The RD200 and RD2800 Series
chart recorders are 100 mm,
multipoint, hybrid recorders that can
print/display multichannel and alarm
data simultaneously. A total of
56 ranges can be programmed for
each channel: 35 thermocouple,
11 RTD, and 10 DC voltage. The
user can also program up to 4 levels
of alarms for each channel; optional
alarm outputs are available. The
recorders display alarm settings
(status and channels), and can print
channel alarms and alarm numbers.
A data acquisition software
package, “KIDS”, lets the user
create real-time and historical
data/trend displays. The RD200 and
2800 Series chart recorders are
manufactured in ISO9001-certified
facilities and conform to CE, UL,
and CSA safety standards.
Universal input, DC voltage,
thermocouple, DC current (requires
shunt resistor)
Range Setup: Programming of input
types and ranges by keys
Scale Setup: Programming of
maximum value, minimum values,
and engineering units by keys
Accuracy Rating:
Refer to the table of inputs
RD204, $2630, shown
smaller than actual size.
RD206, $1560, shown
smaller than actual size.
Measuring Interval:
About 5 s/6 points
About 10 s/12 points
About 20 s/24 points
About 100 ms (pen-type)
Reference Junction Compensation
Accuracy:
K, E, J, T, N, Platinel II:
±0.5°C (0.9°F) or less
R, S, NiMo-Ni, CR-AuFe,
WRe5-WRe26, W-Wre, U, L:
±1.0°C (1.8°F) or less
At measurements higher than 0°C
(32°F), the above errors are added to
the accuracy ratings for internal
reference junction compensation.
Burnout: With function to detect input
signal disconnection for thermocouple
inputs and resistance thermometer
inputs; up-scale burnout, down-scale
burnout or burnout disable is selectable
for each input
Terminal Board: Detachable type,
removable on wirings
Alarm Specifications
Alarm Display: “ALARM” illumination
and flashing of measured value at an
alarm-activated channel.
Alarm Types: Absolute value alarm,
differential alarm, rate-of-change alarm
Alarm Programming: Individual
programming for each channel,
maximum 4 levels (alarm points)
per channel
Unit: Same color as trace printing
Month/Day or Year/Month/Day,
Time, Time Line, Chart Speed:
Black
S-XX
List Printing:
Programmed Parameters:
Same color as trace printing
Others: Black
Programming Change Mark: Black
Alarm Printing: Red
Chart: Fan-fold
Chart Speed: 1 to 1500 mm/hr
Chart Speed Default:
RD200—20 mm/hr, RD2800—25 mm/hr
Periodic Data Printing: Digital printing of
time, channel numbers and measured
values on trace printing interval time
(hour, minute); optional programming
(limited by chart speeds)
Digital Data Printing: Digital printing of
time and measured values by
interrupting trace printing on demand
Alarm Printing:
Alarm-Activated: Time, channel
number, alarm type and level
(alarm point number)
Alarm Reset: Time, channel number
and level (alarm point number)
Subtract Printing: Printing of difference
between 2 channels or between a
channel and a reference value
(programmed value)
Fixed-Time Printing: Printing of
month/day, time, time line, ranges
(scales), tags and engineering units
every fixed-time (interlocking to chart
speed)
S
Display Specifications
(Multipoint)
Digital Display: -9999 to 99999
Display Items: Simultaneous display of
6 or 12 measured values per channel,
or time(year/month/day/hour/minute),
alarm-activated channel and chart speed
Status Display: Printing status, key
lock and alarm-activation condition
Printing Specifications
(Pen-Type)
Printing System:
Analog Tracing:
Disposable cartridge pen
Digital Printing: Plotter pen
