Automatic Upper Arm
Irregular Heartbeat
Normal Heartbeat
Pulse
Blood pressure
Blood pressure
Short
Long
Pulse
: Make sure to press the connection button to pair with the smart
device or transfer the data before taking a measurement when
the SYNC symbol
“” appears.
START INFLATING
DEFLATING
COMPLETED
DO NOT adjust medication based on measurement results from this blood
pressure monitor. Take medication as prescribed by your physician. Only a
physician is qualified to diagnose and treat high blood pressure.
Always consult your physician. Self-diagnosis of measurement results and
self-treatment are dangerous.
Blood Pressure Monitor
EVOLV (HEM-7600T-E)
Instruction Manual
Introduction
Thank you for purchasing the OMRON EVOLV Automatic Upper Arm Blood
Pressure Monitor.
Your new automatic upper arm blood pressure monitor uses the oscillometric
method of blood pressure measurement. This means the monitor detects your
blood movement through your brachial artery and converts the movements into a
digital reading. An oscillometric monitor does not need a stethoscope so the
monitor is simple to use.
Intended Use
This device is a digital monitor intended for use in measuring blood pressure and
pulse rate in adult patient population. The device detects the appearance of
irregular heartbeats during measurement and gives a warning signal with
readings. It is mainly designed for general household use.
Please read the Important Safety Information in this instruction manual
before using the unit.
Please follow this instruction manual thoroughly for your safety.
Please keep for future reference. For specific information about your own blood
pressure, CONSULT YOUR PHYSICIAN.
Important Safety Information
IM-HEM-7600T-E-EN-01-01/2016
9545878-0B
• If battery fluid should get in your eyes, immediately rinse with plenty of clean
water. Consult a physician immediately.
• If battery fluid should get on your skin, wash your skin immediately with plenty of
clean, lukewarm water. If irritation, injury or pain persists, consult a physician.
• Do not use batteries after their expiration date.
• Periodically check your batteries to ensure they are in good working condition.
• Only use batteries specified for this monitor. Use of unsupported batteries may
damage and/or may be hazardous to the monitor.
1. Know Your Device
Contents:
Monitor, battery set, instruction manual, setup instructions, soft case
Monitor:
A. START/STOP button
B. Connection button
A
D
BC
C. Display
D. Arm cuff
E. Battery compartment
F. Systolic blood pressure
G. Diastolic blood pressure
H. Pulse display
2. Insert 4 “AAA” batteries as
indicated, into the battery
compartment.
3. Replace the battery cover.
When the batteries have been securely placed in this device, the message
and SYNC symbol below will appear on the screen:
Notes:
• When the depleted battery symbol “ ” appears on the display, turn the
monitor off and remove all the batteries. Replace with 4 new batteries at
the same time. Long life alkaline batteries are recommended.
• The measurement values continue to be stored in the memory even after
the batteries are replaced.
• The supplied batteries may have a shorter life than new batteries.
• Disposal of used batteries should be carried out in accordance with the
national/local regulations for the disposal of batteries.
3.2 How to Sit Correctly
To take a measurement, you need to be relaxed and
comfortably seated, at a comfortable room
temperature.
• Sit in a chair with your legs uncrossed and your feet flat on
the floor.
• Sit upright with your back straight.
• Sit with your back and arm being supported.
• The arm cuff should be placed on your arm at the same level as your heart.
3.3 Taking a Measurement
Notes:
• To stop the measurement, press the [START/STOP] button once to
deflate the arm cuff.
• Remain still and do not talk while taking a measurement.
• This device will store 100 measurements in the device. Stored
measurements cannot be viewed on the monitor's display. You will need
to use the “OMRON connect” app to view all measurements prior to the
most recent reading taken.
• Do not use the monitor with other electrical devices simultaneously. This
may result in incorrect operation of the monitor.
1. Press the [START/STOP] button.
The arm cuff will start to inflate.
