Omega Products PHB-900 Installation Manual

Page 1
http://www.omega.com
e-mail: info@omega.com
User ’s Guide
PHB-900 Series
pH Benchtop Meters
Page 2
Servicing North America:
USA: One Omega Drive, Box 4047
Tel: (203) 359-1660 FAX: (203) 359-7700 e-mail: info@omega.com
Canada: 976 Bergar
Laval (Quebec) H7L 5A1 Tel: (514) 856-6928 FAX: (514) 856-6886 e-mail: canada@omega.com
For immediate technical or application assistance:
USA and Canada: Sales Service: 1-800-826-6342 / 1-800-TC-OMEGA
SM
Customer Service: 1-800-622-2378 / 1-800-622-BEST
SM
Engineering Service: 1-800-872-9436 / 1-800-USA-WHEN
SM
TELEX: 996404 EASYLINK: 62968934 CABLE: OMEGA
Mexico and Latin America:
Tel: (95) 800-TC-OMEGA
SM
FAX: (95) 203-359-7807
En Espan÷ol: (203) 359-7803 e-mail: espanol@omega.com
Servicing Europe:
Benelux: Postbus 8034, 1180 LA Amstelveen, The Netherlands
Tel: (31) 20 6418405 FAX: (31) 20 6434643 Toll Free in Benelux: 06 0993344 e-mail: nl@omega.com
Czech Republic: ul. Rude armady 1868, 733 01 Kavrine-Hranice,
Czech Republic Tel: 420 (69) 6311627 FAX: 420 (69) 6311114 e-mail: czech@omega.com
France: 9, rue Denis Papin, 78190 Trappes
Tel: (33) 130-621-400 FAX: (33) 130-699-120 Toll Free in France: 0800-4-06342 e-mail: france@omega.com
Germany/Austria: Daimlerstrasse 26, D-75392 Deckenpfronn, Germany
Tel: 49 (07056) 3017 FAX: 49 (07056) 8540 Toll Free in Germany: 0130 11 21 66 e-mail: germany@omega.com
United Kingdom: 25 Swannington Road, P.O. Box 7, Omega Drive,
ISO 9002 Certified Broughton Astley, Leicestershire, Irlam, Manchester,
LE9 6TU, England M44 5EX, England Tel: 44 (1455) 285520 Tel: 44 (161) 777-6611 FAX: 44 (1455) 283912 FAX: 44 (161) 777-6622
Toll Free in England: 0800-488-488
e-mail: uk@omega.com
omega.comomega.com
OMEGAnetSMOn-Line Service Internet e-mail
http://www.omega.com info@omega.com
It is the policy of OMEGA to comply with all worldwide safety and EMC/EMI regulations that apply. OMEGA is constantly pursuing certification of its products to the European New Approach Directives. OMEGA will add the CE mark to every appropriate device upon certification.
The information contained in this document is believed to be correct but OMEGA Engineering, Inc. accepts no liability for any errors it contains, and reserves the right to alter specifications without notice. WARNING: These products are not designed for use in, and should not be used for, patient connected applications.
