (24 VDC ±10%), and an AC voltage must not be
used.
• Reverse connection of the power supply is not allowed. Do not short the load of the open collector output.
• The load must be within the rated range.
• High-voltage lines and power lines must be wired separately from this product. Wiring them together or
placing them in the same duct may cause induction, resulting in malfunction or damage.
• Use the products within the power supply voltages specified in this manual.
• Use the specified size of crimp terminals to wire connections. Do not connect wires that have been simply
twisted together directly to the power supply or terminal block.
• Use a DC power supply with safety measures against high voltages (safety extra low-voltage circuit).
• Use independent power sources for the products. Do not use a shared power source.
• Tighten mounting screws at the torque specified in this manual.
• Always turn OFF the power supply before connecting or disconnecting cables or the power supply wiring.
3. Battery
• Do not short the positive and negative terminals of the Battery.
• Do not use the Touch Finder in an environment that exceeds the operating temperature range of the Battery.
If the Touch Finder is used at temperatures that exceed the operating temperature range, the protective
device may activate and prevent charging.
• Do not connect the Battery directly to a power supply or car cigarette lighter socket.
• Do not use the Touch Finder with any other type of battery.
• Turn OFF the power supply immediately if the Battery leaks or produces an odor. Electrolyte leaked from the
Battery may ignite, possibly causing smoke, rupture, or fire.
• If during usage, charging, or storage, the Battery produces an odor, heats, becomes discolored, becomes
misshapen, or exhibits any other unusual conditions, remove it and do not use it. Continuing to use such a
Battery may result in the Battery heating, smoking, rupturing, or igniting.
• If the Touch Finder (FQ-D31) will be installed permanently or semi-permanently, remove the Battery (FQ-
BAT1). If the rated temperature is exceeded with the Battery inserted, the protective circuit may activate and
stop the Touch Finder.
4. AC Adapter
• Use an AC cable that is suitable for the power supply and power voltage you are using.
• Do not touch the power plug with a wet hand. Doing so may result in electrical shock.
• If you notice an abnormal condition, such as smoke, abnormal heating of the outer surface, or a strange
odor, immediately stop using the AC Adapter, turn OFF the power, and remove the power plug from the
outlet.
Consult your dealer, as it is dangerous to attempt to repair the AC Adapter yourself.
• If the AC Adapter is dropped or damaged, turn OFF the power, remove the power plug from the outlet, and
contact your dealer. There is a risk of fire if you continue using the AC Adapter.
FQ User’s Manual
5
Page 8
5. Other
• Do not use this product in safety circuits associated with nuclear power and human life.
• Do not disassemble, repair, modify, deform by pressure, or incinerate this product.
• Dispose of this product as industrial waste.
• Connect the special products (Sensor, Touch Finder, Cables). The product might break down or malfunction
if you use a part not included in the special products.
• If you notice an abnormal condition, such as a strange odor, extreme heating of any product, or smoke,
immediately stop using the product, turn OFF the power, and consult your dealer.
• The Sensor surfaces become hot during use. Do not touch them.
• Do not drop or subject the products to shock.
• Use the special Sensor (FQ-CR1), Touch Finder (FQ-D), Cables (FQ-WN and FQ-WD), Battery (FQ-BAT1),
and AC Adapter (FQ-AC). Using other than the specified products may cause fire, burning, malfunction or
failure.
• If the product has a lock mechanism, always make sure it is locked before using the product.
6. Laws and Regulations, Standards
• This product complies with the following EC Directives and EN Standards:
EC Directive No.2004/104/EC
EN Standards EN61326
6
FQ User’s Manual
Page 9

Precautions for Correct Use

Observe the following precautions to prevent failure to operate, malfunctions, or undesirable effects on product performance.
1. Installation Site
Do not install this product in locations subjected to the following conditions:
• Ambient temperature outside the rating
• Rapid temperature fluctuations (causing condensation)
• Relative humidity outside the range of 35 to 85%
• Direct vibration or shock
• Strong ambient light (such as other laser beams, light from arc-welding machines, or ultraviolet light)
• Direct sunlight or near heaters
• Strong magnetic or electric field Also, do not install this product in locations subjected to the following conditions to ensure its protective performance as described in the specifications:
• Presence of corrosive or flammable gases
• Presence of dust, salt, or iron particles
• Water, oil, or chemical fumes or spray, or mist atmospheres
2. Power Supply, Connection, and Wiring
• When using a commercially available switching regulator, make sure that the FG terminal is grounded.
• If surge currents are present in the power lines, connect surge absorbers that suit the operating
environment.
• Before turning ON the power after the product is connected, make sure that the power supply voltage is
correct, there are no incorrect connections (e.g. load short-circuit) and the load current is appropriate. Incorrect wiring may result in breakdown of the product.
• For cables, use only the special products specified in this manual.
p.185, p.186
• Use only combinations of the Sensor and Touch Finder specified in this manual. Using other combinations
may cause malfunction or damage.
• Do not turn the power OFF in the following instances. Doing so will damage data that is in the process of
being saved.
- While data is being saved in internal memory
- While data is being saved on the SD card
• The LCD panel has been made using precision technology, and sometimes a few pixels are missing in the
panel. This is due to the structure of the LCD panel, and is not a malfunction.
• Connector cover
Always attach the covers of I/O cable connector and Ethernet cable connector. This prevents extraneous material from making malfunction of the Sensor.
3. Battery
• Do not use or charge the Battery with other than the specified products.
• Do not charge the Battery with other than the specified AC adapter.
• When using the Touch Finder, the battery cover screw must be tightened.
4. AC Adapter
• During maintenance and when not using the Touch Finder for an extended time, remove the power plug
from the outlet.
• Do not bend the power cable past its natural bending radius.
• Do not use the AC Adapter with other than the specified products.
• If a voltage higher than 380 V is applied, there is a risk that the capacitor will be damaged, the pressure
valve will open, and vaporized gas will be emitted. If there is a possibility that a voltage higher than 380 V
will be applied, use a protective device.
FQ User’s Manual
7
Page 10
5. Maintenance and Inspection
Important
Note
Do not use thinner, benzene, acetone or kerosene to clean the Sensor and Touch Finder. If large dust particles adhere to the Camera, use a blower brush (used to clean camera lenses) to blow them off. Do not use breath from your mouth to blow the dust off. To remove dust particles from the Camera, wipe gently with a soft cloth (for cleaning lenses) moistened with a small amount of alcohol. Do not use excessive force to wipe off dust particles. Scratches to the Camera might cause error.
Editor's Note
Meaning of Symbols
Menu items that are displayed on the Touch Finder LCD screen, and windows, dialog boxes and other GUI elements displayed on the PC are indicated enclosed by brackets "[ ]".
Visual Aids
Indicates points that are important to achieve the full product performance, such as operational precautions.
Indicates application procedures.
Indicates pages where related information can be found.
8
FQ User’s Manual
Page 11
Table of Contents

1. Introduction

1-1 FQ-CR1 Fixed Mount Multi Code Readers . . . . . . . . . . . . . . . . . . . . . . . . 12
1-2 Measurement Process . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
1-3 Startup Display and Display Elements . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
1-4 Basic Operational Flow. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16

2. Installation and Connections

2-1 System Configuration. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
2-2 Part Names and Functions. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
2-3 Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
2-4 Wiring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
2-5 Setting Up Ethernet. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31

3. Taking Images

3-1 Selecting a Sensor for Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
3-2 Adjusting Image Quality . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
3-3 Adjusting the Object Position . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
3-4 Preventing Mutual Interference of Multiple Sensors . . . . . . . . . . . . . . . . 44
Table of Contents

4. Setting Up Inspections

4-1 Inspection Item Selection Guide . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46
4-2 Setup Procedure for Inspection Items . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
4-3 Configuring Inspection Items. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 48
4-4 Reading Bar Codes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50
4-5 Reading 2D-codes. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56
4-6 Calculations and Judgements Using Inspection Item Data . . . . . . . . . . 62

5. Testing and Saving Settings

5-1 Performing Test Measurements. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 72
5-2 Shortening the Measurement Takt Time. . . . . . . . . . . . . . . . . . . . . . . . . . 74
5-3 Adjusting the Judgement Parameters. . . . . . . . . . . . . . . . . . . . . . . . . . . . 75
5-4 Checking a List of All Inspection Item Results . . . . . . . . . . . . . . . . . . . . 76
FQ User’s Manual
9
Page 12
5-5 Saving Data to the Sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 77

6. Operation

6-1 Starting Operation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 80
6-2 Configuring the Run Mode Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 82
6-3 Checking the Trend of Measurement Results with Graphs . . . . . . . . . . 84
6-4 Adjusting Judgement Parameters during Operation. . . . . . . . . . . . . . . . 87

7. Convenient Functions

7-1 Changing the Scene to Change the Line Process . . . . . . . . . . . . . . . . . . 90
7-2 Display Functions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 92
7-3 Monitoring the Signal I/O Status . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 95
7-4 Logging Measurement Data and Image Data . . . . . . . . . . . . . . . . . . . . . . 96
7-5 Saving Sensor Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 103
7-6 SD Card Operations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 104
7-7 Convenient Functions for Operation. . . . . . . . . . . . . . . . . . . . . . . . . . . . 106
7-8 Convenient Functions for Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 108
7-9 Functions Related to the System. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 109

8. Communications with External Devices

8-1 Controlling/Outputting in Parallel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 112
8-2 Outputting/Controlling with Ethernet . . . . . . . . . . . . . . . . . . . . . . . . . . . 129

9. Troubleshooting

9-1 Error Table . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 156
9-2 Basic Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 158

10. Appendices

10-1 Menu Tables. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 160
10-2 External Reference Parameters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 172
10-3 Specifications and Dimensions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 176
10-4 Updating the Software . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 187
10-5 LED Safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 188
10-6 Requirements from Regulations and Standards . . . . . . . . . . . . . . . . . . 189
Index. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 193
Revision History . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 196
10
FQ User’s Manual
Page 13

Introduction

1-1 FQ-CR1 Fixed Mount Multi Code Readers . . . . . . . . . . . . . . . . . . . . . .12
1-2 Measurement Process . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .13
1-3 Startup Display and Display Elements . . . . . . . . . . . . . . . . . . . . . . . . .14
1-4 Basic Operational Flow . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .16
1
Introduction
Page 14

1-1 FQ-CR1 Fixed Mount Multi Code Readers

The same functions as those that are provided by the Touch Finder can be performed from a computer. The PC Tool is available free of charge.
Includes the camera, lighting, measurement processor, and I/O functions. After the Sensor has been set up, it can be operated alone to perform measurements without the Touch Finder or PC Tool.
PC Tool
Used to check images and set the judgement parameters. It can also be used to save measurement results and check status during operation.
FQ-CR1 Sensor
Touch Finder
Setup, Image Confirmation, and Logging Tools
The FQ-CR1 is a Fixed Mount Multi Code Reader created by adding bar code and 2D-code inspection functions to the integrated processing FQ-series Sensor. Once configured, they are used stand-alone for quality inspection of presence, position, and other product characteristics. To set up or monitor the sensors, either the touch screen based console 'Touch Finder' or a 'PC Tool' can be used.
12
FQ-CR1 Fixed Mount Multi Code Readers
FQ User’s Manual
Page 15

1-2 Measurement Process

Logging
This section describes the basic flow of the measurement process.
Trigger input
Take image
urement
Meas
O
utput
• The measurement is started by inputting a trigger signal from an external device.
• Images are taken according to the trigger.
• The image is measured to see if it matches the configured settings.
• You can also perform calculations based on the measurement results from inspection items.
• The overall judgement of all inspection items are output using OR logic.
• You can output detailed measurement results from the inspection items. (An Ethernet connection is required.)
• Measurement data and image data can be logged in memory in the Sensor or in an SD card.
1
Introduction
FQ User’s Manual
Measurement Process
13
Page 16

1-3 Startup Display and Display Elements

Startup Display

The Sensor is automatically detected by the Touch
1
Finder when power supply to the Sensor and Touch Finder is turned ON.
The Auto Connect Display will appear if the Sensor can­not be detected. Check that cables are connected cor­rectly to the Sensor and Touch Finder, and then press [Auto connect].
Note
If the Sensor is still not detected after pressing [Auto Connect], refer to the following information.
The Sensor cannot be detected: p. 158
2 When the Sensor is detected, the following display will appear.
• The Setup Mode will appear if a Sensor that has not been set up is connected.
• The Run Mode will appear if a Sensor that has been set up is connected.
Note
When the Touch Finder is started, IP addresses are automatically set for each Sensor. To allocate specific IP addresses, set the IP address of each Sensor and the Touch Finder.
Setting Up Ethernet: p. 31
14
Startup Display and Display Elements
FQ User’s Manual
Page 17

