Omega Products PHUTX601 User Manual

1
DIN pH, ORP, and Conductivity Controllers and Transmitters
4
PHUCN601
U For pH, ORP, Conductivity/Resistivity
1
4 DIN, NEMA 4X (IP66),
Polycarbonate Enclosure
U User-Friendly Design with Interior LCD Menu U Easy to Read with Large, Bright LED Display U Quick and Easy to Calibrate U Reduces Chemical Consumption
with Cycle-Timed Relays on Controller
U Two Control Relays with Third Relay
for High/Low Alarm on Controller
The PHUCN600 Series multi-parameter controller is a microprocessor-based controller capable of measuring pH, ORP, or conductivity.
There are two displays on the PHUCN600 Series: a bright LED numeric display with bar graph on the outside front panel, and a 2-line, 16-character LCD display on the inside. The LED readout on the outside panel can be seen several yards away. The distinctive, color-coded bar graph will immediately indicate if you are within the process parameters that you set (green), if the control relays are on (yellow), and if you are in alarm condition (red). This makes diagnosing pump and alarm malfunctions easy. All configuration and control functions are performed on the LCD menu on the inside front panel.
A universal mounting kit is included for surface, panel, and pipe­mount applications. The ¹⁄₄ DIN enclosure makes panel-mount cutouts and engineering simple.
Complete and versatile, the PHUTX600 is a ¹⁄₄ DIN two-wire transmitter, which can measure conductivity, pH, or ORP. The PHUTX600 comes complete with a universal mounting kit for surface, panel, and pipe-mount applications. The NEMA 4X (IP66) ¹⁄₄ DIN enclosure is perfect for stand-alone or panel-mount operation.
PHUCN600 and PHUTX600 Series are packaged in a rugged NEMA 4X (IP66) polycarbonate enclosure, making them ideally suited for heavy-duty applications such as industrial wastewater neutralization, municipal water and wastewater, pulp and paper, and process control.
When the PHUCN600 Series is powered up for the first time, it will display the meter selection screen where the meter type is selected. This meter selection screen will only be displayed when the PHUCN600 Series is powered up for the first time.
After the user selects a meter type, the PHUCN600 Series will remain set to that meter type until it is changed with the meter selection menu function in the Utilities menu.
To return the PHUCN600 Series to its factory settings, the user re-selects the current meter type from the meter selection menu function. This will override all setpoints and return all settings to the factory settings.
The PHUCN600 Series User’s menu has been divided into five main categories:
 •Calibration,usedtocalibratethePHUCN600Series
with the selected sensor
 •Utilities,usedtomanuallycontroloroverridetheoutputs
 •Setup,usedtoconfigurethePHUCN600Series’manyoptions
 •Diagnostics,usedtotroubleshootanyproblems
 •Outputs,usedtoconfigurethePHUCN600Seriesoutputs
D-X
For Sales & Service
U.S.A. and Canada
PHUCN601
ALARM
ALARM
RELAY B
RELAY A
CAL UP DOWN ENTER
TEMP
MINIMUM
INTERIOR
CLEARANCE
FOR PANEL MOUNTING
162 (6.37)
MOUNTING
BRACKET
97
(3.80)
TOP VIEW
PANEL CUTOUT
91
(3.60)
91
(3.60)
FRONT VIEW
SIDE VIEW
128
(5.05)
1
2 CONDUIT HOLES
4 PLACES
29
(1.15)
184
(7.23)
145
(5.70)
10
(0.38)
7.0
(0.26)
4 PLACES
64
(2.5)
5.0
(0.20)
4 PLACES
MOUNTING
BRACKET
117
(4.63)
MOUNTING CONFIGURATIONS
PANEL MOUNT SURFACE MOUNT HORIZONTAL
PIPE MOUNT
VERTICAL
PIPE MOUNT
3.80 (97)
omega.com
PHUCN601 Enclosure dimensions: mm (inches)
MOUNTING
97
(3.80)
TOP VIEW
PHUCN601
ALARM
RELAY B
RELAY A
ALARM
TEMP
CAL UP DOWN ENTER
FRONT VIEW
BRACKET
MINIMUM
INTERIOR CLEARANCE FOR PANEL
MOUNTING
162 (6.37)
3.80 (97)
91
(3.60)
PANEL CUTOUT
184
(7.23)
145
(5.70)
1
2 CONDUIT HOLES
SIDE VIEW
29
(1.15)
4 PLACES
128
(5.05)
(3.60)
91
PHUCN601 inside control panel shown smaller than actual size.
117
(4.63)
Mounting configurations
(0.38)
10
(0.26)
4 PLACES
(0.20)
4 PLACES
7.0
5.0
64
(2.5)
MOUNTING
BRACKET
SPECIFICATIONS FOR PHUCN600 AND PHTX600 SERIES
pH ORP Conductivity Display, PHUCN600/610 Series Front Panel: 4 x 7 segment 12.7 mm (1⁄2") LED display, 1 LED indicator online, 7 LED bar graph
Inside Panel: 2 x 16 alpha-numeric LCD display
Display, PHUTX600 2 x 16 alpha numeric LCD display Power, PHUCN600/610 Series 120 or 240 Vac 50/60 Hz (less than 12 VA) Power, PHUTX600 4 to 20 mA loop-powered 16 to 32 Vdc Measuring Range 0.01 to 14.00 pH -1999 to 1999 mV µS/cm3: 0 to 2.000,
-5 to 95°C -5 to 95°C 0 to 20.00, 0 to 200.0, 0 to 2000, (23 to 203°F) (23 to 203°F) mS/cm: 0 to 20.00, 0 to 200.0 MΩ/cm: 0 to 19.99 0 to 100°C (32 to 212°F)
Temperature Compensation Automatic or manual, Not required Automatic or manual;
-5 to 95°C (23 to 203°F) User-selectable temperature compensation slope 0.0 to 10.0%/°C. 0 to 100°C (32 to 212°F).
Temperature Units °C or °F, selectable Temperature Sensor User selectable: 300 Ω NTC thermistor, 3000 Ω NTC thermistor or Pt. 1000 RTD Calibration Modes Auto-calibration Manual calibration Dry calibration
Manual calibration Temperature Sample calibration Temperature calibration calibration Temperature calibration
Ambient Conditions -20 to 60°C (-4 to 140°F), 0 to 90% RH (non-condensing) Menu Access Front Panel Auto-calibration Manual calibration Dry calibration
Manual calibration Temperature calibration Sample calibration Temperature display Temperature calibration
Menu Access Inside Panel Full Access to all parameters of operations menu Sensor to Transmitter Distance Differential Sensor: 914 m (3000') 914 m (3000') 91 mm (300')
Combination Sensor: 3.0 m (10') 3.0 m (10') 3.0 m (10') Relay Outputs Two Control Relays: 10 A/NO, 5 A/NC @ 240 Vac or 28 Vdc.
for PHUCN600/610 Series Mode: Process control; adjustable parameters: process direction, (rising or falling) on-setpoint, off setpoint, (0 to 100% of full scale), cycle timer (on/off, 0 to 600 seconds), failsafe (on/off). One Alarm Relay: 10 A/NO, 5 A/NC @ 240 Vac or 28 Vdc. Mode: high/low alarm; adjustable parameters: low on/low off setpoint (0 to 100% of full scale, low on must be less than low off), high on/high off setpoint (0 to 100% of full scale, high on must be greater than high off).
Analog Outputs 4 to 20 mA channel 1 isolated output, range expand 0 to 100% of full scale (min segment 10% of full scale), max load 800 V