OMEGA
TM
®
Page 3
Table Of Contents
Quick Reference . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .ii
Getting Started
Front panel controls . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1
Digital display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1
Rear panel connectors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2
Connecting to a power source . . . . . . . . . . . . . . . . . . . . . . .2
Installing and Maintaining Electrodes . . . . . . . . . . . . . . . . . . . .3
Preparing glass electrodes . . . . . . . . . . . . . . . . . . . . . . . . . . .3
Installing the combination glass pH electrode . . . . . . . . . . . .3
Installing the FET pH electrode . . . . . . . . . . . . . . . . . . . . . . . .4
Installing ORP or ISE electrodes . . . . . . . . . . . . . . . . . . . . . . . .4
Rinsing electrodes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5
Storing electrodes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5
Standardizing for pH Measurement . . . . . . . . . . . . . . . . . . . . . .6
Entering buffers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6
The diagnostic electrode test . . . . . . . . . . . . . . . . . . . . . . . . .7
Selecting buffers based on 20oC or 25oC . . . . . . . . . . . . . . .8
Using Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9
Clearing Buffers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9
Reviewing electrode standardization . . . . . . . . . . . . . . . . . . .9
Selecting buffer sets . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10
Standardizing for Millivolt Measurement . . . . . . . . . . . . . . . . .11
(Relative Millivolts) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .11
Entering a millivolt standard . . . . . . . . . . . . . . . . . . . . . . . . .11
Clearing a millivolt standard . . . . . . . . . . . . . . . . . . . . . . . . .12
Measuring pH or Millivolts . . . . . . . . . . . . . . . . . . . . . . . . . . . . .12
Understanding pH Theory . . . . . . . . . . . . . . . . . . . . . . . . . . . . .13
Defining pH . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .13
Measuring pH . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .14
Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .15
The “Error”Icon . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .15
The “CAL Error”Icon . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .15
The “Err oC”Icon . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .15
The “Electrode Error”Icon . . . . . . . . . . . . . . . . . . . . . . . . . . .15
Electrode testing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .16
Meter Operation testing . . . . . . . . . . . . . . . . . . . . . . . . . . . .16
Meter Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .17
i
Page 4
1. Connect power cable to meter power jack and to AC power source.
power
ref
input
FET
ATC
2. Connect the glass electrode to the input and ATC connectors, or con­nect the FET electrode to the FET con­nector.
rel mV pH
3. Press pH/mV until the dis­play indicates the appro­priate measurement mode.
4. Standardize the meter using up to three buffers by immers­ing the electrode in a buffer, then pressing Standardize to enter each buffer.
Standardizing
PHB900 Series pH Meter Quick Reference
5. The display shows the current reading in pH,mV,or relative mV units.
6. Press Setup to review elec­trode calibration and to clear or select buffer sets.
%slope
Clear
Set Buffers
Measuring
S
ii
pH/mV Standardize
pH/mV Standardize
Setup Enter
Setup Enter
pH/mV Standardize
pH/mV Standardize
Setup Enter
Setup Enter
Page 5
Getting Started
1
pH/mV Standardize
Setup Enter
pH/mV - button toggles between pH and mV modes.
Standardize - the meter rec­ognizes and displays buffers.
Setup - use to clear buffers and to select new set of buffers.
Enter - use to enter buffers.
Digital Display
100.0°C
%slope rel mV pH
-18.0.0.0
Good Electrode Error FET Standardizing Clear Measuring Set Buffers
2 4 7 10 12 1 3 5 8 10 13
1.68 4.01 6.86 9.18 12.46
1.08 3.06 4.86 6.79 9.23 12.76
S
Stability
Icon
Temperature
Mode
Front Panel Controls
Result
Electrode
Check
Icons
Prompts
Buffer Icons
Page 6
power
ref
input
FET
ATC
Rear Panel Connectors
Power Cable
Connector
Reference
Electrode
Connector
Glass pH or
ISE
Electrode
Connector
FET
Electrode
Connector
Temperature
Probe
2
AC
Adapter
Power
Connector
Page 7
Installing and Maintaining Electrodes
This meter allows you to use two types of pH electrodes: the pH glass electrode and the field effect transistor (FET) pH elec­trode.If both types of electrodes are installed,the meter will read the FET electrode.
NOTE: If both electrodes are connect­ed to the meter, do not put them in a solution together because you will get inaccurate measurements.
3
1.Remove the protective end
cover from electrode.
3.Remove the shorting cap on the BNC connector.Install the com­bination glass pH electrode by plugging it into the input connec­tion (twist-lock) and the ATC connector into the ATC jack.
2.Before first use of your glass pH electrode,or whenever the electrode is dry,soak over-night in a filling or KCI solution.
ATC
FET
input
ref
er
w
o
p
Page 8
4
ATC
4.
