VERY-HIGH-ACCURACY MILLIVOLT
ARTWORK/PRODUCT ART/ OMEGADYNE/OD-PX01,PX02MV
38
(1.50)
DIA.
Wiring Code
Pin
A:
B:
C:
D:
E:
F:
Connection
+ OUTPUT
– OUTPUT
– INPUT
+ INPUT
SHUNT
SHUNT
PX02:
19 HEX
(0.75)
0 to 10,000 psi: 7/16-20 UNF or 1/4 NPT
PX02: 10,000 psi and higher 1/4 HP tube
PX01, PX02 mV Output
T ART/ OMEGADYNE/OD-PX01,PX02MV
OUTPUT PRESSURE TRANSDUCER
STANDARD AND METRIC MODELS
mV/V Output
0-10 to 0-30,000 psi-
Standard Models
0-700 mbar to 0-400 bar-
Metric Models
1 bar = 14.5 psi
1 kg/cm2 = 14.22 psi
1 atmosphere = 14.7 psi = 29.93
inHg = 760.2 mmHg = 1.014 bar
PX01/PXM01 Series
PX01K1-100GV, 7⁄16-20 UNF
female connection, shown
Available in Cable
or Connector
Styles
actual size.
PXM01MC1-010BARGV, G
male, shown actual size.
PRESSURE TRANSDUCERS
MILLIVOLT OUTPUT
1
⁄4
B
Standard
U 0.05% Linearity
U High Operating
Temperature: 163°C (325°F)
U 3 mV/V Output
U All Stainless Steel
Construction
U Shunt Calibration for Fast
System Setup
U Hermetically Sealed
Design for Harsh
Environments
(Not Gage Units)
U Internal Shunt for Easy
OMEGA’s PX01/PXM01 are veryhigh-accuracy, 30-millivolt output
industrial pressure transducer.
Their all stainless steel construction
and hermetically sealed case make
them suitable for harsh environments.
The PX01/PXM01 has a high
operating range, a broad
compensated range, and excellent
temperature compensation that gives
stable readings in harsh applications.
System Checks
SPECIFICATIONS
mV/V Output
Excitation: 10 Vdc (15 Vdc max)
Output: 3 mV/V ±0.25%
Input Resistance: 350 ±10 Ω
Output Resistance: 350 ±10 Ω
Zero Balance: 1.0% FS
Linearity: ±0.05% FS
Hysteresis: ±0.05% FS
Repeatability: ±0.05% FS
Shunt Calibration: Value specified
on calibration sheet
Sensing Element: 4-active-arm bridge,
using thick-film strain gages in a
hermetically sealed chamber
Operating Temp Range:
-73 to 163°C (-100 to 325°F)
Compensated Temp Range:
16 to 71°C (60 to 160°F)
Thermal Effects:
Span: 0.0015% rdg/°F
Zero: 0.0015% FS/°F
Female Connection
Wiring Code:
DIN Pin Wire Conn
2 A: GN + Out
3 B: WT - Out
GND C: BK - Ex
1 D: RD +Ex
E: BU Shunt
F: OR Shunt
11
(0.45)
PTIH-10-6P
OR EQUAL
B-62
38
(1.5)
DIA.
70 MAX
(2.75)
PX01
PX02:
19 HEX
19 HEX
(0.75)
(0.75)
Proof Pressure: 150% of rated pressure
Burst Pressure: 300% of rated pressure
Wetted Parts: 17-4 PH SS
Pressure Port: See custom
configurations
Electrical Connection: See custom
configurations
Mating Connector (Style 1):
PT06F10-6S (sold separately)
Weight: 255 g (9 oz)
Male Connection
11
(0.45)
ELECTRICAL CONNECT OR:
PTIH-10-6P OR EQUAL
PXM02:
21 HEX
16
(0.625)
PXM02: G1/4
7/16-20 TH’D
Internal TH'D
22
(0.875)
PXM01
PXM02:
18 HEX
21 HEX
(0.71)
(0.83)
Dimensions: mm (inch)
70
(2.75)
0 to 7500 psi: 7/16-20 UNF or 1/4 NPT
0 to 10,000 psi: 7/16-20 UNF or 1/4 NPT
(0.83)
PX01: 10,000 psi and higher 1/4 HP tube
PX02: 10,000 psi and higher 1/4 HP tube
38
(1.50)
DIA.
