United Kingdom:One Omega Drive, River Bend Technology Centre
ISO 9002 CertifiedNorthbank, Irlam, Manchester
M44 5EX United Kingdom
TEL: +44 (0)161 777 6611FAX: +44 (0)161 777 6622
Toll Free in United Kingdom: 0800-488-488
e-mail: sales@omega.co.uk
It is the policy of OMEGA to comply with all worldwide safety and EMC/EMI regulations that
apply. OMEGA is constantly pursuing certification of its products to the European New Approach
Directives. OMEGA will add the CE mark to every appropriate device upon certification.
The information contained in this document is believed to be correct, but OMEGA Engineering, Inc. accepts
no liability for any errors it contains, and reserves the right to alter specifications without notice.
WARNING: These products are not designed for use in, and should not be used for, patient-connected applications.
• Remove power to unit before wiring input and output connections.
• Follow instructions carefully to avoid personal injury.
ENGLISH
Contents
1. Power Connections
2. Compatible Electrode/Preamp Wiring
3. 4 - 20 mA Current Output Connections
4. Relay Connections
5. Relay Operation
Remove terminal blocks for easy wiring
1.
Std. Sensor
U
L
1 2
1. Power Connections
CAUTION!
Never connect 110 VAC or 220 VAC to rear power
terminals. High voltage AC will damage instrument and
void warranty.
= Double Insulated
= DC or AC power
1 2 3 4 5 6
Freq. IN
LISTED
77CJ
Freq. IN
Sen. Pwr.
Open Collector
Sensor
12-24 V
12-24 V
10 W
+
10 W
+
+
Iso. Gnd
output
-
Gnd
PLS
-
1 2 3 4
2
1
Gnd
+
Total
AUX
reset
output
PHCN-5700
Terminals
-
+
Flow
Patent No. D376,328
External power
supply
12 - 24 VDC
OR
12 - 24 VAC
6. Menu Functions
7.Parts and Accessories
8.Specifications
9.Quick Reference Menu Parameters
10. Troubleshooting
11. Maintenance
2.
1 2
3.
Std. Sensor
U
L
1 2
LISTED
77CJ
1 2 3 4 5 6
Freq. IN
Freq. IN
Sen. Pwr.
Open Collector
Sensor
12-24 V
10 W
+
2. Compatible Electrode/Preamp Wiring
Silver (earth ground)
Brown
Blue
Green
White
Red
Black
+
V -
Black wire
4-20
-
mA
Brown wire
Iso. Gnd
mV Input
Blue wire
T +
Green wire
T -
White wire
V +
Red Wire
1 2 3 4
+
Gnd
+
Gnd
AUX
PLS
Iso. Gnd
output
output
-
pH/ORP
PHCN-5700
Terminals
2
1
Total
Flow
reset
Patent No. D376,328
Preamp:
OMEGA PH-2720-PA
ORP Electrodes:
OMEGA ORE-2715
OMEGA ORE-2717
Technical Notes:
• To reduce the possibility of noise interference, isolate AC power
lines from signal lines.
• Maximum 4-20 mA loop impedance (sec. 3) is affected by the
supply voltage.
3. 4 - 20 mA Current Output Connections
To isolate output and prevent ground loop problems:
1. Use monitor device with isolated inputs, or
2. Use separate DC supply for PHCN-5700 and monitor device, or
3. Power PHCN-5700 with12 - 24 VAC step down transformer
Technical Notes:
**1/8A fuse recommended (customer supplied)
* 4-20 mA output is internally powered (non-isolated), maximum loop
impedance 350 Ω with a 12 V instrument supply voltage, 950 Ω
with a 24 V instrument supply voltage.
Technical Notes:
• Use six conductor shielded cable for cable extensions up to 120 m
(400 ft) max.
• Cable shield must be maintained through cable splice
Fuse**
1/8A
+
4 - 20 mA
*
-
+
-
4-20
mA
pH/ORP
PHCN-5700
Terminals
4. Relay Connections
Two internal relay contact sets (COM, NO, and NC) may be used for
external device control. Front panel LED annunciators indicate the
activation status of each relay. Each relay can control up to two
devices simultaneously, as shown. Relay operation modes include Low
alarm, High alarm, and Proportional Pulse (sec. 5).
Common device connections include:
• Pulse mode - metering pump control
• Pulse mode - solenoid valve control
• Low or High mode - warning lamps
• Low or High mode - bells or sirens
• Low or High mode - external heavy-duty relay
PHCN-5700
Terminals
NO
C
NC
+
External AC/DC
Relay 1Relay 2
NO
power
Relay
De-energized
C
NO
NC
C
NC
Device
+
A
-
Relay
Energized
C
NO
NC
-
-
Device
B
+
Wiring Example Right
Device A IS powered when relay 2 is de-energized (front panel LED
“off”). Power is discontinued when the relay 2 setpoint is reached
(front panel LED “on”). Device B IS NOT powered when the relay 2 is
de-energized. Power is applied after the relay 2 setpoint is reached.