Option:
Install the optional FET electrode by plugging it into the FET jack on the back of the meter panel.Allow the FET electrode to warm up for one minute prior to use.
5.
Option:
Install ORP or Ion Selective Electrode pairs by removing the BNC shorting cap and plugging the BNC connector (twist-lock) into the BNC jack.If a combination electrode is not used, plug the reference electrode into the ref pin.
FET
input
ref
er
w
o
p
ATC
FET
input
ref
er
w
o
p
Page 9
5
6. Rinse and blot-dry elec­trodes between each measurement (do not wipe).Rinse electrodes with distilled water or deionized water,or part of the next solution to be measured.
7. Store glass and FET pH elec­trodes in electrode filling solu­tion or KCI solution.Always leave the filling hole open and refill with filling solution when the internal solution level gets low.
Page 10
6
E
Standardizing for pH Measurement
Because electrodes vary in their response,you must standard­ize your pH meter and electrode to compensate for electrode variation.The more frequently you standardize,the more accurate your measurements.Standardize daily,or more often,for accurate results.
1.Immer se electrode in a buffer solution.Stir gently.
pH
4.12
2.Press and release the pH/mV button until your digital dis­play indicates pH mode.This button toggles between pH and mV modes.
pH
Clear
Buffers
3.Clear existing buffers when doing a new 2 or 3 point stan­dardization.Use the Setup button to clear existing buffers and to select a new set of buffers.
4.Press Standardize.The meter recognizes the buffer and flashes a buffer icon.When the signal is stable,or when you press Enter,the buffer is entered.
pH
4.00
Standardizing
4
nter
pH/mV Standardize
Setup Enter
pH/mV Standardize
Setup Enter
pH/mV Standardize
Setup Enter
Page 11
7
pH/mV Standardize
5.The meter displays the %slope of the electrode as 100.0% On entering a second or third buffer,the meter performs a diagnostic check on the elec­trode.
%slope
100.0
4
6.To enter a second buffer, place the electrode in the second buffer solution and press Standardize again.The meter recognizes the buffer and displays the first and sec­ond buffer icons.
pH
7.00
Standardizing
4 7
7.Next, the meter performs a diagnostic test of the elec­trode.The display indicates either Good Electrode or Electrode Error.The meter dis­plays the % slope of the elec­trode.
8.Electrode Error indicates that your electrode is not working properly. The electrode response must be between 90 and 105% slope.Measure­ments causing Electrode Error are not accepted,used or stored by the meter.
%slope
99.4
Good Electrode
4
%slope
82.3
Electrode Error
4
Setup Enter
pH/mV Standardize
Setup Enter
pH/mV Standardize
Setup Enter
pH/mV Standardize
Setup Enter
Page 12
8
pH/mV Standardize
9.To set a third standard,place the electrode in the third buffer solution and press Standardize.The results will be the same as in steps 6 and 7, except the display will show three buffer icons.
10.After enter ing each buffer, the
Standardizing
icon goes
off and the
Measuring
icon appears on the display to indicate that the meter returns to
Measuring
.
22.5°C
pH
10.03
Standardizing
4 7 10
11.The first set of buffers in the meter is used at 25°C in North America and, typically, at 20°C in Europe.In pH mode,press Standardize and Setup together to show the current buffer temperature setting.Press both Standardize and Setup again to toggle between the tem­perature settings.Press Enter to select the displayed tem­perature setting and to return to
Measuring.
12.Standardize your meter and electrode using at least two buffers with pH values brack­eting the expected pH of your samples.Stirr ing with a magnetic stir bar and stirrer provides faster electrode response.
25°C
Set Buffer
2 4 7 10 12
pH
10.00
Standardizing
4 7 10
NOTE: The meter continually adjusts for temperature. Therefore, buffers may vary slightly from the nominal values because of tem­perature variations.
