Omega Products TAC80B-K Installation Manual

TAC80B-J, K, T
Thermocouple to Analog Converter
AT
u
oc
o
m
r
e
hT
no r
C
golan
A
-
o
t
-elp etrev
K-
B0
8
C
INSTRUCTION
M1710/0109
SHEET
Shop online at: omega.com e-mail: info@omega.com For latest product manuals: omegamanual.info
GENERAL DESCRIPTION
The OMEGA
®
TAC80B-J, K, T Thermocouple to Analog Converter can turn any chart recorder, analog or digital voltmeter into an accurate, wide range temperature measuring instrument. It is powered by either an internal 9 volt battery or an optional power adapter (TAC80B-AC). The TAC80B is a universal thermocouple amplifier and linearizer which provides a precision 1mV/°C or °F signal for type J, K or T thermocouples. Cold junction compen­sation is built in. Each unit is supplied with mating con­nector, standard-to-miniature connector adaptor, and 9V battery.
UNPACKING
Remove the Packing List and verify that you have received all items. If you have any questions about the shipment, call the Customer Service Department. When you receive the shipment, inspect the container and equipment for any signs of damage. Note any evidence of rough handling in transit. Immediately report any damage to the shipping agent. NOTE: The carrier will not honor any claims unless all shipping material is saved for their examination. After examining and removing contents, save packing material in the event reshipment is necessary.
INSTALLATION
1. Connect the TAC80B to the strip chart recorder or
meter. The H
I plug connects to the HI (+) receptacle
and the LO plug to the LO (-) receptacle.
2. Plug the thermocouple into the SMP socket on the
TAC80B.
OPERATION
CAUTION: When the range switch is in the TEST position
(center), full battery voltage (9V) is applied to the output. The power switch must be in the OFF position before switching from °F to °C or vice-versa (see Figure 1-1).
+ OUTPUT
°F °C
Figure 1-1. °C/°F Switch
®
Multimeter Use
1. For temperatures less than 200°C or 200°F, set the multimeter range to 200 mV.
2. For temperatures greater than 200°C or 200°F, set the multimeter range to 2 V.
Temperature is read directly on the multimeter in °C or °F; the TAC80B converts the mV signal into a temperature measurement displayed in °C.
Strip Chart Recorder Use
To use the TAC80B with a strip chart recorder, set the recorder span to the mV range that corresponds to the mV output at that temperature to be measured. For example, to read between 50°C and 150°C, set the range to 100 mV in the CALIBRAT
E mode and suppress the zero by
50 mV.
NOTE: Output is 1 mV per °C or 1 mV per °F.
SPECIFICATIONS
RANGES: Type J: -4° to 1112°F
Type K
Type T
OPERATING TEMPERATURE: 0° to 50°C
POWER: 9V alkaline INPUT CONNECTION: SMP connector, standard
OUTPUT CONNECTION: Standard banana plug
STORAGE TEMPERATURE: 14° to 122°F
OUTPUT: 1 mV per °C or °F ACCURACY: Type J: ±2.6°C, ±4.3°F
Type K Type T
COLD JUNCTION COMPENSATION: 0.05°C/°C
DIMENSIONS: H: 2.25" (57mm) x
WEIGHT: 6 oz.
(-20
° to 600°C)
: -4° to 1832°F
(-20
° to 1000°C)
: -4 to 572°F
(-20
° to 300°C)
P adaptor supplied
to SM
or jack
° to +50°C)
(-10
: ±3.6°C, ±6.1°F
: ±1.8°C, ±3.0°F
W: 2.4" (71mm) x D: 1" (25mm)
CALIBRATION
Equipment required: 3-1/2 or 4-1/2 digit multimeter with ±0.1% accuracy, stable voltage source, TRC III Ice Point Cell, TRP (J, K, T) Reference Probe.
Set up the equipment as shown below. Set the Function Switch to “°C” or “°F” position. Turn power on. Set voltage source and adjust potentiometers as follows.
