Omega Products PHB-220 Installation Manual

http://www.omega.com
e-mail: info@omega.com
User ’s Guide
PHB-215 pH Meter
PHB-220 pH/Conductivity Meter
PHB-225 pH/Ion Meter
PHB-250 pH/Ion/Conductivity Meter
PHB-215
eter
Servicing North America:
USA: One Omega Drive, Box 4047
ISO 9001 Certified Stamford, CT 06907-0047
Tel: (203) 359-1660 FAX: (203) 359-7700 e-mail: info@omega.com
Canada: 976 Bergar
Laval (Quebec) H7L 5A1 Tel: (514) 856-6928 FAX: (514) 856-6886 e-mail: canada@omega.com
For immediate technical or application assistance:
USA and Canada: Sales Service: 1-800-826-6342 / 1-800-TC-OMEGA
SM
Customer Service: 1-800-622-2378 / 1-800-622-BEST
SM
Engineering Service: 1-800-872-9436 / 1-800-USA-WHEN
SM
TELEX: 996404 EASYLINK: 62968934 CABLE: OMEGA
Mexico and Latin America:
Tel: (95) 800-TC-OMEGA
SM
FAX: (95) 203-359-7807
En Espan÷ol: (203) 359-7803 e-mail: espanol@omega.com
Servicing Europe:
Benelux: Postbus 8034, 1180 LA Amstelveen, The Netherlands
Tel: (31) 20 6418405 FAX: (31) 20 6434643 Toll Free in Benelux: 06 0993344 e-mail: nl@omega.com
Czech Republic: ul. Rude armady 1868, 733 01 Kavrine-Hranice,
Czech Republic Tel: 420 (69) 6311627 FAX: 420 (69) 6311114 e-mail: czech@omega.com
France: 9, rue Denis Papin, 78190 Trappes
Tel: (33) 130-621-400 FAX: (33) 130-699-120 Toll Free in France: 0800-4-06342 e-mail: france@omega.com
Germany/Austria: Daimlerstrasse 26, D-75392 Deckenpfronn, Germany
Tel: 49 (07056) 3017 FAX: 49 (07056) 8540 Toll Free in Germany: 0130 11 21 66 e-mail: germany@omega.com
United Kingdom: 25 Swannington Road, P.O. Box 7, Omega Drive,
ISO 9002 Certified Broughton Astley, Leicestershire, Irlam, Manchester,
LE9 6TU, England M44 5EX, England Tel: 44 (1455) 285520 Tel: 44 (161) 777-6611 FAX: 44 (1455) 283912 FAX: 44 (161) 777-6622
Toll Free in England: 0800-488-488
e-mail: uk@omega.com
omega.comomega.com
OMEGAnetSMOn-Line Service Internet e-mail
http://www.omega.com info@omega.com
It is the policy of OMEGA to comply with all worldwide safety and EMC/EMI regulations that apply. OMEGA is constantly pursuing certification of its products to the European New Approach Directives. OMEGA will add the CE mark to every appropriate device upon certification.
The information contained in this document is believed to be correct but OMEGA Engineering, Inc. accepts no liability for any errors it contains, and reserves the right to alter specifications without notice. WARNING: These products are not designed for use in, and should not be used for, patient connected applications.
OMEGA
TM
®
Quick Start Guide for pH Measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . Page ii
Introduction
Menu Keys,Softkeys and Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
Electrode Connectors and Inputs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
LCD Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
Function Keys . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Electrodes
Preparing pH and Ion Selective Electrodes. . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Connecting Electrodes. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Using and Storing Electrodes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
pH Electrodes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Solid-State FET Electrodes. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Ion Selective Electrodes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Meter Setup
Meter Setup Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
pH Mode
pH Mode Standardization Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Cal Reminder Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Select Custom Buffer Set. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
pH Mode Options Menu. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Standardizing and Measuring pH. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Clearing Buffers. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
mV Mode
mV Mode Standardization Menu. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
mV Mode Options Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Clearing Relative mV Mode. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Quick Start Guide for Ion Measurement. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Ion Mode
Ion Mode Standardization Menu. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Standardizing and Measuring Ion . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Measuring Ion using a Known Addition type (incremental ion) method. . . 17
Quick Start Guide for Conductivity/Resistivity/Salinity/TDS Measurement. . . . 18
Conductivity/Resistivity/Salinity/TDS Modes
Conductivity/Resistivity/Salinity/TDS Standardize Menu . . . . . . . . . . . . . . . . 19
Conductivity/Resistivity/Salinity/TDS Options Menu . . . . . . . . . . . . . . . . . . . . 20
Data logging . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
Meter Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
pH Theory . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
Ion Selective Electrode Theory . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
Determining Isopotential Point . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
RS-232 Serial Interface Meter Command Set. . . . . . . . . . . . . . . . . . . . . . . . . . . 26
Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
Table of Contents
i
1.Connect power cable to meter connector on the rear panel marked “power”and to AC power source.