Setup Enter
pH/mV Standardize
Setup Enter
pH/mV Standardize
Setup Enter
Denver Instrument
22.2¡C
pH
S
6
0
.
4
g
n
i
r
u
s
a
e
M
4
Enter
Setup
Standardize
pH/mV
BASIC
Page 13
9
Using Setup
The Setup button lets you clear all the buffers that you have entered,review calibration information,or select the buffer set that you want.
NOTE: You can escape setup mode at any
time by pressing pH/mV.
1.Press Setup and the meter dis­plays a flashing
Clear Buffers
icon.
Use this step only when you wish to clear all buffers you have entered
.To clear all existing buffers, press Enter.The meter clears all buffers and returns to
Measuring.
2.Press Setup again to show electrode performance.If the meter has accepted an elec­trode and buffers, it will dis­play
Good Electrode
,display the slope between the first and second buffers and dis­play the two buffer icons.
3.Pressing Setup again shows the electrode slope between the second and third buffers (if three buffers have been entered) and shows the sec­ond and third buffer icons.
4.Press Setup again to display a flashing
Set Buffers
icon and to display the first buffer set icons.
%slope
99.4
Good Electrode
4
pH
Set Buffers
2 4 7 10 12
pH
Clear
Buffers
%slope
98.7
Good Electrode
7 10
pH/mV Standardize
Setup Enter
pH/mV Standardize
Setup Enter
pH/mV Standardize
Setup Enter
pH/mV Standardize
Setup Enter
Page 14
10
5.Press Enter to select the set of buffers shown on the display
or Press Setup again to view the next set of buffers. Continue pressing Setup to view the third and fourth buffer sets.
6.Press Enter to select the dis­played buffer set that con­tains the buffer you want to use.Press Setup again,or press pH/mV at any time,to return to
Measuring
.
NOTE: You may mix buffers from different sets.
pH
Set Buffers
1.68 4.01 6.86 9.18 12.46
pH
Set Buffers
1 3 5 8 10 13
pH/mV Standardize
Setup Enter
pH/mV Standardize
Setup Enter
Page 15
11
E
Standardizing for Millivolt Measurement (Relative Millivolts)
You will use millivolt measurement for measuring ion concentration and for measuring redox potential (also called ORP, oxidation reduc­tion potential).You will normally use an ion selective electrode (ISE), combined with a reference electrode,to measure ion concentration. The ISE senses the ion concentration and responds with a millivolt potential.The millivolt readings are then used to calculate ion con­centrations.You will normally use a platinum indicator electrode, combined with a reference electrode,to measure redox potential (ORP).ORP measurements indicate the oxidizing or reducing capa­bility of a solution.You can use ORP values to monitor or control solu­tions requiring a set amount of oxidants or reductants.
1.Immer se electrode in a stan­dard solution.
mV
100.0
2.Press the pH/mV button until your digital display indicates mV mode.
3.Press Standardize to enter a mV standard and read rela­tive mV.
NOTE: Relative mV mode is not allowed with the FET electrode.
4.When the signal becomes sta­ble,or when you press Enter, the current absolute mV value becomes zero relative millivolts.
rel mV
0.0
Measuring
mV
100.0
Standardizing
nter
pH/mV Standardize
Setup Enter
pH/mV Standardize
Setup Enter
pH/mV Standardize
Setup Enter
Page 16
pH/mV Standardize
5.To clear a mV offset and return to absolute millivolt mode,press Setup.The meter displays a flashing
Clear
icon, and the current relative milli­volt offset.
6.To clear your previous mV standard,press Enter.You then return to absolute mV mode.
rel mV
0.0
Measuring
mV
100.0
Clear
12
Measuring pH or Millivolts
1.Standardize your meter. See page 10.
2.Rinse electrode and immerse in sample solution.Stir gently.
3.Press pH/mV until your display indicates the correct mode.
4.The display shows the current reading in pH, mV,or relative mV units.When the signal is stable, the meter displays the S icon.The S icon means the signal is changing less than 0.007 pH or 0.08 mV from the prior reading.
