Dear Embrace No Code Owner
Thank you for choosing the Embrace No Code Blood Glucose Monitoring
System to help you manage your diabetes. We designed the Embrace
No Code Blood Glucose Monitoring System to be accurate and easy
to use. The added convenience of the bi-lingual talking feature can be
used as a verbal aid for using the meter. This manual contains all of
the information needed to use and maintain your new blood glucose
meter. Please read it carefully before use.
Your Embrace No Code Blood Glucose Monitoring System provides an
easy and precise way to measure blood glucose using fresh capillary
whole blood taken from fingertips, palm, or forearm. The added
convenience of auto-coding can help reduce the chance of inaccurate
readings caused by incorrect coding of the meter. Testing is done
outside the body (in vitro diagnostic use). The test results are plasmacalibrated for easy comparison to lab results. The Embrace No Code
Blood Glucose Monitoring System is a portable battery operated meter
intended for use as an aid by persons with diabetes and by health care
professionals to monitor glucose concentration in whole blood.
If you have questions or simply need more information, please call
Omnis Health at 877-979-5454 or visit our web site at
www.embracemeter.com. Again thank you for choosing the Embrace
No Code Blood Glucose Monitoring System.
NO Code

NO Code
2
Table of Contents
Table of Contents
Understanding Your New Blood Glucose Kit
The Embrace No Code Meter 5
The Embrace No Code Blood Glucose Test Strip 7
Meter Display Screen 8
Setting Up Your New System
The Set Up Screen 10
Setting the Clock 11
Setting the Alarm 14
Talking and Horn Function 16
Talking Function Table 19
Control Solution Testing 24
Performing a Control Solution Test 25
Control Solution Trouble Shooting 28
Testing Your Blood Glucose 29
Preparing Your Lancing Device 30
Performing a Blood Glucose Test 31
Important Information on Alternate Site Testing (AST) 34

3
Understanding Your Blood Glucose Test Results 36
Viewing Stored Readings and 39
7, 14, 30 Days Average
Caring For The Meter
Changing the Battery 41
Cleaning the Meter 43
Storage and Precautions 43
Solving Problems 44
Product Warranty 47
Specifications 47
Table of Contents
NO Code

NO Code
4
Understanding Your New Blood Glucose Kit
Understanding Your New Blood Glucose Kit
The Embrace No Code Blood Glucose Monitoring System
comes as a starter kit or simple kit. Each kit may include
the following items:
CONTENTS KIT CONFIGURATION
Starter Kit Simple Kit
Embrace No Code Meter
Instruction Manual
Carrying Case
Log Book
Lancing Device
With Clear Cap
Test Strips
Lancets
Glucose Control Solution

5
Front of the Embrace No Code Meter
Understanding Your New Blood Glucose Kit
NO Code
Test Strip Port
Insert test strip to
turn on the Meter.
Meter Display
Shows your blood
glucose test results,
memory values,
averages, language
selection, data
port connection,
and alarm times.
(See page 8)
Meter Buttons
Press and hold the middle button located
on the left side of the
meter to turn off and on.
Press and release middle button for
meter option screen.
(See page 10)
Control Solution Test
(see page 23).

NO Code
6
Understanding Your New Blood Glucose Kit
Battery Compartment
Holds two AAA batteries
battery cover.
Data Port
Connect an
Omnis Health
data cable here
to send data
into a PC*
*purchased separate l y
Back of the Embrace No Code Meter
BATTERY COVER TABS
USED FOR REMOVING
BATTERY COVER.

The Embrace Test Strips are glucose specific, biosensor-based
test strips that requires very little blood sample to quickly test
glucose in fresh capillary whole blood. The test results are
plasma-calibrated for an easy comparison to lab results, but
may be 10 to 15% higher then results obtained from wholeblood referenced blood glucose analysis. The test strip is also
equipped with under-fill detection to alert you when there is
not enough blood to perform a test, so you can be assured that
each reading you get is an accurate and meaningful result.
IMPORTANT:
• Immediately close the vial cap of the test strip bottle
tightly after each use. Keep the unused test strips in
the original bottle. DO NOT leave any test strips
outside the bottle while not in use.
• DO NOT reuse test strips. Test strips are for single use only.
• Carefully discard used test strips and lancets in proper
waste containers.
7
The Embrace No Code Blood Glucose Test Strip
Understanding Your New Blood Glucose Kit
NO Code
Contact Points
Insert this end of
the test strip into the
top of the meter.
Sampling End
Dip this end of the
strip into the blood or
control solution drop.

