Omega Products PHUCN600 Installation Manual

User’s Guide
Shop online at
omega.com
e-mail: info@omega.com
For latest product manuals:
omegamanual.info
PHUCN600 Multi-parameter
Controller & Analyzer
OMEGAnet®Online Service Internet e-mail
omega.com info@omega.com
Servicing North America:
U.S.A.: One Omega Drive, P.O. Box 4047
ISO 9001 Certified Stamford, CT 06907-0047
TEL: (203) 359-1660 FAX: (203) 359-7700 e-mail: info@omega.com
Canada: 976 Bergar
Laval (Quebec) H7L 5A1, Canada TEL: (514) 856-6928 FAX: (514) 856-6886 e-mail: info@omega.ca
For immediate technical or application assistance:
U.S.A. and Canada: Sales Service: 1-800-826-6342 / 1-800-TC-OMEGA
Customer Service: 1-800-622-2378 / 1-800-622-BEST Engineering Service: 1-800-872-9436 / 1-800-USA-WHEN TELEX: 996404 EASYLINK: 62968934 CABLE: OMEGA
Mexico: En Espan˜ ol: (001) 203-359-7803 e-mail: espanol@omega.com
FAX: (001) 203-359-7807 info@omega.com.mx
®
®
®
Servicing Europe:
Benelux: Postbus 8034, 1180 LA Amstelveen, The Netherlands
TEL: +31 (0)20 3472121 FAX: +31 (0)20 6434643 Toll Free in Benelux: 0800 0993344 e-mail: sales@omegaeng.nl
Czech Republic: Frystatska 184, 733 01 Karviná, Czech Republic
TEL: +420 (0)59 6311899 FAX: +420 (0)59 6311114 Toll Free: 0800-1-66342 e-mail: info@omegashop.cz
France: 11, rue Jacques Cartier, 78280 Guyancourt, France
TEL: +33 (0)1 61 37 2900 FAX: +33 (0)1 30 57 5427 Toll Free in France: 0800 466 342 e-mail: sales@omega.fr
Germany/Austria: Daimlerstrasse 26, D-75392 Deckenpfronn, Germany
TEL: +49 (0)7056 9398-0 FAX: +49 (0)7056 9398-29 Toll Free in Germany: 0800 639 7678 e-mail: info@omega.de
United Kingdom: One Omega Drive, River Bend Technology Centre
ISO 9002 Certified Northbank, Irlam, Manchester
M44 5BD United Kingdom TEL: +44 (0)161 777 6611 FAX: +44 (0)161 777 6622 Toll Free in United Kingdom: 0800-488-488 e-mail: sales@omega.co.uk
It is the policy of OMEGA Engineering, Inc. to comply with all worldwide safety and EMC/EMI regulations that apply. OMEGA is constantly pursuing certification of its products to the European New Approach Directives. OMEGA will add the CE mark to every appropriate device upon certification.
The information contained in this document is believed to be correct, but OMEGA accepts no liability for any errors it contains, and reserves the right to alter specifications without notice. WARNING: These products are not designed for use in, and should not be used for, human applications.
Table of Contents
Subject Page No.
Introduction 1
Section 1 - Specifications 2
Section 2 - Installation 3 - 5
Section 3 - Electrical Connections & Setup 6 - 14
Differential Probe connection & setup 7 Combination Probe connection & setup 8 Conductivity Cell (Contacting style) connection & setup 9 Relay connections 10 Relay A & B setup 11 Alarm relay setup 12 Manual test mode & Relay override 13 4-20mA Isolated Outputs Channel 1 & Channel 2 14 Service & Fuse Replacements 15
Section 4 - Using the PHUCN600 in pH Mode 16
LCD Menu 17 - 43 LED Menu 44 - 45
Section 5 - Using the PHUCN600 in ORP Mode 46
LCD Menu 47 - 72 LED Menu 73
Section 6 - Using the PHUCN600 in Conductivity Mode 74
LCD Menu 75 - 101 LED Menu 102
Introduction
The PHUCN600 series multi-parameter controller is a microprocessor based controller capable of measuring one of the following parameters, pH, ORP or conductivity.
When shipped from the factory, the PHUCN600 series is not set to measure any one parameter. When the PHUCN600 series is powered up for the first time, it will display the meter selection screen where the meter type must be selected. (refer to section 4.6 Meter Selection)
This meter selection screen will only be displayed when the unit is powered up for the first time.
After the user selects a meter type the PHUCN600 series will remain set to that meter type until it is changed with the meter selection menu function in the Utilities menu.
To return the PHUCN600 series to its factory settings, the user must re-select the cur­rent meter type from the meter selection menu function. This will override all set-points and return all settings back to the factory settings.
The PHUCN600 User’s menu has been divided into five main categories
- Calibration, used to calibrate the PHUCN600 series with the selected sensor
- Utilities, Used to manually control or override the outputs.
- Setup, used to configure the PHUCN600 series’ many options
- Diagnostics, used to troubleshoot any problems with the PHUCN600 series
- Outputs, used to configure the PHUCN600 series’ outputs.
There are two displays on the PHUCN600 series. A bright LED numeric display with bar graph on the outside front panel, and a 2-line, 16-character LCD display on the inside. The LED readout on the outside panel can be seen several yards away. The distinc­tive,color-coded bar graph will immediately indicate if you are within the process param­eters that you set (green), if the control relays are on (yellow) and if you are in alarm condition (red). This makes diagnosing pump and alarm malfunctions easy. All configu­ration and control functions are performed on the LCD menu on the inside front panel.
A universal mounting kit is included for surface,panel and pipe-mount applications. The 1/4 DIN enclosure makes panel-mount cutouts and engineering simple.
PHUCN600 series is packaged in a rugged NEMA 4X polycarbonate enclosure making it ideally suited for heavy-duty applications such as industrial wastewater neutralization, municipal water and wastewater, pulp and paper, and process control.
Introduction Page 1
Section I - Specifications
pH ORP Conductivity Flow
Display
Power Requirements 120Vac (±10%) 50/60Hz (less than 12VA) or 240Vac (±10%) 50/60Hz (less than 12VA)
Measuring Range pH: 0.01 to 14.00 ORP: -1999 to +1999mV
Temperature Automatic or Manual
Compensation 0 to 100°C (32° to +212°F)
Temperature Unit °C or °F Not required
Temperature Sensor User selectable: 300NTC Thermistor, 3000NTC Thermistor or Pt. 1000 RTD Not required
Calibration Modes Manual Calibration
Ambient Conditions Temperature:-20°C to +60°C or -4°F to +140°F Humidity: 0 to 90% RH (non-condensing)
Menu Access
Front Panel
Menu Access Inside Panel Full Access to all parameters of operations menu
Sensor to PHUCN600 Differential Sensor: 3000 ft (914 m)
Distance Combination Sensor: 10 ft (3 m)
Relay Outputs
Analog Outputs 4 to 20mA Channel 2
Memory Back-up All user settings are retained indefinitely in memory (EEPROM)
Mechanical
Sensor Input
Invalid Entries Invalid entries cannot be stored
Manual Test Mode Process value can be simulated with arrow keys to verify correct setup of outputs
Manual Relay Override Relays can be set to on / off / auto, to verify correct wiring of auxiliary devices, or to manually adjust process
Output Hold All outputs are placed on hold when PHUCN600 is in Menu mode
Calibration Data
Auto Return User selectable auto return if unit is left in menu mode or if relays are left in manual override mode for more than 10 min.