2.Connect the glass pH/ATC electrode to the channel A BNC connector marked “ch.A”and to the channel A temperature connector marked “temp A”.
3.Verify the meter is in pH mode on channel A.Use the Mode key and Channel key to set the meter to the correct mode and channel if necessary. (See Function keys).
4.Standardize the electrode by immersing the electrode in a buffer,pressing Standardize, press­ing 1) Auto-enter a buffer and following the prompts.Repeat this step to enter each buffer. The meter will check the electrode and buffers, and give an error message if there is a problem.
Press Standardize to enter or clear buffers,select buffers, set resolution, or set other parameters for the current mode and channel.
5.The display shows the current measurement,and indicates a stable reading with the indicator. Press Cal Data to review and graph the elec­trode calibration data.
ii
Quick Start Guide for pH Measurement
A 7.000
pH
25.0°C
No buffers.
Mode
Channel A: pH mode
Standardize Menu
1) Auto-enter a buffer
2) Manual buffer entry
3) Clear buffers
4) Options menu
5) Cal reminder menu
6) Select buffer set
channel indicator
units indicator
Channel
Standardize
A 7.000
pH
25.0°C
4.008 pH
7.000
entered buffers
“power”
“ch. A”
“temp A”
s
s
stability indicator
s
c
h
. A
re
f. A
c
h
T
e
m
c
p
o
. A
n
d
. C
T
e
m
p
. B
R
S
2
3
2
C
p
o
w
e
r
c
h
. A
c
h
re
f. A
T
c
e
o
m
n
p
d
. A
. C
T
e
m
p
. B
R
S
2
3
2
C
p
o
w
e
r
This Omega meter is a powerful,versatile and accurate instrument. It features easy menu-based operation with easy to understand prompts and electrode/standard error checking.
These meters feature many advanced options,such as programmable stability criteria, programmable standardization reading delay times, multi-channel oper­ation,fast reading update rates of twice per second for all channels,program­mable alarms, programmable data logging of 500 data points and a superb RS­232 serial interface for controlling the meter and obtaining data.
Omega meters use flash programmable ROM for the operating code.The meter operating code can easily be upgraded as new features are made available.
Direct Menu Keys, Softkeys and Display
The meter uses six Direct Menu keys to access the menus and operations (such as selecting pH mode,standardizing,checking electrode calibration data, selecting the electrode channel).
There are four Softkeys that provide additional operations; these Softkeys change their function as needed and each Softkey has an icon to indicate its current function.
The display is a backlit quarter-VGA screen capable of displaying all four elec­trode channels (of a PHB-250) simultaneously. The backlight will turn off after a 45 minute period of non-use; pressing any key will automatically turn the backlight on again.
A S
7.001
pH
25.0¡C
4.008 pH
7.000
6/30/1998 02:28 PM
Denver Instrument
S
1
Introduction
Softkeys
Direct Menu
Keys
Direct Menu
Keys
Numeric
Keypad
PHB-215
eter
Electrode Connectors and Inputs BNC ("ch.A" or "ch.B") connectors: pH,Ion
Selective Electrodes and ORP (redox) elec­trodes attach to the meter through a BNC ("round twist-on") connector to channel A or channel B (PHB-225 and PHB-250 only).
Temperature ("temp A" or "temp B") connec­tors: use to connect the 2.5mm mini-phone
plug from the temperature sensor (built into the pH electrode with the Omega standard pH/ATC electrode) for Automatic Temperature Compensation (ATC).