ctl
AM
DAY
AVG
PM
mg /dL
mem
Understanding Your New Blood Glucose Kit
Meter Display Screen
CABLE
Indicates the Meter is in
data transfer mode to PC*.
*Purchased Separately
MEM
Blood glucose readings are
stored (see page 38).
ALARM
To set Alarm or indicates
Alarm has been set (see page
14).
CTL
This icon indicates a control
solution test (see page 24).
To view the control solution
test (see page 38).
BATTERY
Appears when
batteries need to
be changed
(see page 41).
HORN
Beeps when voice is
turned off.
Set sound volume (see
page 16).
Display of the date, time, am/pm of
past and present blooding readings.
As well as the day averages 7, 14, or
30 (see page 38).
{

9
Understanding Your New Blood Glucose Kit
NO Code
TEMPERATURE
Appears when it is either too hot or too cold to test
(outside the ranges of 50ºF~104ºF).
CLOCK
Indicates the Meter is ready to set up the
year, date and time (see page 11).
LANGUAGE SELECTION
Indicates the language selection (see page 17).
UNITS
Unit of measurement for your blood glucose is MG/DL.
BLOOD DROP and ARROW
Indicates the Meter is ready for testing.
TEST STRIP
Flashes to prompt you to insert a test strip for testing.

NO Code
10
Setting Up Your New System
Setting Up Your New System
The Setup Screen
The Setup Screen of the Embrace No Code Meter allows you
take advantage of the advance functions of this Meter.
Simply press and release the middle button for setup screen.
Use the ▲▼buttons to navigate to the desired function icon.
When the icon is flashing press and release middle button to
enter that function. Use the up or down arrows for the desired
setting then press and release the middle button to set . To
exit use the up or down button to navigate to the END of
Function mode press and release the middle button. The
meter will announce each function when
the voice feature is on.
MEM ..............Stored blood and control solution
test (see page 38).
CABLE:...........Press the buttons to enter PC
Link Mode for data transfer to PC*
*Software purchased separately
CLOCK:...........For setting the year, date & time (see page 11).
ALARM CLOCK
..For setting up the Alarms (see page 14). The meter has 3
alarm settings (see page 14).
HORN:............Set sound volume or turn off the voice feature. Setup (see
page 16).

11
LANGUAGE SELECTION:
.Press the buttons to enter Language Selection
(see page17).
End: ..........................Exit Setup Screen and return to testing screen.
Setting the Clock
You will need:
Your Embrace No Code Meter
NOTE:
If the date and time on the Embrace No Code Meter has not
been set, test or control result will not be stored in memory.
Step 1: Press and hold the middle
button to turn on the meter.
Step 2: Press and release middle button for setup screen and
then press and release the up or down buttons until the
clock (3rdicon on the top row)is flashing. When the clock
icon is flashing press and release the middle button to enter
the clock setup function. If the talking feature is on the
meter will say “set date and time”.
Step 3: Set Year. If the talking feature is on, the
meter will say, “Set Year.” Press the up or
down button until the desired year is reached
and then press and release the middle button.
Setting Up Your New System
NO Code

NO Code
12
Setting Up Your New System
Step 4: Set Month: The current month will flash.
If the talking feature is on, the meter will say,
“Set Month.” Press the up or down button to
reach the desired month. Once the desired
month is reach press and release the middle
button.
Step 5: Set Day: The current date will flash.
If the talking feature is on the meter will say
“Set Day”. Press and release the up or down
button to reach the desired day. When the
desired day is reached press and release the
middle button.
Step 6: Set Time Format to 12-hours
or 24-hours: The current time format will flash.
If the talking feature is on, the meter will say
“Set 12 hour/Set 24-hour. When the desired
format is reached press and release the middle
button.(12 hour is standard time/24 hour is
military time).
Step 7: Set Hour: The current hour will flash. If
the talking feature is on, the meter will say, “Set
Hour.” Press the up or bottom button to reach
the correct hour.
(AM or PM will be determine by the HOUR)
By pressing & releasing the up or down buttons

13
and passing the 12 this will change (AM to PM)
or (PM to AM). When the desired hour and AM
or PM is reached press and release the middle
button to set the correct hour and if it’s AM or
PM.
Step 8: Set Minutes: The current minutes will
flash. If the talking feature is on, the Meter
will say, “Set Minutes”. Press and release
the top or bottom button to the desired
minute. Press and release middle button to
set the minutes. If the talking feature is on
the meter will say. The Year, Date, Time and AM or PM.
• Anytime during set up, inserting a Embrace No Code Test
Strip will begin testing. Any changes made so far are
will be stored.
• You may need to reset the Year, Date and Time after
changing the batteries. After date and time have been set,
the meter will announce the year, date and time if the
talking feature is on. Begin testing by inserting an Embrace
Blood Glucose Test Strip (see page 29, Testing Your Blood
Glucose), or press and hold the middle button to turn off
the meter, or see the next section on how to set up the
alarm feature on your meter.
Setting Up Your New System
NO Code