Display Damping User can select rate at which the PHUCN600 updates display. Enables display damping of unstable process
Net Weight 2.2lbs (1kg)
Front Panel:4 x 7 segment 1/2” LED display, 1 LED indicator 0n-line,7 LED Bar Graph Inside Panel: 2 x 16 alpha-numeric LCD display
M/cm30 to 19.99 0.01
0 to 2.000 0.01 0 to 20.00 0.1
3
(Dependent on sensor)
Temp: 0 to 100°C or Temp: 0 to 100°C or
32° to +212°F 32° to +212°F (
uS/cm
mS/cm
Temp:
0 to 200.0 0.1 0 to 2000 1.0 0 to 20.00 10
3
0 to 200.0 50
0 to 100°C or 32° to +212°F
Flow: 0 to 9999 with selectable
flow rate units
Volume: 0 to 9999 with Auto Range
Flow rate units: Gallons (GP), Cubic Feet (CF), Liters (LP), Cubic Meters
CM), custom by entering factor
related to Gallons
Time units: Seconds (S), Minutes (M), Hours (H)
Automatic or Manual
Not required
User selectable temperature compensation slope 0.0 to 10.0%/°C.
Not required
0 to 100°C (32° to +212°F)
Auto-Calibration
Temperature Calibration Temperature Calibration
Auto-Calibration,Manual Calibration,Temperature Display
Manual Calibration Temperature Calibration
Manual-Calibration, Manual Calibration Temperature Display Temperature Display
Dry Calibration Sample Calibration K factor Input
Not available
300 ft (91.4 m) 2000 ft
Two Control Relays, 10A / NO, 5A / NC @ 240VAC or 28VDC. Mode: Process control, Adjustable parameters: process direction,(rising or falling) on-set-point, off set-point, (0 to 100% of full scale), cycle timer (on / off, 0 to 600 seconds), failsafe (on / off). One Alarm Relay, 10A / NO, 5A / NC @ 240VAC or 28VDC. Mode: High / Low Alarm, Adjustable parameters: Low on / Low off set-point (0 to 100% of full scale, low on must be less than low off), High On / High Off set-point (0 to 100% of full scale, High on must be greater than High off). 4 to 20mA Channel 1 Isolated Output, Range expand 0 - 100% of full scale (min segment 10% of full scale),max. load 800
Isolated Output, Range expand 0 - 100% of full scale (min segment 10% of full scale),max. load 800 Can be set to track temperature if sensor is equipped with a temperature sensor
Enclosure: NEMA 4X, 1/4 DIN, polycarbonate enclosure with four 1/2”conduit holes Mounting: Universal Mounting kit for surface, pipe and panel mount, is included Probe: -600 to +600mV Probe: -1999 to +1999mV Cell:0 to 9999 Temp. Sensor: 0 to 9999 Temp. Sensor: 0 to 9999 Temp. Sensor: 0 to 9999
Recall data from last calibration, calibration mode, 1st & 2nd accepted buffer value and probe mV output, calibration temperature, calibration slope, and probe efficiency
Recall data from last calibration, calibration buffer accepted value, and cell Recall store K factor. resistance, calibration temperature
Paddle: 0 to 2000Hz
Section I - Specifications Page 2
Section 2 - Installation
2.1 Unpacking
Save the shipping carton and packing material in case the instrument needs to be stored or returned. Inspect the instrument and packing material for shipping damage and report any problems immediately.
2.2 Location
Locate the controller/analyzer close to the sensor. The list below gives typical maximum dis­tances for various sensors. Refer to the sensor specifications for exact information.
• PHE-600/610 Series Differential pH Probe 3000 ft (914 meters)
• ORE-600/610 Series Differential ORP Probe 3000 ft (914 meters)
• CDE-600 Series Conductivity Probe 300 ft (91 meters)
2.3 Mounting
Fig 2.1 Controller dimensions
Section 2 - Installation Page 3
Section 2 - Installation
MIN. PIPE DIA.
1" PIPE (NOMINAL)
MAX. PIPE DIA.
2" PIPE (NOMINAL)
NUT 10-24 QTY.4
SCREW 10-24 X 3-1/2" QTY.4
UNIVERSAL MOUNTING BRACKET QTY.1
SCREW 1/4-20 X 6" QTY.4
UNIVERSAL MOUNTING BRACKET QTY.1
PANEL (CUSTOMER SUPPLIED)
EXTERNAL PANEL GASKET QTY.1
UNIVERSAL MOUNTING CLAMP QTY.2
NUT 1/4-20 QTY.4
SCREW 10-24 X 1/2" QTY.4
Panel Mount – The PHUCN600 series can be panel mounted to a panel using the hardware kit provided. The panel cutout dimensions are shown in fig. 2.1.
Figure 2.2 Panel Mount
Pipe Mount – The PHUCN600 series can also be mounted to a hor­izontal or vertical pipe with:
• a minimum outside diameter of 1.30” (33mm) (for example 1” CPVC pipe)
• and a maximum of
2.375” (60mm) (for example 2” CPVC pipe)
Fig. 2.3 Vertical Pipe Mount
Section 2 - Installation Page 4
Section 2 - Installation
UNIVERSAL MOUNTING BRACKET QTY.1
SCREW 10-24 X 1/2" QTY.4
HOLES ¯1/4" FOR MOUNTING SCREWS QTY.4 (CUSTOMER SUPPLIED)
MIN. PIPE DIA.
1" PIPE (NOMINAL)
MAX. PIPE DIA.
2" PIPE (NOMINAL)
SCREW 10-24 X 3-1/2" QTY.4
UNIVERSAL MOUNTING BRACKET QTY.1
NUT 10-24 QTY.4
Figure 2.4 Horizontal Pipe Mount
Surface Mount – The PHUCN600 series can be surface mounted using the hardware kit provided with the unit.
Figure 2.5 Surface Mount
Section 2 - Installation Page 5
Section 3 - Electrical Connections and Setup
AC POWER CONNECTIONS FOR 120 VAC
AC POWER CONNECTIONS FOR 240 VAC
LINE
LINE
GROUND
PHUCN610
NEUTRAL
LINE GROUND
PHUCN601
L L
P3
G N
L G
P3
3.1 Conduit Connections
The PHUCN600 series has four 1/2” conduit holes, 2 on each side of the enclosure as shown on fig. 2.1. The unit is shipped with these holes plugged with liquid tight conduit seals. These must be left in unused holes to maintain the NEMA 4X integrity. Use approved conduit hubs to connect the conduit, connect these to the conduit before connecting to the enclosure.
Wire Specification
: Size and fuse wire according to local electrical code. Maximum current not
to exceed 10A when used to power auxillary decvices powered via internal connections.
3.2 A.C. Power Connections
Caution: This instrument uses 120 or 240 50/60 Hz AC power. Opening the enclosure door exposes you to potentially hazardous line power voltage which might be pres­ent on the terminals of plug P3 and P4. Always remove line power before working in this area. If the relay contacts on P4 are powered from a seperate source from the line power on P3, be sure to disconnect that power before proceeding. The PHUCN600 flip out door contains only low voltage and is safe to handle.
The PHUCN600 series is available in two power models. The PHUCN610 is designed to operate at 240 VAC. The PHUCN601 is designed to operate at 120 VAC. To connect power to the PHUCN600, remove the terminal block plug P3 and connect the wiring as shown below.