Warning
• DO NOT adjust medication based on measurement results from this blood
pressure monitor. Take medication as prescribed by your physician. Only a
physician is qualified to diagnose and treat high blood pressure.
• DO NOT use this monitor on an injured arm or an arm under medical
treatment.
• DO NOT wear the arm cuff on your arm while on an intravenous drip or blood
transfusion.
• DO NOT apply or inflate the arm cuff around any body parts other than your
upper arm.
• DO NOT use this monitor in areas containing high frequency (HF) surgical
equipment, magnetic resonance imaging (MRI) equipment, computerized
tomography (CT) scanners or in oxygen rich environments.
• DO NOT use this monitor on infants, toddlers, children or persons who cannot
express themselves.
• Consult your physician before using this monitor if you have common
arrhythmias such as atrial or ventricular premature beats or atrial fibrillation;
arterial sclerosis; poor perfusion; diabetes; pregnancy; pre-eclampsia; or renal
disease.
• Always consult your physician. Self-diagnosis of measurement results and
self-treatment are dangerous.
• Contains small parts that may cause a choking hazard if swallowed by children,
toddlers or infants.
• Keep the BATTERY out of the reach of children, toddlers and infants.
• This product emits radio frequencies (RF) in the 2.4 GHz band. DO NOT use
this product in locations where RF is restricted, such as on an aircraft or in
hospitals.
Caution
• Stop using this monitor and consult your physician if you experience skin
irritation or discomfort.
• Consult your physician before using this monitor on an arm with an
arterio-venous (A-V) shunt.
• Note that PATIENT motion, trembling, shivering may affect the measurement
reading.
• Consult your physician before using this monitor if you have had a mastectomy.
• Consult your physician before using the monitor if you have severe blood flow
problems or blood disorders as cuff inflation can cause bruising.
• Do not take measurements more often than necessary because bruising, due
to blood flow interference, may occur.
• Remove the arm cuff if it does not start deflating during the measurement.
• Do not use this monitor for any purpose other than measuring blood pressure.
• Do not use a mobile phone or other electrical devices that emit electromagnetic
fields near this monitor except when transferring your readings to your smart
device. This may result in incorrect operation of this monitor.
• Do not disassemble or attempt to repair the monitor or components. This may
cause an inaccurate reading.
• Do not use in a location where there is moisture or a risk of water splashing the
monitor. This may damage the monitor.
• Do not use this monitor in a moving vehicle such as in a car or on an airplane.
• Do not drop, or subject this monitor to strong shocks or vibrations.
• Do not use this monitor in places with high or low humidity or high or low
temperatures. See the “6. Specifications” section of the manual.
• Please check, by observation of the arm being used, that the monitor is not
causing any impairment of patient blood circulation.
• Do not use this monitor in high-use environments such as medical clinics or
physician offices.
• Avoid bathing, drinking alcohol or caffeine, smoking, exercising and eating for
30 minutes before taking a measurement.
• Rest for at least 5 minutes before taking the measurement.
• Remove tight-fitting or thick clothing from your arm while taking a
measurement.
• Remain still and do not talk while taking a measurement.
• Read and follow the “Important information regarding Electro Magnetic
Compatibility (EMC)” in the “6. Specifications”.
• Read and follow the “Correct Disposal of This Product” in “6. Specifications”
when disposing of the device and any used accessories or optional parts.
Data Transmission
• Do not replace the battery while your reading is being transferred to your smart
device. This may result in the incorrect operation of your monitor and failure to
transfer your blood pressure data.
• Do not place integrated circuit cards, magnets, metal objects, or other devices
that emit electromagnetic fields near this monitor while your reading is being
transferred to your smart device. This may result in the incorrect operation of
your monitor and failure to transfer your blood pressure data.
Battery Handling and Usage
• Do not insert the batteries with their polarities incorrectly aligned.
• Use only 4 “AAA” alkaline batteries with this monitor. Do not use other types of
batteries. Do not use new and used batteries together.
• Remove the batteries if the monitor will not be used for a long period of time.