Technical Notes:
• Maximum relay contact ratings: 5 A @ 30 VDC, 5 A @ 125 VAC,
or 3 A @ 250 VAC
• An external heavy-duty relay must be used for devices with surge
currents or operating currents that exceed the above specifications.
5. Relay Operation
A. LOW alarm mode
The relay is energized when the ORP drops below the setpoint, and is
de-energized when the ORP rises above the setpoint plus hysteresis
(sec. 6.3D, 6.3E).
ORP
Low setpoint=
Adjustable hysteresis=
Relay energized=
Relay de-energized=
C. Proportional PULSE mode
The proportional pulse relay configuration is primarily designed for metering pump control. The operator is prompted to enter a minimum and
maximum ORP setpoint and maximum pulse rate for the assigned relay (sec. 6.3F, 6.3G). Relay pulse width is fixed at 130 ms. Refer to the
operation examples below.
• Metering pump chemical addition (dry contact activation type required)
B. HIGH alarm mode
The relay is energized when the ORP rises above the setpoint and is
de-engergized when the ORP falls below the setpoint plus
hysteresis (sec. 6.3D, 6.3E).
High setpoint=
Adjustable hysteresis=
Relay energized=
Relay de-energized=
ORP
300 pulses
Example 1 (right):
As the process ORP rises above the minimum pulse setpoint (-500 mV) the relay
begins pulsing; triggering the metering pump for chemical addition. As the ORP
continues to rise, pulsing accelerates proportionally until the maximum programmed
pulse rate of 300 pulses/minute and setpoint (+500 mV) are reached, forcing the
process ORP back down to intended levels (e.g. ≤ -500 mV).
Example 2 (right):
As the process ORP falls below the minimum pulse setpoint (+500 mV) the relay
begins pulsing; triggering the metering pump for chemical addition. As the ORP
continues to decrease, pulsing accelerates proportionally until the maximum
programmed pulse rate of 300 pulses/minute and setpoint (-500 mV) are
reached, forcing the process ORP back up to intended levels (e.g. ≥ +500 mV).
ORP
-1000
mV
-1000
mV
0 pulses
30
-500 mV
300 pulses
-500 mV
Relay 1 or
Relay 2
30
Relay 1 or
Relay 2
+500 mV
+500 mV
+1000
0 pulses
ORP
+1000
mV
mV
6. Menu Functions
ENTER
To access either CALIBRATE or OPTIONS menus, press and hold the
ENTER key as illustrated below:
Press &
hold for
ENTER
access:
VIEW
2s5s
CALIBRATE
OPTIONS
Menus:
• VIEW menu (sec. 6.1): The VIEW menu is displayed during
standard operation. The operator can navigate freely through the
menu by pressing either UP or DOWN arrow keys.
• CALIBRATE Menu (sec. 6.3): The CALIBRATE menu contains all
critical display setup and output parameters. A simple security code
feature prevents unauthorized tampering. The operator is required
to enter a simple access code for menu access. The same code
also unlocks OPTIONS menus.
• OPTIONS Menu (sec. 6.4): The OPTIONS menu contains setup and
display features that are seldom accessed for minor display or
output adjustments.
0 - 14 pH
A
L
A
R
M
1
pH
A
L
A
R
M
2
• Reversible Dial
Your 5700 includes a
reversible dial face for
pH use. See enclosed
pH manual for operation
details.
BA
0
D
14
C
A
L
A
R
M
1
ORP
A
L
A
R
M
2
-
1000
+1000
6.1 VIEW Menu
1
Choose:
↓G
A.
ORP mV/Raw mV
B.
Loop output
C.
Range
D.
Relay 1
E.
Relay 2
2
Change:
3
Save:
Menu Displays A - G:
(Factory default displays shown in menu
column 1)
A. Shows calibrated solution ORP mV
and Raw (uncalibrated) input mV
from ORP electrode/preamplifier
B. Loop output display: shows the loop
current output level.
C. ORP range display: shows the
programmed min and max meter
range (sec. 6.3A)
D. Relay 1 display: this display shows
the programmed operation mode
and setpoint for relay 1 (sec.
6.3D-E)
E. Relay 2 display: this display shows
the programmed operation mode
and setpoint for relay 2 (sec.
6.3D-E)
F.
G.
Last calibration
Buffer calibration
↑A
Press and release to access
EASY CAL Procedure
See EASY CAL Procedure (Section 6.2)
F. Last calibration: this display shows a
user defined setup date for
maintenance records. This feature is
not an internal timer or calender.
G. EASY CAL: press the right arrow key
to access the EASY CAL buffer
calibration procedure (sec. 6.2).
Loading...
+ 11 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.