Omega Products RD8901 User Manual

Paperless Recorder
RD8900 Series
Starts at
$
2375
Up to 18 Isolated
Analog Inputs
6.1" Color TFT DisplayPlug and Play I/O Cards,
6 Slots
User Friendly with
Auto LCD Shut Off
via Infrared Detector, Prolongs Life and Saves Power
Stores Data on Supplied
128 MB Flash Card
Ethernet Interface
Standard with RS232/422/485 Optional
Windows Graphical
Software Standard and Optional Software for Real Time View
Portable Benchtop Option
The OMEGA®RD8900 Series paperless recorder offers real time display of data in a variety of formats on a high resolution (VGA) color TFT display. The user friendly unit with plug & play cards can easily be set to monitor, record, and evaluate any application. The user can access data on the screen, as well as from a remote site via the standard ethernet or optional RS232/422/485 serial interface. The historical data may be stored in a flash ROM, compact flash card, or collected in a remote host PC for evaluation and print out. The unit's compact size saves space and requires less than 7" behind a panel.
Specifications
Display: 6.1" TFT LCD, 640X480 pixel
resolution, 256 colors
Analog Input Card:
RD8900-AI1, AI2, AI3
Resolution: 18 bits Sampling Rate: 5 times/sec Maximum Rating:
-2 Vdc minimum, 12 Vdc maximum (1 min for mA input)
Temperature Effect: ±1.5 µV/°C for all inputs except mA input ±3.0 µV/°C for mA input
Sensor Lead Resistance Effect: T/C:
0.2 µV/ 3-wire RTD: 2.6°C/ of resistance
difference of 2 leads 2-wire RTD: 2.6°C/ of resistance sum
of 2 leads
Burn-Out Current: 200 nA Common Mode Rejection
Ratio (CMRR): 120 dB Normal Mode Rejection
Ratio (NMRR): 55 dB Isolation Breakdown Voltage Among
Channels: 430 Vac min Sensor Break Detection: Sensor open
for TC, RTD and mV inputs, below 1 mA for 4 to 20 mA input, below 0.25V for 1 to 5V inputs, unavailable for other inputs
S-19
RD8906, $2730.
Sensor Break Responding Time:
Within 10 seconds for TC, RTD and mV inputs, 0.1 second for 4 to 20 mA and 1 to 5V inputs
Digital Input Card (RD8900-DI6)
Channels: 6 per card Logic Low: -30V min, 0.8V max Logic High: 2V min, 30V max External Pull-Down Resistance:
1 k max
External Pull-Up Resistance:
1.5 M min
Digital Output Card (RD8900-DO6)
Channels: 6 per card Contact Form: N O (form A) Relay Rating: 5 A/240 Vac, life cycles
200,000 for resistive load
Comm Module (RD8900-RS)
Interface: RS232 (1 unit), RS485 or
RS422 (up to 247 units)
Protocol: MODBUS®protocol RTU mode Address: 1 to 247
Baud Rate: 0.3 to 38.4 Kbits/sec
Extended Warranty
Program
SM
Data Bits: 7 or 8 bits Parity Bit: None, even or odd Stop Bit: 1 or 2 bits
Standard Ethernet Communication
Protocol: Mod Bus TCP/1P,
10 BaseT; auto polarity correction for 10 BaseT
Ports: AUI (attachment unit Interface) and RJ-45 auto-detect capability
Memory
Storage Memory on Board: 8 MB CF Card: 128 MB standard;
optional 1 GB
Infrared Detector
Distance: Detect moving human body
within 2 m Power: 90 to 264 Vac, 47 to 63 Hz,
60 VA, 30 W max (standard); 11 to 18 or 18 to 36 Vdc 60 VA, 30 W max (optional)
Environmental and Physical
Operating Temperature:
5 to 50°C (40 to 122°F)
Storage Temperature:
-25 to 60°C (-12 to 140°F) Humidity: 20 to 80% RH
(non-condensing) Insulation Resistance: 20 M min
(at 500 Vdc) Dielectric Strength: 3000 Vac
50/60 Hz for 1 min Vibration Resistance: 10 to 55 Hz,
2
10 m/S Shock Resistance: 30 m/S
for operation, 100 g for transportation
Dimensions:
166 W x 144 H x 174 mm D (6.54 x 5.67 x 6.85") for panel mount
Panel Cutout: 138 x 138 mm (5.44 x 5.44")
for 2 hr
2
(3 g)
Approval Standards
Safety: CE; EN61010-1 (IEC1010-1)
Overvoltage category II, pollution degree 2
Protective Class: IP-30 front panel, indoor use, IP-20 housing and terminals
EMC
Emission: EN50081-1, EN61326
(EN55011 class B, EN61000-3-2, EN61000-3-3)
Immunity: EN50082-2, EN61326 (EN61000-4-2, EN61000-4-3, EN61000-4-4, EN61000-4-5, EN61000-4-6, EN61000-4-11, EN50204)
OMEGACARESMextended warranty program is available for models shown on this page. OMEGACARE covers parts, labor, and equivalent loaners. Ask your sales representative for full details when placing an order.