Display Elements

Tool Button: Used to call functions, such as saving data or select scenes.
Capture Button: Used to capture the current screen to the SD card.
This button menu is always displayed.
[Image]: Used to adjust the image.
[In/Out]: Used to set the I/O.
[Run]: Used to switch to Run Mode.
[Test]: Used to test and adjust the set measurements.
[Inspect]: Used to set the inspection items.
The setup flow is shown by these five tabs.
If the [ ] Button appears, pressing it will display the sub-menu or commands.
Buttons will appear on the right according to the mode.
The menu changes according to the selected tab page.
The selected scene number is displayed.
The name of the Sensor being set up is displayed.
Only-image Button: Used to select either displaying the camera image and messages, or
only the camera image.
Display Button: Used to select the source of the image or to zoom the image.
The name of the mode or the menu hierarchy is displayed.
p. 107
Display Functions: p. 92
This Sensor has a Setup Mode and a Run Mode. Refer to the following information for menu items.
p. 160
Setup Mode
In Setup Mode, you can set the image conditions, judgement parameters, and I/O settings for the Sensor.
1
Introduction
Note
The Display Button can be used to switch between the following images.
• Live: The live image is displayed.
• Freeze: The image that was taken last is displayed.
• Log: An image saved in internal memory is displayed.
• File: An image saved on an SD card is displayed.
Run Mode
In Run Mode, measurements are performed, and measurement results are output.
p. 79
FQ User’s Manual
Startup Display and Display Elements
15
Page 18

1-4 Basic Operational Flow

Operation
*2
Starting Operation (Run Mode)
Saving the Settings
Adjusting Judgement
Parameters
Test Measurement and
Results Verification
Testing ([Test] Tab Page)
Output Settings
([In/Out] Tab Page)
Setup Evaluation
*1
Setting Judgement
Parameters
Inspection Setup
([Inspect] Tab Page)
Teaching
Registering
Inspection Items
Image Setup
([Image] Tab Page)
Starting the Sensor
Connections and Wiring
Judgement
Expression
Inspection
Calculation
Section 2 Installation and Connections
Section 1 1-3 Startup Display and Display Elements
Section 3 Taking Images
Section 4 Setting Up Inspections
Section 8 Communications with External Devices
Section 5 Testing and Saving Settings
Section 6 Operation
The following flow shows the basic operation of FQ-CR1 Sensors.
*1: In Setup Mode, the Sensor can be set up and adjusted, but it does not output signals on the I/O lines. *2: In Run Mode, the Sensor performs measurements and outputs signals on the I/O lines.
16
Basic Operational Flow
FQ User’s Manual
Page 19

Installation and Connections

2-1 System Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
2-2 Part Names and Functions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
2-3 Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
2-4 Wiring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .26
2-5 Setting Up Ethernet . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .31
2
Installation and Connections
Page 20

2-1 System Configuration

Connecting to Ethernet
PLC
24-VDC power supply
24-VDC power supply
FQ Ethernet Cable
I/O Cable
Touch Finder or PC Tool
Setup Tool
FQ-CR1 Sensor
Parallel Interface Connection
Trigger Sensor
Touch Finder or PC Tool
Setup Tool
Switching Hub
Standard RJ45 Ethernet Cable
Standard RJ45 Ethernet Cable
FQ-CR1 Sensor
24-VDC power supply
Standard RJ45 Ethernet Cable
I/O Cable
PLC to control Sensor
24-VDC power supply
Switching Hub Switching Hub
Switching Hub
FQ-CR1 Sensor
Multiple Connection Sensors
Standard RJ45 Ethernet Cable
Standard RJ45 Ethernet Cable
Standard RJ45 Ethernet Cable
Touch Finder or PC Tool
Setup Tool
FQ Ethernet Cable FQ Ethernet Cable
Important
Product Model number Remarks
FQ-CR1 Sensor FQ-CR1@@@@@ This is the Sensor.
Touch Finder FQ-D@@ This is a setup console. (Software version must be 1.3 or higher)
PC Tool --- The PC Tool can be used instead of the Touch Finder. (Software version must be 1.3 or
FQ Ethernet Cable FQ-WN0@@ Connects the Sensors to external devices such as the Touch Finder, computers, and PLCs.
Standard RJ45 Ethernet Cable
I/O Cable FQ-WD0@@ Connects the Sensor to the power supply and external devices.
Switching Hub W4S1-0@@ Used to connect multiple Sensors to one Touch Finder or PC Tool.
*1: The shape and dimensions of the Ethernet connector plug and jack are specified in ISO/IEC8877:1992 (JIS X 5110:1996) and RJ-45 of the
Do not connect network devices other than PLCs on the same network as the Touch Finder or computer. If another device is connected, the responsiveness of displays and settings of the Touch Finder or computer may become slow.
18
System Configuration
FCC regulations. To prevent connector connection failures, the structure of the jack of this product does not allow insertion of plugs that do not comply with the standard. If a commercially available plug cannot be inserted, it is likely that the plug is non-compliant.
--- Connects the Switching Hub to the Touch Finder, computers, and PLCs. Use a connector
*1
higher) If you register as a member, you can download the free PC Tool as a special ser­vice to purchasers. Refer to the Member Registration Sheet that is enclosed with the Sensor for the member registration procedure and the download procedure for special member software.
that complies with the FCC RJ45 standard. (STP (shielded twisted-pair) cable, category 5e or 6, impedance: 100 )
FQ User’s Manual
Page 21

2-2 Part Names and Functions

(2)
(3)
(5)
(1)
(6)
(7)
(4)
FQ-CR1 Sensor
No. Name Description
(1) Lighting LEDs for illumination
(2) Camera lens This lens can be focused.
(3) I/O Cable connector An I/O Cable is used to connect the Sensor to the power supply and exter-
nal I/O.
(4) Ethernet cable connector An Ethernet cable is used to connect the Sensor to external devices such as
PLCs, the Touch Finder, or computers.
(5) Focus adjustment screw Used to adjust the focus of the image.
(6) Operation
indicators
(7) Mounting Bracket Used to mount the Sensor.
OR Lights orange when the OR output signal turns ON.
ETN Lights orange during Ethernet communications.
ERROR Lights red when an error occurs.
9-1 Error Table p. 156.
BUSY Lights green when the Sensor is executing a process.
The Mounting Bracket can be attached to the front, left side, right side, or back of the Sensor.
2
Installation and Connections
FQ User’s Manual
Part Names and Functions
19
Page 22
Touch Finder
(1)
(3)
(4)
(6)
(7)
(5)
(2)
(12)
(10)
(11)
(8)
(9)
No. Name Description
(1) Operation
indicators
(2) LCD/touch panel Displays the setting menu, measurement results, and images input by the
(3) SD card slot An SD card can be inserted.
(4) Battery cover
(5) Power supply switch Used to turn the Touch Finder ON and OFF.
(6) Touch pen holder The touch pen can be stored here when it is not being used.
(7) Touch pen Used to operate the touch panel.
(8) DC power supply connector Used to connect a DC power supply.
(9) Slider Used to mount the Touch Finder to a DIN Track.
(10) Ethernet port Used when connecting the Touch Finder to the Sensor with an Ethernet
(11) Strap holder This is a holder for attaching the strap.
(12) AC power supply connector*1Used to connect the AC adapter.
*1: Applicable to the FQ-D31 only.
POWER Lights green when the Touch Finder is turned ON.
ERROR Lights red when an error occurs.
9-1 Error Table p. 156.
SD ACCESS Lights yellow when an SD card is inserted.
Flashes yellow when the SD card is being accessed.
CHARGE
*1
*1
Lights orange when the Battery is charging.
camera.
The Battery is inserted behind this cover. Remove the cover when mounting or removing the Battery.
p. 28
cable. Insert the connector until it locks in place.
20
Part Names and Functions
FQ User’s Manual
Page 23

2-3 Installation

Mounting Bracket
6
8
10
12
14
0
20
40
60
50
130
210
0 100
200
300
35
45
55
200
600
1,000
200 400
0
0
400
Horizontal field of view (mm)
Horizontal field of view (mm)
Installation distance (L) (mm)
Installation distance (L) (mm)
FQ-CR10050F, FQ-CR15050F
FQ-CR10100N, FQ-CR15100N
Horizontal field of view (mm)
Horizontal field of view (mm)
Installation distance (L) (mm)
Installation distance (L) (mm)
FQ-CR10010F, FQ-CR15010F
FQ-CR10100F, FQ-CR15100F
Installation distance (L)
Vertical field of view
The optical chart indicates the horizontal field of view. The vertical field of view will be approximately 60% of the horizontal field of view.
Horizontal field of view
Note: The tolerance is ±10%.
Installing the Sensor

Installation Procedure

1 Align the tabs on one side of the Mounting Bracket with
the slot on the Sensor.
The FQ-XL Mounting Bracket can be attached to the back, side, or front of the Sensor.
2 Press the Mounting Bracket onto the Sensor until the oth-
er tabs click into place.
3 Use the following optical charts to check the field of view
and installation distance of the Sensor so that it is mount­ed at the correct position.
Tightening torque (M4): 1.2 N·m
2
Installation and Connections
FQ User’s Manual
Installation
21
Page 24
• There is a certain amount of deviation among Sensors in the center of the optical axis. For this reason, when install-
Important
Important
15 mm min.
15 mm min.
ing the Sensor, check the center of the image and the field of view on the LCD monitor of the Touch Finder and in the PC Tool.

Removal Procedure

1 Insert a flat-blade screwdriver between the Mounting Brack-
et and the Sensor case on either side and remove the Mounting Bracket.
Mounting Bracket
Installing the Touch Finder

Installation Precautions

Install the Touch Finder in the following orientation to allow sufficient heat dissipation.
Do not mount it in the following orientations.
• To improve ventilation, leave space on both sides of the Touch Finder. The distance between the Touch Finder and other devices should be at least that shown in the following diagram.
• Make sure that the ambient temperature is 50C or lower. If it exceeds 50C, install an cooling fan or an air condi­tioner and maintain the temperature at 50C or lower.
• To prevent interference by noise, do not mount the Sensor on panels which contain high-voltage devices.
• To keep the level of noise from the surrounding environment to a minimum, install the Sensor and Touch Finder at least 10 m away from power lines.
22
Installation
FQ User’s Manual
Page 25

Mounting to DIN Track

Important
Important
Installation Procedure
1 Press the slider on the Touch Finder to the top. 2 Hook the clip at the top of the Touch Finder on to the DIN
Track.
3 Press the Touch Finder onto the DIN Track until the bottom
clip clicks into place.
3
• Attach End Plates (sold separately) on the sides of the Touch Finder on the DIN Track.
• If other devices will be installed next to the Touch Finder on the same DIN Track, make sure that sufficient space is kept between the devices as indicated on previous page.
• Always hook the clip at the top of the Touch Finder on the DIN Track first. If the lower clip is hooked on first, the Touch Finder will not be mounted very securely.
Removal Procedure
2
3
1
1 Pull down on the slider on the Touch Finder. 2 Lift the Touch Finder at the bottom and remove it from the
DIN Track.
2
1
2
Installation and Connections

Mounting to a Control Panel

The Touch Finder can be mounted on a panel using the FQ-XPM Panel Mounting Adapter.
• Always turn OFF the Touch Finder power before attaching or detaching the Panel Mount Adapter. Attaching or detaching with the power turned ON may cause a failure.
1 Set the Touch Finder in the Panel Mount Adapter.
FQ User’s Manual
Installation
23
Page 26
2 Press the slider up on the Touch Finder.
Hand Strap
Neck Strap
3 Create holes in the panel for mounting.
Refer to the following page for hole dimensions.
p. 183
4 Connect the cable to the Touch Finder. 5 Mount the Touch Finder with the Panel Mount Adapter from
the front of the panel.
6 Hook the hooks on the Mounting Bracket in the four holes
of the Panel Mount Adapter and secure them with screws. (Tightening torque: 1.2 N·m)
7 Check that the Touch Finder is attached properly to the
Panel.
Mounting Bracket

Using the Touch Finder as a Portable Device (with Battery)

