Omega Products PHTX-45 User Manual

2-Wire Isolated pH/ORP Transmitter Systems
PHTX-45
$
1175
Meter and Electrode
PEEK Sensor Body ConstructionDual-Glass Style SensorReplaceable Sensor Saltbridge Electrode Breakage Diagnostic Universal Mounting ConfigurationsMicroprocessor Based SystemLarge Dual Display FormatLoop Powered, Fully Isolated
Sensor Features
Sensor housings are constructed of PEEK, a high performance thermoplastic that provides outstanding mechanical strength and chemical resistance. Multiple sealing materials are used to preserve sensor integrity over a wide range of applications.
A large volume, dual junction saltbridge is used to maximize the in-service time of the sensor. The annular junction provides a large surface area to minimize the chance of fouling. Large electrolyte volume and dual reference junction minimizes contamination of the reference solution. The replaceable saltbridge allows for easy sensor regeneration.
The reference element of this sensor is a second pH electrode immersed in a reference buffer solution. This glass reference system allows the sensor to be used in applications that poison conventional pH sensors.
An integral preamplifier is encapsulated in the body of the sensor. This creates a low impedance signal output which ensures stable readings in harsh environments, and maximize the distance between sensor and transmitter. Sensor diagnostics are used to alarm the user in the event of electrode breakage, loss of sensor seal integrity, or integral temperature sensor failure.
Sensor electrodes can be user-specified to ensure measurement reliability and maximum sensor lifetime. The type of glass used in the pH electrode can be selected for optimal performance. The metal electrode used for ORP measurements can be platinum or gold, depending on chemical makeup of the process solution.
PHTX-45, transmitter, $625 shown with PHE-45P electrode, $550, shown smaller than actual size.
MOST POPULAR MODELS HIGHLIGHTED!
To Order (Specify Model Number)
Model No. Price Description PHTX-45 $625 pH transmitter PHE-45P 550 pH electrode ORTX-45 625 ORP transmitter ORE-45P 570 ORP sensor, -1000 mV to 2000 mV U24Y101 169 24 Vdc power supply PHTX-45-SMH 95 Submersion mounting hardware, 1.8 m (6') PHTX-45-RK 75 pH/ORP sensor regeneration kit: 1 saltbridge plus
1, 125 mL bottle of reference cell solution, pH 7.00 (for Models PHE-45P and ORTX-45E sensors only)
PHA-4 8 pH 4.01 buffer solution, 475 ml bottle PHA-7 8 pH 7.01 buffer solution, 475 ml bottle PHA-10 8 pH 10.01 buffer solution, 475 ml bottle
Comes complete with operator’s manual. Ordering Example: PHTX-45, pH transmitter, PHE-45P, electrode and PHA-4, buffer solution, $625 + 550 + 8 = $1183.
Recommended Reference Book: Advanced pH Measurement and Control
FW-332, $89. Visit omega.com/bobi for Additional Books
,
D-X
For Sales & Service
U.S.A. and Canada
Transmitter Features
omega.comomega.com
®
The microprocessor-based transmitter is loop-powered and fully isolated for high service reliability. The transmitter includes devices to protect the system from power surge and brownout events.
The large, high contrast, super-twist display provides excellent readability over a wide operating temperature range, even in low light conditions. The main display line consists of large, segmented characters with measurement units. The secondary display line utilizes easily readable dot matrix characters for clear display of calibration and diagnostic messages. Two of four measured parameters may be displayed simultaneously.
Four-button programming provides intuitive navigation through the menu driven user interface. The 4 to 20 mA transmitter output can be configured to represent any portion of the measurement range. Output HOLD, ALARM and SIMULATION features provide the user with complete control of the system output under any condition.
Diagnostic messages provide a clear description of system condition, which eliminates confusing error codes that must be looked up in the operator’s manual.