SM
RD8906, $2730, rear view featuring plug and play I/O cards, shown smaller than actual size.
Characteristics
Accuracy Input
Type Range @25°C Impedance
-120 to 1000°C ±1°C 2.2 M
(-184 to 1832°F)
-200 to 1370°C ±1°C 2.2 M
(-328 to 2498°F)
-250 to 400°C ±1°C 2.2 M
(-418 to 752°F)
-100 to 900°C ±1°C 2.2 M
(-148 to 1652°F)
0 to 1820°C ±2°C 2.2 M
(32 to 3308°F) (200 to 1820°C)
0 to 1767.8°C ±2°C 2.2 M
(32 to 3214°F)
0 to 1767.8°C ±2°C 2.2 M
(32 to 3214°F)
-250 to 1300°C ±1°C 2.2 M
(-418 to 2372°F)
-200 to 900°C ±1°C 2.2 M
(-328 to 1652°F)
PT100 -210 to 700°C ±0.4°C 1.3 k (DIN) (-346 to 1292°F)
PT100 -200 to 600°C ±0.4°C 1.3 k (JIS) (-328 to 1112°F)
mV -8 to 70 mV ±0.05% 2.2 M mA -3 to 27 mA ±0.05% 70.5 0~1V -0.12 to 1.15V ±0.05% 32 k 0~5V -1.3 to 11.5V ±0.05% 332 k 1~5V -1.3 to 11.5V ±0.05% 332 k 0~10V -1.3 to 11.5V ±0.05% 332 k
S-20
S
Trend Mode
Historical Mode
View max 6 real time data trends verticallyRecognize data trends easily by different colors
and tag names
Switch to other configured pages easily by “Page” function keyDisplay current “Time/Date” informationRemind the user of “Alarm” or “Memory Full”
Bar Graph Mode
View max 6 real time data in bar graphsScale individually by user in “configuration”Display data value and tag name in different colors
together with each bar graph
Mark “Hi/Lo” alarm limitsDisplay current “Time/Date” informationRemind the user of the “Alarm” or “Memory Full”
Display max 6 sets of historical data simultaneouslyView desired data section by “Access precise data value at a point selected by
moving the “ruler”
“Zoom” to expand/contract the display time spanView historical data trends and their respective
data values
Recognize trends easily by different colors and
individual tag names
Alarm List
List all the alarm records clearly with useful informationBrowse through the alarm list or “acknowledge” alarm
easily by function keys on the vertical bar
Remind the user of the alarm status in different colors
” and “佣” function keys
Numerical Mode
View max 6 real time data in numbersDisplay data value and tag name in different colorMark “Hi/Lo” alarm limitsDisplay current “Time/Date” informationRemind the user of the “Alarm” or “Memory Full”
Configuration Mode
Configure pen (input/output, pen name, event, job.....)
Configure page (color, pen, decimal, pen width.....)
Configure timerConfigure instrument (storage media, display,
communication, time/date....)
S-21
Realtime trend screen.
RD8906, $2730, with benchop conversion kit, RD8900-BPS, $219.
To Order (Specify Model Number)
Model No. Price Description RD8901 $2375 Recorder, 1-input, color, 5 spare I/O slots RD8902 2415 Recorder, 2-input, color, 5 spare I/O slots RD8903 2430 Recorder, 3-input, color, 5 spare I/O slots RD8904 2630 Recorder, 4-input, color, 4 spare I/O slots RD8905 2689 Recorder, 5-input, color, 4 spare I/O slots RD8906 2730 Recorder, 6-input, color, 4 spare I/O slots RD8909 3050 Recorder, 9-input, color, 3 spare I/O slots RD8912 3365 Recorder, 12-input, color, 2 spare I/O slots RD8915 3575 Recorder, 15-input, color, 1 spare I/O slot RD8918 3785 Recorder, 18-input, color, no spare I/O slots
MOST POPULAR MODELS HIGHLIGHTED!
Options (Field Installable in Spare I/O Slots, One Slot Required for Each Item)
Model No. Price Description RD8900-AI1 $155 Analog input card, 1-channel RD8900-AI2 255 Analog input card, 2-channels RD8900-AI3 399 Analog input card, 3-channels RD8900-DI6 280 Digital input card, 6 inputs RD8900-DO6 280 Digital output card, 6 relays/alarms RD8900-TPS 210 Transmitter power supply (24 Vdc 180 mA)
Accessories
Model No. Price Description RD8900-BPS $219 Kit to convert to benchtop style RD8900-SW2 199 Software, real time graphical via comm port 1GB-FC 65 Memory card,1 GB compact flash (optional); standard with recorder: 128 MB
Comes complete with operator’s manual, panel mounting brackets, basic observer software for non-communication application and 128 MB flash card.
