Omega Products PHH-SD1 User Manual

omega.com
PHH-SD1
U LCD with Green Light Backlighting U 0 to 14 pH or ±1999 mV for ORP Measurement U Real Time SD Memory Card Data Logger
(SD Card Included)
U Automatic Buffer Calibration (Buffers Sold Separately) U RS232/USB Interface (Optional Cable) U Automatic Temperature Compensation (Optional Probe)
The OMEGA® PHH-SD1 is innovative and easy to operate. What sets the PHH-SD1 apart from other pH/ORP meters is that it incorporates a real-time SD card data logger. The PHH-SD1 measures pH or mV (ORP) using standard electrodes with BNC connectors. Download data from the SD card directly into an Excel spreadsheet. This allows the user to do further data or graphical analysis without special software. The PHH-SD1 has a wide variety of applications from water conditioning, aquariums, beverage manufacturing and fish hatcheries to food processing, photography, paper and plating industries, educational and other laboratories.
ELECTRONICS
Specifications
Display: 52 x 38 mm (2.04 x 1.5") LCD
with switchable green backlight
Measurement: 0 to 14 pH or -1999 mV to 1999 mV (ORP) Sampling Time:
Auto: 1 second to 8 hrs 59 min 59 sec Manual: Push the data logger button once will save data one time
Memory Card: SD memory card Temperature Compensation: Automatic (with optional
temperature probe) or manual
Display Update: At 1 second Data Output (RS232/USB): With optional cable/software Operating Temperature: 0 to 50°C (32 to 122°F) Operating Humidity: Less than 85% RH Power: 6 “AAA” alkaline batteries (included),
or 9V power adaptor (optional) Current Draw: 14 mA DC (without SD card or backlight), 49 mA (all functions)
Weight: 489 g (1.08 lb) Dimensions: 177 H x 68 W x 45 mm D (6.96 x 2.67 x 1.77") Input Impedance: 10^12 Ω Manual Temperature Compensation: Pushbutton on front panel Automatic: With the optional temperature measurement
probe from 0 to 65°C (32 to 149°F) Calibration: pH4, pH7, and pH10, 3 points calibration provides the best linearity and accuracy
Measurement Range Resolution Accuracy
pH* 0 to 14 pH 0.01 pH ± (0.02 pH + 2 d) mV (ORP) 0 to 1999 mV 1 mV ± (0.5% + 2 d)
* pH accuracy is based on calibrated meter only.
PHH-SD1, pH/ORP meter with real-time data logger
PHH-SD1 includes manual, carrying case, batteries, and 2 GB SD card.
(electrode sold separately).
To Order Visit omega.com/phh-sd1 for Pricing and Details
Model No. Description
PHH-SD1 Data logging pH/ORP meter with SD card PHE-1304 Standard pH laboratory electrode ORE-1311 Standard ORP laboratory electrode SW-U101-WIN Software for data logging from meters, with USB
HC-SD Replacement hard case SC-SD Soft case ADAPTER-SD AC adaptor USB-SD Spare USB cable (SW-U101-WIN
RS232-SD Spare RS232 cable (SW-U101-WIN
2GB-SD Spare 2 GB SD memory card PHA-4 Calibration buffer solution pH 4 PHA-7 Calibration buffer solution pH 7 PHA-10 Calibration buffer solution pH 10 PHAT-222 Temperature probe
Comes complete with operator’s manual, hard carrying case, 2 GB SD card, and 6 “AAA” batteries. Electrodes are sold separately.
Ordering Example: PHH-SD1, data logging pH/ORP meter with 2 GB SD card, and PHE-1304 pH electrode.
and RS232 cables
software required)
software required)
Order Online
Over 100,000 Products Available!
®
To download information and to order pH Field and Lab Instrument products online, visit omega.com