Figure 3.1 A.C. Power Connections
Section 3 - Electrical Connections and Setup Page 6
Section 3 - Electrical Connections and Setup
BLACK (4)
YELLOW (5)
SHIELD (8)
WHITE (10)
RED (3)
GREEN (2)
P6
843
2
1
7
65
11109
1514
1312
17
16
DIFFERENTIAL pH or ORP PROBE
SELECT pH SEC. 4.6
MANUAL CALIBRATE
SELECT DIFFERENTIAL PROBE
pH PROBE SEC. 4.1 or ORP PROBE SEC. 5.1
pH SEC. 4.7 or ORP SEC. 5.6 (IF NECESSARY)
or ORP SEC. 5.5 (IF NECESSARY)
MANUAL CAL PH
7.15pH 25.0C
PROBE SELECT
METER SELECTION
RUN MODE
3.3 pH and ORP Differential Probe connections and setup
The drawing shows the connections for the 5-wire Differential probes. The cable should be run in a conduit separate from AC power wires, and via a separate conduit hole.
Note: Leave 4” to 6” slack for all wires connected to the terminals of P6. Slack required so that wires do not interfere with open­ing or closing of the front door.
If the cable of the differential probe is cut, the blue wire is not used.
CAUTION:
Always remove line power before unplugging or plugging in the P6 connector
Once connected, step through the LCD menus to select the probe in the order shown. The first two steps may be skipped if the meter is already con­figured for pH or ORP and a Differential Probe. When using a pH probe, it is important to ensure that the PHUCN600 is reading the probe temper­ature correctly for accurate temperature compen­sation. The ORP probe does not require tempera­ture compensation, although the PHUCN600 can display process temperature measured by the probe. The factory temperature calibration is usu­ally accurate enough that no adjustments are nec­essary.
Section 3 - Electrical Connections and Setup Page 7
Figure 3.2 Connections for Differential (5 wire) pH or ORP probe
Section 3 - Electrical Connections and Setup
These wires are only present with 4-wire combination probe.
JUMPER 2-4 MUST BE INSTALLED FOR COMBINATION PROBE (CUSTOMER SUPPLIED)
COAX SHIELD (4)
COAX CENTER (3)
WHITE (5)
GREEN (4)
P6
8
1
2
34 65
7
1091112 13
14 15
17
16
COMBINATION pH or ORP PROBE
SEC. 5.8
SEC. 5.5
SEC. 5.1
ORP PROBE
SELECT COMBINATION PROBE
SELECT ORP METER
MANUAL CALIBRATE
PH PROBE
MANUAL CALIBRATE
SELECT PH METER
SELECT COMBINATION PROBE
SEC. 4.1
SEC. 4.6
SEC. 4.7
SEC 4.10
MANUAL PROBE TEMPERATURE SETUP
DISABLE TEMPERATURE SEC 5.10
DISPLAY
MANUAL CAL PH
METER SELECTION
PROBE SELECT
7.15pH 25.0C
T.COMP OVERRIDE
7.15pH 25.0C
PROBE SELECT
METER SELECTION
MANUAL CAL PH
T.DISP OVERRIDE
RUN MODE
TWO WIRE PROBE WITH NO TEMP SENSORTWO WIRE PROBE WITH NO TEMP SENSOR
RUN MODE
ORPpH
3.4 pH or ORP Combination Probe connections and setup
The drawing shows the connections for the Combination style probes. The cable should be run in a conduit separate from AC power wires, and via a separate conduit hole. The cable length should not exceed 10 feet (3 meters).
Fig. 3.3 Connections for the 2 and 4 wire Combination Probe
The 2 wire version has no temperature sensor and is connected via a coaxial wire. In a pH meter, the user should set the T COMP
OVERRIDE
(Section 4.11) and adjust the temperature setting to the actual probe temperature. In an ORP meter, the user should set the
to ON (Section 5.10) to
RIDE
blank the temperature reading on the display.
menu to ON
T.DISP OVER-
4 wire version has two
The additional wires for the probe internal temperature sensor. Ensure that the
T COMP OVERRIDE or T.DISP OVERRIDE
Note
: Leave 4” to 6” slack for all wires connected to the ter­minals of P6. Slack required
is OFF.
so wires do not interfere with opening/closing of front door.
CAUTION:
Always remove line power before unplugging or plugging in the P6 connector
Once connected, step throught the LCD menus to select the probe in the order shown. The first two steps may be skipped if the meter is already configured for a Combination Probe. If a two wire pH probe is used, which has no temperature sensor, ensure that the Temp. Comp. Override is set to same temperature as the buffer before calibrating. If a two wire ORP probe is used, you can blank the Temp display with the T DISP OVERRIDE menu.
Section 3 - Electrical Connections and Setup Page 8
Section 3 - Electrical Connections and Setup
WHITE (CELL) (1)
BLACK (CELL) (4)
RED (TEMP. SENSOR) (4)
GREEN (TEMP. SENSOR) (5)
P6
8
1
2
34 65
7
1091112 13
14 15
17
16
CONTACTING
CONDUCTIVITY
CELL
SET TEMPERATURE COMPENSATION
3Kohm NTC (thermistor) 1Kohm RTD
ENSURE SENSOR IS CORRECT TYPE
SEC. 6.7
SELECT CONDUCTIVITY RANGE
SELECT COND METER
SEC. 6.13
OR
FOR PROCESS
SEC 6.9
SEC. 6.6
1000uS 25.0C
MANUAL CAL COND
TEMP COMP CURVE
TEMP SENSOR
COND RANGE
METER SELECTION
CALIBRATE WITH
RUN MODE
REFERENCE SOLUTIONS SEC 6.1
DRY CAL COND
CALIBRATE WITH FACTORY SPECIFIED CELL CONSTANT SEC 6.2
3.5 Conductivity Cell (Contacting style) connections and setup
The drawing shows the connections for the Conductivity Cells (Contacting style). The cable should be run in a conduit seperate from the AC power wires, and via a seperate conduit hole.The cell cable length should not exceed 300ft. (91 meters).
Note: Leave 4” to 6” slack for all wires connected to the terminals of P6. Slack required so that wires do not interfere with open­ing or closing of the front door.
CAUTION:
Always remove line power before unplugging or plugging in the P6 connector
Figure 3.4 Connections for Conductivity Cells
Once connnected, step through the LCD menus to select the cell in the order shown. The TEMP COMP CURVE setup default is
1.8%/deg C. This is accept­able for most process appli­cations. If your process is significantly different from this, change the setting in the TEMP COMP CURVE menu.
Section 3 - Electrical Connections and Setup Page 9
Section 3 - Electrical Connections and Setup
N.C.
N.O.
N.C.
N.O.
N.C.
N.0
GND
RELAY B
ALARMRELAY A
NEUTRAL
GROUND
LINE
LINE OUT (21)
connections
Internal
NEUTRAL OUT (22)
RELAY CONNECTIONS FOR 120VAC
PHUCN601
P3
G
L
N
21
P4
22
23
24 25
26
27
28
29
30
31
WARNING
PIN 21 must be connected to relays for power.
PIN 22 cannot be connected to the relays.
N.C.
N.O.
N.C.
N.O.
N.C.
N.0
GND
RELAY B
ALARMRELAY A
RELAY CONNECTIONS FOR 240vac
PHUCN610
LINE
GROUND
LINE
LINE OUT (21)
connections
Internal
LINE OUT (22)
G
L
L
P3
21
P4
22
23
24 25
26
27
28
29
30
31
WARNING
PIN 21 must be connected to relays for power.
PIN 22 cannot be connected to the relays.
3.7 Relay connections
The PHUCN600 series controller has three internal relays. Relays A and B are for control, the Alarm Relay can be configured for alarm functions or as an additional control relay.