Figure 1-2. Potentiometer Locations
TM
Set Voltage Source Adjust for Multimeter Reading
Model TAC80B-J
°C 0.000 mV P1 00.00 mV
33.102 mV P2 600.0 mV
°F 0.000 mV P4 32.00 mV
33.102 mV P3 1112.0 mV
Model TAC80B-K
°C 0.000 mV P1 00.00 mV
37.325 mV P2 900.0 mV
°F 0.000 mV P4 32.00 mV
37.325 mV P3 1652.0 mV
Model TAC80B-T
°C 0.000 mV P1 00.00 mV
14.862 mV P2 300.0 mV
°F 0.000 mV P4 32.0 mV
14.862 mV P3 572.0 mV
THERMOCOUPLE LEADS
COM
DIGITAL MULTIMETER
V
COM
A
Figure 1-3. Calibration Set-up
OMEGAnet®On-Line Service Internet e-ma il
omega.com info@omega.com
U.S.A.: Omega Engineering, Inc., One Omega Drive, P.O. Box 4047
ISO 9001 Certifie
Canada: 976 Bergar
For immediate technical or application assistance:
U.S.A. and Canada:
Mexico: En Español: 001 (203) 359-7803 FAX: (001) 203-359-7807
Benelux: Managed by the United Kingdom Office
Czech Republic: Frystatska 184
France:
Germany/Austria: Daimlerstrasse 26
United Kingdom: OMEGA Engineering Ltd.
ISO 9001 Certifie
It is the policy of OMEGA Engineering, Inc. to comply with all worldwide safety and EMC/EMI regulations that apply. OMEGA is constantly pursuing certification of its products to the European New Approach Directives. OMEGA will add the CE mark to every appropriate device upon certification.
The information contained in this document is believed to be correct, but OMEGA Engineering, Inc. specifications without notice. WARNING: These products are not designed for use in, and should not be used for, human applications.
Servicing North America:
d Stamford, CT 06907-0047
Tel: (203) 359-1660
X: (203) 359-7700
FA e-mail: info@omega.com
Laval (Quebec), Canada H7L Toll-Free: 1-800-826-6342 TEL: (514) 856-6928
X: (514) 856-6886 e-mail: info@omega.ca
FA
Sales Service: 1-800-826-6342/1-800-TC-OMEGA Customer Service: 1-800-622-2378/1-800-622-BEST Engineering Service: 1-800-872-9436/1-800-USA-WHEN
info@omega.com.mx e-mail: espanol@omega.com
5A1
®
Servicing Europe:
Toll-Free: 0800 099 3344 TEL: +31 20 347 21 21
X: +31 20 643 46 43 e-mail: sales@omega.nl
FA
733 01 Karviná, Czech Republic Toll-Free: 0800-1-66342 TEL: +420-59-6311899
X: +420-59-6311114 e-mail: info@omegashop.cz
FA
Managed by the United Kingdom Office Toll-Free: 0800 466 342 TEL: +33 (0) 161 37 29 00
X: +33 (0) 130 57 54 27 e-mail: sales@omega.fr
FA
D-75392 Deckenpfronn, Germany Toll-Free: 0 800 6397678 TEL: +49 (0) 7059 9398-0
X: +49 (0) 7056 9398-29 e-mail: info@omega.de
FA
d One Omega Drive, River Bend Technology Centre, Northbank
Irlam, Manchester M44 5BD England Toll-Free: 0800-488-488 TEL: +44 (0)161 777-6611
X: +44 (0)161 777-6622 e-mail: sales@omega.co.uk
FA
accepts no liability for any errors it contains, and reserves the right to alter
+
INPUT
THERMOCOUPLE CONNECTOR
®
®
K TYPE
V
K TYPE
COPPER CONNECTOR
TRC III
COPPER LEADS
0 0 4 0
100mV 10mV 1mV 100µV 10µV
SERIES TRP REFERENCE PROBES
ICE POINT GENERATOR
9 5
1µV
STABLE VOLTAGE SOURCE
WARRANTY/ DISCLAIMER
OMEGA ENGINEERING, INC. warrants this unit to be free of defects in materials and workmanship for a period of 13 months from date of pu rchase. OMEGA’s WARRAN TY adds an additiona l one (1) warranty to cover handling and shipping time. This ensures that OMEGA’s customers receive maximum coverage on each product.