Reference ("ref.A" or "ref.B") connectors: use for attaching a reference electrode tip-pin plug when a separate reference electrode is used.
Conductivity DIN (“ch.C conductivity”) connector: use to connect a 4-band conduc-
tivity/ATC cell.These 4-band cells offer improved linearity and stability over older 2­band conductivity cells. (PHB-220 and PHB-250 only).
FET DIN ("ch.F FET") connector: use to connect the Denver Solid-State Field Effect Transistor (FET) pH/ATC electrode. These non-glass pH electrodes offer certain advantages over con­ventional glass pH electrodes.
Serial port ("RS232") DB-9 connector: use to connect a serial printer or Personal Computer. This bi-directional interface outputs data and receives meter commands.
Power (“power”) connector: use to connect a
5.5mm OD x 2.1 ID coaxial connector with 12VDC at 500mA (center pin negative).
2
Introduction
Connectors for PHB-250
ch. A
ref. A
ref
temp
A
ch. F
FET
RS232C
power
D
Note: Not all of the following will display at the same time.
A
. Result: current measurement.
B
.Units: displays the units for the current measure-
ment.Examples: pH, mV, mg/L F-,µS/cm or W-cm.
C
.Softkey icons: show the current function
assigned to each softkey.
D
.Calibration due reminder: the icon means
a calibration is now due.
E
. Date and time: displayable in different formats.
F
.Da talogging: the icon indicates datalog-
ging is active.
G
.Buffers/Standards: in single channel mode,all
entered buffers or standards are displayed.A “!”symbol beside a buffer indicates that buffer is out of the entered calibration valid time (See Calibration reminder,page 9).
H
.Temperature: displays the measured tempera-
ture when an electrode with ATC or separate temperature probe is attached.Shows “M” when a manually entered temperature is being used.
I
.Alar m: “*”indicator means data is outside the
set alarm limits.
J
Channel: indicates which electrode channel (input) is being displayed.Channel A and Channel B (PHB-225 only) are BNC inputs. Channel F is the FET electrode input.
K
.Stability: the indicates the electrode is sta-
ble to the selected criteria.
L
.Multiple Channel: display can show two
(PHB- 215),three (PHB-220 and PHB-225) or four (PHB-250) electrode measurements with tem­perature simultaneously.
M
.Out-of-range or non-valid reading: dashes indi-
cate a measurement is not available.This usual­ly means the reading is out of range,or can mean in ion mode that no standards have been entered,or strict calibration has been set and the calibration expired.
3
Display
*
A 7.000
pH
25.0°C
4.008 pH
7.000
9.999
6/30/1998 02:28 PM
A B
J I H G
E
C
Single channel display
K
F
A 6.998
pH
24.6°C
B 1.02
mg/L
23.4°C F-
30-6-1998 14:28:05
Dual channel display
L
A pH
23.8
°
C
1998.6.30 14:28:30
M
s
s
s
CAL!
Log
S
A
.Mode: Selects the mode: pH,mV
(PHB-215),Ion (PHB-225 and PHB-250), Conductivity - Resistivity-NaCl Salinity­Practical Salinity-TDS (PHB-220 and PHB-250) to use for the currently selected channel (electrode input).
B
.Standardize: Enters buffers or standards
for the currently selected channel and mode.Use to enter pH buffers,relative mV offset,ion standards or conductivity /resis­tivity standards.
Also used to change other settings which affect the measurement.
C
.Cal Data: Displays and graphs buffers or
standards with time and date stamp and electrode calibration data for the select­ed channel and mode.
D
.Channel: Selects the channel(s) (elec-
trode inputs) to display.The PHB-215 can display one or two channels simultane­ously (Channels A and F).The PHB-225 can display one,two or three channels simultaneously (Channels A,B and F). The PHB-220 can display up to three channels (Channels A,C and F).The PHB-250 can display up to four channels (Channels A, B,C and F).
E
.Setup: The Setup menu is used to set vari-
ous general meter settings,such as date and time,display contrast,keypress beep and serial port.
F
.Da ta Log: Displays the datalogging menu
used to set datalogging and view the stored Data Log (see Datalogging).
G
.Clear: Exits from the current menu and
returns to the previous menu,cancels the current operation or clears a number entry.