Printing Color:
Analog Tracing: 1st pen red,
2nd pen green, 3rd pen blue,
4th pen brown
Digital Printing: Purple periodic data
printing, digital data printing (analog
tracing continuance/interruption), date
and time printing (at power-on, every
hour), chart speed printing, scale, unit
and tag printing, alarm activation/reset
printing, programming change mark,
pen offset correction (POC)
Chart Speed:
1 to 600 mm/hr, 1 to 200 mm/min
Default:
RD200: 20 mm/hr
RD2800: 25 mm/hr
Phase Synchronizing Correction:
Time-axis POC
(POC) Subtract Printing: Printing of
difference between 2 channels or
between a channel and a reference value
(programmed value)
Message Printing: Pre-programmed
letters are printed by a key or a remote
contact (optional); 5 kinds of message
(time + message of maximum 15 letters)
Pen-Lift Function: By RECORD OFF
key, all pens are lifted up
simultaneously; at power-off, the last
pen status is kept; a lever for manually
lifting up/down of all pens is provided
Display Specifications
(Pen-Type)
Analog Indication:
RD200 Series: 100 mm bar graph
for each input point (51 segments,
same color as analog tracing is
indicated at each 5 segments)
RD2800 Series: 180 mm bar graph
for each input point (101 segments,
same color as analog tracing is
indicated at each 10 segments)
Digital Display: -9999 to 99999
[optional decimal place, with cursor
(by each analog tracing color)]
Display Items:
RD200 Series: Simultaneous
display of 4-channel measured
values, hour/minute, chart speed
and alarm-activated channel
RD2800 Series: Simultaneous
display of 4-channel measured
values or year, month/day,
hour/minute, chart speed and
alarm-activated channel lock and
alarm-activation condition
RD2800-KIDS
KIDS data acquisition
and communications
setup software, $270.
S-XX
RD2800-PASS
Extended Warranty
Program
SM
Pass parameter
programming and
communications setup
software, $270.
General Specifications
Rated Power Supply:
100 to 240 Vac, 50/60 Hz
Rated Power Consumption:
Multipoint: Max 45 VA
Pen-Type: Max 60 VA
Normal Operating Conditions:
Ambient Temp/Humidity:
Multi-Point:
0 to 40°C (32 to 104°F),
20 to 80% RH
Pen Type: 0 to 50°C (32 to 122°F),
20 to 80% RH
Power Voltage: 90 to 264 Vac
Power Frequency: 50/60 Hz ±2%
Altitude:
Left/right 0 to 10°, forward tilting 0°,
backward tilting 0 to 30°
Power Failure Protection:
Multipoint: Programmed
parameters stored in EEPROM
memory; clock circuit sustained
for minimum 10 years by a lithium
battery (at operation of more
than 8 hours/day)
Pen-Type: Programmed
parameters stored in EEPROM
memory; clock circuit and POC
data sustained for minimum 8 years
by a lithium battery (at operation
of more than 8 hours/day)
Case Assembly Material:
Door: ABS resin (frame) with glass
Enclosure: Steel
OMEGACARESMextended warranty program
is available for models shown on this page.
Ask your sales representative for full details
when placing an order. OMEGACARE
covers parts, labor and equivalent loaners.
RD2804, $3850, shown smaller than actual size.
SM
S-XX
S
RD2812, $2295.
RD204, $2630.RD206, $1506.
RD2804, $3850.
All models shown
smaller than actual size.
Options (Pen-Type)
OptionsExplanations
Alarm Outputand 12 points (RD2800 only); maximum
Printing Format*channel is compressed or expanded;
Communications(user can specify); two kinds of protocol,
Basic Mathematics square root, logarithm, natural
Totalizer/Totalizing of measured data or
Flow Correctioncalculated results and correction of
One kind of printing format specified by user.