The connections to the relays are shown in the drawing. Note that the AC power is internally connected to the relay terminal plug P4. This is used to provide 120V or 240V AC power for the relay contacts.
Wire Specification
: Size and fuse wire accroding to local
electrical code. Wire size not to exceed 14 AWG.
Figure 3.6 Connections for Relay A, B and Alarm
DISCONNECT POWER FROM
CONTROLLER AND LOADS WHILE
CONNECTING TO THE RELAY
OUTPUT TERMINAL PLUG.
WARNING
The contacts are rated at 10 amp N.O. and 5 amp N.C. Do not exceed this rating.When switching larger currents, use an auxillary relay switched by the controller relay to extend the controller relay life. If the relays are controlling an inductive load, use appropriate transient suppression at the load.
Section 3 - Electrical Connections and Setup Page 10
Caution:
Section 3 - Electrical Connections and Setup
3.8 RELAY A and B Setup
(LCD MENU SECTIONS - pH: 4.18 & 4.19, ORP: 5.17 & 5.18, Conductivity: 6.18 & 6.19)
Relay A & Relay B on the PHUCN600 series are SPDT dry contact relays. They are configurable to operate in response to rising or falling process val­ues. Each relay has independently adjustable on and off set-points, cycle times, and fail-safe options.
The operator would use the control relays if the device to be controlled is a simple on/off device. For example a pump, solenoid valve, fan, or an indicating light.
The control relays have 6 user configurable set­tings:
CYCLE OFF time amount of time in seconds that the relay will be de­energized, it can be set between 0 and 600 sec­onds. To disable the cycling feature set the cycle off time to 0.
VERFEED TIMER: The overfeed timer is
O designed to help safeguard against a process or instrumentation error causing one of the control relays to remain energized for extended periods of time.
When enabled, the overfeed timer will time out if the control RELAY OFF set point is not reached inside the overfeed time out. The control relays will de-energize, the alarm relay will energize and an LED will flash at the front.
: The CYCLE OFF time is the
DIRECTION
: The relay can be set to control either a rising or falling process. If for example the relay is set to control a falling process, the ON set-point must be set lower than the OFF set-point. If the relay is set to control a falling process, the PHUCN600 will not allow the RELAY OFF set point to be set lower than the RELAY ON set-point. This rule will also apply to a rising process.
Y ON set-point: This is the process value at
RELA which the relay will energize.This value can be set anywhere between 0-100% of the range.
Y OFF set-point: This is the process value at
RELA which the relay will de-energize.Depending on the direction for which the relay is configured, the RELAY OFF set-point will only be setable in a lim­ited range.
CYCLE ON time
: To obtain a tighter process con­trol, and limit over-shoot, the control relay can be set with the cycling feature. This feature, if enabled, will cause the control relay to cycle when the process is between the RELAY ON set-point and RELAY OFF set-point. The cycle on time is the amount of time in seconds that the relay will be energized. It can be set between 0 and 600 sec­onds.
AILSAFE: The FAILSAFE feature is designed to
F reverse the normal action of the control relay.
When the relay is set to FAILSAFE OFF the relay will operate as a normal relay. When the relay is de-energized the NO contacts are open and the NC contacts are closed. Thus the device connect­ed via the NO contacts will be off. When the relay becomes energized the device will be on.
When the relay is set to FAILSAFE ON, the normal action of the relay is reversed. Thus the NO con­tacts act as the NC contact and the NC act as the NO. The device connected to the NC contacts will be energized when the RELAY ON set-point is reached. The relay will be de-energized but because it is acting in reverse the device will be energized. When the RELAY OFF set-point is reached the relay will energize and the device connected to the NC contact will de-energize.
The purpose of the Fail Safe option is to have the device turned on in the event of a power interrup­tion.
The factory default for FAILSAFE is OFF.
Section 3 - Electrical Connections and Setup Page 11
Section 3 - Electrical Connections and Setup
3.9 ALARM RELAY Setup
(LCD MENU SECTIONS - pH: 4.20, ORP: 5.19, Conductivity: 6.20)
The third relay (Relay C) is used as an alarm relay. The alarm relay on the PHUCN600 is a SPDT dry contact relay.
This relay will respond to both a rising and falling process. The alarm relay will act as a low alarm (falling process) and a high alarm (rising process). Both relays will have independently adjustable on and off set-points. The ALARM ON set-points will always be set before the ALARM OFF set-points. The controller will not let the user input a value below the ALARM ON set-point. The same rule holds true for the high alarm.
The control relays have 5 user configurable set­tings:
ALARM LO value that will cause the relay to energize. This value can be set anywhere between 0-100% of the range.
ALARM LOW OFF set-point: This is the value that the process must reach in order to de-energize the alarm relay after it has dropped below the ALARM LOW ON set-point.This value must be higher than the ALARM LOW ON set-point.
ALARM HIGH ON set-point value that will cause the relay to energize. This value can be set anywhere between 0-100% of the range.
ALARM HIGH OFF set-point the process must reach in order to de-energize the alarm relay after it has increased over the ALARM HIGH ON set-point. This value must be lower than the ALARM HIGH ON set-point.
W ON set-point:This is the low process
: This is the process
: This is the value that
ALARM SET ON set-point is set higher than the factory default ALARM LOW OFF set-point, when the user advances from the ALARM LOW ON set-point to the ALARM LOW OFF set-point the shark will adjust the ALARM LOW OFF set-point to be equal to the ALARM LOW ON set-point. If the user then tries to decrease the ALARM LOW OFF set-point the controller will display the ALARM LOW ALARM setup error screen.
This screen will be displayed for 10 seconds, then return back to the setup screen that was previous­ly displayed. If the user presses the down key again the error message will be displayed again for 10 seconds. The user must accept the LOW OFF set-point, equal to, or greater than the LOW ON set-point.
The same conditions apply to the ALARM HIGH set-points. Except the ALARM HIGH OFF set­point must be lower than the ALARM HIGH ON set-point. If the user tries to increase the ALARM HIGH OFF set-point higher than the ALARM HIGH ON set-point the High Alarm setup error screen will be displayed.
ALARM RELA ALARM LOW ON set-point and the ALARM LOW OFF set-point equal to 0% of the range. It will dis­able the low alarm relay.
If the user sets the ALARM HIGH ON set-point and the ALARM HIGH OFF set-point equal to 100% of the range. It will disable the high alarm relay.
-POINT ERROR: If the ALARM LOW
Y DISABLE: If the user sets the
AILSAFE: This option can be turned on or off. It
F reverses the normal action of the relay. (see description under control relay)
Section 3 - Electrical Connections and Setup Page 12
Section 3 - Electrical Connections and Setup
3.10 MANUAL TEST MODE
(LCD MENU SECTIONS - pH: 4.4, ORP: 5.3, Conductivity: 6.4)
Once the relays are configured, the setup can be tested using Manual Test Mode to simulate process changes.
MANUAL TEST MODE is used to simulate a process reading in order to verify the correct response of the outputs. When in the MANUAL TEST MODE, the relays and outputs are no longer placed on hold as they are when in the rest of the menu. The relays and outputs will react to the simulated change in process as if the controller was in RUN MODE.
Note that when the user exits the MANUAL TEST MODE, the relays and outputs will remain in the MANUAL TEST MODE state until the user enters RUN MODE.
3.11 RELAY OVERRIDE
(LCD MENU SECTION - pH: 4.5, ORP: 5.4, Conductivity: 6.5)
Relay Override is used to manually override the state of the relays. The user is able to set the operating mode of the relay as AUTO/ON/OFF (the default and RUN MODE states are AUTO).