The flexible two-point and sample calibration options include auto-buffer recognition from thirteen built-in buffer tables. Manual override of the automatic buffer values allows the user to customize calibration values. To ensure high accuracy, all calibration methods include stability monitors that check temperature and main parameter stability before accepting data.
Specifications
PHE-45P and ORE-45P Sensor Specifications
Sensor Cable: 3 m (10') Measuring Range:
PHE-45P: 0 to 14.00 pH
ORE-45P: -1000 to 2000 mV Sensitivity: 0.002 pH, 0.2 mV (ORP) Stability: 0.02 pH or 2 mV per
24 hours, non-cumulative
Wetted Materials:
PHE-45P: PEEK, ceramic, titanium,
glass, FKM, EDPM (316 stainless
steel with 316SS body option)
ORE-45P: PEEK, ceramic, titanium,
glass, FKM, EDPM, platinum or gold
Temperature Compensation: Pt1000 Sensor Cable:
6 Conductor plus 2 shields
Temperature Range:
-5 to 95°C (23 to 203°F)
PHTX-45 in Line Application Solution
ORP
ORTX-45, $625, shown with ORE-45P electrode reading ORP, $570.
Pressure Range: 0 to 100 psig Maximum Flow Rate:
3 m (10') per second
Sensor to Transmitter Distance:
914 m (3000')
Mounting Options: 1 NPT convertible Weight/Shipping Weight:
0.45 kg (1 lb)
PHTX-45 Transmitter Specifications
Enclosure: NEMA 4X, IP65,
polycarbonate, stainless steel hardware, weatherproof and corrosion resistant, 112 H x 112 W x 89 mm D (4.4 x 4.4 x 3.5") Mounting Options: Wall, panel, pipe, Din rail, integral-sensor Conduit Openings: 2-PG9 openings, 1 to 1 NPT center opening, cord grips and plug included
Weight/Shipping Weight:
0.45 kg (1 lb) Display: Large, high-contrast, super-Twist (STN) LCD; 4-digit main display with 19.1 mm (0.75") seven-segment character, 12-digit secondary display,
7.6 mm (0.3") 5 x 7 dot matrix character Keypad: 4-key membrane type with tactile feedback, polycarbonate with UV coating, integral EMI/static shield and conductivity coated window
Ambient Temperature:
Service: -20 to 60°C (-4 to 140°F) Storage: -30 to 70°C (-22 to 158°F)
Ambient Humidity:
0 to 95%, non-condensing Location: Designed for hazardous and non-hazardous areas
pH
PHTX-45, $625 shown with PHE-45P electrode reading pH, $550.
CDTX-45, $650,
shown with CDE-45P
electrode reading
conductivity, $450. Visit
omega.com/cdtx45
EMI/RFI Influence:
Designed to EN61326-1
Voltage Range:
16 to 35 Vdc (two-wire device)
Output Isolation:
600 V galvanic isolation
Transmitter Cable Type:
Belden twisted-pair, shielded Filter: Adjustable 1 to 99 seconds additional damping to 90% step input Temperature Input: Selectable Pt1000 or Pt100, automatic compensation
µS
PHE-45P Performance Specifications
Displayed Parameters: Main Input: 0 to 14.00 pH; Sensor Voltage: ±500 mV; Loop Current: 4 to 20 mA; Sensor Temperature:
-10 to 110°C (14 to 230° F)
Main Parameter Range: 0 to 14.00 pH Input Impedance: Greater than 1013 Repeatability: 0.1% of or better Sensitivity: 0.05% of span Non-Linearity: 0.1% of span Stability: 0.1% of span
per 24 hours, non-cumulative Warm-Up Time: 4 seconds to rated performance
Supply Voltage Effects: ±0.05% span Transmitter Response Time:
4 seconds to 90% of step input at lowest setting
Temperature Drift:
Span or zero, 0.02% of span/°C
Sensor to Transmitter Distance:
914 m (3000') w/preamp,
9.1 m (30') w/o preamp
Order Online
Over 100,000 Products Available!
D-X
Loading...
+ 1 hidden pages