The Touch Finder with a Battery can be used as a portable device. Use the strap when carrying it to prevent dropping it.
There are two types of straps (FQ-XH, sold separately), a Neck Strap and a Hand Strap.
1 Attach the Mini-strap to the Touch Finder.
There are a total of four holes for attaching the Mini-strap on the left and on the right of the Touch Finder.
24
Installation
FQ User’s Manual
Page 27
2 Connect the Neck Strap or Hand Strap to the Mini-strap.
Mini-strap
Neck Strap or Hand
Strap
2
Installation and Connections
FQ User’s Manual
Installation
25
Page 28

2-4 Wiring

GND OUT0 (OR) OUT1 (BUSY) OUT2 (ERROR)
IN0
TRIG
IN1 IN2 IN3 IN4 IN5
Yellow
Purple
White
Red
Green
FQ-WD0@@ I/O Cable
Gray
Pink
Light blue
Orange
Black
Blue
Power supply
Brown
Important
Important
Wiring the Sensor
Connect the I/O Cable to the I/O Cable connector located at the bottom of the Sensor.
26
Cut off lines that are not required so that they do not come into contact the other signal lines.
Classifi­cation
Powe r supply
Inputs TRIG This terminal is the trigger signal input.
Outputs OUT0 (OR) By default, this is the OR output signal (overall judgement).
The assignments of I/O signals can be changed.
Wiring
Signal Application
Power supply (24 V)
GND
IN0 to IN5 These are the command input terminals.
OUT1 (BUSY) By default, this is the BUSY output signal.
OUT2 (ERROR) By default, this is the ERROR output signal.
Note
Section 8 Communications with External Devices: p. 111
These terminals are for the external power supply (24 V).
Wire the power supply separately from other devices. If the wiring for other devices is placed together or in the same duct as the wiring for the Sensor, the influences of electromagnetic induction may cause the Sensor to malfunction or may damage it.
The assignment can be changed to an individual judgement signal from OR0 to OR31.
The assignment can be changed to an individual judgement signal from OR0 to OR31.
The assignment can be changed to an individual judgement signal from OR0 to OR31.
FQ User’s Manual
Page 29

I/O Signal Circuit Diagrams

GND (0V)
OUT0 (OR)
OUT1 (BUSY)
OUT2 (ERROR)
TRIG IN0 IN1 IN2 IN3
IN4 IN5
PNPNPN
IN0 IN1 IN2
IN3 IN4 IN5
TRIG
GND (0V)
OUT0 (OR)
OUT1 (BUSY)
OUT2 (ERROR)
Brown
Power supply (24 VDC)
24 VDC
Yellow
Blue
Load
Purple
White
Black
Red
Orange
Light blue
Green
Gray
24 VDC
Pink
Blue
Yellow
Internal circuits
Internal circuits
Purple
White
Light blue
Red
GreenOrange
Black
Gray
Load
Pink
Brown
Power supply (24 VDC)
Important
Important
Attachment Example
Warning Label
Warning Label
Preventing Chattering
• The Sensor is equipped with an anti-chattering function, but if the chattering is 100 s or longer, a faulty input may occur. (Input signals of 99 s or shorter are ignored. Signals of 100 s or longer are treated as input signals.)
• Use no-contact output devices (e.g., SSR or PLC transistor output) for the input signals. If contacts (e.g., relay) are used, chattering may cause the trigger to be input again during execution of a measurement.
2
Installation and Connections

Power Supply Specifications When a Switching Regulator Is Connected

Use a power supply that meets the following specifications. (The power supply is sold separately.)
Item Description
Power supply voltage 24 VDC (21.6 to 26.4 V)
Output current 3.75 A min. Recommended Power Supply S8VS-09024@ (24 VDC, 3.75 A)
External power supply terminal screws M4 (tightening torque: 1.2 N·m)
Supply power from a DC power supply for which measures have been applied to prevent high voltages (e.g., a safety extra low voltage circuit).
If UL certification is required for the overall system, use a UL Class II DC power supply.

Attaching the LED Warning Label

Attach the enclosed LED warning label to the cable or other location. The LED warning label must be attached to a location that is readily visible from the Sensor.
FQ User’s Manual
10-5 LED Safety p. 188
Wiring
27
Page 30
Wiring the Touch Finder
1
2
ON
OFF
Important
Power Supply Wiring
Connecting the Power Supply
1 Loosen the two terminal screws using a Phillips
screwdriver.
2 Attach crimp terminals to the power lines.
Secure the positive and negative lines as indicated using M3 screws.
Power supply tightening torque: 0.54 Nm
3 In environments where there is excessive noise, at-
tach a ferrite core (ZCAT1730-0730 from TDK or the equivalent) to the power supply cable.
Turning ON the Touch Finder
24 VDC
When you attach the ferrite core to the power supply cable, wrap the cable only one time.
+
28
1 Remove the cover from the power switch on the left side of
the Touch Finder.
2 Press the switch toward ON.
Power Supply Specifications
Use a power supply that meets the following specifications. (The power supply is sold separately.)
Item Description
Power supply voltage 24 VDC (21.6 to 26.4 V)
Output current 0.65 A min. Recommended Power Supply S8VS-01524@ (24 VDC, 0.65 A)
External power supply terminal screws M4 (tightening torque: 1.2 Nm)
Recommended power line wire size AWG16 to AWG22 (length of 5 m max.)
• Supply power from a DC power supply for which measures have been applied to prevent high voltages (e.g., a safety extra low voltage circuit). If UL certification is required for the overall system, use a UL Class II DC power supply.
• When using the FQ-D31, do not connect a switching regulator and AC Adapter (FQ-AC@) at the same time.
Wiring
FQ User’s Manual
Page 31
Charging the Battery
Important
This section describes how to charge and install the FQ-D31 Battery and provides applicable precautions.
Charge the Battery while it is attached to the Touch Finder.
Use the AC adapter to charge the battery.
Mounting the Battery in the Touch Finder
1 Remove the screw from the battery cover on the top of the
Touch Finder, slide the cover in the direction of the arrow, and open the battery cover.
2 Face the rounded side of the battery toward the back of the
Touch Finder and insert the battery.
Do not insert the battery in the wrong orientation.
3 Close the battery cover, slide the battery cover in the direc-
tion of the arrow, and tighten the screw on the battery cov­er.
2
Installation and Connections
4 Attach the AC adapter to the Touch Finder to start changing
the battery.
The CHARGE indicator will be lit while the battery is being charged. It will go out when charging the battery has been com­pleted.
Note
The Touch Finder will operate even if the AC adapter is connected when no battery is mounted in the Touch Finder.
FQ User’s Manual
CHARGE indicator
Wiring
29
Page 32
• If the Touch Finder (FQ-D31) will be installed permanently or semi-permanently, remove the Battery (FQ-BAT1). If
Important
EU
Li-ion00
Japan Taiwan
the rated temperature is exceeded with the Battery inserted, the protective circuit may activate and stop the Touch Finder.
• The battery complies with the following recycling regulation.
• California regulations concerning perchlorate: This product is a lithium battery that contains perchlorate, which is regulated by the State of California. Please com­ply with these regulations. For details see the following URL: www.dtsc.ca.gov/hazardouswaste/perchlorate/
30
Wiring
FQ User’s Manual
Page 33

2-5 Setting Up Ethernet

Connecting to Sensors from the Touch Finder

When the Sensor is used with a Touch Finder, IP addresses are automatically assigned. No settings are required to use Ethernet. However, if a Sensor or Touch Finder is connected to a network where a PLC or computer is already connected, the Ethernet must be set to be compatible with the existing network.
• Sensor
(Setup Mode)  [Sensor settings]  [Network]  [Ethernet]
1 Set [Auto connection] to [OFF]. 2 Set the IP address and subnet mask according to the network settings.
• Touch Finder
(Setup Mode)  [TF settings]  [Ethernet]
1 Set the IP address and subnet mask according to the network settings.

Connecting to Sensors from External Devices Such as PLCs

2
Installation and Connections
Set the IP address of the Sensor according to the network where the external devices, such as PLCs, are connected.
(Setup Mode)  [Sensor settings]  [Network]  [Ethernet]
1 Set [Auto] to [OFF]. 2 Set the IP address and subnet mask according to the network where the external devices, such
as PLCs, are connected.
Note
If you connect OMRON CS/CJ-series PLCs to the Ethernet, the following default IP addresses are assigned to the
PLCs.
• IP address: 192.168.250.node_address

Connecting to Sensors from a Computer Using the PC Tool

When connecting the Sensor directly to a computer using an Ethernet Cable, set the network settings on the computer as given below. Setting a fixed IP address is not required if there is a hub between the computer and Sensor and a DHCP server is used. The following procedure is for Windows XP.
1 Select [Control Panel] from the Windows Start Menu. 2 Click [Network and Internet Connections] in the control
panel and then double-click [Network Connections].
FQ User’s Manual
Setting Up Ethernet
31
Page 34
3 Right-click the [Local Area Connection] Icon and select
[Properties].
4 On the [General] Tab Page, double-click Internet Proto-
col (TCP/IC).
5 Select the Use the following IP address Option and en-
ter the following IP address and subnet mask.
• IP address: 10.5.5.101
• Subnet mask: 255.255.255.0
6 Click the [OK] Button. This completes the settings.
32
Setting Up Ethernet
FQ User’s Manual
Page 35

Taking Images

3-1 Selecting a Sensor for Configuration. . . . . . . . . . . . . . . . . . . . . . . . . . 34
3-2 Adjusting Image Quality . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
3-3 Adjusting the Object Position . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
3-4 Preventing Mutual Interference of Multiple Sensors . . . . . . . . . . . . . 44
3
Taking Images
Page 36

3-1 Selecting a Sensor for Configuration

If multiple Sensors are connected to a single Touch Finder or computer, you can select the Sensor that you want to set up.
1 Press [Run].
This will enable setting the current Sensor into RUN Mode before selecting another Sensor.
2 Then press [Switch to Run mode].
3 Press [Yes].
4 Press [Switch Sensor]. 5 Press the image of the Sensor to be set up.
will be displayed for Sensors that are not yet set.
Note
Once the Touch Finder detects and records a Sensor, the display or­der for showing more than one Sensor is fixed. Even if the system configuration is changed to reduce the number of Sensors, the pre­vious display location will remain for Sensors that were removed. To update displays of multiple Sensors to the current connection sta-
tus, press [ ] - [Auto connect] on the right of the display in step 5, above, to automatically reconnect.
6 Press  [Sensor settings] to return to Setup
Mode.
7 Press [Yes].
34
Selecting a Sensor for Configuration
FQ User’s Manual
Page 37

3-2 Adjusting Image Quality

(The field of view will widen.)
Turn the screw counterclockwise to focus on objects at a distance.
(The field of view will narrow.)
Turn clockwise to focus on closer objects.
Focus adjustment screw
Important

Adjusting the Focus

[Image] [Camera setup]
1 Display the Camera Setup Display.
The focus can be seen as a numerical value. The higher the value, the better the focus.
Focus Level
2 Manually adjust the focus using the focus adjust-
ment screw on the Sensor while checking the image and focus value on the Touch Finder.
In the default settings, the field of view is set to the nar­rowest setting.
3 Press [Back].
3
Taking Images
• Turn the focus adjustment screw clockwise or counterclockwise a little bit to make sure that it has not already reached the dead stop. Do not force the screw if it does not rotate anymore. This will damage the Sensor.
• Do not turn the focus adjustment knob with a force that is greater than 0.1 N·m. This may damage it.
FQ User’s Manual
Adjusting Image Quality
35
Page 38

Adjusting the Brightness

100
0
Brightness of image
Bright
Dark
Important
OFF
ON
TRIG signal (Brightness
Correction ON)
OFF
ON
BUSY signal
Timing Chart When the Brightness Correction Mode Is ON
ON for 1 ms min.
Taking images begins 25 ms after the TRIG signal is input.
ON while the measurement process is being executed.
To achieve stable measurements, the brightness of the camera image must be adjusted so that the characteristic to be measured is clearly visible.
[Image] [Camera setup]
1 Press [ ]  [Brightness] on the right side of the dis-
play.
2 Move the bar to the left or right to adjust the bright-
ness.
Moving it to the right will make the image brighter, while moving it to the left will reduce the brightness of the im­age.
Press [AUTO] to automatically adjust the brightness ac­cording to the image.
3 Press [OK].
• The exposure time will be longer for larger values. This may cause the image to blur if the object is moving fast. If the Sensor is used on a high-speed line, check that the images are not blurred under actual operating conditions.
• If the brightness changes inconsistently with each image, turn ON the Brightness Correction Mode.
[Image] [] [Brightness correction]
When the Brightness Correction Mode is ON, the brightness will be consistent but the timing when images are taken will be delayed by 25 ms. Make sure that appropriate images of the measured objects are taken when the Brightness Correction Mode is ON.
Reference Timing Chart When the Brightness Correction Mode Is OFF: p. 113
36
Adjusting Image Quality
FQ User’s Manual
Page 39

Taking Clear Images of Moving Objects

1000
1/30,000 1/250
Moving speed
Slow
Fast
Exposure time
Important
Inputting Images with a Limit Range of Brightness
Combining Images to Create an Image with a Wide Dynamic Range
Dark
Bright
For quick moving objects, the effect of blurring can be reduced by decreasing the exposure time.
[Image] – [Camera setup] – [ ] – [Brightness]
Adjusting the Brightness: p. 36
The image becomes darker the smaller the exposure time. If the Sensor is used in a dark environment, make sure that the darkness of the image does not cause the measurements to be unstable.