This feature can be used to energize or de-energize the relays to manually correct the process, or to shut down an ancillary device to perform maintenance. When in the RELAY OVERRIDE mode, the relays are no longer placed on hold as they are when in the other menus.
Note that if the RELAY AUTO RETURN is set to "ON", the controller will place all the relay settings back to AUTO 10 minutes after the controller returns to the run mode..
Section 3 - Electrical Connections and Setup Page 13
Section 3 - Electrical Connections and Setup
P6
54
P1
P4
P3
423
1
7
895 6
15
1311 12
141017
16
24 23
21 22
29 28
26 25
27
6364 6267
66
65 57 56 55
61
60 59 58
CH1 4-20mA -
CH1 4-20mA +
Primary Output
CH2 4-20mA -
CH2 4-20mA +
Secondary Output
31 30
52 53 51
3.12 4-20 mA Isolated Outputs Channel 1 and Channel 2
(LCD MENU SECTIONS - pH: 4.21 & 4.22, ORP: 5.20 & 5.21, Conductivity: 6.21 & 6.22)
The PHUCN600 Series Controller has two 4 to 20mA outputs, electri­cally isolated from each other and ground. Either output can source current into a maximum of 800 ohms.
Channel 1 (the primary output) is located on the flip out door, terminal plug P6. Channel 1 is dedicated to track the process and has fully inde­pendent and fully adjustable 4 & 20 mA output setpoints.This will enable the operator to span the output over the desired range.
Figure 3.7 Connections for the 4-20mA outputs
Channel 2 (the secondary output) is located in the enclosure terminal plug P1. Channel 2 can be selected to track the process value or tem­perature and has fully independent and adjustable 4 & 20 mA output setpoints.
Both Channel 1 and 2 can be pre­cisely trimmed through the LCD menu for precision applications.
The drawing shows the connections for both outputs.
Wire Specification
: 22 AWG 7/30,
insulation 0.010”
Note: Leave 4” to 6” slack for all wires connected to the terminals of P6. Slack required so that wires do not interfere with opening or closing of the front door.
Section 3 - Electrical Connections and Setup Page 14
Section 3 - Electrical Connections and Setup
FUSE
P6
54
P1
P4
P3
423
1
7
895 6
15
1311 121410
17
16
24 23
21 22
29 28
26 25
27
6364 6267
66
65 57 56 55
61
60 59 58
31 30
52 53 51
3.13 Service
PHUCN600 SERVICE TO BE PERFORMED BY QUALIFIED PERSONNEL ONLY.
3.14 Fuse Replacement
WARNING:
DISCONNECT LINE POWER TO THE UNIT
TO AVOID THE POSSIBILITY
OF ELECTRICAL SHOCK.
Figure 3.8 Fuse Location
1. Proceed after disconnecting line power from the instrument.
2. Open the front panel by rotating the quarter-turn fasteners, using a flat blade screwdriver, to expose the relay board.
3. The fuse, F1, is located in the middle of the relay board, directly above the three terminal connectors.
4. Remove the open fuse and replace it only with a fuse of the same type and rating. REFER TO THE FUSE RATING TABLES BELOW.
5. Close the front panel and secure using the quarter-turn fasteners.
6. Restore power to the unit.
Fuse Rating Table for
Fuse Type: Slo-Blo fuse 5 x 20mm
120 volt operation
Fuse Ratings: 250 VAC, 100mA
Fuse Rating Table for
Fuse Type: Slo-Blo fuse 5 x 20mm
Fuse Ratings: 250 VAC, 50mA
240 volt operation
Section 3 - Electrical Connections and Setup Page 15
Section 4 - Using the PHUCN600 in pH Mode
RUN LED (GREEN)
LED WILL BE ILLUMINATED WHEN THE CONTROLLER IS IN THE RUN MODE. IT WILL BE OFF WHEN THE CONTROLLER IS IN THE MENU MODE.
4 DIGIT, 7 SEGMENT LED DISPLAY
DISPLAYS PROCESS VALUE IN RUN MODE. DISPLAYS CALIBRATION DATA IN CALIBRATION MODE.
RELAY C (ALARM RELAY)
STATUS LED's (RED)
BOTTOM LED WILL ILLUMINATE
WHEN ALARM RELAY IS
ENERGIZED BECAUSE OF LOW
ALARM CONDITION.
TOP LED WILL ILLUMINATE WHEN
ALARM RELAY IS ENERGIZED
BECAUSE OF HIGH ALARM
CONDITION.
BAR GRAPH LED's (GREEN)
LINEAR INDICATOR OF PROCESS
VALUE.
(REFER TO SECTION 4.13)
1/4" TURN SCREWS
RELAY A STATUS LED (YELLOW)
LED WILL BE ILLUMINATED WHEN RELAY A IS ENERGIZED.
REMOVABLE TERMINAL BLOCK
CONNECTORS
"SNAP-ON" CONNECTORS FOR EASY
WIRING OF YOUR pH SENSOR AND
PRIMARY 4-20mA OUTPUT.
(REFER TO SECTIONS 3.3, 3.4 & 3.12)
SIMPLE THREE-BUTTON INTERFACE
FOR FAST & EASY SETUP (REFER TO SECTIONS 4.0 TO 4.21)
RELAY B STATUS LED (YELLOW)
LED WILL BE ILLUMINATED WHEN RELAY B IS ENERGIZED.
UNIT LABEL
LABEL TO INDICATE UNIT OF MEASURE THE CONTROLLER IS CONFIGURED FOR.
AUXILLARY LED's
NOT USED IN pH MODE.
CALIBRATION FROM THE
FRONT PANEL
4 PUSH BUTTONS USED TO
CALIBRATE THE CONTROLLER
FROM THE FRONT PANEL.
(REFER TO SECTIONS 4.22 & 4.23)
2 LINE, 16 CHARACTER
LCD DISPLAY
MAIN MENU INTERFACE
SCREEN
TEMPERATURE DISPLAY
WHEN THE 'UP' AND 'DOWN' ARROW KEYS ARE PRESSED SIMULTANEOUSLY, THE LED WILL DISPLAY THE CURRENT PROCESS TEMPERATURE FOR 5 SECONDS.  THIS FEATURE WILL BE DISABLED IF THE 'TEMPERATURE COMPENSATION OVERRIDE' IS TURNED ON IN THE SETUP MENU. IN THIS CASE THE LED WILL DISPLAY
---- IN PLACE OF THE
TEMPERATURE.
P6
843
2
1
7
65
11109
1514
1312
17
16
FRONT PANEL
INSIDE PANEL
NOTE
WHEN THE CONTROLLER IS TAKEN INTO MENU MODE VIA THE INSIDE LCD SCREEN, THE FRONT 7-SEGMENT LED
WILL DISPLAY
----. THE STATUS &
BAR GRAPH LED's WILL BE TURNED
OFF. THE 4 PUSH BUTTONS ON THE
FRONT WILL NOT RESPOND.