Indicates a potentially hazardous situation which, if not
avoided, could result in death or serious injury.
Indicates a potentially hazardous situation which, if not
avoided, may result in minor or moderate injury to the
user or patient or damage to the equipment or other
property.
Connection symbol Battery symbol (low)
Connection OFF symbol Battery symbol (depleted)
Connection ON symbol Deflation symbol
Movement error symbol
SYNC symbol Cuff Wrap Guide Symbol
Irregular heartbeat symbol
Heartbeat symbol
(Flashes during measurement.)
1.1 Display symbols
Irregular Heartbeat Symbol ( )
When the monitor detects an irregular rhythm
two or more times during the measurement, the
irregular heartbeat symbol will appear on the
display with the measurement values.
An irregular heartbeat rhythm is defined as a
rhythm that is 25% less or 25% more than the
average rhythm detected while the monitor is
measuring the systolic and diastolic blood
pressure.
If the irregular heartbeat symbol displays with your measurement results,
we recommend you consult your physician. Follow the directions of your
physician.
Movement Error Symbol ( )
The movement error symbol is displayed if you move your body during the
measurement. Please remove the arm cuff, and wait 2 - 3 minutes.
Take another measurement and remain still during measurement.
Cuff Wrap Guide Symbol ( / )
If the cuff was wrapped too loosely, it may cause unreliable results. If the
wrapping of cuff is too loose, the cuff wrap guide symbol “ ” appears.
Otherwise “ ” appears. This is the function which is used as an aid in
determining if the cuff is wrapped snugly enough.
SYNC Symbol ( )
• When the SYNC symbol appears on the display, make sure that your
monitor has been pairing with the smart device or transfer the data.
• When the SYNC symbol appears on the display, transfer the data before
the measurements are deleted. Maximum internal memory is for
100 measurements. The oldest value will be deleted when this is reached
before downloading.
2013 ESH/ESC Guidelines for the management of arterial hypertension
Definitions of hypertension by office and home blood pressure levels
Office Home
Systolic Blood Pressure ≥140 mmHg ≥135 mmHg
Diastolic Blood Pressure ≥90 mmHg ≥85 mmHg
These are from statistical values for blood pressure.
1.2 Before Taking a Measurement
To help ensure an accurate reading, follow these directions:
1. Avoid bathing, drinking alcohol or caffeine, smoking, exercising and
eating for 30 minutes before taking a measurement.
2. Rest for at least 5 minutes before taking the measurement.
3. Stress raises blood pressure. Avoid taking measurements during
stressful times.
4. Measurements should be taken in a quiet place.
5. Remove tight-fitting clothing from your arm.
6. Keep a record of your blood pressure and pulse readings for your
physician. A single measurement does not provide an accurate
indication of your true blood pressure. You need to take and record
several readings over a period of time. Try to measure your blood
pressure at the same time each day for consistency.
2. Preparation
2.1 Battery Installation
1. Remove the battery cover.
Gently slip into the gap of the
hook and pull downward.
2.2 Pairing Your Monitor with a Smart Device
The date and time on your monitor will automatically be set when you pair
it with your smart device.
To begin using the “OMRON connect” app for the first time, please visit
omronconnect.com/setup
Review the list of compatible smart devices at:
omronconnect.com/devices
For details, visit: omronconnect.com
1. Turn on the Bluetooth
for the initial set-up instructions.
®
feature of the smart device.
2. Download and install the free “OMRON connect” app on
your smart device.
* If you already have the “OMRON connect” app, go to:
Menu > Devices > Add Device
3. Open the app on your smart device and follow the set-up
and pairing instructions.
4. Confirm the monitor is connected successfully.
If the monitor is connected successfully to the smart device, “OK” will
appear on the display.
If “Err” appears, refer to “Connection failure. / Data is not being
transmitted.” in section 4.2 for more detail.
Notes:
• If you use any other app instead of “OMRON connect”, the date and time
may not be automatically set.