Improving the Image Quality of Metallic and other Shiny Surfaces

When objects with shiny surfaces are being measured, the lighting may be reflected off the surface and affect the image. To remove reflections, one of the following two functions can be used.
Function Description
HDR (High Dynamic Range) If objects have contrasting light and dark areas, the dynamic range can be made wider to
improve the quality of the images.
Polarizing filter Specular reflections can be eliminated from an image by attaching a polarizing filter to the
Sensor.
• Selection Tips
• When the measurement object is not moving Use the HDR function.
• When the measurement object is moving Use a polarizing filter.
HDR Function
The HDR function is used for objects that have a large difference between light and dark areas. For this kind of object, clear images cannot be achieved with the standard brightness setting. The HDR function combines several images of different brightnesses (exposure times) so that the resulting image has a lower degree of contrast and can be measured stably for the desired characteristic.
3
Taking Images
FQ User’s Manual
Adjusting Image Quality
37
Page 40
Observe the following precautions.
Level 1 Level 2 Level 3 Level 4
Dark
Bright
• Use the HDR function only for objects that are not moving to avoid image blurring. Several images are taken with different shutter speeds and combined. If the object moves while the image is being taken, the image will become blurred.
• Images with different brightnesses are combined, so the resulting image will have a lower degree of con­trast.
[Image] [Camera setup]
1 Press [ ] [Brightness] on the right side of the display. 2 Press [ ] [HDR] on the right side of the display.
The best HDR mode will be selected automatically. The enabled range will appear in blue on the brightness adjust­ment bar.
3 Press the [OK] Button.
Note
• If the measurement object is changed after setting the HDR function, press the [AUTO] Button to automatically set the HDR mode again.
• If the automatic selection does not achieve the desired results, press [ ] [HDR] on the right side of the display and manually set the best HDR mode. As shown below, the higher the level, the wider the combined dynamic range will be.
• If the reflections cannot be sufficiently removed using the HDR function, use a polarizing filter as well.
Using a Polarizing Filter
Specular reflections can be eliminated from an image by attaching a FQ-XF1 Polarizing Filter to the Sensor.
Observe the following precautions.
• The image will be darker compared to when no filter is used.
• If the image becomes too dark, adjust the brightness.
p. 36
• Mounting the Filter
1 Hook the filter in the hole at the top of the Sensor.
38
Adjusting Image Quality
FQ User’s Manual
Page 41
2 Using the top section as a pivot point, pull down the
filter so that it attaches to the Sensor.

Adjusting the Colors of the Image (White Balance)

If external lighting is used, the image may appear as having different colors than the actual object. If this is the case, adjust the white balance. If the lighting built in to the Sensor is used, the white balance is already adjusted. No setting is required in this case.
[Image] [Camera setup]
1 Input a picture of white paper or cloth. 2 Press [ ] – [White balance] on the right side of the
display.
3 Press the [Auto] Button. The Sensor will automatical-
ly adjust the colors.
4 Move the bar to the left (light) or right (dark) to fine-
tune the colors.
5 Press [OK].
3
Taking Images
FQ User’s Manual
Adjusting Image Quality
39
Page 42

3-3 Adjusting the Object Position

Position compen­sation applied...
The object is measured properly within the measurement region.
The image is aligned so that the object is returned to the reference position before performing measurements.
Object is out of place: The measurement object is outside the measurement region.
Object
Measurement region
The measurement object is properly within the measurement region.
Reference position (as set by the operator)
Drag a corner to size the rectangle.
Drag the rectangle to move it.
If objects are moving, the position in the image of the characteristic that is to be measured will vary according to the timing of the trigger signal. The FQ-CR1 Sensor offers two different ways to adjust this position variation.
Function Description Reference
Position compensation If the timing of the trigger is accurate, the FQ-CR1 Sensor can correct variations in the
Trigger delay A delay can be applied from when the trigger (the TRIG signal) is input until when the
position of the object for each measurement with the Position Compensation function.
image is input, to synchronize the timing of image input with the speed of the moving objects.

Position Compensation

The FQ-CR1 Sensor performs measurements in measurement regions that are set to a fixed shape and position by the operator. A measurement is not performed properly if the object is not positioned inside this measurement region (e.g., the result will be NG even if the object is OK). In machines in which the position of the object varies for each image, stable measurements would be impossible. To cope this problem, the Position Compensation function aligns the whole image so that the object to be measured is moved exactly to the measurement region.
p. 40
p. 42
[Image] – [Position compensation] – [Mode on/off]
1 Press [ON] for [Mode on/off] 2 Press [Settings]. 3 Press [Teach]. 4 Adjust the image so that the measurement object is
in the center.
5 Press [ ] – [Model region]. 6 Move the rectangle so that the characteristic part for
position compensation is inside it.
7 Press [OK]. 8 Press [Teach].
The characteristic part and reference position for position compensation will be registered.
9 Press [OK] to save the settings.
40
Adjusting the Object Position
FQ User’s Manual
Page 43
Adjusting the Judgement Parameters
Set how similar the image must be to the registered image for position compensation to be applied.
[Image] – [Position compensation] – [Settings] – [Judgement]
1 Set the correlation range for an OK judgement.
• Measurement Data That Can Be Used for External Outputs and Calculations
The following values can be output to external devices or used in calculations as measurement data.
Expression text string Data name Description Data range
JG Judgement This is the judgement result. 2: No judgement (not measured),
X Position X This is the X coordinate of the position
Y Position Y This is the Y coordinate of the position
CR Correlation This value indicates how similar the
SX Reference position X This is the X coordinate of the position
SY Reference position Y This is the Y coordinate of the position
DX Scroll X X coordinate of measured position
DY Scroll Y Y coordinate of measured position
where the model was found.
where the model was found.
image is to the image registered as reference.
where the model was registered.
where the model was registered.
X reference position
Y reference position
0: Judgement is OK,
1: Judgement is NG,13: Teaching not performed error,14: Figure not registered error,15: Out of range error
99999.9999 to 99999.9999
99999.9999 to 99999.9999
0 to 100
99999.9999 to 99999.9999
99999.9999 to 99999.9999
99999.9999 to 99999.9999
99999.9999 to 99999.9999
3
Taking Images
4-6 Calculations and Judgements Using Inspection Item Data: p. 62
FQ User’s Manual
Adjusting the Object Position
41
Page 44
Measures to Stabilize Position Correction
• Low Mark Contrast and Inconsistent Correlations
Areas with high contrast are registered as model registration sections. Or, adjust the brightness to increase the contrast of the mark.
Making dark images brighter: p. 36
• Inconsistent Marks and Inconsistent Correlations
Inconsistent portions can be masked so that are omitted from matching.
Increasing the Speed of Position Compensation
The following method can be used to reduce processing time.
• Reduce the range in which searches are performed for the model.

Adjusting the Image Timing

The internal timing for taking an image can be set to be delayed in relation to the external trigger signal. This can be used to adjust the object position in the image, e.g., if an external trigger sensor is used. If the object position still varies in the image the Position Compensation function must also be used.
Delay time
Delay from when the trigger is input until when Sensor input is started.
[Image] [Trigger setup] [Trigger delay]
1 A TRIG signal is input.
Images are input continuously.
2 Select the image with the measurement object in the
center using [ ] and [ ].
3 Press the image. 4 Press [OK].
42
Adjusting the Object Position
Trigger input Sensor
FQ User’s Manual
Page 45
Note
Directly input the delay time.
Or
Move the bar to the left or right.
The delay time can be set using the adjustment bar or by directly entering a value.
3
Taking Images
FQ User’s Manual
Adjusting the Object Position
43
Page 46

3-4 Preventing Mutual Interference of Multiple Sensors

Sensor 0
Sensor 1
Trigger Sensor
Important
When the same trigger signal is input to multiple Sensors, the lighting from one Sensor may affect the measurements of the other Sensors. This is called mutual interference. This kind of interference can be prevented offsetting the image input timing of each Sensor from when the trigger signal is received. Example: A trigger (i.e., the TRIG signal) is input to Sensor 0 and Sensor 1 at the same time.
TRIG signal
Timing of taking the image
Sensor 0
Sensor 1
Sensor 0 immediately begins image input when the trigger is input. Sensor 1 begins image input after the specified time has passed.
1 Change to the setup for to Sensor 1.
p. 34
2 Press [Image] – [Trigger setup] – [Trigger delay]. 3 Set the trigger input delay time for Sensor 1.
p. 42
No trigger delay
Sensor 1
trigger delay
Timing of taking the image
• The delay time for preventing mutual interference must be longer than the shutter time.
44
When the lighting built into the Sensor is used, the shutter time is 4 ms max. Therefore make the delay at least 4 ms.
Preventing Mutual Interference of Multiple Sensors
FQ User’s Manual
Page 47

Setting Up Inspections

4-1 Inspection Item Selection Guide . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46
4-2 Setup Procedure for Inspection Items . . . . . . . . . . . . . . . . . . . . . . . . 47
4-3 Configuring Inspection Items . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 48
4-4 Reading Bar Codes. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50
4-5 Reading 2D-codes. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56
4-6 Calculations and Judgements Using Inspection Item Data . . . . . . . 62
4
Setting Up Inspections
Page 48

4-1 Inspection Item Selection Guide

OK
12345 12346
NG
OK
12345 12346
NG
The FQ-CR1 Sensor uses inspection items to judge measurement objects. There are five different measurement objects. Select the best inspection items for the characteristics of the measurement object that are being judged.
Inspection Example Inspection
items used
Judging according to bar code Checking product information Bar code p. 50
Judging according to 2D-code Checking product information 2D-code p. 56
Refer­ence
46
Inspection Item Selection Guide
FQ User’s Manual
Page 49

4-2 Setup Procedure for Inspection Items

Step 3
Step 2
Step 1
Setting Judgement Parameters
Teaching
Configuring Inspection Items
Step 5
Step 4
Re-teaching
Setting Detailed Items
The basic steps for setting up inspection items are shown below.
If measurements are unstable
4
Setting Up Inspections
FQ User’s Manual
Setup Procedure for Inspection Items
47
Page 50

4-3 Configuring Inspection Items

Adding New Inspection Items

Press [Inspect] [Inspection].
1
2 Press an unused inspection item number.
3 Press [Add item.] on the menu.
4 Select an inspection item, such as [2D-code].
When registering multiple inspection items, press the inspection item number after 1.--- and set it in the same way.
Note
If more than six inspection items are set, drag the icon at the bottom of the menu up to display the next inspec­tion item numbers.
48
Configuring Inspection Items
FQ User’s Manual
Page 51

Modifying Existing Inspection Items

1 Press the number of the inspection item to be set.
2 Press [Modify] on the menu.

Deleting Inspection Items

4
Setting Up Inspections
1 Press the number of the inspection item to be delet-
ed.
2 Press [Delete] on the menu.
Note
Executing Similar Measurements in Different Places
Copy an inspection item that is already registered: [Copy]. Change the name of an inspection item: [Rename].
FQ User’s Manual
Configuring Inspection Items
49
Page 52

4-4 Reading Bar Codes

Setup Measurement
Measurement region
One barcode is read in a single measurement region. When the measurement region contains multiple bar­codes, the result of the first barcode to be read is out­put.
Draw the measurement region, then execute AUTO setting. The barcode type and some of the decode conditions are automatically set.