Section 4 - Using the PHUCN600 in pH Mode Page 16
pH - Menu Overview 4.0
DOWN
DOWNDOWN
DOWN
DOWN
DOWN
DOWN
DOWN
DOWN
DOWN
DOWN
DOWN
DOWN
DOWN
DOWN
DOWN
DOWN
DOWN
DOWN
DOWN
DOWNDOWN
DOWNDOWN
DOWN
DOWN
DOWNDOWN
DOWN
DOWN
DOWN
DOWN
DOWN
DOWNDOWN
DOWN
DOWN
DOWN
DOWNDOWN
SELECT DEG C OR DEG F
ENABLE TIME OUT
PROBE W/O TEMP. SENSOR
TEMPERATURE SETUP FOR
SETUP DISPLAY AND OUTPUT RESPONSE TIME
SELECT TYPE OF
4-20 mA CH 1 OUTPUT
4-20 mA CH. 2 OUTPUT
SETUP ALARM RELAY SEC 4.20
SETUP
SEC 4.21 SETUP
SEC 4.22
SETUP RELAY A SEC 4.18
SETUP RELAY B SEC 4.19
FROM MENU
DISPLAY FIRMWAVE REVISION SEC. 4.15
DISPLAY PREVIOUS CALIBRATION DATA SEC 4.16
DISPLAY UNCALIBRATED PROBE DATA SEC 4.17
SEC. 4.11
SEC 4.12
SEC 4.13 SETUP RESOLUTION
OF FRONT PANEL BAR GRAPH SEC 4.14
SEC. 4.8
SEC 4.9
SELECT TYPE OF TEMP SENSOR SEC 4.10
pH PROBE
RUN MODE
SEC. 4.3
UP
NOTE
PRESS THE AND KEYS
TOGETHER TO GO IMMEDIATELY BACK TO
RUN MODE
DOWN
RELAYS
SELECT TYPE OF METER
SEC. 4.5
SEC. 4.6 RESETS RELAY
OVERFEED TIMER SEC. 4.7
VERIFY OUTPUTS
SIMULATE PROCESS
MANUALLY OVERRIDE
SEC. 4.4
AUTO CALIBRATE
MANUAL CALIBRATE
SENSOR IN pH PROBE
CALIBRATE TEMP.
pH PROBE SEC. 4.1
pH PROBE SEC. 4.2
DIAGNOSTICS
EXIT
EXIT
ALARM RELAY
4-20mA CH1
4-20mA CH2
RELAY A
RELAY B
SENSOR INPUT
CALIBRATION DATA
FIRMWARE REV
EXIT
OUTPUTS
DISPLAY DAMPING
BAR GRAPH O/R
T.COMP OVERRIDE
EXIT
AUTO RETURN
TEMP UNIT
TEMP SENSOR
PROBE SELECT
7.15pH 25.0C
UTILITIES
SETUP
CALIBRATION
MANUAL TEST MODE
METER SELECTION
RESET OVERFD TMR
RELAY OVERRIDE
EXIT
EXIT
TEMP CALIBRATION
MANUAL CAL PH
AUTO CAL PH
pH - Menu Overview 4.0 Page 17
pH - Calibration Menu - Manual Calibrate 4.1
DOWN
DOWN
DOWN
UP DOWN
DOWN
DOWN
UP DOWN
DOWN
DOWN
DOWN
DOWN
DOWN
DOWN
UP
DOWN
DOWN
UP
DOWN
DOWN
Please wait for the controller to complete the measurement
After 5 seconds, the controller will compute the slope of the calibration, the estimated probe efficiency and the probe temperature.
Then press to store the value and complete the Manual Calibration
Then press to store the value and move to BUFFER2
Use the and keys to adjust the reading until it agrees with the actual buffer pH value
When complete, the controller will report the measured value
The controller will read the pH value, averaging a number of results to get a stable calibration value.
Place the probe in the second buffer. Be sure to clean and rinse the Probe first with D.I. water and then insert it in the 4.00 buffer.
Press
Use the and keys to adjust the reading until it agrees with the actual buffer pH value
When complete, the controller will report the measured value
Please wait for the controller to complete the measurement
The controller will read the pH value, averaging a number of results to get a stable calibration value.
Place the probe in the first buffer solution, be sure to clean and rinse the Probe first with D.I. water and then insert it in the 7.00 buffer.
Press
If the calibration is OK, use the key to move the cursor over the Y text and press the down key.
If the calibration did not appear to be correct, press the key which will return back to the Manual Cal menu.
RUN MODE
UP
NOTE
PRESS THE AND KEYS
TOGETHER TO GO IMMEDIATELY BACK TO
RUN MODE
DOWN
Press to store the calibration data and return back to the Manual Cal menu so the user can select another function.
Then press to move the cursor to the RH position
Then press to move the cursor to the RH position
STORE? Y N STORE? Y N
IF BUFFER1 READYCALIBRATION MANUAL CAL PH
MANUAL CAL PH
MANUAL CAL PH BUFFER2 4.00 >
SLOPE 61.22MV/PH EFF 95% 24.8C
BUFFER1 7.00 >
MANUAL CAL PH BUFFER2 4.4 0 >
MANUAL CAL PH BUFFER2 4.0 0 >
BUFFER2 WAIT...
RUNNING MANU CAL
IF BUFFER2 READY PRESS 'DOWN'
MANUAL CAL PH
MANUAL CAL PH BUFFER1 7.0 0 >
RUNNING MANU CAL BUFFER1 WAIT...
PRESS 'DOWN'
BUFFER1 7.3 5 >
MANUAL CAL PH
7.15pH 25.0C
MANUAL CAL PH
This example shows a MANUAL calibration with buffers of 7pH & 4pH.
When performing a manual calibration, any two known buffer solutions can be used.
pH - Calibration Menu - Manual Calibrate 4.1 Page 18
pH - Calibration Menu - Auto Calibrate 4.2
DOWN
DOWN DOWN
DOWN
DOWN
DOWN
DOWN
DOWN
DOWN
DOWN
DOWN
DOWN
If an errors occurs, the controller will indicate a "BUFFER ERROR" alarm. This could be caused by
- using a defective buffer
- incorrect probe wiring
- defective probe Press to return to the Auto cal menu, try to
correct the problem and calibrate again, or go to the diagnotics menus to test the probe see Section 4.17
If an errors occurs, the controller will indicate a "BUFFER ERROR" alarm. This could be caused by
- using the same buffer twice
- using a defective buffer
- incorrect probe wiring
- defective probe Press to return to the Auto cal menu, try to
correct the problem and calibrate again, or go to the diagnotics menus to test the probe
If the calibration did not appear to be correct, press the key which will return back to the Auto cal menu.
UP
NOTE
PRESS THE AND KEYS
TOGETHER TO GO IMMEDIATELY BACK TO
RUN MODE
DOWN
This will store the calibration data and return back to the Auto cal setup so the user can select another function.
If the calibration is OK, use the key to move the cursor over the Y text and press the down key.
After 5 seconds, the controller will compute the slope of the calibration, the estimated probe efficiency and the probe temperature.
When complete, the controller will report the measured value
When complete, the controller will report the measured value
Press the key to move to Buffer 2
Please wait for the controller to complete the measurement
The controller will read the pH value, averaging a number of results to get a stable calibration value.
Place the probe in the second buffer solution.Be sure to clean and rinse the Probe first with D.I. water and then insert it in the 4.00 buffer.
Press the key
Please wait for the controller to complete the measurement
The controller will read the pH value, averaging a number of results to get a stable calibration value.
Place the probe in the first buffer solution, be sure to clean and rinse the Probe first with D.I. water and then insert it in the 7.00 buffer.
Press
RUN MODE
AUTO CAL PH
STORE? Y N
STORE? Y N
AUTO CAL PH
AUTO CAL PH
EFF 95% 24.8C
SLOPE 61.22MV/PH
BUFFER2 4.00 >
AUTO CAL PH
AUTO CAL PH
BUFFER ERROR CONTINUE >
RUNNING AUTO CAL BUFFER2 WAIT...