If you need to set the date and time, pair your monitor or transfer the data
by using “OMRON connect” app.
• The date/time will only be set for future measurements taken after the
measurement that was transferred, it will not be saved to the
measurement that was just transferred.
• Please be aware that OMRON will not be responsible for the loss of data
and/or information.
• If you use the monitor without pairing, refer to the “Using your Monitor
without Pairing” in the “3.3 Taking a Measurement”.
3. Using the Device
3.1 Applying the Arm Cuff
Remove tight-fitting clothing from your left upper arm.
Do not place the arm cuff over thick clothes.
1. Apply the arm cuff to your left upper
arm.
The bottom edge of the arm cuff should be
1 to 2 cm above the elbow. Positioning the
display in clear view of the user makes it
easier to read the results.
1 - 2 cm
2. Securely close with the fabric
fastener.
Notes:
• The blood pressure can differ between
the right arm and the left arm, and the
measured blood pressure values can
be different. Omron recommends to always use the same arm for
measurement. If the values between both arms differ substantially, please
check with your physician as to which arm to use for your measurements.
• Positioning the monitor with the [START/STOP] button towards the body
makes it easier to use the device.
: The arm cuff was wrapped snugly enough.
: The arm cuff was loose or not wrapped correctly.
Note: The cuff wrap guide result appears on the display during the
measurement.
The following symbols indicate incorrect measurement
• If any of these symbols below are displayed, please take another
measurement and try again.
2. Remove the arm cuff.
3. Transfer your readings.
Please make sure your monitor is within 5 m (16 ft.) of your smart device
and your “OMRON connect” app is active.
As soon as the measurement is complete, your reading will be
automatically transferred to the app.
Transferring
1. Open the app on your smart device and then follow the instructions.
2. Press connection button to transfer your readings.
• If your readings are successfully transferred to your smart device,
“OK” will flash on the display.
If “Err” appears, refer to “Connection failure. / Data is not being
transmitted.” in section 4.2 for more detail.
Notes:
• When the SYNC symbol “ ” appears on the display, transfer the data
immediately before the readings are deleted.
• When the Connection OFF symbol “ ” appears on the display, turn on
the Bluetooth feature of your monitor. (Refer to section 3.4)
• Once you send your readings to the app, the data cannot be resent to the
app or smart device.
your readings later
4. After your readings have been transferred to the “OMRON
connect” app, press the [START/STOP] button to turn off.
Notes:
• The monitor will automatically turn off after 2 minutes.
• Wait 2-3 minutes before taking another measurement. Waiting between
measurements allows the arteries to return to their prior condition to
taking the measurement.
• If you use any other app instead of “OMRON connect”, the data may not
be displayed correctly after sending.
Data Transfer With Multiple Smart Devices
Once you transfer your data to a smart device, that data cannot be
transferred again to the same smart device, nor to another smart device.
About Multiple Users
This monitor is recommended for personal use.
When this monitor is used by more than one person, the measurement data
will be combined.
Using your Monitor without Pairing
You can use this blood pressure monitor without pairing to a smart device.
1. Apply the arm cuff. (Refer to section 3.1)
2. Press [START/STOP] button. (Refer to step 1 to 2 in section 3.3)
3.
Write down your SYS/DIA and Pulse readings, along with the date and time
of your measurement on a piece of paper to keep a record of your history.
Note: When using this blood pressure monitor without a smart device, you
will not be able to view your prior readings stored in the memory.
3.4 Turning “OFF” Bluetooth
®
Turn off the Bluetooth® in your monitor in the following areas where use
of wireless equipment is prohibited.
• On aircraft • In hospitals • While abroad
Your blood pressure monitor comes with the Bluetooth
®
already turned
on.
1. Turn your monitor power off if it is turned on.
2. Press and hold the connection button for more than
10 seconds.
3. Release the connection button when “ ” appears on
the display.
4. Press the [START/STOP] button to turn the monitor off.
Note: The monitor will automatically turn off after 2 minutes.