*
Three kinds of output (alarm, fail and
chart-end) are possible; output: 6 points
1 second/6 points, 2 second/12 points,
4 seconds/24 points; alarm judgment
interval: same as measuring interval,
multipoint simultaneous display only;
printing interval: about 5 seconds/point,
Measuring Interval conforming to CE, UL, CSA; the
indication equivalent to maximum 25°C
or 2 mV may vary under the test
environment requested by EMC
directive; by signals of 4-point
contacts and 2-point common
Output: 6-, 12- or 24-point individual
Alarm Outputoutput possible; maximum contact
rating: 100 Vac/0.5 A, 240 Vac/0.2 A,
100 Vdc/0.3 A
Alarm OutputPrinting area is divided into 4 zones
(RD2800) or 2 zones (RD200)
Compressed/A part of printing area of each channel
Expanded Printing
is printing compressed or expanded
AutomaticPrinting range is automatically changed
Range-Shiftinto a new printing compressed or
Printing Format*
expanded
RS232C, RS422A, RS485
Communications(user can specify); parameter
programming, operation and data
logging (MODBUS protocol)
Totalizing of measured data or
Totalizercalculated data; interval: 1 minute to
24 hours or no interval
S-XX
Input Table
InputMeasuringReference AccuracyDisplay
Signals RangesRangesRatings
-13.8 to 13.8 mV 13.8 mV10 µV
-27.6 to 27.6 mV ±27.6 mV10 µV
-200 to 200 mV ±200 mV100 µV
DC-500 to 500 mV ±500 mV100 µV
Voltage-2 to 2V±2V1 mV
-5 to 5V±5V1 mV
-10 to 10V±10V10 mV
-20 to 20V±20V10 mV
-50 to 50V±50V10 mV
-200 to 300°C ±13.8 mV ±0.1% ±1 digit0.1°C
K-200 to 600°C ±27.6 mV0.1°C
-200 to 1370°C ±69.0 mV1°C
-200 to 200°C ±13.8 mV0.1°C
E-200 to 350°C ±27.6 mV0.1°C
-200 to 900°C ±69.0 mV1°C
-200 to 250°C ±13.8 mV0.1°C
J-200 to 500°C ±27.6 mV0.1°C
-200 to 1200°C ±69.0 mV1°C
T-200 to 250°C ±13.8 mV0.1°C
-200 to 400°C ±27.6 mV0.1°C
R0 to 1200°C±13.8 mV1°C
0 to 1760°C±27.6 mV1°C
S0 to 1300°C±13.8 mV1°C
0 to 1760°C±27.6 mV1°C
B0 to 1820°C±13.8 mV1°C
-200 to 400°C ±13.8 mV ±0.15% ±1 digit0.1°C
N-200 to 750°C ±27.6 mV0.1°C
-200 to 1300°C ±69.0 mV1°C
W-0 to 2315°C±69.0 mV1°C
WRe26
WRe5-0 to 2315°C±69.0 mV1°C
WRe26
Thermocouples
PrRh40-0 to 1888°C±13.8 mV1°C
PrRh20
-50 to 290°C±13.8 mV0.1°
NiMo-Ni -50 to 600°C±27.6 mV0.1°C
-50 to 1310°C ±69.0 mV1°C
CR-AuFe
Platinel II
U-200 to 500°C ±27.6 mV0.1°C
L-200 to 500°C ±27.6 mV0.1°C
Pt100-200 to 300°C220 Ω0.1% ±1 digit 0.1°C
JPt100 -200 to 300°C 220 Ω0.1% ±1 digit0.1°C
Pt50-200 to 649°C220 Ω0.1°C
Resistance Thermometer
Pt-Co4 to 374 K220 Ω
0 to 280 K±13.8 mV0.1 K
0 to 350°C±13.8 mV ±0.15% ±1 digit0.1°C
0 to 650°C±27.6 mV0.1°C
0 to 1390°C±69.0 mV1°C
-200 to 250°C ±13.8 mV0.1°C
-200 to 600°C ±69.0 mV0.1°C
-200 to 250°C ±13.8 mV 0.1% ±1 digit 0.1°C
-200 to 900°C ±69.0 mV1°C
-140 to 150°C160 Ω
-200 to 850°C400 Ω0.1°C
-140 to 150°C160 Ω
-200 to 649°C400 Ω0.1°C
±0.1% ±1-digit
±0.1% ±1-digit
±0.15% ±1-digit
±0.2% ±1-digit
±0.15% ±1-digit
±0.1% ±1-digit
±0.15% ±1-digit
±0.1% ±1-digit
±0.15% ±1-digit
±0.1% ±1-digit
±0.15% ±1-digit
Resolutions
RD206, $1560, shown smaller than actual size.