Bar Codes

Bar codes are read and verified.You can output the result of reading and/or verifying bar codes to external devices.The following nine code types can be read. JAN/EAN/UPC, Code39, Codabar (NW-7), ITF (Interleaved 2 of 5), Code 93, Code128/GS1-128, GS1 DataBar* (Truncated, Stacked, Omni-directional, Stacked Omni-directional, Limited, Expanded, Expanded Stacked), Pharmacode, GS1-128 Composite Code (CC-A,CC-B,CC-C) * GS1 DataBar is a code in which the name has been changed from the previous RSS.
Note
Important
Reading will end in error if the barcode type is incorrectly set. Set the barcode type before starting reading. When AUTO setting is executed, the captured barcode is read, and the following items can be set:
• Code type Setting Decode Settings: p. 54
• Direction (for Pharma only) Setting Decode Settings: p. 54
• Reverse decode (for Pharma only) Setting Decode Settings: p. 54
50
Reading Bar Codes
FQ User’s Manual
Page 53

Setup Procedure for Bar Code

Step 1 Selecting the Inspection Item
[Inspect] [Inspection]
1 Press an unused inspection item number and press
[Add item.].
2 Press [Bar code].
4-3 Configuring Inspection Items: p. 48
Step 2 Teaching
Teaching means to store the region and the bar code in the region as master data for the measurement.
1 Press [Teach]. 2 Input an image of the bar code to be registered. 3 Register master data.
Master data can be registered automatically or manually. Auto: Master data is registered using actual codes.Input an image of the applicable bar code and press [TEACH] to read the code and register the result as master data.
Manual: Up to 32 characters can be registered in a text string.Control codes can also be registered. Register master data of a text string of up to 32 characters using a software keyboard. Input a text string and press [TEACH].Check the ASCII table for characters that can be used. Double-byte characters, single-byte kana characters or control codes cannot be input. The following characters can be used as wild cards. *: Wild card of a text string that includes 0 digit. ?: Single-character wild card
ASCII code table: p. 52
4
Setting Up Inspections
FQ User’s Manual
Reading Bar Codes
51
Page 54
Note
Upper 4 bits
Lower 4 bits
ASCII code table
The following table shows the ASCII codes that can be used for manual registration of master data and also for registering characters of Limits.
52
If more than one master data exist, register each by repeating the procedure above.
Step 3 Adjusting Judgment Parameters
[Inspect] [Inspection] [Add item.] [Bar code] [Settings] Tab Page
1 Press [Judgment]. 2 Set the range within which the number of characters
and text strings are judged as OK.
Continuous measurements will be performed for the im­ages that are displayed.
3 Press [OK] to enter the value.
Item Paramet er Setting Description
Judgment Parameter Num. of characters 0 to 1024 Set the upper and lower limits for an OK
Characters 32 characters max. (default *) Set a text string for an OK judgment of
Reading Bar Codes
judgment of the number of characters read.
the text string read.
FQ User’s Manual
Page 55

Detail Settings

OMRO N 2 0 0 8
2nd digit
6th digit
[Inspect] [Inspection] [Add item.] [Bar code] [Details] Tab Page
Press [Function settings].
Setting Function Settings
Set the function settings (verification on/off, verified master data, partial verif. on/off, partial verification setup, string output on/off, error string, partial output on/off and output string setup). Specify these settings as necessary.
Setup Item Setting Value Description
Verification on/off OFF (default value) When set to ON, verification with the master data is executed.
ON
Verified master data All master data (default
Partial verif. on/off OFF (default value) The number of digits in the read results to be verified with the master
Partial verification setup 1 to 1024
value)
Master data 0 to 31
ON
(Default First digit: 1, last digit: 1024)
Selects the master data to be used for verification.
data can be limited. When [Partial verification] is set to ON, set the first and last compared digit positions. Up to 32 characters can be set as the number of digits. Example: First compared digit 2, last compared digit 6
4
Setting Up Inspections
The text string used for verification becomes "MRON2".
String output on/off OFF (default value) Sets whether or not to output the reading result via Ethernet.
ON
Error string Max. 20 digits (default value:
Partial output on/off OFF (default value) Sets whether or not to specify a range in the read data to output from
Output string setup 1 to 1024 (default value
NG)
ON
first digit:1 last digit:1)
Sets the text string to be output when a read error occurs.
ASCII code table: p. 52
when [Output characters] is set to ON.
Sets the first and last digit positions to output. If the read text string con­tains non-targeted digits, the error string is output.
FQ User’s Manual
Reading Bar Codes
53
Page 56
Note
Output function of the read results
• The output order
If both the calculation and reading results are being output via Ethernet, the reading result (text string) is output after the calculation result is output. Example: Output of read results together with two expressions Read result: ABC [Expression] - [Data0]: 123 [Expression] - [Data1]: 456
The output order is as follows: 123 (field separator) 456 (record separator) ABC (record separator) CR
• Code conversion
The following character codes need to be converted for output. The texts after conversion are output.
Character code Before conversion After conversion Space &h20 &h8540 CR &h0D &h8541 LF &h0A &h8542 NULL &h00 &h8543 FF &hFF &h8544
Setting Decode Settings
Press [Decode settings]. Be sure to always specify the code type. Specify other items as necessary. If master data is automatically registered, the code type, the composite codes on/off setting, and the read direction are automatically set.
Setup Item Setting Value Description
Code type JAN/EAN/UPC (default
Timeout 1 to 9999 ms (default value:
Composite codes on/off OFF (default value) Sets whether or not to support composite codes.
Check digit on/off OFF Selects whether or not to insert a check digit. This item is enabled only
Direction Horizontal mode (default
Reverse on/off OFF (default value) Sets whether or not to use the reverse mode. This item is enabled only
value)
Code39
Codebar
ITF
Code93
Code128/GS1-128
GS1 DataBar
Pharmacode
9999)
ON
ON (default value)
value)
Vertical mode
ON
Selects the type of barcode to be read.
Sets the timeout time in read processing. A read error occurs if measure­ment does not end after the preset timeout time is exceeded.
This item is enabled only when the following barcode types are selected: Code128/GS1-128, GS1-DataBar
when the following barcode types are selected: Codebar, Code39, ITF (Interleaved 2 of 5). When the other codes are selected, this item is set to ON.
Selects the reading direction. This item is enabled only when Pharma­code is selected as the barcode type.
when Pharmacode is selected as the barcode type.
54
Reading Bar Codes
FQ User’s Manual
Page 57

Unstable Reading Results

Reading Is Unstable Due to Low Contrast
Adjust the brightness to improve the contrast of the bar code.
Adjusting the Brightness: p. 36

Measurement Data That Can Be Used for External Outputs and Calculations

Expression text string
JG Judgment This is the judgment result. 2: No judgment (not measured),
IN Index The verification result (master data No.) is out-
N Num. of characters The number of characters read is output. 0 to 1024
Data name Description Data range
0: Judgment is OK, 1: Judgment is NG, 13: Teaching not performed error, 14: Figure not registered error, 15: Out of range error
put.
2: Verification OFF, or reading error1: Verification is NG,
0 to 31: Master data No.

Measurement Data That Can Be Logged (Bar Code)

The following values can be logged as measurement data.
Measured item Range of value Description
Num. of characters 0 to 1024 The number of characters read is output.

Errors

Errors in Teaching
4
Setting Up Inspections
A teaching error message is displayed If bar codes cannot be read during an automatic registration. The reading may be unstable due to low contrast.Adjust the brightness to improve the contrast of the bar code.
Adjusting the Brightness: p. 36
FQ User’s Manual
Reading Bar Codes
55
Page 58

4-5 Reading 2D-codes

Setup Measurement
Measurement region
One 2D barcode is read in a single measurement region. When the measurement region contains multiple 2D codes, the result of the first 2D code to be read is output.
Draw the measurement region, then execute AUTO setting. The code type and some of the decode conditions are automatically set.

2D-codes

2D-codes are read and verified.You can output the result of reading and/or verifying 2D-codes to external devices.The following eight code types can be read. DataMatrix (EC200), QR Code, MicroQR Code, PDF417, MicroPDF417, Maxi Code, AZtec Code, Codablock
Note
Important
Reading will end in error if the 2D Code type is incorrectly set. Set the code type before starting reading. When AUTO setting is executed, the captured 2D barcode is read, and the following items can be set:
• Code type Setting Decode Settings: p. 60
• Mirror Setting Decode Settings: p. 60
• Code color (for Data Matrix and PDF417 only) Setting Decode Settings: p. 60
• Shape (for Data Matrix only) Setting Decode Settings: p. 60
56
Reading 2D-codes
FQ User’s Manual
Page 59

Setup Procedure for 2D-code

Step 1 Selecting the Inspection Item
[Inspect] [Inspection]
1 Press an unused inspection item number and press
[Add item.].
2 Press [2D-code].
4-3 Configuring Inspection Items: p. 48
Step 2 Teaching
Teaching means to store the region and the 2D-code in the region as master data for the measurement.
1 Press [Teach]. 2 Input an image of the 2D-code to be registered. 3 Register master data.
Master data can be registered automatically or manually. Auto: Master data is registered using actual codes.Input an image of the applicable 2D-code and press [TEACH] to read the code and register the result as master data.
Manual: Up to 32 characters can be registered in a text string.Control codes can also be registered. Register master data of a text string of up to 32 characters using a software keyboard. Input a text string and press [TEACH].Check the ASCII table for characters that can be used.Double-byte characters, single-byte kana characters or control codes cannot be input. The following characters can be used as wild cards. *: Wild card of a text string that includes 0 digit. ?: Single-character wild card
ASCII code table: p. 58
4
Setting Up Inspections
FQ User’s Manual
Reading 2D-codes
57
Page 60
Note
Upper 4 bits
Lower 4 bits
ASCII code table
The following table shows the ASCII codes that can be used for manual registration of master data and also for registering characters of Limits.
58
If more than one master data exist, register each by repeating the procedure above.
Step 3 Adjusting Judgment Parameters
[Inspect] [Inspection] [Add item.] [2D-code] [Settings] Tab Page
1 Press [Judgment]. 2 Set the number of characters and text strings that are
to be judged as OK.
Continuous measurements will be performed for the im­ages that are displayed.
3 Press [OK] to enter the value.
Item Paramet er Setting Description
Judgment Parameter Num. of characters 0 to 1024 Set the upper and lower limits for an OK
Characters 32 characters max. (default *) Set a text string for an OK judgment of
Reading 2D-codes
judgment of the number of characters read.
the text string read.
FQ User’s Manual
Page 61

Detail Settings

OMRO N 2 0 0 8
2nd digit
6th digit
[Inspect] [Inspection] [Add item.] [2D-code] [Details] Tab Page
Press [Function settings].
Setting Function Settings
Set the function settings (verification on/off, verified master data, partial verif. on/off, partial verification setup, string output on/off, error string, partial output on/off and output string setup). Specify these settings as necessary.
Setup Item Setting Value Description
Verification on/off OFF (default value) When set to ON, verification with the master data is executed.
ON
Verified master data All master data (default
Partial verif. on/off OFF (default value) The number of digits in the read results to be verified with the master
Partial verification setup 1 to 1024 (default value
value)
Master data 0 to 31
ON
first digit:1 last digit:1024)
Selects the model (master data) to be used for verification.
data can be limited. When [Partial verification] is set to ON, set the first and last compared digit positions. Up to 32 characters can be set as the number of digits. Example: First compared digit 2, last compared digit 6
4
Setting Up Inspections
The text string used for verification becomes "MRON2".
String output on/off OFF (default value) Sets whether or not to output the reading result via Ethernet.
ON
Error string Max. 20 digits (default value:
Partial output on/off OFF (default value) Sets whether or not to specify a range in the read data to output from
Output string setup 1 to 1024 (default value
NG)
ON
first digit:1 last digit:1024)
Sets the text string to be output when a read error occurs.
ASCII code table: p. 58
when [Output characters] is set to ON.
Sets the first and last digit positions to output. If the read text string con­tains non-targeted digits, the error string is output.
FQ User’s Manual
Reading 2D-codes
59
Page 62
Note
Character code Before conversion After conversion Space &h20 &h8540 CR &h0D &h8541 LF &h0A &h8542 NULL &h00 &h8543 FF &hFF &h8544
Output function of the read results
• The output order
If both the calculation and reading results are being output via Ethernet, the reading result (text string) is output after the calculation result is output. Example: Output of read results together with two expressions Read result: ABC [Expression] - [Data0]: 123 [Expression] - [Data1]: 456
The output order is as follows: 123 (field separator) 456 (record separator) ABC (record separator) CR
• Code conversion
The following character codes need to be converted for output. The texts after conversion are output.
Setting Decode Settings
Press [Decode settings]. Be sure to always specify the code type. Specify other items as necessary. If master data is automatically registered, the code type, the composite component, and the direction (Pharma) are automatically set.
Setup Item Setting Value Description
Code type Data Matrix Selects the type of 2D Code to be read.
QR Code
MicroQR Code
PDF417
MicroPDF417
Maxi Code
AZtec Code
Codablock
Timeout 1 to 9999 ms
Reverse Normal (default value) Sets normal or reverse as the image orientation.
Code color Black Specifies the code color.
Shape (Data Matrix) Square (default value) Specifies the code shape.
(default value: 9999)
Reverse
White
Auto (default value)
Square or Rectangle
Sets the timeout time in read processing. A read error occurs if measure­ment does not end after the preset timeout time is exceeded.
This item is enabled only when the following barcode types are selected: Data Matrix, PDF417
This item is enabled only when the following barcode types are selected: Data Matrix
60
Reading 2D-codes
FQ User’s Manual
Page 63

Unstable Reading Results

Reading Is Unstable Due to Low Contrast
Adjust the brightness to improve the contrast of the bar code.
Adjusting the Brightness: p. 36

Measurement Data That Can Be Used for External Outputs and Calculations

Expression text string
JG Judgment This is the judgment result. 2: No judgment (not measured),
IN Index The verification result (master data No.) is out-
N Num. of characters The number of characters read is output. 0 to 1024
Data name Description Data range
0: Judgment is OK, 1: Judgment is NG, 13: Teaching not performed error, 14: Figure not registered error, 15: Out of range error
put.
2: Verification OFF, or reading error 1: Verification is NG, 0 to 31: Master data No.