How to turn “ON” Bluetooth
®
Operate the following step when the power has been off.
Press and hold the connection button for more than 2 seconds and
then release the connection button when
“” appears on the
display.
3.5 Restoring Your Monitor to the Default Settings
To delete all the information stored in your monitor, follow the instructions
below. Make sure that your monitor is turned off. It will be off when the
screen is not showing numbers or symbols.
1. While holding the connection button down, the Connection
symbol “ ” and “P” appear on the display, and then press
the [START/STOP] button for more than 5 seconds.
2. When the “Clr” symbol appears on the display, release the
connection button and [START/STOP] button.
The “Clr” will appear on the display when your monitor settings have
reverted back to their original.
3. Press the [START/STOP] button to turn the monitor off.
Notes:
• The monitor will automatically turn off after 2 minutes.
• Reverting to initial settings of this monitor will not delete the information
in the app.
• When using this monitor again, please delete the registration of the
monitor in the app first, and then pair your monitor.
4. Error Messages and Troubleshooting
4.1 Error Messages
Error Display Cause Solution
Remove the arm cuff. Wait
2 - 3 minutes and then take
another measurement.
Repeat the steps in section
3.1. If this error continues to
appear, contact your
physician.
Carefully read and repeat the
steps in section 3.3.
Apply the arm cuff tighter.
Refer to section 3.1.
Recommend replacement of
all 4 batteries with new ones.
Refer to section 2.1.
Immediately replace
4 batteries with new ones.
Refer to section 2.1.
Confirm the batteries are
properly inserted with
polarities correctly aligned.
Refer to section 2.1.
Refer to “Connection failure.
Data is not being
transmitted.” in section 4.2.
Apply the arm cuff tighter.
Refer to section 3.1.
Do not touch the arm cuff
while taking a measurement.
Refer to section 3.3.
Repeat measurement.
Remain still and do not talk
during measurement.
Refer to section 3.3.
Remove any clothing
interfering with the arm cuff.
Refer to section 3.1.
Contact your local OMRON
representative.
* : “XX” will change to
display an error
number, 6 and
higher.
Irregular heartbeat is
detected.
Movement during
measurement.
Arm cuff is applied
too loosely.
The batteries are
low.
The batteries are
depleted or battery
polarities are not
properly aligned.
Connection failure.
Data is not being
transmitted.
Arm cuff is applied
too loosely.
The arm cuff was
inflated exceeding
the maximum
allowable pressure,
and then deflated
automatically.
Movement during
measurement.
Clothing is interfering
with the arm cuff.
Device error.
4.2 Troubleshooting
Problem Cause and Solution
No power.
No display appears on the
monitor.
Replace all batteries with new ones.
Check the battery installation for proper
placement of the battery polarities.
Refer to section 2.1.
Blood pressure varies constantly. Many
Measurement values appear
too high or too low.
factors including stress, time of day, and how
you apply the arm cuff, may affect your blood
pressure.
Review the section, 3.2 and 3.3.
The blood pressure monitor might not be
properly placed within the smart device's
transmission range and is too far from the
smart device. If there are no causes of data
transmission interference found near the
blood pressure monitor, move the blood
pressure monitor within 5 m (16 ft.) of the
smart device and try again.
The Bluetooth feature of the smart device
is turned off. Turn on the Bluetooth feature
of the smart device and your monitor, and try
sending the data again.
The Bluetooth feature of the blood
pressure monitor is turned off.
Refer to “To turn “ON” the Bluetooth ” in
Connection failure. /
Data is not being transmitted.
“3.4 Turning “OFF” the Bluetooth ”.
The blood pressure monitor did not pair
successfully to the smart device. Try to pair
the devices again.
Refer to “Pairing Your Monitor with a Smart
Device” in section 2.2.
For more detail, go to “Help” in the “OMRON
connect” app.
The application on the smart device is not
ready. Check the application then try sending
the data again.