0.1°C
0.1°C
S
RD2812, $2295, shown smaller than actual size.
0.1 K
S-XX
237 (9.33)
144 (5.6)
)
93.5( 731
)
43.6( 161
67)
.5( 441
18 (0.71)63 (2.48)
24
(0.94)
260 (10.24)
144 (5.6)
)
93.5( 731
)
44.6( 161
)
76.5( 441
18 (0.71)63 (2.48)
24
(0.94)
260 (10.24)
297 (11.69)
288 (11.34)
11( 882
)
43.
01( 772
)
19.
24 (0.94)
RD200 Series
Multi-Points
External Dimensions
Dimensions: mm (in)
RD200 Series
Pen-Type
RD2800 Series
Note: Accuracy ratings are of measuring ranges at reference operation conditions. The reference junction compensation accuracy is
not included with the accuracy ratings of thermocouple inputs. The indication equivalent to 200 µV or 5°C (41°F) may vary under the test
environment requested by EMC directive. Reference operating conditions:
Ambient Temperature/Humidity Range:
21 to 25°C (70 to 77°F), 45 to 65% RH
Power Voltage: 100 Vac ±1%
Power Frequency: 50/60 Hz ±0.5%
Altitude: Left/right 0°, forward tilting 0°, backward tilting 0°
Warm-Up Time: More than 30 minutes
S-XX
MOST POPULAR MODELS HIGHLIGHTED!
To Order (Specify Model Number)
Model No.PriceDescription
RD201$1430100 mm, 1-pen
RD202 1830100 mm, 2-pen
RD2032230100 mm, 3-pen
RD204 2630 100 mm, 4-pen
RD206 1560100 mm, 6-dot points
RD2801 1850180 mm, 1-pen
RD2802 2490180 mm, 2-pen
RD2803 3150180 mm, 3-pen
RD2804 3850 180 mm, 4-pen
RD2806 2130 180 mm, 6-dot points
RD2812 2295180 mm, 12-dot points
RD28243130180 mm, 24-dot points
Comes complete with user’s manual, 1 pen per channel and 1 pack of chart paper.
Calibrators, Connectors, General Test and Measurement
Instruments, Glass Bulb Thermometers, Handheld Instruments
for Temperature Measurement, Ice Point References,
Indicating Labels, Crayons, Cements and Lacquers, Infrared
Temperature Measurement Instruments, Recorders Relative
Humidity Measurement Instruments, RTD Probes, Elements
and Assemblies, Temperature & Process Meters, Timers and
Counters, Temperature and Process Controllers and Power
Switching Devices, Thermistor Elements, Probes and
Assemblies,Thermocouples Thermowells and Head and Well
Assemblies, Transmitters, Wire
Flow and Level
Air Velocity Indicators, Doppler Flowmeters, Level
Measurement, Magnetic Flowmeters, Mass Flowmeters,
Pitot Tubes, Pumps, Rotameters, Turbine and Paddle Wheel
Flowmeters, Ultrasonic Flowmeters, Valves, Variable Area
Flowmeters, Vortex Shedding Flowmeters
pH and Conductivity
Conductivity Instrumentation, Dissolved Oxygen
Instrumentation, Environmental Instrumentation, pH
Electrodes and Instruments, Water and Soil Analysis
Instrumentation
Data Acquisition
Auto-Dialers and Alarm Monitoring Systems,
Communication Products and Converters, Data
Acquisition and Analysis Software, Data Loggers
Plug-in Cards, Signal Conditioners, USB, RS232, RS485
and Parallel Port Data Acquisition Systems, Wireless
Transmitters and Receivers
Band Heaters, Cartridge Heaters, Circulation Heaters,
Comfort Heaters, Controllers, Meters and Switching
Devices, Flexible Heaters, General Test and Measurement
Instruments, Heater Hook-up Wire, Heating Cable
Systems, Immersion Heaters, Process Air and Duct,
Heaters, Radiant Heaters, Strip Heaters, Tubular Heaters
click here to go to the omega.com home page
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.