2500
⻬ Employing Clear 144 mm
(5.67") TFT Color LCD
⻬ Large Capacity of Data
Memory and Various
Recording Method
⻬ Multi-Points Recording
with High-Speed/
Accuracy
⻬ Easy Operating and
Programming
Without Manuals
⻬ LAN Network Capability
⻬ Safety System
and Reliability
⻬ Analyzing/Data
Acquisition Application
Software
RD9900 Series are networkcompatible paperless recorders
with high performance and high
operating function employed high
visibility 5.6" TFT color LCD. High
speed of sampling rate 100 ms for
12 points and high accuracy of
±0.1% were realized, and measured
data is stored into internal memory
and maximum 2 GB compact flash
card. As it can be monitored by a
web browser display on several
computers on intranet or internet,
FTP transfer of data file and e-mail
notification are also available. The
recorder can be used for various
applications such as data
management, research,
development, remote and wide
range monitoring by utilizing internet
environment.
Specifications
Input Specifications
Number of Measuring Points:
RD9906: 6 points
RD9912: 12 points
Input Types: Universal (refer to
the table of inputs for RD200/2800
Accuracy Rating: Refer to the table
of inputs for RD200/RD2800 at
omega.com/rd200_rd2800
RD9912, $3200, shown
smaller than actual size.
Reference Junction Compensation
Accuracy: Type K, E, J, T, N, Platinel II;
±0.5°C or less; R, S, NiMo-Ni, CRAuFe, WRe5-WRe26, W-Wre, U, L;
±1.0°C or less
Sampling Rate: Approximately 100 ms
for all points
Burnout: Disconnection of input signal
is detected on thermocouple and
resistance thermometer input;
UP/DOWN disable is selectable for
each input
Scaling: Range/scale is selectable
when DC voltage/current is programmed
Digital Filter: Programming FIR filter for
each point (common to all points)
Allowable Signal Source Resistance:
Thermocouple Input (Burnout
Disable)/DC Voltage Input
(±2V or less): 1 kΩ or less
DC Voltage Input (±5 to ±50V):
100 Ω or less
Resistance Thermometer: Per wire
10 Ω or less (same resistance for
3 wires)
Input Resistance: DC voltage,
thermocouple input; approximately
1 MΩ
S-9
Maximum Input Voltage: DC voltage
input (±2V or less)/thermocouple input
(burnout disable), ±10 Vdc
DC Voltage Input (±5 to ±50V): ±60 Vdc
Dielectric Strength Between Channels:
1000 Vac or more between each channel
(high strength semiconductor relay used)
Display Specifications
Display: 144 mm (5.67") TFT color LCD
Display Types:
Measured Data Display: Trend
screen, data screen, bar-graph screen
Historical Trend Display:
Simultaneous display with real-time
trend is available
Information Display: Alarm display,
marker list, file list
Setting Screen: Alarm, computation,
memory, system, maintenance,
communication, etc.
Trend Screen: 12 colors selectable
Display Screen: 5 screens (5 groups)
Display Points: Maximum 44 points/
screen/
Time Axis Direction: Vertical or
horizontal
Line Width: 1/3/5 dot selectable
Scale Display: 4 scales
Tag/Data Display (Show/hide
selectable)
Marker Display
Data Screen: Display screen, 5 screens (5 groups)
Display Points: Maximum 44 points/screen
Display Contents: Measured value, channel/tag, unit,
alarm status
Bargraph Screen: 12 colors selectable
Display Screen: 5 screens (5 groups)
Display Points: Maximum 44 points/screen
Display Direction: Vertical or horizontal
Scale Display: 1 scale
Information Display:
Alarm Display (Alarm activation/released history display)
Marker List
File List (Group data file list display)
LCD Back Light: Auto/manual OFF function
LCD Brightness: 4 levels adjustment
Recording Specifications
Internal Memory: Flash memory, 4 MB capacity
External Memory: CF card, 32 MB to 2 GB capacity;
128 MB CF card is included as standard
Recording Cycle: 100, 200, 500 ms; 1, 2, 3, 5, 10, 15, 20,
30 s; 1, 2, 3, 5, 10, 15, 20, 30, 60 min
Numbers of Logging Files: 250/numbers of used groups
Logging Data: Measured data; file name (group name),
time of day, month and year of recording start, tag,
measured data, alarm status/types
Storing Types: Binary/CSV type
Storing Methods:
Manual Start/Stop: Dedicated key operation
Schedule: Designation for time of day and date
Trigger Signal: Alarm event pre-trigger is selectable
Measuring Numbers of Pre-Trigger: Maximum 950 data
Recording Cycle
be programmed for 100, 200 and 500 ms recording; 5 groups
of 44 points/group for recording with 1s or more (total of
100 points programming possible)
Recording Duration (CF Card): When 6 channels recorded
in sampling mode (real data)
Recording
Cycle* 128 MB 256 MB 512 MB 1 GB 2 GB
0.1 sec 6.32 days 12.6 days 25.3 days 50.6 days 101 days
1 sec 63.2 days 126 days 253 days 1.4 yrs 2.8 yrs
60 sec 10 yrs 21 yrs 42 yrs 83 yrs 166 yrs
When 12 channels recorded in sampling mode (real data)
Recording
Cycle 128 MB 256 MB 512 MB 1 GB 2 GB
0.1 sec 3.16 days 6.32 days 12.6 days 25.3 days
1 sec 31.6 days 63.2 days 126 days 253 yrs 1.4 yrs
60 sec 5.2 yrs 10 yrs 21 yrs 42 yrs 83 yrs
*: Up to 3 groups of 12 points/group can
Computation Specifications
Computation Points: Maximum 44 points
Computation Types:
Arithmetic Operations: Addition, subtraction,
multiplication, division, remainder, exponential
Comparison Operations: Equality, inequality, great, less,
equality/great, equality/less
Logical Operations: AND, OR, XOR, NOT
General Functions: Round-up, round-down, absolute
value, square root, exponent of e, natural logarithm,
common logarithm
Integration Operations: Analog integration,
digital integration
Channel Data Operations: Measured data computation,
calculated data computation
50.6 days
Alarm Specifications
Setups: Up to 4 alarms can be
programmed per channel
Alarm Types: Upper limit, lower limit,
differential higher limit, differential
lower limit (deadband is selectable),
abnormal data
Delay Function: Setup range of alarm delay, 1 to 3600 sec
Alarm Settings: AND/OR selectable
Communication Functions Network
FTP Server: Data file can be read from the network computer
Web Server: Conformed to HTTP1.0; displays the alarm,
information of maintenance by browser software (Internet
Explorer 5.0 or later, Net Scape 6.0 or later, Opera 7 or later)
User’s ID and password registration available
E-Mail: E-Mail notification at specified time for alarm
activation; report data at specified time is selectable from
all registered
USB Communications**
USB: Communication type, USB1.1
Transfer Systems: Bulk transfer, control transfer
Communication Contents: File transfer by virtual drive
connection
General Specifications
Rated Power Voltage: 100 to 240 Vac (universal power
supply); 50/60 Hz
Maximum Power Consumption: 50 VA (DO: all points
ON, 240 Vac)
Reference Operating Condition:
Ambient Temperature/Humidity Range: 21 to 25°C,
45 to 65% RH
Power Voltage: 100 Vac, ±1.0%
Power Frequency: 50/60 Hz ±0.5%
Attitude: Left/right 0°, forward/backward 0°
Warm-Up Time: Longer than 30 minutes
Normal Operating Condition:
Ambient Temperature/Humidity Range: 0 to 50°C,
20 to 80% RH
Power Voltage: 90 to 264 Vac
Power Frequency: 50/60 Hz ±2%
Attitude: Left/right 0°, forward tilting 0°, backward
tilting 0° to 20°
Transportation Condition (At the Packed Condition on
Shipment from Our Factory):
Ambient Temperature/Humidity Range: -20 to 60°C,
5 to 90% RH (note: no dew condensation)
Vibration: 10 to 60 Hz, 0.5 G (4.9 m/S2) or less
Impact: 40 G (392 m/ S2) or less
Storage Condition: Ambient temperature/humidity range
-20 to 60°C, 5 to 90%RH (note: no dew condensation)
Power Failure Protection: Setups and data are backed up
by flash memory
Clock: Lithium battery backs up RAM (minimum 5 years)
Insulation Resistance:
Secondary Terminals and
Protective Conductor Terminals:
20 MΩ or more at 500 Vdc
Primary Terminals and Protective
Conductor Terminals: 20 MΩ or more at 500 Vdc
Primary and Secondary Terminals: 20 MΩ or more at
500 Vdc
Dielectric Strength:
Secondary Terminals and Protective Conductor
Terminals: 1 minute at 500 Vac
Primary Terminals and Protective Conductor
Terminals: 1 minute at 1500 Vac
Primary and Secondary Terminals: 1 minute at 2300 Vac
See page S-11 for additional information.
**