Refer to “Pairing Your Monitor with a Smart
Device” in section 2.2. If the “Err” symbol still
displays after checking the application,
contact your OMRON retail outlet or
distributor. Confirm your smart device
compatibility with this monitor at
omronconnect.com
5. Maintenance and Storage
5.1 Maintenance
To protect your device from damage, please follow the directions below:
• Store the device and the components in a clean, safe location.
• Do not use any abrasive or volatile cleaners.
• Do not wash the device and any components or immerse them in water.
• Do not use gasoline, thinners or similar solvents to clean the device.
• Use a soft dry cloth, or a soft cloth moistened with neutral soap to clean
on the monitor and the arm cuff, and then wipe it with a dry cloth.
• Changes or modification not approved by the manufacturer will void the
user warranty. Do not disassemble or attempt to repair the device or
components.
Calibration and Service
• The accuracy of this blood pressure monitor has been carefully tested
and is designed for a long service life.
• It is generally recommended to have the device inspected every 2 years
to ensure correct functioning and accuracy. Please consult your OMRON
retail outlet or distributor.
5.2 Storage
Do not store the device in the following situations:
• If the device is wet.
• Locations exposed to extreme temperatures, humidity, direct sunlight,
dust or corrosive vapors such as bleach.
• Locations exposed to vibrations, shocks or where it will be at an angle.
6. Specifications
Product category
Product description
Model (code)
Display OLED display
Cuff pressure range Pressure: 0 to 299 mmHg
Measurement range Pressure: 40 to 260 mmHg
Accuracy Pressure: ±3 mmHg
Inflation Fuzzy-logic controlled by electric pump
Deflation Automatic rapid deflation
Measurement method Oscillometric method
Transmission method
Wireless communication Frequency range: 2.4 GHz (2400 - 2483.5 MHz)
IP classification IP 22
Power source
Battery life Approximately 300 measurements
Durable period (Service life) Monitor: 5 years
Operating conditions 10°C to 40°C / 15 to 90% RH (non-condensing) /
Storage / Transport
conditions
Weight Monitor: Approximately 240 g
Dimensions Monitor: Approximately 85 mm × 120 mm × 20 mm
Arm circumference 22 to 42 cm
Contents Monitor, battery set, instruction manual, setup instructions,
Applied part Type B F
Protection against electric
shock
Electronic Sphygmomanometers
Automatic Upper Arm Blood Pressure Monitor
EVOLV (HEM-7600T-E)
Pulse: 40 to 180 beats / min.
Pulse: ±5% of display reading
Bluetooth Smart
Modulation: GFSK
Effective radiated power: <20 dBm
4 “AAA” batteries 1.5V
(using new alkaline batteries)
800 to 1060 hPa
-20°C to 60°C / 10 to 90% RH (non-condensing)
(not including batteries)
(not including the arm cuff)
storage case
Internally powered ME equipment
Notes:
• These specifications are subject to change without notice.
• In the clinical validation study, K5 was used on 85 subjects for determination of
diastolic blood pressure.
• This device is clinically investigated according to the requirements of
ISO81060-2:2013.
• This device has not been validated for use on pregnant patients.
• IP classification is degrees of protection provided by enclosures in accordance
with IEC 60529. This device is protected against solid foreign objects of 12 mm
diameter and greater such as a finger, and against oblique falling water drops
which may cause issues during a normal operation.
• This device fulfils the provisions of EC directive 93/42/EEC (Medical Device
Directive).
• This blood pressure monitor is designed according to the European Standard
EN1060, Non-invasive sphygmomanometers Part 1: General Requirements
and Part 3: Supplementary requirements for electromechanical blood pressure
measuring systems.
• Hereby, OMRON HEALTHCARE Co., Ltd., declares that the radio equipment
type EVOLV is in compliance with Directive 1999/5/EC.
• This OMRON product is produced under the strict quality system of OMRON
HEALTHCARE Co., Ltd., Japan. The Core component for OMRON blood
pressure monitors, which is the Pressure Sensor, is produced in Japan.