NO Code
14
Setting Up Your New System
Setting the Alarms
You can set up to three alarms on the Embrace No Code
Meter. You must set the clock before you can set the alarms.
When an alarm sounds, the Meter will turn on, and a melody
will sound for 30 seconds. Pressing the middle button or
inserting a test strip will silence the alarm. You can set up
one or all three alarms.
You will need:
Embrace No Code Meter
Step 1: Press and hold the middle
button to turn on the meter.
Step 2: Press and release the middle button to enter the set
up screen. Then press and release the up or down button
until the on the second row starts to flash. If the talking
feature is on, the meter will say, “Set Alarm.” Press and
release the middle button to enter the alarm setting.
Step 3: The display screen will indicate alarm
1 ON or OFF. If the alarm is OFF and the
voice feature is on the meter will say “Set
Alarm 1” Press and release the top button to
turn the alarm on. The meter will now show 1
ON. Press and release the middle button to set alarm 1.
Press and release the up or down button to desired HOUR.

15
(AM or PM) will be determined by setting the
hour. By pressing and releasing the top or
bottom and passing the 12 will change (AM to
PM) or (PM to AM).
Press and release the middle button to set the
hour. The minutes will begin flashing. Press and release the
top or bottom button to desired minute. Press and release
the middle button to set the minute. Repeat step 3 to Set
Alarms 2 and 3.
Step 4: You may set 1 alarm or all 3 alarms. If you choose to
only set one alarm and have completed that desired alarm
press and release the middle button until you have exited
the alarm set up screen.
If you choose to set all 3 alarms and the desired alarm times
have been set press and release the middle button to exit the
alarm set up screen.
NOTE:
• When the alarm or alarms have been set the icons ( and
) should be displayed on the display screen of the meter.
NOTE:
• Alarm will not sound during testing.
• Alarm will follow the time format (12-hour or 24-hour) you
choose when setting the clock.
• Alarm will not be erased when changing the batteries.
Setting Up Your New System
NO Code

NO Code
16
Setting Up Your New System
Begin testing by inserting an Embrace Blood Glucose Test
Strip (see page 29, Testing Your Blood Glucose) or press and
hold the middle button to turn off the meter, or see the next
section on how to turn on or off the talking feature on your
meter.
Talking and Horn Function
Your Meter has a Language Feature Option for English or
Bilingual which can be used as a verbal aid for using the
Meter or hearing test results. The default Language is English
which can be changed to Bilingual if desired. If the speaking
feature is turned off the meter will beep instead of speak.
You will need:
Your Embrace Meter
Setting The Volume
Step 1: Press and hold the middle button to turn on the meter.
Step 2: Press and release the middle button.
Step 3: Press and release the top button until
the starts to flash. Press and release the
middle button to enter volume setting.

17
Press and release the top or bottom button to
turn UP or DOWN the volume. (1 being lowest
volume and 5 being loudest volume) Once the
desired volume is reached press and release the
middle button to set the desired volume.
To turn off the voice press and release the bottom button until
O appears. Press and release the middle button to set VOICE
FEATURE OFF. The meter will now beep instead of speaking.
To turn the voice feature on if turned off follow steps 1-3.
Language Selection
Step 1: Press and hold the middle
button to turn on the meter.
Step 2: Press and release the middle button to enter set up
screen. Press and release the bottom button until the
language icon is flashing. If the voice feature is on the meter
will say “Language Selection”. Press and release the middle
button to enter the Language Selection Option.
Step 3: The default language ENGLISH is
announced (graph). To change the Language
from the default (English) to another
Language. Press and release the top button.
The meter will announce the Language change
and the number 1 (English) will change to the
Setting Up Your New System
NO Code

NO Code
18
Setting Up Your New System
number 2, and will announce that Language. Once the
desired Language has been selected press and release the
middle button to set that Language and exit the Language
Selection Option.
NOTE:
• Anytime during set up you can insert an Embrace No Code
Test Strip and begin testing. Any changes made so far are
stored.
• Your choice of language and volume settings will not be
erased when changing batteries.
Begin testing by inserting an Embrace No Code Blood
Glucose Test Strip (see page 29, Testing Your Blood Glucose)
or press and hold the middle button to turn off the meter.

19
Talking Function Table
NO Code
Prompt user to insert test strip
“Please insert test strip”
Test strip is inserted and
the Meter is ready to test (with
flashing arrow and blood drop)
“Blood glucose test.
Please apply blood
onto the test strip”
Blood Glucose test is
completed and the result is
shown on screen
“Your blood glucose reading
is (number) milligram (mg)
per deciliter”(dL).
Blood Glucose test result is
ABOVE the measurement
range of 600 mg/dL
“Your blood glucose reading is
above 600mg/dL”.
Talking Function Table
The talking function can be used as an aid for you to hear
the test results and use of the meter. The following table lists
WHEN the meter will speak and WHAT the meter will say.
Testing Blood and Control Solutions (see page 24, Control
Solution Testing, and page 29, Testing Your Blood Glucose).
WHEN the Meter will speak... WHAT the Meter will say...
Turned on Greeting Tone
Blood Glucose test result is
BELOW the measurement
range of 20 mg/dL
“Your blood glucose reading is
below 20mg/dL”.