Measurement Data That Can Be Logged

The following values can be logged as measurement data.
Measured item Range of value Description
Num. of characters 0 to 1024 The number of characters read is output.

Errors

Errors in Teaching
4
Setting Up Inspections
If 2D-codes cannot be read during an automatic registration, a teaching error message is displayed. The reading may be unstable due to low contrast.Adjust the brightness to improve the contrast of the 2D-code.
Adjusting the Brightness: p. 36
FQ User’s Manual
Reading 2D-codes
61
Page 64

4-6 Calculations and Judgements Using Inspection Item Data

You can set inspection item judgement results and measurement data with the Calculation menu command to use them in basic arithmetic operations and functions. The judgement results of the calculations are reflected in the overall judgement.

Calculation

Use the Calculation menu command to set the calculation expressions and the judgement parameters for the calculation results.
Expression
You can get up to 32 expressions. You can also combine expressions. You can use the following values in calculations.
• Inspection item data (measurement data, reference values, and judgement results)
•Constants
• Other calculation results
Judgement
Upper and lower limit values are used for the judgement of calculation results (D0 to D31). Each calculation judgement result (J0 to J31) is turned ON if the result falls within the upper and lower limits. The OR logic of these results will be the overall judgement result JG.
62
Calculations and Judgements Using Inspection Item Data
FQ User’s Manual
Page 65
Outputting the Calculation Results
Logging p. 96
No-protocol Output p. 129
The overall judgement (JG) of the calculations are reflected in the overall judgement of the inspection item. The calculation results (D0 to 31) can be output as Ethernet outputs or it can be output by using logging.
Inspection items*
Calculation Process
Calculation (Z0)
Expression 0
Expression 31
(I0 to I31)
Data
* The position compensation values
can also be referenced as data.
Ethernet output
The results can be allocated to data 0 to 31 for Ethernet output.
• Data logging
• Reference values
Calculation
results (D0)
Calculation
results (D31)
• Judgement results
• Reference values
Upper and lower limit judgement
Upper and lower limit judgement
SD card/ Computer
The results can be recorded as logging data.
Individual
judgement results
(OR0 to OR31)
Calculation judgement result (J0)
OR logic
Calculation judgement result (J31)
Reflected in the overall judgement result.
Overall judgement result of calcula­tions (JG)
Overall judgement
result (OR)
4
Setting Up Inspections
FQ User’s Manual
Calculations and Judgements Using Inspection Item Data
63
Page 66

Examples for Calculation

Region 0
Region 1
Calculating the Sum of the Num. of Characters Read for Two 2D-codes in the Region
The sum of the num. of characters read is calculated with the expression below using the two 2D-codes of inspection item 0 and inspection item 1.
• Region 0 (Num. of characters of the 2D-code of inspection item 0): (I0.N)
• Region 1 (Num. of characters of the 2D-code of inspection item 1): (I1.N)
• Total sum of the num. of characters of the 2D-codes = I0.N+I1.N

Procedure (Calculation)

1 Press [Inspect] [Calculation].
64
Setting Expressions
1 Press [Expression] on the [Settings] Tab Page.
Calculations and Judgements Using Inspection Item Data
FQ User’s Manual
Page 67
2 Press the expression number that you want to use.
Performing Similar Calculations At Different Locations
• [Rename] The name of the calculation can be changed. (16 characters max.)
• [Copy] Previously registered calculation expressions can be copied.
Note
DIST(I1.X,I1.Y,I2.X,I2.Y)
Inspection item data
Function
3 Press [Modify] on the menu.
4 Set the expression by selecting items from the [Da-
ta], [Const.], and [Math.] Tab Pages.
The expression will be displayed in the space under [Ex­pression settings].
Do not exceed 255 characters in the expression.
Item Description
Const. Press this to input constants or a mathematical operator. The following operators can be used:
+ (addition), (subtraction), (multiplication), and / (division).
Data Press this to use measurement data, reference values, and judgement results of other items.
Math. Press this to use functions.
Expression Notation
Expressions must have the following notation.
I0.X
Text string corresponding to the item: position correction data, inspection item, or calculation settings.
Position correction data: P0 Inspection item: I + inspection item number calculation settings data: Z0
Example: Finding the distance between the centers of gravity of inspection item 1 and inspection item
2 using a function.
4
Setting Up Inspections
FQ User’s Manual
Calculations and Judgements Using Inspection Item Data
65
Page 68
Example: Subtracting 120 from the calculation result of expression 0.
Constant
Mathematical operator
Result of expression 0
Judgement result for inspection item 1
Mathematical operator
Judgement result for inspection item 0
Example: Adding the judgement result of inspection items 0 and 1.

Function List

The following functions can be used in calculations.
Function Description
SIN Finds the sine. The result is a value between 1 and 1.
The angle in the expression is in degrees.
SIN(angle)
COS Finds the cosine. The result is a value between 1 and 1.
The angle in the expression is in degrees.
COS(angle)
ATAN Finds the arctangent of the value (Y component, X component). The result is a radian value between
 and .
ANGL(Y_component,X_component)
Example: Finding the angle between the straight line joining the centers of region 0 and region 1 and
horizontal.
ATA N (R1.Y-R0.Y,R1.X-R0.X)
If the two arguments are both 0, the result is 0 and the judgement is NG.
AND Finds the logical AND.
If one of the arguments is 0, the calculation result is 0. Otherwise it is 1.
AND(argument_1,argument_2)
OR Finds the logical OR.
If both of the arguments are 0, the calculation result is 0. Otherwise it is 1.
OR(argument_1,argument_2)
NOT Applies a logical NOT operation.
If the argument is 0, the calculation result is 1. Otherwise it is 0.
NOT(argument)
ABS Finds the absolute value.
ABS(argument)
MAX Returns the larger of the two arguments.
MAX(argument_1, argument_2)
MIN Returns the smaller of the two arguments.
MAX(argument_1, argument_2)
66
Calculations and Judgements Using Inspection Item Data
FQ User’s Manual
Page 69
Function Description
First point
(Horizontal)
Second point
(R1.X−R0.X)
2
+
(R1.Y−R0.Y)
2
ANGL Finds the angle of the straight line joining two points (the center of gravity and center of the model).
The angle against the horizontal is found. The result is a value between 180 and 180.
ANGL(Y_component,X_component)
Example: Finding the angle of the straight line joining the centers of region 0 and region 1
ANGL(R1.Y-R0.Y,R1.X-R0.X)
If the two arguments are both 0, the result is 0 and the judgement is NG.
MOD Finds the remainder after dividing a non-ordinal number with an ordinal number.
MOD(non-ordinal, ordinal) If any of the arguments are real numbers, the decimals are rounded off before calculating the remain­der. The remainder is the result of dividing integers.
Example: MOD(13,4) ... Result: 1 (remainder when 13 is divided by 4)
MOD(25.68,6.99) ... Result: 5 (remainder when 26 is divided by 7)
SQRT Finds the square root.
If the argument is negative, the result is 0. The judgement will be NG.
SQRT(argument)
DIST • Finds the distance between two points.
DIST(first_position_X, first_position_Y, second_poistion_X, second_position_Y) Example: Finding the distance between the centers of gravity of region 0 and region 1
DIST(R0.X,R0.Y,R1.X,R1.Y)
The following calculation is performed internally.
4
Setting Up Inspections
• Finds the length of a perpendicular line from point (x,y) to line ax + by + c = 0. DIST (X_coordinate_of_point, Y_coordinate_of_point, coefficient_a_of_line, coefficient_b_of_line,
coefficient_c_of_line)
FQ User’s Manual
Calculations and Judgements Using Inspection Item Data
67
Page 70
Setting Judgement Parameters for Expressions
1 Press [Judgement] on the [Settings] Tab Page.
2 Press an expression between 0 to 31 and set the cor-
responding judgement parameters using the slider.
3 Press the [OK] Button.
68
Calculations and Judgements Using Inspection Item Data
FQ User’s Manual
Page 71
Inspection Item Data That Can Be Used in Expressions
Inspection item Data name Expression text
Position compensation Judgement JG 2: No judgement (not measured)
Position X X 99999.9999 to 99999.9999 0
Position Y Y 99999.9999 to 99999.9999 0
Correlation CR 0 to 100 0
Reference position X SX 99999.9999 to 99999.9999 0
Reference position Y SY 99999.9999 to 99999.9999 0
Scroll X DX 99999.9999 to 99999.9999 0
Scroll Y DY 99999.9999 to 99999.9999 0
string
0: Judgement is OK
1: Judgement is NG13: Teaching not performed error14: Figure not registered error15: Out of range error
Bar code Judgement JG 2: No judgement (not measured),
Data range Default
2
2
0: Judgement is OK,
1: Judgement is NG, 13: Teaching not performed error, 14: Figure not registered error, 5: Out of range error
Index IN
2: Verification OFF,
2
or reading error,
1: Verification is NG,
0 to 31: Master data No.
Num. of characters N 0 to 1024 0
2D-code Judgement JG
2: No judgement (not measured),
2
0: Judgement is OK,
1: Judgement is NG, 13: Teaching not performed error, 14: Figure not registered error, 15: Out of range error
Index IN
2: Verification OFF,
2
or reading error
1: Verification is NG,
0 to 31: Master data No.
Num. of characters N 0 to 1024 0
The following values can be specified as calculation data to output them.
Data name Expression text string Description
Judgement JG This is the judgement result. It is the OR logic of the judgement results of all
Data 0 D00 This is the result of expression 0.
Data 1 D01 This is the result of expression 1.
Data 2 D02 This is the result of expression 2.
Data 3 D03 This is the result of expression 3.
Data 4 D04 This is the result of expression 4.
Data 5 D05 This is the result of expression 5.
Data 6 D06 This is the result of expression 6.
Data 7 D07 This is the result of expression 7.
Data 8 D08 This is the result of expression 8.
expressions.
4
Setting Up Inspections
FQ User’s Manual
Calculations and Judgements Using Inspection Item Data
69
Page 72
Data name Expression text string Description
Data 9 D09 This is the result of expression 9.
Data 10 D10 This is the result of expression 10.
Data 11 D11 This is the result of expression 11.
Data 12 D12 This is the result of expression 12.
Data 13 D13 This is the result of expression 13.
Data 14 D14 This is the result of expression 14.
Data 15 D15 This is the result of expression 15.
Data 16 D16 This is the result of expression 16.
Data 17 D17 This is the result of expression 17.
Data 18 D18 This is the result of expression 18.
Data 19 D19 This is the result of expression 19.
Data 20 D20 This is the result of expression 20.
Data 21 D21 This is the result of expression 21.
Data 22 D22 This is the result of expression 22.
Data 23 D23 This is the result of expression 23.
Data 24 D24 This is the result of expression 24.
Data 25 D25 This is the result of expression 25.
Data 26 D26 This is the result of expression 26.
Data 27 D27 This is the result of expression 27.
Data 28 D28 This is the result of expression 28.
Data 29 D29 This is the result of expression 29.
Data 30 D30 This is the result of expression 30.
Data 31 D31 This is the result of expression 31.
Measurement Data That Can Be Logged
The following values can be logged as measurement data.
Paramet er Setting Description
Results 0 to 31 999999999.9999 to 999999999.9999 This is the results of expressions 1 to 31.
70
Calculations and Judgements Using Inspection Item Data
FQ User’s Manual
Page 73

Testing and Saving Settings

5-1 Performing Test Measurements.............................................................72
5-2 Shortening the Measurement Takt Time...............................................74
5-3 Adjusting the Judgement Parameters...................................................75
5-4 Checking a List of All Inspection Item Results ....................................76
5-5 Saving Data to the Sensor......................................................................77
5
Testing and Saving Settings
Page 74

5-1 Performing Test Measurements

Note
After completing the settings in the [Image], [Inspect], and [In/Out] Tab Pages, move to the [Test] Tab Page. The displayed image is measured automatically. This is called a test measurement. A test measurement is used to verify that the settings that have been made will produce stable results and, if necessary, to fine-tune the settings. An overall judgement of all inspection items can be performed. Test measurements can be performed for through images (default) or saved images.