H
.Enter/Pr int: Accepts numeric values,menu
selections or pending operations.In the main measure screen,acts as a Print key, sending all current measurements to a printer/ computer through the serial port and stores the measurements in the Data Log.
I
.Softkeys: These four keys access different
operations at different times.Most menus offer a “Help” softkey and the “Measure” softkey, w hich allows a direct return to the main measuring screen,exiting all menus immediately. The “Up Arrow”and “Down Arrow”softkeys offer one way to select a menu item.The “Left Arrow”key is a back­space,active during number entry.
J
.Numer ic Keys: Pressing a number key
selects a numbered item in a menu.The number keys also allow entering values for buffers, standards, and various meter set­tings.
4
Function Keys
Help
Measure
Scroll up
Scroll down
Backspace
Graph
Exponent number entry
Incremental ion method
Measure lock
Measure unlock
Softkeys
E
+
5
Function Keys
A
B
C
D
E
F
G
H
I
J
Channels
The Channel key is used to turn on or off each available channel.In single-channel operation, additional information for the selected channel is provided,including a display list of all entered buffers or standards. In multi-channel operation, the Mode,Standardize and Cal Data menus ask for the channel before accessing the menu.
The PHB-215 offers single or dual channel opera­tion using channel A ( pH and ORP electrodes) and channel F (Solid-state FET pH electrode).The PHB-225 offers up to three-channel simultaneous operation of channel A and channel B (pH,ORP and Ion Selective Electrodes) and channel F (Solid-state FET pH electrode).
The PHB-220 has Channel A (pH and ORP elec­trodes),Channel C (Conductivity cells) and Channel F (solid-state FET pH electrode).The PHB-250 provides up to four channel measure­ment with Channel A and Channel B (pH and ORP and Ion Selective electrodes),Channel C (Conductivity cells) and Channel F (solid-state FET pH electrode).
Configure Display
1) Turn channel A on/off: ON
2) Turn channel B on/off: ON
3) Turn channel F on/off: OFF
4) Turn channel C on/off: ON
5) Measurement screen
PHB-250 Select Channel
screen
Channel
?
Mode
Standardize
Data Log
Cal Data
Channel
Enter/Print
Setup
Clear
The meter allows you to use a variety of glass mem­brane (“glass”) pH/ATC electrodes,ion selective elec­trodes,Conductivity/ATC cells,the Omega Field Effect Transistor (FET) Solid-State pH/ATC electrode,tempera­ture (ATC) probes, combination electrodes using a BNC connector,or separate electrode pairs with BNC con­nector and reference pin.
Preparing Electrodes and Conductivity Cells
Remove the wetting cap or storage cap from the elec­trode.Before first using your pH electrode or whenever the electrode is dry,soak it several hours in an elec­trode filling or storage solution (3 Molar KCl solution) or in a buffer for pH electrodes. Condition ISE’s in the rec­ommended solutions. Rinse Conductivity cells with deionized water before use.
Connecting Electrodes pH, ORP or ISE electrodes (with BNC connector):
Connect the electrode to the BNC input,either chan­nel A or channel B (PHB-225 and PHB-250 only),located at the rear of the meter.Push in and rotate the elec­trode’s BNC connector until it locks in place.Connect the ATC connector to the temp.A or temp.B connec­tor. To disconnect, twist the BNC connector in the opposite direction and pull.
Electrode Pair Using a Reference Electrode (with Reference Pin Plug):
Connect the indicating electrode to the BNC input. Connect the reference electrode to the Reference input.Push the electrode’s tip pin plug into the input to connect and pull out to disconnect.
Conductivity Cells (with DIN connector):
Align and push in the DIN connector fully to the chan­nel C input (PHB-220 and PHB-250 only).Pull carefully to disconnect.
6
Electrodes
To measure Use channel (connector)
pH A (BNC)*
or
B (BNC)*
mV (ORP) A (BNC)*
or
B (BNC)*
ion (ISE) A (BNC)*
or
B (BNC)* Conductivity C (DIN) Resistivity C (DIN) Salinity C (DIN) TDS C (DIN) pH (FET) F (mini-DIN)
*Separate reference electrodes can be used with “Ref A”or “Ref B”connectors
BNC Connector
Channel C Conductivity Input
ATC Connector
Channel F FET pH Input
ch. A
ref. A
ref
temp
A
ch. F
FET
Using and Storing Electrodes
pH Electrodes
• Provide moderate stirring for faster electrode response.