5.You may receive an out of range error, Err,if your electrode is
not immersed in a solution.To correct the error,immerse the electrode in a solution.
6.Separate calibrations for the glass and FET electrodes are stored in memory.Plugging the FET in recalls it’s calibration, unplugging the FET recalls the glass electrode calibration.
Measuring
S
rel mV pH
Err
Electrode Error Measuring
Setup Enter
pH/mV Standardize
Setup Enter
pH/mV Standardize
Setup Enter
pH/mV Standardize
Setup Enter
Page 17
13
Understanding pH Theory
Defining pH
The measurement of pH plays an important role in identifying and controlling acidity and alkalinity levels for industry and research. pH is a measure of the acidity or alkalinity of a solu­tion and can be represented by this equation:
pH = -log [H+]
with [H+] representing the concentration of hydrogen ions in the solution. pH is sometimes referred to as the power of the hydrogen ion in a solution.
By using a pH meter,you can determine exact pH levels of solutions. For example, rather than say that lemon juice is quite acidic,you can say that lemon juice has a pH of 2.4. An exact pH value can be used to control or measure acidity lev­els for manufacturing processes or for basic research.
pH values generally range from 0 to 14,with a pH value of 7 being the neutral point,or the value of pure water. The pH values above the neutral point represent increasing alkalinity, whereas pH values below the neutral point represent increas­ing acidity (Figure 1).
Strong Acid
Lemon Juice
Tomato Juice
Coffee
Pure water
Baking Soda
Ammonia
Strong Base
Neutral
0 1 2 3 4 5 6 7 8 9 10 11 12 13 14
more acidic
more basic
pH
Figure 1.
pH Scale showing the relative acidity or basicity
of some common substances.
Page 18
Measuring pH
To measure pH with a conventional glass pH electrode,the meter uses a pH-sensing glass bulb electrode that is sensitive to hydrogen ions.The potential developed at the glass bulb is directly related to the pH of the solution.
The glass bulb electrode is paired with a reference electrode which completes the electrical measuring circuit and provides a stable reference point.These two electrodes are joined to create a combination electrode.The combination glass elec­trode is connected to the pH meter which reads the voltage, converts it to pH units,and displays the result.
Combination Glass pH Electrode
This meter can also use a field effect transistor (FET) electrode for measuring pH.The FET uses an ion-sensitive solid state membrane attached to the transistor to measure the hydro­gen ion concentration of a solution.The FET is paired with a reference electrode and counter electrode that maintain a constant potential while the FET responds to the sample.
14
Reference Electrode (Outer Body)
Glass Membrane
Glass Electrode (Inner Body)
Porous Junction
Glass Membrane
Temperature Sensor
Reference Electrode
pH-Sensitive Membrane
Counter Electrode
FET
Temperature Sensor
Annular Porous Junction
Page 19
Troubleshooting
15
Err
Electrode Error
1. If the signal from the elec­trode is
out of range
,the dis­play will show Err.This may happen when the electrode is not in a solution.
Electrode Error
2. The meter will display
Electrode Error
when it detects an error in electrode response.During standardiza­tion,the message indicates that the electrode is less than 90% or more than 105% of the correct response.The
Electrode Error
message can indicate either a bad elec­trode or bad buffer(s).
Err°C
Error
3. If the meter detects an error in the temperature probe,the display shows
Err°C
.If you do not use a temperature probe, the meter uses the default temperature that you set to either 20°C or 25°C.
4. To test the pH electrode, place it in a good pH 7 buffer. Press pH/mV to use the mV mode,and note the milli­volt reading.Repeat for either a pH 4 or pH 10 buffer. The electrode signal must be within the limits shown below (when temperature is near 25°C).
