ELECTRICAL CONNECTOR
TWIST-LOCK PCIH0-10-6P (101)
OR EQUIVALENT
61 (2.40)
MAX.
17.47 ±0.25
(0.688 ±0.01)
(WRENCH FLATS)
6.35 ±0.25
(0.25 ±0.01)
PRESSURE FITTING
MS33656-4
Functions
POSITIVE INPUT
POSITIVE OUTPUT
NEGATIVE OUTPUT
NEGATIVE INPUT
SHORTED TO "D"
SHORTED TO "C"
Pins
A
B
C
D
E
F
SCHEMATIC
25.7 (1.01) DIA. MAX.
61 (2.40)
MAX.
17.47 ±0.25
(0.688 ±0.01)
(WRENCH FLATS)
6.35 ±0.25
(0.25 ±0.01)
PRESSURE FITTING
MS33656-4
Functions
POSITIVE INPUT
POSITIVE OUTPUT
NEGATIVE OUTPUT
NEGATIVE INPUT
GROUND (CASE)
SCHEMATIC
25.7 (1.01) DIA. MAX
Cable Color
RED STRIPE
YELLOW STRIPE
BLUE STRIPE
WHITE
BROWN STRIPE
762 (30.0)
MIN.
HIGH- AND VERY-HIGH-TEMPERATURE
PRESSURE TRANSDUCERS
PX1004/PX1009 Series
mV/V Output
0-15 to 0-10,000 psi
0-1 to 0-700 bar
1 bar = 14.5 psi
1 kg/cm2= 14.22 psi
1 atmosphere = 14.7 psi = 29.93
inHg = 760.2 mmHg = 1.014 bar
Starts at
$
1850
⻬ Two Versions That
Operate from -54 to
232°C (-65 to 450°F)
(PX1004) or -54 to 343°C
(-65 to 650°F) (PX1009)
⻬ High Performance
⻬ Proven High Reliability
⻬ Exceptional Calibration
Stability
Applications
⻬ Engine and Test Stands
⻬ High-Temperature Testing
⻬ Process Monitoring
⻬ Ground Support
Equipment
⻬ High-Temperature
Geothermal Generators
PX1009L0-1KAV, $2500,
shown actual size.
tandard
S
OMEGA’s thin-film technology
makes this premium performance
possible. The strain gages are
sputter-deposited, forming a
molecular bond with the substrate.
There is virtually no shift, drift,
or creep to cause the transducer’s
calibration to change. The all-welded
stainless steel pressure cavity
and double-isolated case ensure
pressure integrity and reliability
in adverse environments.
These high-temperature
transducers are available in many
standard ranges from 15 psi to
10,000 psi. A NIST-traceable
calibration record is available for
these units.
SPECIFICATIONS
Excitation: 10 Vdc
Full Range Output:
30 mV nominal, 26 mV minimum
Residual Unbalance:
0 mV ±5% FSO at zero pressure
Input Resistance: 400 Ω nominal
Output Resistance: 400 Ω nominal
PX1004
PX1004L1-1KAV, $1850,
shown actual size.
Insulation Resistance:
PX1004:
100 MΩ or greater @ 45 Vdc
PX1009: 50 MΩ or greater @ 45 Vdc
Electrical Connections:
PX1004: PCIH0-10-6P or
equivalent
PX1009: Color-coded pigtail leads,
0.8 m (30") long, minimum
Sensing Element:
4-active-arm bridge using sputterdeposited thin-film elements.
Accuracy: Combined linearity,
hysteresis and repeatability: ±0.25%
FSO, BSL (over the compensated
temperature range)
[Note: Performance is determined at
25°C ±1°C (77°F ±2°F), open circuit,
rated excitation, unless otherwise
specified]
Operating Temp Range:
PX1004: -54 to 232°C (-65 to 450°F)
PX1009: -54 to 343°C (-65 to 650°F)
Compensated Temp Range:
PX1004: 24 to 204°C (75 to 400°F)
PX1009: 24 to 316°C (75 to 600°F)
PX1009
OMEGA’s PX1004 and PX1009
sputtered thin-film pressure
transducers are designed for
high-temperature service. These
instruments can operate in
temperatures from -54 to 232°C or
343°C (-65 to 450°F or 650°F).
Yet even in these extreme
temperatures, they provide
outstanding accuracy, long-term
calibration stability, and reliability.
Static accuracy is ±0.25%, and
thermal zero and sensitivity shifts
over their compensated ranges are
less than ±0.01%/°F.
Dimensions: mm (in)
B-75
HIGH AND VERY HIGH TEMPERATURE PRESSURE TRANSDUCERS
Thermal Effects:
Span: ±0.01% FSO/°F
Zero: ±0.01% FSO/°F
Natural Frequency:
50 kHz at 5000 psi decreasing
logarithmically to 5 kHz at 15 psi
Vibration Sensitivity: At 35 g peak
from 10 Hz to 2000 Hz (limited to
1
⁄2" D.A.), the output shall not exceed:
15 psi 0.04% FSO/g 0.15% FSO/g
50 psi 0.02% FSO/g 0.07% FSO/g
100 psi 0.015% FSO/g 0.05% FSO/g
500 psi 0.006% FSO/g 0.02% FSO/g
1000 psi 0.003% FSO/g 0.02% FSO/g
PX1004 PX1009
Shock: 100 g, 11 ms half sine wave
without calibration shift or damage
Humidity:
Per MIL-STD-810, Method 507.1
Standard ranges: 0 to 15, 25, 50, 100,
250, 500, 1000, 1500, 2000, 2500, 5000
and 10,000 psia. Sealed gage available
in ranges of 100 psi and above.