Performing Test Measurements with Samples

[Test] [Continuous test]
1 Press [Graphics+Details]. 2 Input an image of a previously prepared object.
Check the judgement results.
3 When you finish checking the results, press [Back].
The same five types of displays are available for the [Continuous test] on the [Test] Tab Page, i.e., [Graphic], [Graph­ics + Details], [All results/region], [Trend monitor], and [Histogram]. Press the [Back] Button to access the menu to change the display.
Changing the Run Mode display: p. 82

Performing Test Measurements with Saved Images (Re-measuring)

This Sensor can save measured images in the Sensor’s built-in memory or on an SD card. Test measurements can be performed using these saved images. This function is useful for adjusting the judgement parameters when objects are not available.
[Test] [Continuous test] (Any display)
1 Press  .
72
Performing Test Measurements
FQ User’s Manual
Page 75
2 Select [Log] or [File].
S
3 The display switches to the saved image and mea-
surements are taken again.
Saving images: p. 108
Images in the memory: Press Log.
Images on the SD card: Press File.
ensor's built-in
5
Testing and Saving Settings
FQ User’s Manual
Performing Test Measurements
73
Page 76

5-2 Shortening the Measurement Takt Time

Measurement time
Display
Measurement
Image input
File
logging
Measurement time = Measurement takt time
Inputting a trigger
Important

Checking the Measurement Takt Time

The measurement time of this Sensor can be checked from the Setup or Run Mode display.
The measurement time is the time taken from when a trigger is input until when all measurement processes are executed. During the measurement time, this Sensor will not accept the next trigger. This means that the measurement time is the basic measurement takt time.
With the partial input function, it is possible to input only images that are in the region that is necessary for measurements. The image measurement region becomes smaller and thus the image input time is shortened.
If you use partial input, perform teaching again.
74
Shortening the Measurement Takt Time

Increasing Image Input Speed

[Image] [Camera setup]
1 Press [ ] – [Partial input] on the right side of the dis-
play.
2 Change the input size. 3 Press [OK]. 4 Press [Back].
FQ User’s Manual
Page 77

5-3 Adjusting the Judgement Parameters

Adjusting Judgement Parameters While Looking at Measurement Results

If correct judgements are not possible, you can move directly from the Setup Mode display to the judgement parameters display to make adjustments.
[Test] – [Continuous test] – (Either display)
1 Press [ ]  [Adjust judgement] on the right of the
display.
2 Adjust the judgement parameters.
Lower limit Upper limit
5
Testing and Saving Settings
FQ User’s Manual
Adjusting the Judgement Parameters
75
Page 78
5-4
Note
Individual judgement results for all inspection items can be checked in a list. The individual inspection items can be selected to change the judgement parameters.

Checking a List of All Inspection Item Results

[Test] [Continuous test]
1 Press [All results/region] to display the list.
Judgement parameters can also be changed from this display. Select an inspection item and press [ ] – [Adjust judgement].
76
Checking a List of All Inspection Item Results
FQ User’s Manual
Page 79

5-5 Saving Data to the Sensor

Important
Note
Until you have saved your settings explicitly to the memory in the FQ-CR1 Sensor, the settings are only stored temporarily. They will be lost if the power is turned OFF. Execute [Save data] after you have finished making your settings. The FQ-CR1 Sensor will remind you to do so with a message if you switch from Setup Mode to Run Mode. You can use this feature to keep the previous settings and discard the new settings if desired, but keep in mind that all settings that are not saved explicitly are replaced by the settings that are stored in the memory of the FQ-CR1 Sensor the next time you turn ON the FQ-CR1 Sensor.
Do not turn the power supply OFF while data is being saved. The data that is being saved may become corrupted.
[Test]
1 Press [Save data]. 2 Press [Yes].
5
• Scene data and system data can be saved in this way.
Scene data and system data details: p. 103
• Measurement data and image data cannot be saved in this way.
Logging measurement data: p. 96
• Settings data can also be backed up to an external memory.
Saving settings: p. 103
Testing and Saving Settings
FQ User’s Manual
Saving Data to the Sensor
77
Page 80
MEMO
78
Saving Data to the Sensor
FQ User’s Manual
Page 81

Operation

6-1 Starting Operation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 80
6-2 Configuring the Run Mode Display . . . . . . . . . . . . . . . . . . . . . . . . . . . 82
6-3 Checking the Trend of Measurement Results with Graphs. . . . . . . . 84
6-4 Adjusting Judgement Parameters during Operation. . . . . . . . . . . . . 87
6
Operation
Page 82

6-1 Starting Operation

The present display name.
Overall judgement
Measurement processing time The time taken from when an
image is input until all measurements have been completed.
Inspection results can be displayed in six formats, such as in a list or as a trend graph.
These buttons are displayed both in Setup and Run Mode.
Sub-menu
Tool Button or OK Button
Selected inspection item
Connected Sensor name
Selected scene number
p. 82
p. 14
When test measurements and adjustments in Setup Mode have been finished, the display moves to Run Mode and actual measurements begin. In Run Mode, the Sensor operates stand-alone and outputs the measurement judgement results on the I/O lines accordingly to the settings. If the Touch Finder or the PC Tool is connected via network to the Sensor, the operation of the Sensor can be monitored in the following ways.

Run Mode Display

Moving to Run Mode

You can move from Setup Mode to Run Mode by using the following procedure.
1 Press [Run]. 2 Press [Switch to Run mode.].
3 Press [Yes].
If you press [No], the setting will not be saved and you will move to Run Mode.
80
Starting Operation
FQ User’s Manual
Page 83
Note
ON
OFF
The BUSY signal that was always ON will turn OFF.
BUSY signal
Display
Run Mode
Setup Mode
• Returning to Setup Mode
Press and press [Sensor settings].
• Signal Status When Moving to Run Mode When moving to Run Mode, the signal will change as shown below and data can be input from and output to an
external device.
6
Operation
FQ User’s Manual
Starting Operation
81
Page 84

6-2 Configuring the Run Mode Display

There are six types of displays that can be used, as shown below. Select the display as desired.
Checking the Judgement Results of Inspection Items Checking the Overall Judgement Result History
Graphics
Graphics + Details
Statistical data
The image and region currently being measured will appear.
Checking the Judgements of All Inspection Items in a List
All results/region Trend monitor Histogram
The judgement results of all inspection items can be checked in a list.
In addition to [Graphics] display, indi­vidual judgement results and measure­ment values of selected inspection items will appear.
Displaying Measurement Result Histories
The statistical data for the currently selected inspection item can be checked against time.
The currently measured image and history of the overall judgement results (measurement count, NG count, and NG rate) will appear.
The distribution of measurement results of the currently selected inspection item can be checked.
p. 84
(Run Mode) [Select display]
The following displays are convenient if more than one Sensor is connected.
Multi sensor NG sensor
p. 85
Displays the measurement results of all connected Sensors. Green display: OK, Red display: NG
(Run Mode) [Sensor monitor]
82
Configuring the Run Mode Display
Automatically changes to the display for any Sensor with an NG result.
FQ User’s Manual
Page 85
Specifying the Startup Run Mode Display
Switches to the next inspection item.
Switches to the previous inspection item.
The display that appears when power supply is turned ON can be set. The default setting is [Graphics].
(Setup Mode or Run Mode) [TF settings] [Startup display] [Display pattern]
Note
You can set the scene to be displayed when the power supply is turned ON.
Setting the Startup Scene: p. 91
Displaying the Inspection Item Results
You can scroll though the measurement results of all the configured inspection items by using the following operations.
Note
The following are also displayed in addition to the measurement results for each inspection item.
• Camera input: The image that is being measured is displayed.
• Position comp.: The result of position compensation is displayed.
• All Region: The measurement regions for all inspection items are displayed.
6
Operation
FQ User’s Manual
Configuring the Run Mode Display
83
Page 86

6-3 Checking the Trend of Measurement Results with Graphs

Changes in the measured value of the selected inspection item are displayed against time in a graph.
Measurement values
Switches the display to the previous inspection item number.
Switches the display to the next inspection item number.
Measurement result histories can be checked using the trend monitor and histograms.

Trend Monitor

Changes in the measurement values of the selected inspection item against time can be observed from the graph. It becomes possible to predict when malfunctions may occur or to analyze the cause of the malfunction by checking the trends in the measurement values. The most recent 1,000 measurement values are displayed on the graph.
• [Trend Monitor] Display
(Run Mode) [Select display] [Trend monitor]
Arranging the Trend Monitor Display
The display range for the vertical axis and display conditions for the horizontal axis can be changed. However, the display range and the number of data settings are disabled when a display other than the trend monitor is displayed.
• Disabling Automatic Selection of the Display Range
1 Press [ ] [Auto display] on the right of the trend monitor. 2 Press [OFF].
• Changing the Display Range of the Vertical Axis
1 Press [ ] [Display range] on the right of the trend monitor. 2 Set the minimum and maximum values of the measurement values.
• Changing the Number of Values That Are Displayed
1 Press [ ] [Number of data] on the right of the trend monitor. 2 Select the number of values from 200, 400, and 1,000.
84
Checking the Trend of Measurement Results with Graphs
FQ User’s Manual
Page 87
Note
This is equivalent to the judgement value of the selected inspection item.
The distribution of measurement values is displayed in a histogram.
Measurement values
Number of measurement values
Switches the display to the next inspection item number.
Switches the display to the previous inspection item number.
• Trend monitor data is held until the power supply is turned OFF. You can select whether to display all data on the trend monitor or only data for which the overall judgement is NG.
• Logging settings are applied to the trend monitor as well. However, they are not applied to trend monitor when it is displayed in Setup Mode.
Check recent measurement trends (recent results logging): p. 99

Histograms

The distribution of each measurement value can be checked on a histogram. The most recent 1,000 measurement values are displayed on a graph.
(Run Mode) [Select display] [Histogram]
Arranging Histogram Display
The display range on the horizontal axis and the number of data on the vertical axis of the histogram can be changed.
• Disabling Automatic Adjustment of the Display Range
1 Press [ ] [Auto display] on the right of the histogram. 2 Press [OFF].
• Changing the Display Range of the Horizontal Axis
1 Press [ ] [Display range] on the right of the histogram. 2 Select the maximum measurement value, the minimum measurement value, and the class.
• Changing the Number of Data on the Vertical Axis
1 Press [ ] [Number of data] on the right of the histogram. 2 Select the maximum number of data to display.
6
Operation
FQ User’s Manual
Checking the Trend of Measurement Results with Graphs
85
Page 88
Note
• Histogram data is held until the power supply is turned OFF.
• You can select whether to display all data in the histogram or only data for which the overall judgement is NG. Logging settings are applied to the histogram as well. However, they are not applied to histograms displayed in Setup Mode.
Check recent measurement trends (recent results logging): p. 99
86
Checking the Trend of Measurement Results with Graphs
FQ User’s Manual
Page 89

6-4 Adjusting Judgement Parameters during Operation

Important
This Sensor enables judgement parameters to be adjusted while measurements are being performed. Downtime can be eliminated with this feature because the production line does not have to be stopped while making adjustments.

Preparations

This function is switched OFF as a default to prevent it from inadvertently working during operation. Turn ON the function if you want to use it.
(Setup Mode)  [Sensor settings]  [Adjustment mode in Run]
1 Press [ON].