BUFFER1 7.00 >
PRESS 'DOWN'
IF BUFFER2 READY
BUFFER ERROR CONTINUE >
PRESS 'DOWN'
IF BUFFER1 READY
RUNNING AUTO CAL BUFFER1 WAIT...
CALIBRATION
7.15pH 25.0C
MANUAL CAL PH
This example shows an AUTO calibration with buffers of 7pH & 4pH.
When performing an auto calibration, any two standard buffers of 4.00, 7.00 or 10.00 pH can be used. The buffers will be automatically recognized if the probe measures them within +/- 1 pH of their nominal value.
pH - Calibration Menu - Auto Calibrate 4.2 Page 19
pH - Calibration Menu - Temperature Calibration 4.3
DOWN
DOWN DOWN
DOWNUP
DOWN
DOWN
DOWN
UP
DOWN
DOWN
DOWN
If the Temperature Compensation Override is set to ON (see section 4.11), the controller cannot calibrate the temperature sensor. This display will appear to alert the user to the condition.
If the calibration did not appear to be correct, press the key which will return back to the Temp Calibration menu.
UP
NOTE
PRESS THE AND KEYS
TOGETHER TO GO IMMEDIATELY BACK TO
RUN MODE
DOWN
This will store the calibration data and return back to the Temp Calibration setup so the user can select another function.
Use the and to adjust the reading until it agrees with the known temperature of the probe
Press the key to accept the reading and move to the store function
"T FACTOR SHIFTED" will be displayed for 5 seconds to indicate that the temperature sensor has been calibrated. The controller will then move to the store function.
The controller displays the current probe temperature. If this incorrect, the controller can be adjusted to compensate.
Press the key once which will move the cursor back to the RH side of temperature display.
Press the key once which will move the cursor over the least sign. digit of the temperature display.
RUN MODE
AUTO CAL PH
STORE? Y N
TEMP CALIBRATION
STORE? Y N
TEMP CALIBRATION
TEMP CALIBRATION
25.0C >
T FACTOR SHIFTED
25. 0 C >
TEMP CALIBRATION
TEMP CALIBRATION
TEMP CALIBRATION
TEMP CALIBRATION
22. 4 C >
22.4C >
TEMP CALIBRATION TEMP O/R ON >
7.15pH 25.0C
CALIBRATION MANUAL CAL PH
pH - Calibration Menu - Temperature Calibration 4.3 Page 20
DOWN
DOWN
DOWN
UP DOWN
DOWN
UP
DOWN
Manual Test Mode is used to simulate a process reading in order to verify the correct response of the outputs. When in the Manual Test Mode, the outputs are no longer placed on hold as they are when in the rest of the menu.
NOTE: When the user exits the Manual Test Mode, the relays and 4-20mA outputs will remain in the Test Mode state until the operator enters the run mode. The relays and
outputs will then revert back to the previous On-line state.
The CH1 4-20 mA output will track the change in process
When a relay is energized, the characters a,b,c will change to upper case A,B,C
RUN MODE
UP
NOTE
PRESS THE AND KEYS
TOGETHER TO GO IMMEDIATELY BACK TO
RUN MODE
DOWN
As the value is changed, the state of the relays will change depending on their settings in the OUTPUTS menu. a/A - state of Relay A b/B - state of Relay B c/C - state of Alarm relay
Press the key once to move the cursor to the RH side of the display
As well, Channel 1 4-20 ma output will also follow the process value change. The actual change will depend on how the CH1 output was scaled in the OUTPUTS menus. (See Section 4.21)
Press the key to return to the MANUAL TEST MODE menu
Use the and keys to change the simulated process value.
Press the key once which will move the cursor over the least digit of the simulated process value.
TEST 7.0 0 >
TEST 7.00 >
7.15pH 25.0C
CALIBRATION
UTILITIES
a b c 19.0mA
a b c 19.0mA
TEST 13.1 5 >
TEST 13.15 >
MANUAL TEST MODE
a b c 12.0mA
a b c 12.0mA
a B c 19.0mA
TEST 13.1 5 >
DOWN
pH - Utilities Menu - Manual Test Mode 4.4
pH - Utilities Menu - Manual Test Mode 4.4 Page 21
pH - Utilities Menu - Relay Override 4.5
UP
UP
UP
To Alarm Relay (Relay C) Override
UP
UP
DOWN
DOWN
UP
DOWN
DOWN
DOWN
DOWN
DOWN
DOWN
DOWN
DOWN
DOWN
DOWN
DOWN
DOWN
DOWN
DOWN
DOWN
DOWN
DOWN
DOWN
DOWN
DOWN
DOWN
Relay Override is used to manually override the state of the relays, the user is able to set the operating mode of the relay as AUTO/ON/OFF (the default and run mode state are AUTO). This feature can be used to turn the relays ON or OFF to manually correct the process, or to shut down an ancillary device to perform maintenance. When in the Relay Override mode, the outputs are no longer placed on hold as they are when in the other menus.
If you wish to Override Relay B, Press the key once which will move the cursor over to the AUTO/ON/OFF text
If you wish to Override Relay A, Press the key once which will move the cursor over to the AUTO/ON/OFF text
If you do not wish to Override Relay B, then use the key to the Alarm Relay
If you do not wish to Override Relay A, then use the key to move to Relay B
RUN MODE
UP
NOTE
PRESS THE AND KEYS
TOGETHER TO GO IMMEDIATELY BACK TO
RUN MODE
DOWN
If you wish to accept the Override setting, Press the key once to move the cursor to the RH of the display
Use the key to change the Override state from AUTO - controlled by process settings ON - relay forced on OFF- relay forced off The relay will react immediately.
Then use the key to move to the Override setting for Relay B
CAUTION:
Note that if the Relay Auto Return Menu is set to "ON", the controller will place all these settings back to AUTO 10 minutes after the controller returns to the run mode.
Use the key to change the Override state from AUTO - controlled by process settings ON - relay forced on OFF- relay forced off The relay will react immediately.
If you wish to accept the Override setting, Press the key once to move the cursor to the RH of the display
To use the Relay Override function, press the key
Then use the key to move to the Override setting for Relay A
RELAY OVERRIDE RELAY B A UTO >
RELAY OVERRIDE RELAY B O N >
RELAY OVERRIDE RELAY B O FF >
RELAY A O FF >
RELAY OVERRIDE RELAY A O N >
RELAY OVERRIDE
RELAY OVERRIDE
RELAY A A UTO >
RELAY OVERRIDE
MANUAL TEST MODE
RELAY OVERRIDE RELAY B AUTO >
RELAY A AUTO >
RELAY OVERRIDE
CALIBRATION
7.15pH 25.0C
UTILITIES
RELAY B ON >
RELAY B AUTO >
RELAY B OFF >
RELAY OVERRIDE
RELAY OVERRIDE
RELAY OVERRIDE
RELAY A ON >
RELAY A AUTO >
RELAY A OFF >
RELAY OVERRIDE
RELAY OVERRIDE
RELAY OVERRIDE
pH - Utilities Menu - Relay Override 4.5 Page 22
pH - Utilities Menu - Relay Override 4.5
DOWN
DOWN
DOWN
DOWN
DOWN
DOWN
UP DOWN
DOWN
DOWNUP
UP
DOWN
DOWN
DOWN
DOWN
UP DOWN
Use the or key to change the Override state from AUTO - controlled by process settings ON - relay forced on OFF- relay forced off The relay will react immediately.