Proof Pressure:
PX1004: 200% of rated pressure
PX1009: 150% of rated pressure
Burst Pressure:
PX1004: 300% of rated pressure
PX1009: 200% of rated pressure
Wetted Parts: 17-4 PH vacuum melt or
15-5 stainless steel
MOST POPULAR MODELS HIGHLIGHTED!
Pressure Fitting:
7
⁄16-20 male per MS33656-4
Mounting Isolation: Case isolation
ensures that the sensing element will
not be affected by external stresses
Mating Connector (PX1004):
PCS06E-10-6S(SR) (sold separately)
Weight: 145 g (5 oz) maximum
PX1004 Mating
Connector
PCS06E-10-6S(SR)
$135
To Order (Specify Model Number)
RANGE MODEL NO. PX1004 PRICE MODEL NO. PX1009 PRICE COMPATIBLE METERS*
ABSOLUTE PRESSURE (All Ranges Available in Sealed Gage Pressure)
0 to 15 psi 0 to 1.0 bar PX1004L1-015AV $1850 PX1009L0-015AV $2500 DP41-S, DP25B-S
0 to 25 psi 0 to 1.7 bar PX1004L1-025AV 1850 PX1009L0-025AV 2500 DP41-S, DP25B-S
0 to 50 psi 0 to 3.4 bar PX1004L1-050AV 1850 PX1009L0-050AV 2500 DP41-S, DP25B-S
0 to 100 psi 0 to 6.9 bar PX1004L1-100AV 1850 PX1009L0-100AV 2500 DP41-S, DP25B-S
0 to 250 psi 0 to 17.2 bar PX1004L1-250AV 1850 PX1009L0-250AV 2500 DP41-S, DP25B-S
0 to 500 psi 0 to 34.5 bar PX1004L1-500AV 1850 PX1009L0-500AV 2500 DP41-S, DP25B-S
0 to 1000 psi 0 to 68.9 bar PX1004L1-1KAV 1850 PX1009L0-1KAV 2500 DP41-S, DP25B-S
0 to 1500 psi 0 to 103 bar PX1004L1-1.5KAV 1850 PX1009L0-1.5KAV 2500 DP41-S, DP25B-S
0 to 2000 psi 0 to 138 bar PX1004L1-2KAV 1850 PX1009L0-2KAV 2500 DP41-S, DP25B-S
0 to 2500 psi 0 to 172 bar PX1004L1-2.5KAV 1850 PX1009L0-2.5KAV 2500 DP41-S, DP25B-S
0 to 5000 psi 0 to 345 bar PX1004L1-5KAV 1850 PX1009L0-5KAV 2500 DP41-S, DP25B-S
0 to 10,000 psi 0 to 689 bar PX1004L1-10KAV 1850 PX1009L0-10KAV 2500 DP41-S, DP25B-S
Comes complete with 5-point calibration.
To order sealed gage pressure, replace “A” in model number with “S” (no extra charge).
Ordering Examples: PX1004L1-100AV, 100 psi absolute pressure transducer with male pressure connection per MS33656-4 and
PCHI0-10-6P electrical connector, $1850. PCS06E-10-6S(SR), mating connector (sold separately), $135. PX1009L0-1KSV, 1000 psi sealed
gage pressure transducer with MS33656-4
male pressure connection and 0.8 m (30")
high-temperature cable, $2500.
CUSTOM CONFIGURATIONS
PRESSURE ELECTRICAL RANGE
PX1004/1009 PORT CONNECTION UNITS OUTPUT OPTIONS
SERIES [ 1 ] [ 2 ] [ 3 ] [ 4 ] [ 5 ] [ 6 ]
L = MS33656-4
PX1004 only 1 = PCIH0-10-6P
Connector (or equal).
Mating connector
(sold separately), order
PCS06E-10-6S(SR)
$135.
PX1009 only 0 = color-coded
cable, 0.8 m (30") min.
Metric ranges available – Consult Engineering.
Ordering Examples: PX1004L1-100AV, MS33656-4 male pressure port, twist-lock connector, 100 psi absolute pressure range and mV/V
output, $1850. PCS06E-10-6S(SR), mating connector (sold separately), $135. PX1004L1-015SV, MS33656-4 male pressure port, twist-lock
connector, 15 psi sealed gage pressure range and mV/V output, $1850. PCS06E-10-6S(SR), mating connector (sold separately), $135.
PX1009L1-1KSV-CAL11, MS33656-4 male pressure port, 0.8 m (30") high-temperature cable, 1000 psi sealed gage pressure range,
mV/V output and optional 11-point NIST-traceable calibration, $2500 + 250 = $2750.
See section D for compatible meters. Metric ranges available – Consult Engineering.
*
ACCESSORY
MODEL NO. PRICE DESCRIPTION
PCS06E-10-6S(SR) $135 Mating connector for PX1004 transducers
015 psi
025
050
100
250
500
1K
1.5K
2K
2.5K
5K
10K
A
A
A
A, S
A, S
A, S
A, S
A, S
A, S
A, S
A, S
A, S
V = mV/V Output
Base Price PX1004:
$1850
Base Price PX1009:
$2500
To order a custom configuration:
1. Select a pressure port
2. Select electrical connection
3. Select a pressure range
4. Select pressure units
CAL11 = 11-point
traceable calibration
+$250
5. Select output
6. Select options and agency approvals
NIST-
PRESSURE TRANSDUCERS
MILLIVOLT OUTPUT
B
B-76