• Leave the fill hole open during all use.
• Rinse the electrode between each measurement with a portion of the next sample or buffer to be measured,or with deionized or distilled water.
• Keep glass electrodes wet when not being used by placing some electrode filling solution in the wetting cap and storing with the wetting cap on.
• Keeping glass electrodes “wet”will improve their performance.Store electrodes in electrode filling solution or storage solution (3M KCl).
Solid-State FET Electrode
• All models allow use of both standard glass pH/ATC and Solid-State FET (Field Effect Transistor) pH/ATC electrodes.The meter can store a cali­bration for both types of electrodes.Plug the FET electrode into the channel F mini-DIN input.
• Allow the FET about 1 minute to stabilize when first connected.The FET electrode can be stored dry or in electrode storage solution.Provide mod­erate stirring.
Ion Selective Electrodes
• Add proper amount of Ionic Strength Adjuster (ISA) to all standards and samples,usually 1 mL ISA to 50 mL standard or sample.
• Provide moderate stirring for faster electrode response.
• Rinse the electrode(s) between each measure­ment with a portion of the next sample or stan­dard to be measured,or with deionized or dis­tilled water.
• Follow the instruction sheets for the individual electrode.Store as recommended.
Conductivity Cells
• When changing samples or standards,immerse the cell into the new solution,then lift and allow solution to drain out.Repeat two more times.
• Gently tap cell to dislodge air bubbles.
7
Electrodes
Store with wetting cap and Fill Solution (3M KCl)
Provide moderate stirring
Store FET and Conductivity Cells dry
Meter Setup Menu
Press Setup to access the Meter Setup menu:
1.Time and date menu: use to set the time format (HH:MM AM/PM or HH:MM:SS),set the time,set the date format (MM/DD/YY,DD-MM-YY or YYYY.MM.DD), and set the date.
2.Select temperature units: use to select temperature measurement and display in degrees Celsius,degrees Fahrenheit or Kelvin.
3.Select contrast: use to select the display contrast, mak­ing the displayed characters lighter or darker.Select setting “5”for typical conditions.
4.Select baud rate: use to set the ser ial RS232 port baud rate. This must match the baud rate setting of the printer or computer being used with the meter.
5.Setup serial port: use to configure the serial port start bits and parity setting. This must match the settings of the printer or computer being used with the meter.
6.Keypress beep on/off: use to turn on or off a "beep" upon each keypress as an audible signal that a key has been pressed.
7.Select video color scheme: use to set the display to black characters on a white background or white characters on a black background.
8.Show meter information: use to show the meter model,software version and serial number.
9.Enable measure lock: use to enable the measure lock where a stable measurement is locked (frozen) for later review. Sta bility criter ia should be set to SLOW for all channels and modes in use.
0.Enable strict calibration: use to set strict calibration
where no measurements are displayed if the calibra­tion reminder has expired.
“±”Set screen saver timeout: use to set a time for the
backlight to turn off and the screensaver to activate.
"•" Restore factory defaults: use to reset all settings to
factory defaults.On occasion it may be useful to completely reset the meter,for example,if other users have changed a setting.
Warning!
A reset also clears all electrode standardizations.
8
Meter Setup
!
Meter Setup Menu
1) Time and date menu
2) Select temperature units
3) Select contrast
4) Select baud rate
5) Setup serial port
6) Keypress beep on/off
7) Select video color scheme
8) Show meter information
9) Enable measure lock
0) Enable strict calibration ±) Set screen saver timeout
•) Restore factory defaults
Setup
Select Time Format
1) hr : min : sec
2) hr : min AM/PM
Select Date Format
1) mm / dd / yyyy
2) dd – mm – yyyy
3) yyyy.mm.dd
Enable Measure Lock
Display lock/unlock softkey?
1) Yes
2) No
If “measure lock” is enabled, a “lock” soft key will appear on the measurement screen
Enable Strict Calibration
Require calibration?
1) Yes
2) No
If strict calibration is enabled, no measurements will be displayed for a channel with an expired cal reminder.