If you do not wish to Override Relay C, then use the key to the Store function
With "N" highlighted, pressing the key will NOT store the Override states of the relays, but they will revert to the state they were in before the menu was entered. Note that, if any of the relays were in a specific Override state before the menu was entered (i.e. not in AUTO), they will revert to that state again unless the changes are STORED.
TO RELAY OVERRIDE MENU
NOT STORED
FROM RELAY B OVERRIDE MENU
Then press the key to store the Override states of all the relays and return back to the Override menu
STORED
UP
NOTE
PRESS THE AND KEYS
TOGETHER TO GO IMMEDIATELY BACK TO
RUN MODE
DOWN
If you wish to accept the Override setting, Press the key once to move the cursor to the RH of the display
If you wish to store the state of the Relay Overrides, press the key to highlight the Y character
If you wish to Override Relay C, Press the key once which will move the
Then use the key to move to the Override setting for Relay C
NOTE Relay C is the Alarm Relay
STORE? Y N
RELAY C AUTO >
RELAY OVERRIDE
RELAY OVERRIDE
STORE? Y N
RELAY OVERRIDE
RELAY OVERRIDE
RELAY OVERRIDE
RELAY C O N >
RELAY C O FF >
RELAY OVERRIDE RELAY C A UTO >
RELAY OVERRIDE
RELAY OVERRIDE RELAY C ON >
RELAY C OFF >
RELAY OVERRIDE RELAY C AUTO >
pH - Utilities Menu - Relay Override 4.5 Page 23
pH - Utilities Menu - Meter Selection 4.6
DOWNDOWN
DOWN
DOWN
DOWN
DOWN
DOWN
DOWN
DOWN
DOWN
DOWN
DOWN
DOWN
DOWN DOWN
DOWN
If you do not wish to initialize to this type of meter after all, then press key to return back to the Meter Selection menu
To reset the controller back to all the factory default values, reselect the current meter type
Then press The controller will initialize itself for a specific meter
If you do wish to initialize to this type of meter, press the key to highlight the Y character
UP
NOTE
PRESS THE AND KEYS
TOGETHER TO GO IMMEDIATELY BACK TO
RUN MODE
DOWN
Press the key to Exit and return to the Utilities menu
RUN MODE
Press the key to initialize the controller as a Flow meter
Press the key to initialize the controller as a Conductivity meter
The Meter Selection menu is used to select the meter type that the controller is configured for, either pH, ORP or Conductivity. Once selected, the controller will initialize itself for the selected meter's functionality and move to run mode.
Press the key to initialize the controller as an ORP meter
Press the key to initialize the controller as a pH meter.
INITIALIZE? Y N
INITIALIZE? Y N
INITIALIZE? Y N
INITIALIZE? Y N
INITIALIZE? Y N
PH
PH
E XIT
MANUAL TEST MODE
7.15pH 25.0C
UTILITIES
CALIBRATION
RELAY OVERRIDE
METER SELECTION
C ONDUCTIVITY
FLOW
EXIT
F LOW
CONDUCTIVITY
O RP
CONDUCTIVITY
FLOW
ORP
P H ORP
PH
pH - Utilities Menu - Meter Selection 4.6 Page 24
pH - Utilities Menu - Overfeed Timer Reset 4.7
DOWN
DOWN
DOWN
DOWN
DOWN
DOWNDOWN
DOWN
If you wish to reset the overfeed timer, then press the key to proceed and return back to the Utilities menu.
If you do not wish to reset the overfeed timer after all, press the key to highlight the N character and press the key to return back to the Utilities menu.
DOWN
UP
NOTE
PRESS THE AND KEYS
TOGETHER TO GO IMMEDIATELY BACK TO
RUN MODE
DOWN
RUN MODE
Press the key to choose between Y or N.
If the control relay overfeed timer has been enabled, the relay will “time out” after the specified overfeed time. When the relay times out, it must be manually reset. The time out will be signaled by the time out LED flashing on the front panel.
The reset overfeed timer will reset the timers for both relay A & relay B at the same time.
Y N
MANUAL TEST MODE
7.15pH 25.0C
UTILITIES
CALIBRATION
RELAY OVERRIDE
METER SELECTION
RESET OVERFD TMR
RESET OVERFD TMR
Y N
RESET OVERFD TMR
pH - Utilities Menu - Overfeed Timer Reset 4.7 Page 25
pH - Setup Menu - Probe Select 4.8
DOWN
DOWN
DOWN
DOWN
DOWN
DOWN
UP DOWN
DOWN
DOWN
DOWN
UP
DOWN
Then press the key to store the selection and return to the PROBE SELECT Menu.
With "N" selected, pressing the key will NOT store the selection, but simply return to the PROBE SELECT Menu. This function is useful if you wish to view the current selection without making any changes.
Not stored
Stored
Or press the key to highlight the Y character.
Note:
If using the 2 wire Combination Probe (which doesn't have a temperature sensor), T COMP OVERRIDE must be set to ON, and the actual probe temperature set through the T COMP OVERRIDE menu. (see sec. 4.12)
PROBE SELECT will allow the user to select whether the probe is a 2 or 4 wire combination probe, or a 5 wire differential probe.
RUN MODE
UP
NOTE
PRESS THE AND KEYS
TOGETHER TO GO IMMEDIATELY BACK TO
RUN MODE
DOWN
Use the or keys to scroll through the probe types available. In this case, the user can select
- the 2 or 4 wire combination probe
- the 5 wire differential probe
Press the key to accept the setting and move to the store function
Press the key once which will move the cursor back to the RH side of the display.
Once the correct probe type is selected, move to the store function to save the selection
Press the key once which will move the cursor over the first character of the probe type.
STORE? Y N STORE? Y N
PROBE SELECT PROBE SELECT
7.15pH 25.0C
UTILITIES
SETUP
CALIBRATION
PROBE SELECT COMBINATION >
C OMBINATION
PROBE SELECT
D IFFERENTIAL
PROBE SELECT
PROBE SELECT PROBE SELECT
DIFFERENTIAL >
Note: Refer to Appendix A - Probe Configuration Table
pH - Setup Menu - Probe Select 4.8 Page 26
pH - Setup Menu -Temp Unit 4.9
DOWN
DOWN
DOWN
DOWN
DOWN
DOWN
DOWN
DOWN
DOWN
UP DOWN
DOWN
UP
DOWN
With "N" selected, pressing the key will NOT store the selection, but simply return to the TEMP UNIT Selection Menu. This function is useful if you wish to view the current selection without making any changes.
RUN MODE
Or press the key to highlight
UP
NOTE
PRESS THE AND KEYS
TOGETHER TO GO IMMEDIATELY BACK TO
RUN MODE
DOWN
Then press the key to store the selection and return to the TEMP UNIT Selection Menu.
Not stored
Stored
the Y character.
Use the or to select C or F
Press the key to accept the change and move to the store function
Press the key once which will move the cursor over the unit type, C or F.
Press the key once which will move the cursor back to the RH side of the display.
TEMP UNIT allows the user to select either Degrees Centigrade or Fahrenheit units for display
7.15pH 25.0C
CALIBRATION
UTILITIES
SETUP
TEMP UNIT
PROBE SELECT
STORE? Y N
STORE? Y N
TEMP UNIT TEMP UNIT
DEGREE C >
DEGREE F >
TEMP UNIT
DEGREE F >
DEGREE C >
TEMP UNIT
TEMP UNIT
TEMP UNIT
pH - Setup Menu - Temp Unit 4.9 Page 27
Loading...
+ 77 hidden pages