NO Code
20
Talking Function Table
WHEN the Meter will speak... WHAT the Meter will say...
Arrow up
Arrow down
A
r
r
o
w
u
p
A
r
r
o
w
d
o
w
n
A
r
r
o
w
u
p
A
r
r
o
w
d
o
w
n
Performing a Control solution
test.
L1/L2 Control Solution Test
please apply Control Solution
onto the test strips.
Temperature is outside of
testing range( 50ºF~104ºF)
“Temperature is out of range”
When meter is turned on or
when the Test Strip is
inserted into the meter.
When the Meter is turned off.
Melody
Insert Test strip and arrow up or down until CTL
shows next to the test strip icon.
Control solution test is completed
and NOT within range. The result
is shown on screen.
“Your L1/ L2 control solution
test result is (number) milligram
per deciliter, is not within range”

21
Setting Meter Functions via the Setup Screen
(see page 10, “Setting Up Your New System”)
Press and release the middle button to enter the
Meter Set Up Screen.
WHEN the Meter will speak... WHAT the Meter will say...
Talking Function Table
NO Code
MEM
PC LINK
CLOCK (time)
ALARM CLOCK
HORN VOLUME
LANGUAGE SELECTION
END
“Memory Recall”
“Sending Data to PC”
“Set Date and Time”
“Set Testing Alarm”
“Set Sound Volume”
“Language Selection”
“End of Function Mode”

NO Code
22
Talking Function Table
After the year, date and time
has been set up.
The date & time include the Year,
Month, Date, 12-H/24-H, Hour
and Minute
During date and time setup,
setting up the year
“Set Year”
During date and time setup,
setting up the month
“Set Month”
During date and time setup,
setting up the date
“Set Date”
During date and time setup,
setting up the hour
“Set Hour”
WHEN the Meter will speak... WHAT the Meter will say...
WHEN the Meter will speak... WHAT the Meter will say...
During date and time setup,
setting up the Minutes
“Set Minutes”
Setting the Year
Setting the Month
Setting the Day
Setting Hour
“Set Year”
“Set Month”
“Set Day”
“Set Hour”

Meter Failure “Error 1 Meter Failure”
Used Test Strip
“Error 3 Used Test Strip”
Not Enough Blood or Control
Solution
“Error 4 Not Enough Sample”
23
Talking Function Table
NO Code
Meter Can’t Count Down “Error 5 Count Down Error”
Setting 12-H/24-H
Setting Minute
Setting Alarm
When the Set Date and Time
section has been completed
“Set 12th-Hour/Set 24-Hour”
(12-H is Standard Time) (24-H
is Military Time)
“Set Minute”
“Set Alarm (number)”
“The Date and Time Is (Year)
(Month) (Date) (Time) (Minute)
(AM/PM)
OTHER WARNING MESSAGES (see page 44, Solving Problems)
WHEN the Meter will speak... WHAT the Meter will say...

NO Code
24
Control Solution Testing
Control Solution Testing
The purpose of the control solution testing is to make sure the
Embrace No Code Meter and the Test Strip are working
properly. You should perform Control Solution Testing when:
• Using the Meter for the first time
• Using a new bottle of Embrace No Code Blood Glucose
Test Strips
• You left the test strip bottle cap open for a while
• You dropped the Meter
• You suspect your Meter and test strips are not
working properly
• The blood glucose test results do not reflect how you feel
• Practice the testing procedure
IMPORTANT:
• Use only the Embrace Glucose Control Solutions with the
Embrace No Code Test Strips. Other brands of control
solution will produce inaccurate results.
• Always check the expiration date. DO NOT use control
solutions if expired.
• DO NOT FREEZE. Store the control solutions at room
temperature.

25
You will need:
Embrace Control Solution
(Low-L1 or High-L2)
Your Embrace No Code Meter
The Bottle of Embrace No Code Glucose Test Strips
A small clean plate (glass or plastic) or hard surface.
Step 1: Take out the bottle of Control Solution. Check the
expiration date. Do not use the Control Solution if EXPIRED.
Check the Bottle of Control Solution for LOW or HIGH to
determine which Level the Meter will need to be tested on.
(LOW) Meter should be set to L1.
(HIGH) Meter should be set to L2.
Shake the Bottle of Control Solution to ensure liquid is
mixed well.
Place 2 drops of Control Solution on the clean glass/plastic
plate. Replace the cap on the Control Solution Bottle
IMMEDIATELY.
Step 2: Take out an Embrace No Code Test Strip and CLOSE
the LID IMMEDIATELY.
IMPORTANT:
• Check the expiration date printed on the test strip bottle and its
package. Do not use expired test strips.
Control Solution Testing
NO Code