9
pH Mode
pH Mode Standardize Menu
Press Mode and select 1) pH. Press Standardize and the pH Mode Standardize Menu appears:
1.Auto-enter a buffer: use to add a new buffer which is auto recognized by the meter,or update an exist­ing buffer.Follow the prompts.
2.Manual buffer entry: use to enter a buffer value by manually entering the pH of the buffer.
3.Clear buffers: use to clear all buffers entered for the current channel (pH mode).If all entered buffers are being re-entered,it is usually not necessary to clear buffers before re-entering them.
4.Options Menu: A menu of additional specific pH mode settings. (See page 10).
5.Cal reminder menu: use to set a timer reminding you to recalibrate.A icon will appear on the main screen and an exclamation mark will appear beside the buffers for which time has expired.
The calibration reminder is a reminder of when electrode calibration (with buffers) should be redone.It is based on elapsed time from the oldest entered buffer.
If strict calibration is set (see Meter Setup Menu),when a calibration has expired the icon appears, and "- - - " is displayed in place of the measurement. No measurements can be obtained until a calibration is performed.
Note: When strict calibration is set,the calibration reminders for all channels are turned ON,and can't be turned off from the Cal Reminder Menu.
6.Select buffer set: There are five auto-recognition buffer sets and the option to configure and use a custom buffer set of your own.
Select custom buffer set
Use Custom Buffer Set to make a set of buffers con­taining the specific buffers in use (up to five buffers). Select Custom Buffer Set,configure the custom buffer set,then buffers from this set will be automati­cally recognized and entered.
Custom buffers can have any numeric pH value,or can be selected from the built-in temperature cor­rected buffers. Using the built-in buffers allow tem­perature correction of the pH values of the buffers , offering more accuracy.
Channel A: pH mode
Standardize Menu
1) Auto-enter a buffer
2) Manual buffer entry
3) Clear Buffers
4) Options Menu
5) Cal reminder menu
6) Select buffer set
Standardize
Channel A: pH mode
Manual Buffer Entry
Enter the actual pH of this
buffer, after correcting
for temperature:
pH
Press Enter to accept.
Channel A: pH mode
Cal Reminder Menu
1) Turn reminder On/Off: OFF
2) Set reminder interval: 01, 00 : 00 days, hr : min
3) Standardization menu
When the cal reminder expires, a “CAL!” (or “!”) icon will be displayed with the measurement and with each expired stan­dard. If strict cal is ON, reminders are always ON, and expired channels show no measurements.
Channel A: pH mode
Select Buffer Set
1) 2, 4, 7, 10, 12 at 25C
2) 2, 4, 7, 10, 12 at 20C
3) 1, 3, 6, 8, 10, 13
4) 1.68, 4.01, 6.86, 9.18, 12.46
5) 1.09, 3.06, 4.65, 9.23, 12.75
6) Select custom buffer set
Custom Buffer
No buffers
CAL!
Channel A: pH mode
Configure Custom Buffer Set
1) Add a “built-in” buffer
2) Add a manual buffer
3) Delete a buffer
4) Standardize menu
Custom Buffer Set
No buffers
CAL!
pH Mode Options Menu
1.Select resolution: use to set pH readings to 0.1,0.01, or 0.001 pH units.
2.Select stability criteria: use to set stability criteria to slow,medium or fast to match the electrode's speed of response and the variability of the signal allowed for a “stable” ( ) measurement.
3.Select signal averaging: use to set filtering of the electrode signal to very slow (10 readings),slow (8), medium (6),fast (4) or very fast (2). Slower settings give more stable readings,although may require longer times to reach stability.
4.Set standardization delay: use to set a reading delay time for the meter to wait before accepting an elec­trode signal during standardization.Programming a standardization reading delay helps slow responding electrodes reach equilibrium before the electrode signal is accepted.
5.Set pH slope: use to set a known electrode slope used by the meter with a zero- or single-point stan­dardization.The normal default slope is 59.16 mV/pH. The meter allows between 80 and 120 % efficiency to be entered.
6.Standardize menu: Retur ns to the pH mode Standardization Menu.
7.Manual temperature menu: use to set a temperature to be used in the absence of an ATC probe or when manually overriding the ATC.