Electrode Test
pH 7 0 + 30 mV pH 4 159 to 186 mV
more than pH 7
pH10 159 to 186 mV
less than pH 7
pH/mV Standardize
Setup Enter
pH/mV Standardize
Setup Enter
pH/mV Standardize
Setup Enter
Page 20
16
Cal
Error
5. To test the meter for correct operation,install the BNC (input) shorting cap. Press pH/mV to use the mV mode,and note the mV reading.If the meter reads ) + 0.5 mV,it is measur ing correctly.
6. If the meter detects a loss of calibration or a hardware error during its power-up self­test,the display shows
CAL
Error
.This means the mV accuracy may be reduced, but pH accuracy after stan­dardization will be the same. Press Enter to continue using the meter.If recalibration is desired,contact Technical Support.
ATC
FET
input
ref
er
w
o
p
pH/mV Standardize
Setup Enter
Page 21
Meter Specifications
pH -1.99 to 19.99 pH
displayed to 0.01 pH accurate to 0.01 pH
mV -1800.0 to 0.01 mV
displayed to 0.1 mV accurate to 0.5 mV
Temperature -5.0 to 105.0°C
displayed to 0.1°C
accurate to 0.1·°C Standardization 0,1, 2 or 3 buffers Auto buffer recognition 22 buffers
2,4, 7, 10,12
1,3, 6,, 8,10, 13
1.68,4.0, 6.86, 9.18,12.46
1.09,3.06, 4.65, 6.79,9.23, 12.75
Auto Temperature Compensation Automatic electrode slope correvtion for 90-105% Direct reading with both a glass and FET pH electrode
17
Page 22
Page 23
WARRANTY/DISCLAIMER
MEGA ENGINEERING, INC. warrants this unit to be free of defects in materials and
orkmanship for a period of 37 months from date of purchase. OMEGA Warranty adds an dditional one (1) month grace period to the normal three (3) years product warranty to over handling and shipping time. This ensures that OMEGA’s customers receive maximum overage on each product.
the unit should malfunction, it must be returned to the factory for evaluation. OMEGA’s
ustomer Service Department will issue an Authorized Return (AR) number immediately upon hone or written request. Upon examination by OMEGA, if the unit is found to be defective it will e repaired or replaced at no charge. OMEGA’s WARRANTY does not apply to defects resulting
om any action of the purchaser, including but not limited to mishandling, improper interfacing, peration outside of design limits, improper repair, or unauthorized modification. This
WARRANTY is VOID if the unit shows evidence of having been tampered with or shows evidence
f being damaged as a result of excessive corrosion; or current, heat, moisture or vibration;
mproper specification; misapplication; misuse or other operating conditions outside of OMEGA’s
ontrol. Components which wear are not warranted, including but not limited to ontact points, fuses, and triacs.
MEGA is pleased to offer suggestions on the use of its various products. However,
MEGA neither assumes responsibility for any omissions or errors nor assumes liability for ny damages that result from the use of its products in accordance with information pro­ided by OMEGA, either verbal or written. OMEGA warrants only that the parts
manufactured by it will be as specified and free of defects. OMEGA MAKES NO OTHER WARRANTIES OR REPRESENTATIONS OF ANY KIND WHATSOEVER, EXPRESSED OR
MPLIED, EXCEPT THAT OF TITLE, AND ALL IMPLIED W ARRANTIES INCLUDING ANY WAR-
ANTY OF MERCHANTABILITY AND FITNESS FOR A P AR TICULAR PURPOSE ARE HEREBY
ISCLAIMED. LIMITATION OF LIABILITY: The remedies of purchaser set forth herein are xclusive and the total liability of OMEGA with respect to this order, whether based on con-
act, warranty, negligence, indemnification, strict liability or otherwise, shall not exceed
he purchase price of the component upon which liability is based. In no event shall OMEGA
e liable for consequential, incidental or special damages.