NO Code
26
Control Solution Testing
• Use each test strip immediately after removing it from
the bottle.
• Do not use wet or damaged test strips.
• Keep away from direct sunlight and heat. Store the test strip
bottle in a dry, cool place.
• Record the “date opened” on the bottle label. Discard the bottle
and any remaining test strip after 3 months from date of opening.
• Test within the temperature range of 50ºF to 104ºF. Your Meter
will not work outside these ranges.
Step 3: Inset a test strip to turn on the Meter. If the talking
feature is on the Meter will say “Please apply blood onto the
test strip.” A flashing blood drop and arrow will appear in the
meter display screen.
According to the level of Control Solution press and release
the bottom button for (LOW-L1). Press and release the top
button for (HIGH-L2). L1/L2 will appear on the display
screen.
The CTL Icon will appear next to the Test Strip Icon.
If the talking feature is on the Meter will say “L1/L2 Control
Solution Test please apply Control Solution onto the test
strip”.
Step 4: Gently guide the tip of the strip into the
drop of the Control Solution.
When the strip and meter have received the
Control Solution the meter will begin to play a
melody. The meter will start to count down.

27
After 6 seconds the Control Solution reading
will be displayed on the display screen.
Step 5: Compare the reading on the screen to
the Control Solution Range on the bottle of test
strips.
If the number displayed on the meter is with in
the L1 number range printed on the Test Strip
Bottle, the Meter and Strips are working properly
and are ready for testing. If the reading falls
outside the Level of LOW (L1) range printed on the test strip
bottle (see page 28, “Control Solution Trouble Shooting”).
Follow steps 3 through 5 for the Level 2 Control Solution
test and compare the result with the Level 2 range printed
on the test strip bottle.
NOTE:
• Control Solution testing results will be stored in the Meter’s
memory and indicated by CTL icon. Control Solution testing result
will not be used for calculating averages.
• Remove the test strip from the meter into proper waste container.
In approximately 2 minutes the meter will automatically turn off.
IMPORTANT: DO NOT REUSE TEST STRIPS.
Control Solution Testing
NO Code

NO Code
28
Control Solution Testing
Control Solution Trouble Shooting
If your control solution testing is out of range (too high
or too low), it may be caused by the following:
• Wrong brand of Control
Solution being used
• Control Solution not at room
temperature
• Expired or contaminated
control solution
• Damaged test strip
• Meter malfunction
• Make sure you are using Embrace
Control Solution.
• Make sure the testing environment
is between 50ºF to 104ºF.
• Check the expiration and open bottle
date on both the control solution and
test strips.
(Both the Control Solution and Strips
should be used with in 3 months from
date opened).
Repeat testing using a new test strip.
If the result is still out of range, use a
new bottle of control solution and
retest.
• If problem persists, call Omnis Health
@ 877-979-5454.
Possible Causes What you can do

29
Testing Your Blood Glucose
Materials you will need:
Your Embrace No Code Meter
Your Embrace No Code bottle of Test Strips
Lancing device with a clean, unused lancet
(or safety lancets)
Clear cap for Alternate Site Testing (AST) on palm and
forearm (optional)
Before you begin, make sure:
• Set up your Meter properly and run a control test. See
“Setting Up Your New System” on page 4, and “Control
Solution Testing” on page 23 for details.
• Wash your hands and testing site thoroughly with soap and
warm water, and dry well.
• You are testing in an area between the temperature ranges
of 50ºF~104ºF. Your Meter will not test outside of this
range and will display . If the voice feature is on the
meter will say “temperature out of range”. Move the Meter
into an area that is between 50ºF~104ºF, and let it sit for
10 to 15 minutes before testing again.
Testing Your Blood Glucose
NO Code

NO Code
30
Testing Your Blood Glucose
Preparing Your Lancing Device
Embrace Lancing Device included in Starter Kit only (see page 4).
Step 1: Twist off the adjustable cap
counter clock wise and insert a clean
unused lancet as shown.
Step 2: Twist the protective cover off the
lancet exposing the needle. Twist
clockwise adjustable cover back onto the
lancing device.
Step 3. Adjust the puncture setting on
the adjustable cap for the puncture
depth level (1 is the lightest and 5 is
the deepest).
Step 4: With one hand holding middle
way down the lancing device, pull and
release the opposite end of the inserted
lancet (you should hear a click sound).
The lancing device is now ready for
pricking.
IMPORTANT: Use a new sterile lancet every
time you test to avoid cross contamination. If alcohol wipes are used
to cleanse the fingers, make sure the area is dry before the blood
sample is obtained. We recommend washing with soap & warm
water. Alcohol and Hand Sanitizer may alter the blood test.