Symbol descriptions
Please consult this instruction manual
Applied part - Type BF
Degree of protection against electric shock (leakage
current)
To indicate generally elevated, potentially hazardous,
levels of non-ionizing radiation, or to indicate
equipment or systems e.g. in the medical electrical
area that include RF transmitters or that intentionally
apply RF electromagnetic energy for diagnosis or
treatment.
Temperature limitation
Humidity limitation
Atmospheric pressure limitation
or
Artery mark
Product production date is integrated in a Serial or LOT number, which placed
on the Rating Label and sales package: the first 2 digits mean year of
production, the next 2 digits - month of production.
Important information regarding Electro Magnetic Compatibility (EMC)
With the increased number of electronic devices such as PC’s and mobile
(cellular) telephones, medical devices in use may be susceptible to
electromagnetic interference from other devices. Electromagnetic
interference may result in incorrect operation of the medical device and create
a potentially unsafe situation.
Medical devices should also not interfere with other devices.
In order to regulate the requirements for EMC (Electro Magnetic
Compatibility) with the aim to prevent unsafe product situations, the
EN60601-1-2:2007 standard has been implemented. This standard defines
the levels of immunity to electromagnetic interferences as well as maximum
levels of electromagnetic emissions for medical devices.
This medical device manufactured by OMRON HEALTHCARE conforms to
this EN60601-1-2:2007 standard for both immunity and emissions.
Nevertheless, special precautions need to be observed:
• Do not use mobile (cellular) telephones and other devices, which generate
strong electrical or electromagnetic fields, near the medical device. This may
result in incorrect operation of the device and create a potentially unsafe
situation. Recommendation is to keep a minimum distance of 7 m. Verify
correct operation of the device in case the distance is shorter.
Further documentation in accordance with EN60601-1-2:2007 is available at
OMRON HEALTHCARE EUROPE at the address mentioned in this
instruction manual.
Documentation is also available at omron-healthcare.com
.
About a wireless communication interference
This Product operates in the unlicensed ISM band at 2.4GHz. In case this
Product is used around the other wireless devices including microwave and
wireless LAN, which operate same frequency band of this Product, there is a
possibility that interference occurs between this Product and such other
devices. If such interference occurs, please stop the operation of other
devices or relocate this Product before using this Product or do not use this
Product around the other wireless devices.
Use this monitor only in the member countries of EU or the country you
purchased it in. If you use it elsewhere, you may violate the radio laws or
regulations of that country.
Correct Disposal of This Product
(Waste Electrical & Electronic Equipment)
This marking shown on the product or its literature, indicates
that it should not be disposed of, with other household wastes
at the end of its working life. To prevent possible harm to the
environment or human health from uncontrolled waste
disposal, please separate this product from other types of
wastes and recycle it responsibly to promote the sustainable
reuse of material resources.
Household users should contact either the retailer where they purchased this
product, or their local government office, for details of where and how they
can return this item for environmentally safe recycling.
Business users should contact their supplier and check the terms and
conditions of the purchase contract. This product should not be mixed with
other commercial waste for disposal.
7. Trademarks
The Bluetooth® Smart word mark and logos are registered
trademarks owned by Bluetooth SIG, Inc. and any use of
such marks by OMRON HEALTHCARE Co., Ltd. is under
license. Other trademarks and trade names are those of
their respective owners.
Apple and the Apple logo are trademarks of Apple Inc., registered in the U.S.
and other countries. App Store is a service mark of Apple Inc.
Android, the Google Play logo are trademarks of Google Inc.
8. Limited Warranty
Thank you for buying an OMRON product. This product is constructed of high
quality materials and great care has been taken in its manufacturing. It is
designed to give you every satisfaction, provided that it is properly operated and
maintained as described in the instruction manual.
This product is guaranteed by OMRON for a period of 3 years after the date of
purchase. The proper construction, workmanship and materials of this product is
guaranteed by OMRON. During this period of guarantee OMRON will, without
charge for labour or parts, repair or replace the defect product or any defective
parts.