21
(0.81)
PX01 PXM01
PX02
19 HEX 18 HEX
19 HEX
(0.75) (0.71)
(0.75)
PX01
PX02:
19 HEX
19 HEX
(0.75)
(0.75)
PXM01
PXM02:
18 HEX
21 HEX
(0.71)
PXM02
18 HEX
(0.71)
Cable
style.
STANDARD
PX01C0-100GV, 1⁄4 NPT
male, shown smaller than
actual size.
STANDARD CUSTOM CONFIGURATIONS
PX01 PRESSURE ELECTRICAL RANGE
SERIES PORT CONNECTION (psi) UNITS OUTPUT OPTIONS
[ 1 ] [ 2 ] [ 3 ] [ 4 ] [ 5 ] [ 6 ]
Models < 10,000 psi:
C = 1⁄4-18 NPT-M
D = 1⁄4-18 NPT-F
J = 7⁄16-20 UNF-M
K = 7⁄16-20 UNF-F
Models ≥10,000 psi:
S = 1⁄4 HP tube fitting
0 = 3 m (10') cable (std)
1 = PTIH-10-6P connector
(or equal)
Mating connector (sold
separately), order
PT06F10-6S
6 = mini DIN connector*
(*no shunt cal)
2 and 5 psi ranges available only with amplified output.
*
Ordering Examples: PX01C1-100GV, 100 psi gage pressure transducer with 1⁄4-18 NPT male fitting, PTIH-10-6P connector and mV/V output.
PT06F10-6S, mating connector (sold separately).
002*
005*
010
015
020
025
050
075
100
150
200
250
300
500
750
1K
1.5K
2K
3K
3.5K
5K
7.5K
10K
20K
30K
A,G,S
A,G,S
A,G,S
A,G,S
A,G,S
A,G,S
A,G,S
A,G,S
A,G,S
A,G,S
A,G,S
A,G,S
A,G,S
A,G,S
A,G,S
A,G,S
A,G,S
A,G,S
A,G,S
A,G,S
A,G,S
A,G,S
A,G,S
A,G,S
A,G,S
Base Price PX01:
V = mV/V output
For additional options including extended
temperatures ranges, see omega.com
To order a custom configuration:
1. Select a pressure port
2. Select electrical connection
3. Select a pressure range
4. Select pressure units
5. Select output
6. Select options
OL = Overload stops
where available
+ additional cost
OX = Oxygen cleaning
+ additional cost
plus 2 weeks
G1⁄4 Pressure
Snubbers
PS-4G-MG = Gas
PS-4E-MG = Lt Oil
PS-4D-MG = Dense
Liquid
Mating
Connector
PT06F10-6S
METRIC CUSTOM CONFIGURATIONS
PXM01 PRESSURE ELECTRICAL RANGE
SERIES PORT CONNECTION (BAR) UNITS OUTPUT OPTIONS
[ 1 ] [ 2 ] [ 3 ] [ 4 ] [ 5 ] [ 6 ]
MA = G
MB = G1⁄8B Female
MC = G1⁄4B Male
MD = G1⁄4B Female
For additional options including extended
compensated temperatures, see omega.com
Ordering Examples: PXM01MB0-010BARSV, 10 bar sealed gage pressure transducer with G
termination.
PXM01MD1-010BARAV, 10 bar absolute pressure transducer with G
PT06F10-6S, mating connector (sold separately).
1
⁄8B Male
0 = 3 m (10') cable (std)
1 = PTIH-10-6P connector
(or equal)
6 = mini DIN connector*
(*no shunt cal)
1.0
1.6
2.5
4.0
6.0
10
16
25
40
60
100
160
250
400
A, G, S
A, G, S
A, G, S
Base Price PXM01:
V = mV/V output
A, G, S
A, G, S
A, G, S
A, G, S
A, G, S
A, G, S
A, G, S
A, G, S
A, G, S
A, G, S
A, G, S
To order a custom configuration:
1. Select a pressure port
2. Select electrical connection
3. Select a pressure range
4. Select pressure units – absolute, gage
or sealed gage
5. Select output
6. Select options
1
B female connection and 3 m (10') cable
⁄
8
1
B female connection and twist-lock connector.
⁄
4
OX = Cleaned for
OL = Overload stops
( )m = Optional cable
length in meters
oxygen service