If the unit
malfunctions, it mu st be returned to the factory for evalua tion. OMEGA’s Customer S ervice Department will issue an Authorized Return (AR) number immediately upon phone or written request. Upon examination by OMEGA, if the unit is found to be defective, it will be repaired or replaced at no charge. OMEGA’s WARRANTY does not apply to defects resulting from any action purchaser, i ncluding but not limited to mishandling, improper interfacin g, o peration outside of design limit s, improper repair unauthorized modification. This hav ing been dam aged as a re sult of exc essi ve c orrosi on; or c urr ent, hea t, m oistur e or vib ration ; impro per misapplication; m isuse or other operating conditions out side of OMEGA’s control . C omponents in which wear is not warranted, include but are not limited to contact points, fuses, and triacs.
OMEGA is pleased to o ffer suggest ions on the use of its variou s prod ucts. However, OMEGA neither assumes respons ibility for any omissions or er rors nor a ssumes liability for any damages that result from t he use of its products in accordance with information provided by OMEGA, either verbal or written. OMEGA warrants only that the parts manufactured by the company will be as specified and free of defects. OMEGA MAKES NO OTHER WARRANTIE OR REPRESENTATIONS OF ANY KIND WHATSOEVER, EXPRESSED OR IMPLI ED, EXCEPT THAT IMPLIE D WARRA NTIES INCLUD ING ANY WARRANT Y OF MERCHANTABI LITY AND FITNE SS FOR A PARTICULA R PURPOSE excl usiv e, negligence, indemnification, strict liability or otherwise, shall not exceed the purchase price of the component upon which liability is based. In no event shall OMEGA be liable for consequential, incidental or special damages.
CONDITIONS: Equipment sold by OMEGA is not intended to be used, nor shall it be used: (1) as a “Basic Component” under 10 CF 21 (NRC), used in or with any nuclear installation or activity; or (2) in medical applications or used on humans. Should any Product(s be used in or with any nuclear installation or activity, medical application, used on humans, or misused in any way, OMEGA assume no responsibility as set forth in our basic and hold OMEGA harmless from any liability or damage whatsoever arising out of the use of the Product(s) in such a manner
Direc t all warranty and repai r request s/inqu iries to t he OMEGA C ustome r Service D epartme nt. BEFORE RETURNING ANY
PRODUCT( S) TO OMEGA, P URCHASER MUST OBTAIN AN SERVICE DEPART outside of the return package and on any correspondence. The purchaser is responsible for shipping charges, freight, insurance and proper packaging to prevent breakage in transit.
FOR WARRANTY RETURNS, please have the following information available BEFORE contacting OMEGA
1. Purchase Order number under which the product was
2. Model and serial number of the product under warranty, and
3. Repair instructions and/or specific problems relative to the
OMEGA’s policy is to make running changes, not model changes, whenever an improvement is possible. This affords our customers the lat­est in technology and engineering.
OMEGA is a registered trademark of OMEGA ENGINEERING, INC © Copyright 2009 OMEGA ENGINEERING, INC. All rights reserved. This document may not be copied, photocopied, reproduced, translated, or reduced
to any electronic medium or machine-readable form, in whole or in part, without the prior written consent of OMEGA ENGINEERING, INC.
ARE HEREBY DISCLAI MED. LIMITATION OF LI ABILITY: The remedies of p urchaser set fort h herein are
and the total liabili ty of OMEG A wit h res pect to this order, whe ther base d on c ontr act, warr anty,
MENT (IN ORDER T O AVOID PRO CESSING DELAYS). Th e a ssigned AR numb er should then be marked on the
PURCHASED,
product.
WARRANTY is VOID if the unit shows evidence of having been tampered with or shows evidence of
WARRANTY/ DISCLAIMER language, and, additionally, purchaser will indemnify OMEGA
RETURN REQUESTS / INQUIRIES
:
month grace period to the nor mal one ( 1) year produc t
OF TITLE, AND ALL
AUTHORIZ ED RETURN (AR) NUMBER F ROM OMEGA’S CUSTOM ER
FOR NON-WARRANTY REPAIRS, charges. Have the following information available BEFOR contacting OMEGA
1. PurchaseOrder number to cover the COST of the repair
2. Model
3. Repair instructions and/or specific problems relative to the product.
.
:
and serial number of the product, and
consult OMEGA for current repair
of the
, or
spe cifi cation ;
.
E
,
S
R
)
s
Loading...