Notes: Recorder offers Ethernet interface as standard. To include RS232/422/485 communications add “-RS” suffix and add $160 to price. Feature may be field installed by ordering RD8900-RS. To include counter/totalizer math functions add “-MATH” suffix and $340 to price. To operate on DC power add “-DC12” (for 11 to 18 Vdc) or “-DC24” (for 18 to 36 Vdc) suffix, no additional charge
Standard software provided with recorder supports transfer of data and recorder configuration only via the memory card. User must have a memory card reader (not provided) connected with the host computer. The data may be read, analyzed, printed or transferred via an Excel type file to another application. The RD8900-SW2 allows above, plus supports real time monitoring and configuration via the communications port. Software is Windows 98, NT, 2000 and XP compatible.
Ordering Examples: RD8906, recorder with 6 inputs, plus RD8900-DO6, digital output card, plus 1GB-FC, memory card, $2730 + 280 + 65 = $3075. OCW-3, OMEGACARESMextends standard 1-year warranty to a total of 4 years ($350), $2730 + 350 = $3080.
RD8903, recorder with 3 inputs, plus RD8900-DI6, digital input card, plus RD8900-BPS, kit to convert to benchtop style, plus RD8900-SW2, software for real time graphical via communications port, $2430 + 280 + 219 + 199 = $3128. OCW-3, OMEGACARESMextends standard 1-year warranty to a total of 4 years ($350), $2430 + 350 = $2780.
S
S-22
One Omega Drive | Stamford, CT 06907 | 1-888-TC-OMEGA (1-888-826-6342) | info@omega.com
EPG05
www.omega.com
UNITED KINGDOM
www. omega.co.uk
Manchester, England
0800-488-488
UNITED STATES
www.omega.com
1-800-TC-OMEGA
Stamford, CT.
CANADA
www.omega.ca
Laval(Quebec)
1-800-TC-OMEGA
GERMANY
www.omega.de
Deckenpfronn, Germany
0800-8266342
Karviná, Czech Republic
FRANCE
www.omega.fr
Guyancourt, France
088-466-342
CZECH REPUBLIC
www.omegaeng.cz
596-311-899
BENELUX
www.omega.nl
Amstelveen, NL 0800-099-33-44
More than 100,000 Products Available!
Temperature
Calibrators, Connectors, General Test and Measurement Instruments, Glass Bulb Thermometers, Handheld Instruments for Temperature Measurement, Ice Point References, Indicating Labels, Crayons, Cements and Lacquers, Infrared Temperature Measurement Instruments, Recorders Relative Humidity Measurement Instruments, RTD Probes, Elements and Assemblies, Temperature & Process Meters, Timers and Counters, Temperature and Process Controllers and Power Switching Devices, Thermistor Elements, Probes and Assemblies,Thermocouples Thermowells and Head and Well Assemblies, Transmitters, Wire
Flow and Level
Air Velocity Indicators, Doppler Flowmeters, Level Measurement, Magnetic Flowmeters, Mass Flowmeters, Pitot Tubes, Pumps, Rotameters, Turbine and Paddle Wheel Flowmeters, Ultrasonic Flowmeters, Valves, Variable Area Flowmeters, Vortex Shedding Flowmeters
pH and Conductivity
Conductivity Instrumentation, Dissolved Oxygen Instrumentation, Environmental Instrumentation, pH Electrodes and Instruments, Water and Soil Analysis Instrumentation
Data Acquisition
Auto-Dialers and Alarm Monitoring Systems, Communication Products and Converters, Data Acquisition and Analysis Software, Data Loggers Plug-in Cards, Signal Conditioners, USB, RS232, RS485 and Parallel Port Data Acquisition Systems, Wireless Transmitters and Receivers
Pressure, Strain and Force
Displacement Transducers, Dynamic Measurement Force Sensors, Instrumentation for Pressure and Strain Measurements, Load Cells, Pressure Gauges, Pressure Reference Section, Pressure Switches, Pressure Transducers, Proximity Transducers, Regulators, Strain Gages, Torque Transducers, Valves
Heaters
Band Heaters, Cartridge Heaters, Circulation Heaters, Comfort Heaters, Controllers, Meters and Switching Devices, Flexible Heaters, General Test and Measurement Instruments, Heater Hook-up Wire, Heating Cable Systems, Immersion Heaters, Process Air and Duct, Heaters, Radiant Heaters, Strip Heaters, Tubular Heaters
click here to go to the omega.com home page