31
Performing a Blood Glucose Test
IMPORTANT:
• Check the expiration date printed on the test strip bottle and its
package. Do not use expired test strips.
• Use each test strip immediately after removing it from the bottle.
• Do not use wet or damaged test strips.
• Keep away from direct sunlight and heat. Store the test strip
bottle in a dry, cool place.
• Record the “date opened” on the bottle label. Discard the bottle
and any remaining test strip after 3 months from date of opening.
For Finger Tip Testing:
Step 1: Set up the Lancing Device.
Step 2: Wash hands with soap and warm water. Rinse and dry
hands thoroughly.
Step 3: Take out an Embrace No Code Strip from the test
strip bottle and CLOSE the LID IMMEDIATELY.
Step 4: Insert test strip. A flashing blood drop and arrow icon
will appear on the meters display screen. If the talking feature
is turned on the meter will say,“Please apply blood onto the
test strip”.
Testing Your Blood Glucose
NO Code

NO Code
32
Testing Your Blood Glucose
Step 5: Set the lancing device to the desired puncture depth
(1 being less deep-5 being deepest). Pull BLUE end of
lancing device until you hear a click. Hold the lancing device against the side of your finger and press the small BLUE
button located in the middle of the lancing device to create
the puncture.
Step 6: Gently guide the tip of the strip
straight into the blood drop at a slight
angle. The tip of the strip will act as a
straw and pull the blood up into the strip.
Leave the tip of the strip in the blood drop
until the meter begin beeps or plays a
melody. The meter will start to count
down. After 6 seconds the meter will display your results.
If the talking feature is on the meter will say “Your blood
glucose reading is [number] milligrams per deciliter”.
NOTE:
• Your reading will automatically be stored in the
meter’s memory.
CAUTION: IF YOU SEE “HI” OR “LO” DISPLAYED, YOUR BLOOD
GLUCOSE LEVEL MAY BE ABOVE 600 MG/DL OR BELOW 20 MG/DL,
REPEAT TESTING. IF YOU STILL RECEIVE THE SAME RESULT, CALL
YOUR PHYSICIAN OR HEALTHCARE PROFESSIONAL IMMEDIATELY.

33
Step 7: Remove the used test strip and lancet and discard
according to your local disposal regulations where applicable.
The meter will automatically shut off in 2 minutes.
TIP:
• Gently massage your hand and finger toward the puncture
site to form a blood drop (approximately 0.6µl). Do not
“milk,” or squeeze around the puncture site.
• If the testing area is moist this may result in the strip not
being able to pull enough blood up into the strip.
Make sure the drop of blood is in the shape of a dome or
bubble.
• Lance the side of your fingertip to avoid soreness. To avoid
calluses, choose a different lancing site each time.
Testing Your Blood Glucose
NO Code

NO Code
34
Testing Your Blood Glucose
Important Information on Alternate Site Testing (AST)
The Embrace No Code Blood Glucose
Monitoring System can test for blood
glucose other than your fingertip such as
palm and forearm (alternate site testing,
or AST). Alternate site testing can be less
painful than fingertip testing, but because
of the physiological differences between
your fingertip and palm and forearm
(1)
; alternate site test results
may be significantly different than results from fingertip testing
under certain conditions. You should consult with your doctor or
healthcare professional before using alternate site testing.
DO AST ONLY in the following intervals:
• In a pre-meal or fasting state (more than 2 hours since the last meal)
• Two hours or more after taking insulin
• Two hours or more after exercise
Alternate Site Testing SHOULD NOT be used when:
• You have Hypoglycemic unawareness (not able to tell if you have
low blood sugar)
• Within 2 hours of a meal, exercise, or medication
• You will be operating machinery or driving a car
• You are sick
• You think your blood glucose is low
• Your AST results do not match the way you feel
• You are testing for hyperglycemia
• Your routine glucose results are often fluctuating
• If you are pregnant

35
NOTE:
• Rubbing forearm until warm before lancing will help minimize the
difference with fingertip test results.
The palm (at the base of the thumb) does not require rubbing
before testing.
• If results from alternate site testing do not agree with how you
feel, use fingertip testing instead.
PERFORMING ALTERNATE SITE TESTING (AST):
Step 1: Set up the Lancing Device (see page 30, Preparing
Lancing Device).
Step 2: Wash hands with soap and warm water. Rinse and Dry
hands thoroughly.
Step 3: For Testing on Palm and Forearm:
Hold the lancing device (use the clear
cap) against the test site (base of palm or
forearm). Avoid a test site that has veins,
moles, excessive hair, or bone. Press the
lancing device firmly against the test site
and press the button located in the
middle of the lancing device, do not
remove the device from your arm.
Applying pressure, press down and release
with out braking the suction until blood
drop is approximately 0.6µl in size is
Understanding Your Blood Glucose Test Results
NO Code