ONDITIONS: Equipment sold by OMEGA is not intended to be used, nor shall it be used: (1) as
“Basic Component” under 10 CFR 21 (NRC), used in or with any nuclear installation or activity; r (2) in medical applications or used on humans. Should any Product(s) be used in or with any uclear installation or activity, medical application, used on humans, or misused in any way,
MEGA assumes no responsibility as set forth in our basic WARRANTY/DISCLAIMER language, nd additionally, purchaser will indemnify OMEGA and hold OMEGA harmless from any liability r damage whatsoever arising out of the use of the Product(s) in such a manner .
RETURN REQUESTS / INQUIRIES
irect all warranty and repair requests/inquiries to the OMEGA Customer Service Department.
EFORE RETURNING ANY PRODUCT(S) TO OMEGA, PURCHASER MUST OBTAIN AN
UTHORIZED RETURN (AR) NUMBER FROM OMEGA’S CUSTOMER SERVICE DEPARTMENT
N ORDER TO AVOID PROCESSING DELAYS). The assigned AR number should then be
marked on the outside of the return package and on any correspondence.
he purchaser is responsible for shipping charges, freight, insurance and proper packaging to revent breakage in transit.
OR W
ARRANTY RETURNS, please have
he following information available BEFORE
ontacting OMEGA: . P.O. number under which the product was
PURCHASED,
. Model and serial number of the product
under warranty, and
. Repair instructions and/or specific
problems relative to the product.
FOR NON-WARRANTY REPAIRS,
consult OMEGA for current repair charges. Have the following information available BEFORE contacting OMEGA:
1. P.O. number to cover the COST
of the repair,
2. Model and serial number of product, and
3. Repair instructions and/or specific problems
relative to the product.
MEGA’s policy is to make running changes, not model changes, whenever an improvement is possible.
his affords our customers the latest in technology and engineering.
MEGA is a registered trademark of OMEGA ENGINEERING, INC. Copyright 1996 OMEGA ENGINEERING, INC. All rights reserved. This document may not be copied, photocopied,
produced, translated, or reduced to any electronic medium or machine-readable form, in whole or in part, without
ior written consent of OMEGA ENGINEERING, INC.
UUSASA
MADE
ININ
Page 24
M2832/0298
Where Do I Find Everything I Need for
Process Measurement and Control?
OMEGA…Of Course!
TEMPERATURE
MU
Thermocouple, RTD & Thermistor Probes, Connectors, Panels & Assemblies
MU
Wire: Thermocouple, RTD & Thermistor
MU
Calibrators & Ice Point References
MU
Recorders, Controllers & Process Monitors
MU
Infrared Pyrometers
PRESSURE, STRAIN AND FORCE
MU
Transducers & Strain Gauges
MU
Load Cells & Pressure Gauges
MU
Displacement Transducers
MU
Instrumentation & Accessories
FLOW/LEVEL
MU
Rotameters, Gas Mass Flowmeters & Flow Computers
MU
Air Velocity Indicators
MU
Turbine/Paddlewheel Systems
MU
Totalizers & Batch Controllers
pH/CONDUCTIVITY
MU
pH Electrodes, Testers & Accessories
MU
Benchtop/Laboratory Meters
MU
Controllers, Calibrators, Simulators & Pumps
MU
Industrial pH & Conductivity Equipment
DATA ACQUISITION
MU
Data Acquisition & Engineering Software
MU
Communications-Based Acquisition Systems
MU
Plug-in Cards for Apple, IBM & Compatibles
MU
Datalogging Systems
MU
Recorders, Printers & Plotters
HEATERS
MU
Heating Cable
MU
Cartridge & Strip Heaters
MU
Immersion & Band Heaters
MU
Flexible Heaters
MU
Laboratory Heaters
ENVIRONMENTAL MONITORING AND CONTROL
MU
Metering & Control Instrumentation
MU
Refractometers
MU
Pumps & Tubing
MU
Air, Soil & Water Monitors
MU
Industrial Water & Wastewater Treatment
MU
pH, Conductivity & Dissolved Oxygen Instruments
Loading...