The guarantee does not cover any of the following:
a. Transport costs and risks of transport.
b. Costs for repairs and / or defects resulting from repairs done by unauthorised
persons.
c. Periodic check-ups and maintenance.
d. Failure or wear of optional parts or other attachments other than the main
device itself, unless explicitly guaranteed above.
e. Costs arising due to non-acceptance of a claim (those will be charged for).
f. Damages of any kind including personal caused accidentally or from misuse.
g. Calibration service is not included within the guarantee.
Should guarantee service be required please apply to the dealer whom the
product was purchased from or an authorised OMRON distributor. For the
address refer to the product packaging / literature or to your specialised retailer.
If you have difficulties in finding OMRON customer services, contact us for
information.
omron-healthcare.com
Repair or replacement under the guarantee does not give rise to any extension
or renewal of the guarantee period.
The guarantee will be granted only if the complete product is returned together
with the original invoice / cash ticket issued to the consumer by the retailer.
9. Some Useful Information about Blood Pressure
What is Blood Pressure?
Blood pressure is a measure of the force of blood flowing against the walls of the
arteries. Arterial blood pressure is constantly changing during the course of the
heart’s cycle.
The highest pressure in the cycle is called the Systolic Blood Pressure; the
lowest is the Diastolic Blood Pressure. Both pressures, the Systolic and
Diastolic, are necessary to enable a physician to evaluate the status of a
patient’s blood pressure.
What is Arrhythmia?
Arrhythmia is a condition where the heartbeat rhythm is abnormal due to flaws in
the bio-electrical system that drives the heartbeat. Typical symptoms are
skipped heartbeats, premature contraction, an abnormally rapid (tachycardia) or
slow (bradycardia) pulse.
Why is it a Good Thing to measure Blood Pressure at
Home?
Many factors such as physical activity,
anxiety, or the time of day, can influence
your blood pressure. A single
measurement may not be sufficient for an
accurate diagnosis.
Thus it is best to try and measure your
blood pressure at the same time each
day, to get an accurate indication of any
200
150
mmHg
100
50
61218 24
Example: fluctuation within a day (male, 35 years old)
changes in blood pressure. Blood pressure is typically low in the morning and
increases from afternoon to evening. It is lower in the summer and higher in the
winter.
How is Hypertension related to Stroke?
Hypertension (high blood pressure) is the key risk factor for Stroke.
It is estimated that amongst hypertensive patients, effective treatment would
prevent 1 in 4 haemorrhagic strokes (bleeding around the brain).
Hypertension guidelines have endorsed the use of Home Blood Pressure
Monitoring in addition to the measurements in physicians’ offices to help manage
hypertension effectively.
References to above medical claims are available upon request.
Manufacturer
EU-representative
Production facility
Subsidiaries
OMRON HEALTHCARE Co., Ltd.
53, Kunotsubo, Terado-cho, Muko, KYOTO,
617-0002 JAPAN
OMRON HEALTHCARE EUROPE B.V.
Scorpius 33, 2132 LR Hoofddorp,
THE NETHERLANDS
www.omron-healthcare.com
OMRON HEALTHCARE Co., Ltd.
Matsusaka Factory
1855-370, Kubo-cho, Matsusaka-shi,
Mie, 515-8503 Japan
OMRON HEALTHCARE UK LTD.
Opal Drive, Fox Milne, Milton Keynes, MK15 0DG, UK
www.omron-healthcare.com
OMRON MEDIZINTECHNIK HANDELSGESELLSCHAFT mbH
Gottlieb-Daimler-Strasse 10, 68165 Mannheim, GERMANY
www.omron-healthcare.com
OMRON SANTÉ FRANCE SAS
14, rue de Lisbonne, 93561 Rosny-sous-Bois Cedex, FRANCE
www.omron-healthcare.com
Upper curve: systolic blood pressure
Lower curve: diastolic blood pressure
Time of day
Made in Japan