NO Code
36
Understanding Your Blood Glucose Test Results
brought to the surface or blood drop is visible.
When blood is present lift the lancing device straight up not
to smear the blood.
Palm testing (at the base of the thumb) does not require the
suction method.
1. Jungheim K., Koschinsky T.: “Risky delay of hypoglycemia detection by glucose monitoring
at the arm.” Diabetes Care 2001:24(7):1303-04.
IMPORTANT:
• Discard used test strips and lancets according to your local
disposal regulations.
Understanding Your Blood Glucose Test Results
Blood glucose value will vary depending on food intake,
medication, health, stress, and exercise. The American
Diabetes Association suggests the following targets for most
non-pregnant adults with diabetes1:
• Between 70–130 mg/dl before meals
• Less than 180 mg/dl two hours after starting a meal
It is important to consult with your physician or healthcare
professional to determine an appropriate target range for you.

37
What to do if you get a high or low reading
If the Meter displays results that are “HI” or “LO,” or you get
a result that is more than 250 mg/dL or below 50 mg/dL
AND you feel ill:
• Treat your diabetes according to the instructions from your
doctor and/or consult your healthcare provider.
If the Meter displays results that are “HI” or “LO,” or you get
a result that is more than 250 mg/dL or below 50 mg/dL
AND you DO NOT feel ill:
• Test your Meter with control solution, refer to page 27,
Control Solution Testing.
• Test again using fingertip or palm (at the base of thumb).
Do not test on forearm.
If you still get a high or low reading, contact your
health care professional.
Reference: 1. American Diabetes Association. Checking Your Blood Glucose,
www.diabetes.org. Accessed 1/24/2012.
Understanding Your Blood Glucose Test Results
NO Code

NO Code
38
Viewing Your Meter’s Memory
Viewing Your Meter's Memory
Your Embrace No Code Meter automatically stores up to 300
test and control results with date and time. The Meter also
provides 7, 14, and 30 days averaging to help track your
blood glucose trend. You must set up your Meter to use the
memory and day averaging function, see “Setting Up Your
New System” on page 10. The Meter will not memorize test
or control results if the date and time are not set.
NOTE:
• The meter will store up to 300 readings. When meter has stored
300 readings, the next reading will replace the oldest reading in
memory (the Memory can not be cleared).

39
Viewing Stored Readings and 7, 14, 30 Days
Average from Memory
You will need:
Your Embrace No Code Meter
Step 1: Press and hold the middle button
to turn on the meter.
Step 2: Press release the middle button to
enter the set up screen. MEM icon should be
flashing. If the MEM icon is not flashing,
press and release the up or down buttons
until the MEM icon is flashing.
If the talking function is on the meter will say
“Memory Recall.”
Step 3: Press and release the middle button to enter the
Memory (the 3rd picture should show “nil”). Press and
release the top button for the daily readings. When the last
reading is reached, press and release bottom button to
decline to most recent reading. When most recent reading is
reached, press and release the bottom button for the
averages (7, 14 & 30 day averages).
Press and release bottom button for the 7 Day Average, again
Viewing Your Meter's Memory
NO Code

NO Code
40
Viewing Your Meter's Memory
for 14 Day Average and again for 30 Day Average.
The number of results included in the averages is shown right
next to the icon.
Press and release the middle button to exit the Memory
Recall Screen.
nil will appear on the display screen after the last recorded
test result if less then 300 readings are stored.
Control solution results are not included in the day average
calculations.
DAY
AVG

41
NOTE:
• Control testing results are flagged by CTL on the bottom left
corner of the display screen as well as Control Solution Test stored
in Memory.
• When the Meter memory is blank, the display screen will show nil
• When there are no day average data available, the display screen
will show 4 dashes
• If the Meter memory is full, adding a new test result will cause the
oldest one to be deleted
• Stored test results and day averages will not be erased when
changing batteries
• You can begin testing anytime by inserting a test strip
Caring for the Meter
Caring for your Embrace No Code Meter is easy. Follow these
simple guidelines to keep your Meter working properly.
Changing the Battery
Batteries need to be changed when the icon appears on
the Meter display.
Materials you will need:
Two Alkaline AAA Batteries
Your Embrace No Code Meter
Viewing Your Meter's Memory/Caring For The Meter
NO Code

NO Code
42
Caring For The Meter
Step 1. Pull back and up at the same
time on the tabs at the top of the
battery cover to remove the battery
cover.
Step 2. To insert the new batteries. The top battery, The
positive (+) end will face to the left. The bottom battery the
positive (+) end will face to the right. Make sure batteries are
firmly in place.
You should here a click when inserting the batteries, this will
ensure the batteries are in firmly.
NOTE:
• When the batteries have been changed. Push and hold the
middle button to turn on the meter.
• The Meter will prompt you to reconfirm the year, date and
time (see page 11 for “Setting the Clock”).
• The year, date, stored results, alarm settings, volume
setting, and language choice stay in memory when the
batteries are changed.