Changing the Judgement Parameters in Run Mode

This section describes how to change the judgement parameters without stopping measurement in Run Mode.
Run Mode
1 Select the inspection item for which you want to ad-
just the judgement parameters using the and Buttons.
2 Press [ ] – [Adjust judgement].
3 Change the adjustment parameters with the slider. 4 Press [OK].
The judgement results with the changed judgement parameters will appear.
The changed judgement parameters will not be reflected in the measurement result until [OK] is pressed.
6
Operation
FQ User’s Manual
Adjusting Judgement Parameters during Operation
87
Page 90
MEMO
88
Adjusting Judgement Parameters during Operation
FQ User’s Manual
Page 91

Convenient Functions

7-1 Changing the Scene to Change the Line Process ...............................90
7-2 Display Functions ...................................................................................92
7-3 Monitoring the Signal I/O Status............................................................95
7-4 Logging Measurement Data and Image Data .......................................96
7-5 Saving Sensor Settings........................................................................103
7-6 SD Card Operations..............................................................................104
7-7 Convenient Functions for Operation...................................................106
7-8 Convenient Functions for Setup..........................................................108
7-9 Functions Related to the System ........................................................109
7
Convenient Functions
Page 92

7-1 Changing the Scene to Change the Line Process

What Are Scenes?

With an FQ-CR1 Sensor, the inspection items that can be processed at the same time are registered as scenes. A command input from an external device or a touch panel operation can be used to select a certain scene. If a scene is registered for each type of measurement object or inspection, the line process can be changed simply by changing the scene when the measurement object or inspection changes. A maximum of 32 scenes can be created.
Example:
Reading the 2D-code of a container (Inspection item 0: 2D-code)
Reading the 2D-code of a container (Inspection item 1: 2D-code)
Line Process Changes:
Scene 0
Inspection for product A
Scene 1
Inspection for product B
Multiple inspection items in which the 2D-code read conditions have been changed can be created and registered as one scene.
Scene 2
Inspection for product C
Scene 31
Inspection for product Z
• Settings Included with Scenes
The following settings are changed when the scene is changed: Camera image ([Image] Tab Page) and Inspection Items ([Inspect] Tab Page). The settings related to external I/O specifications that are included in the output settings and the system settings for the overall Sensor are used for all of the scenes. Refer to the following information for the data that is included in the scene data.
10-1 Menu Tables: p. 160
90
Changing the Scene to Change the Line Process
FQ User’s Manual
Page 93

Creating New Scenes

The default scene number is 0. To create another scene, use the following procedure to switch the scene and then make the settings.
(Setup Mode)  [Select scene]
1 Press the number of the scene to change to and then
press [Select].
2
The scene will change. Make the settings for the scene.

Changing Scene Names, Copying Scenes, and Deleting Scenes

(Setup Mode)  [Select scene]
1 Press the number of the scene and then press [Re-
name], [Copy], or [Clear].
2
To change the name, enter a new scene name in 15 al­phanumeric characters or less. To copy a scene, press the number of the scene to copy.

Switching Scenes from an External Device

• Controlling with Parallel Inputs
Changing the Scene: p. 123
• Controlling with Ethernet Inputs
Command Details: p. 136

Setting the Startup Scene

(Setup Mode)  [Sensor settings]  [Startup settings]
The following items can be set.
Item Purpose Setting range
Startup mode Select whether the startup scene number is set
manually.
Startup scene Set the scene number to use at startup. 0 to 31
ON OFF (The scene number when the settings were saved will be the startup scene number. The star­tup mode is set to OFF in the default settings.)
7
Convenient Functions
FQ User’s Manual
Changing the Scene to Change the Line Process
91
Page 94

7-2 Display Functions

Reduces the display.
Enlarges the display.
Fits the image to the display size.
The procedures given in this section can be used to make the Sensor easier to use and the display easier to see.

Image Zoom

The display can be zoomed in or out to make the image easier to see.
(Setup Mode or Run Mode)
Press [Back] to end setting the display.

Displaying a Live Image

You can display a live image to check the image that is input by the Sensor in realtime.
(Setup Mode)
1 Press . 2
Press [Camera].
3
Press [Live].
4
Press the [Back] Button to return to the [Display] Dis­play.

Displaying a Frozen Image

You can display a frozen image to stop image refreshing and display the last image that was input.
(Setup Mode)
4
1 Press . 2
Press [Camera].
3
Press [Freeze].
4
Press the [Back] Button to return to the [Display] Dis­play.
92
Display Functions
FQ User’s Manual
Page 95

Displaying a Saved Image

You can display an image that was saved in internal memory in the Sensor or in an SD card. This can be done to configure inspection items or to check measurements using saved images.
(Setup Mode)
1 Press . 2
Images in the Sensor’s built-in memory: Press [Log].
Images on the SD card: Press [File].
3
Press the [Back] Button to return to the [Display] Dis­play.
Note
Refer to the following information for the procedures to save images.
Logging Measurement Data: p. 96

Updating the Display and Measurement Results Only for NG Measurement Results

In Run Mode, you can specify updating the display of the image and measurement results only when the measurement result is NG.
(Run Mode)
1 Press . 2 Press [Last NG image].
3
Press [Back].
Change the following setting to display the last NG image after restarting.
(Setup Mode or Run Mode) - [TF Settings] - [Startup display] - [Display update mode]
1 Press [Last NG image].
7
Convenient Functions
FQ User’s Manual
Display Functions
93
Page 96
Note
If an operation to change the display is performed (e.g., if the display pattern is changed or the inspection item is changed) when displaying images for NG results is set, the display will change to refreshing the most recent mea­surement results and the most recent NG display will disappear. To ensure that you can check the NG results, log the NG results.
Checking Recent Measurement Trends (Recent Results Logging): p. 99

Automatically Changing to the Display for Any Sensor with an NG Result

You can change the settings to automatically display the Sensor for which the measurement result is NG if more than one Sensor is connected.
(Run Mode) [Sensor monitor] [NG Sensor]

Hiding the Menu

You can hide the menu and display only the image on the Touch Finder or PC Tool to check the part of the image hidden behind the menu. If you press the icon again, the menu will be displayed.
(Setup Mode or Run Mode)

Turning ON/OFF the Touch Finder Backlight

You can use Eco Mode to turn OFF the LCD backlight and reduce the power consumed by the Touch Finder whenever there is no operation on the Touch Finder for 30 seconds or longer. The LCD backlight will turn ON whenever any part of the touch panel is pressed.
(Setup Mode or Run Mode) [TF settings] [LCD Backlight] [ECO mode]

Changing the Brightness of the Touch Finder

The brightness of the LCD backlight can be changed to any of five levels.
(Setup Mode or Run Mode) [TF settings] [LCD Backlight] [Brightness]
94
Display Functions
FQ User’s Manual
Page 97

7-3 Monitoring the Signal I/O Status

You can check if the I/O connections are working normally.
[In/Out] [I/O monitor] [I/O Monitor]
1 The I/O status of the external devices will be dis-
played.
2
Press the [OK] Button to return to the [I/O monitor] Display.
Input Signals (TRIG and IN0 to IN5) Signals that are displayed in red are currently being input from the external devices to the Sensor.
Output Signals (OUT0, OUT1, and OUT2) Signals that are displayed in red are currently being output from the Sensor to the external devices. You can turn the signals ON and OFF by pressing them to test the outputs.
FQ User’s Manual
Monitoring the Signal I/O Status
7
Convenient Functions
95
Page 98

7-4 Logging Measurement Data and Image Data

Hard disk drive
Computer*1
Saved for each measurement
SD card
Data logging file
FQ-CR1 Sensor
Image logging file
Touch Finder
There are two ways to log data. Data can be temporarily saved in memory inside the Sensor (called recent results logging) or large amounts of data can be saved in SD cards or other external media (called file logging). The amounts of data that can be logged are given in the following table.
Logged data Recent results logging File logging
Measurement data (Measured val­ues and calculation results)
Image data (Measured values and calculation results)

Logging All Data (File Logging)

Large amounts of measurement and image data can be saved in files in external memory (SD cards or computer).
1,000 measurement values max. Up to the capacity of the external memory
20 images max.
*1: Image data and measurement data can be logged in the same way as for the Touch Finder by installing the PC Tool for FQ.
Only the data for the Sensor that is currently being displayed will be logged even if more than one Sensor is con­nected. If multiple sensors are displayed, or if the most recent NG sensor is displayed, only the results of the sensor that was displayed before changing to the other sensor monitor screen will continue to be logged. Simultaneous logging of the results of multiple sensors is not possible.
You can select the data to be logged.
96
Logging Measurement Data and Image Data
System Configuration: p. 18
Note
Setting the Data To Be Logged
[In/Out] [Log setting]
1 Press the data for which to change the logging pa-
2
rameter.
Change the logging parameter and then press [Back].
FQ User’s Manual
Page 99
Item Description
<Date*1>,<Time*1>,<Measurement value for inspection item 0>,<Measurement value for inspection item 1>,...,<Measurement value for inspection item 31>,
<D0 of expression 0>, ..., <D31 of expression 31>Delimiter
Field separator
Record separator
Image logging • All: All images will be logged regardless of the measurement results.
Data logging • All: All measurement data (Measured values and calculation results) will be logged regardless of the
Note
The logging parameter settings are the same for file logging and recent results logging.
• Only NG: Only images for which the overall judgement was NG will be logged.
• None: No images will be logged (default).
measurement results.
• Only NG: Only measurement data (Measured values and calculation results) for which the overall judgement was NG will be logged.
• None: No measurement data (Measured values and calculation results) will be logged (default).
Starting and Stopping Logging
After logging is started (i.e., set to ON), the specified image data and measurement data will be saved in the SD card or computer hard disk each time measurements are performed.
(Run Mode)
1 Press [Logging]. 2
Press [Image logging] or [Data logging].
3
Press [ON] to start logging.
Press [OFF] to stop logging.
4
Press [Back].
Saved Log Data
• Storage Locations and File Names for Logged File Data
Data Storage location File name
Image logging \sensor_name\
LOGIMAGE\number
Data logging \sensor_name\LOGDATA*2YYYY_MM_DD-HH_MM_SS.CSV
*1, *2
*1: A five-digit number is assigned as a name to the image data storage folder in the order of folder creation as
shown below. Up to 100 images are stored in each folder. 00000
00001
: :
*2: Files are stored in the following folder when the PC Tool is used.
\My Documents\OMRON FQ\ SDCard
• File format Image logging: Image data is saved in a special format for OMRON Vision Sensors. (The file name extension
is IFZ.) Data logging: Measurement data is saved in the following CSV format.
YYYY_MM_DD-HH_MM_SS.IFZ Example: The following name would be used for measurements performed at 10:10:21 pm on March 10, 2010. 2010_03_10-22_10_21.IFZ
Example: The following name would be used for measurements performed at 10:10:21 pm on March 10, 2010. 2010_03_10-22_10_21.CSV
7
Convenient Functions
FQ User’s Manual
Logging Measurement Data and Image Data
97
Page 100
*1: The date is given in the following format: 2010/6/1 (for June 1, 2010). The time is given in the following format: 12:01:20 (for 12:01 and 20 sec-
Data Logging
Display
Measurements
Image input
Sensor
Touch Finder
Trigger signal
Image Logging
BUSY
ON
OFF
onds pm).
Changing the File Format
The output CSV file format can be changed according to the external device.
Item Symbol
Field separator None, comma (default), tab, space, or semicolon
Decimal symbol None, point (default), or comma
Record separator None, comma, tab, space, or CR or CF+LF (default)
Use the following menu command to change the output CSV file format.
(Setup Mode or Run Mode) [TF settings] [File format]
Ensuring That All Measurement Results Are Logged in External Memory
To ensure that all measurement results are actually saved, change the settings so that the BUSY signal remains ON until logging has been completed. During operation, do not input the next trigger until the BUSY signal turns OFF.
[In/Out] [I/O setting] [I/O terminals] [Output] Tab Page [BUSY output]
Change the BUSY output parameter to [Data logging].
Note
• File logging cannot be used when performing continuous measurements.
• If you use the PC Tool, the logging time may vary by up to 100 ms depending on the application conditions of your computer.
• If logging data to an SD card, the write time varies depending on the amount of the available space on the SD card. Reference value: For SDHC class 4, the time required to write image data is approx. 200 to 800 ms.
98
Logging Measurement Data and Image Data
FQ User’s Manual
Loading...