43
Caring For The Meter
NO Code
Cleaning the Meter
• If the Meter gets dirty, use a moist (NOT WET) lint-free
cloth dampened with a mild detergent
• Do not get water inside the Meter or hold it under
running water
• Do not use glass or household cleaners on the Meter
• Do not try to clean the test strip holder
Storage and Precautions
• Handle the Meter with care- severe shock, such as
dropping the Meter, could damage the electronics
• The Meter and test strips are designed to be used
within the temperature ranges between 50ºF~104ºF
• Avoid leaving the Meter in extremely hot or cold places,
such as near a heat source or in an extremely hot or cold
car
• Do not store or use the Meter or test strips in high
humidity levels, such as bathroom or kitchen
• Always close bottle cap immediately after removing a
test strip and make sure it’s closed tightly
• Do not take the meter apart. Doing so will void the
warranty. If there are technical problems or questions,
call Omnis Health at 877-979-5454 for assistance.

NO Code
44
Caring For The Meter
Solving Problems
This section details the significant display screen messages
and error codes you will encounter when using your Embrace
No Code Meter and test strips.
All warranty claims must be directed to Omnis Health, LLC.
The warranty applies only to the original purchaser
of the system.

45
Solving Problems
NO Code
“Error 1 Meter
Failure”
“Error 3 used
test strip”
“Error 4 not
enough
sample”
“Error 5 count
down error”
“Temperature is out
of range”
What What the What it What you
you see Meter says means should do
The Meter
electronics are
damaged
Used or
damaged test
strip
Not enough
sample in the
test strip for
the meter to
read.
Test Strip was
removed
during testing
Temperature out
of range
Remove and put back in the
batteries and turn the Meter
back on.
If problems persists, call us
at 877-979-5454
Repeat the test with a new
test strip. Remember to add
sample only when you see
the prompt on the display
screen or the Meter verbally
prompts you to add sample.
Remove the Test Strip and
repeat the test with a new
test strip. See “Testing Your
Blood Glucose” on page 29.
Remove the Test Strip and
repeat the test with a new
test strip. See “Testing Your
Blood Glucose” on page 29.
Move the Meter into an area
that is between 50ºF~104ºF,
and allow 10 to 15 minutes
for the meter’s temperature
to get back with in range.

NO Code
46
Solving Problems
HI “Your blood
glucose
reading is
above 600
mg/dL”
LO “Your
blood glucose
reading is
below 20
mg/dL”
“Low Battery”
“nil”
What What the What it What you
you see Meter says means should do
Test result is
higher than
600 mg/dL
Test result is
lower than 20
mg/dL
Batteries are
low
There are no
results stored
in the Meter
Wash and dry your hands
and repeat the test on
your fingertip with a new
test strip. If the result is
still “HI,” contact your
physician or healthcare
professional immediately.
Wash and dry your hands
and repeat the test on
your fingertip with a new
test strip. If the result is
still “LO,” contact your
physician or healthcare
professional immediately.
Change the battery
according to page 41,
Inserting (or changing)
the Battery.
Check if the date and
time on you Meter is set
up. See “Setting Up Your
New System” on page 10.
Start testing your blood
glucose, see “Testing Your
Blood Glucose”
on page 29.

47
Warranty/Specifications
NO Code
Product Warranty
Omnis Health, LLC warrants the Embrace Blood Glucose Meter to be free of defects in
workmanship and materials under normal use for a 5 year warranty. The liability of
Omnis Health, LLC is limited to repair or replacement and in no event shall Omnis
Health, LLC be liable for any collateral or consequential damages or loss.Instruments
subjected to misuse, abuse, neglect, unauthorized repair or modification will be
excluded from this warranty. This warranty specifically excludes expendables and
consumables.
All warranty claims must be directed to Omnis Health, LLC.
The warranty applies only to the original purchaser of the system.
Specifications
Test Strips: Embrace No Code Blood Glucose Test Strips
Test Range: 20 - 600 mg/dl
Calibration: Plasma
Test Time: 6 Seconds
Sample Size: 0.6µl
Hematocrit Range: 30-55%
Display Type: Large LCD screen
Memory: 300 blood test results with date and time
Result Averaging: 7, 14, and 30 days averaging
Dimension: 100L x 54W x 20H (mm)
Weight: 66g
Battery: 2 Alkaline AAA Battery
Battery Life: 1000 tests of continuous use or 1 year
Automatic Power-off: After 2 minutes of non-use
Operating Temperature: 50ºF~104ºF
Relative Humidity: Less than 85%
Storage Condition: Meter at 39ºF~122ºF
Test Strips at 39ºF~86ºF
For additional information, refertotheEmbraceNoCode TestStrip insert.