8.Data alarm menu: use to set pH limits.If the limits are exceeded an alarm indication (“*”) is displayed and recorded with any data points placed in the Data Log.
9.Set isopotential point: use to set an isopotential point for use in high accuracy electrode measurements (See Isopotential,page 20).
10
pH Mode
Channel A: pH mode
Options Menu
1) Select resolution
2) Select stability criteria
3) Select signal averaging
4) Set standardization delay
5) Set pH slope
6) Standardize menu
7) Manual temperature menu
8) Data alarm menu
9) Set isopotential point
S
Notes:
1.Auto-recognized buffers are found in the Auto-rec­ognized built-in buffer sets.These buffers are auto­recognized by the meter,and are also automatically temperature corrected for the variation of buffer pH with temperature.
2.When manually entering buffers, the exact pH of the buffer at the current temperature must be entered. All buffers change pH with temperature.For best accuracy, either use the built-in buffers or make sure manually entered buffers are at the expected tem­perature (so that their pH as entered is correct).
3.Auto-recognition Buffer Sets:
1) 2,4, 7, 10, 12 (nominal value adjusted at 25°C)
2) 2,4, 7, 10, 12 (nominal value adjusted at 20°C)
3) 1,3, 6, 8, 10,13
4) 1.68,4.01, 6.86, 9.18, 12.46 (NIST buffers)
5) 1.09,3.06, 4.65, 9.23, 12.75 (DIN buffers)
6) Select custom buffer set
4.Temperature Correction of Electrodes and Buffers
The meter automatically compensates for the temper­ature dependence of the electrode's response when measuring pH.The meter also compensates for buffer change in pH value with temperature.Temperature compensation is based on temperature either from an ATC probe or a manually entered temperature.
Using a Solid-state FET (Field Effect Transistor) pH/ATC Electrode
By turning channel F (FET) on,an Omega FET pH/ATC electrode can be directly used.pH and mV modes are available with the FET electrode.FET devices can have large offset potentials that vary with each transistor chip,so Manual buffer entry must be used to enter the first buffer.After one buffer has been entered using “Manual buffer entry”, following buffers can usually be entered with “Auto-enter a buffer”.
pH Mode
Actual Buffer pH vs. Temperature
pH 4.00(4.01)/7.00/10.00 buffer (nominal 25°C)
Temperature Buffer 4 Buffer 7 Buffer 10
(°C)
30 4.016 6.991 9.947 25 4.008 7.003 10.000 20 4.003 7.020 10.057 15 4.000 7.042 10.119 10 3.998 7.069 10.187
11
12
Standardizing and Measuring pH
1.Immerse the electrode in a buffer and stir moderate­ly. The meter displays the current pH measurement.
2.Allow the electrode sufficient time to reach
equilibrium.
3.Press Standardize,then press either 1) Auto-enter a buffer or 2) Manual buffer entry.
4.Follow the prompts on the display.
5.The meter waits for a stable signal,automatically rec­ognizes the buffer (if using “Auto-enter”),checks the electrode and buffer and enters the buffer.The entered buffer appears in the display.
6.Alternatively,if the signal is not sta ble, you can press Enter when the reading stabilizes according to your tolerance criteria.The meter then enters the buffer.
7.Repeat steps 1 through 4 to enter a second,third, fourth or fifth buffer.With more than one buffer the meter performs a diagnostic check on the electrode. The electrode is considered good if the slope is between 90 to 105%.If a sixth buffer is entered,the buffer farthest away is replaced by the new buffer.
Hints to achieve better accuracy:
• During standardization,allow time for the electrode
to stabilize before entering the buffer into the meter.
• Standardize using at least two buffers , bracketing the
expected pH of your samples.
• Standardize at least daily for the most accurate read-
ings.
• Open the Fill Hole on the electrode.
• Stir all buffers and samples.
• Rinse the electrode with DI water between samples
and buffers.
• Always use fresh buffers.
Clearing Buffers
Press Standardize,then press 3) Clear buffers to clear
buffers. If all previously entered buffers will be re­entered,it is not necessary to clear buffers since the meter will replace the previous values.If re-enter ing only some buffers, all the old buffers should be cleared.
Channel A: pH mode
pH Checklist
Prepare buffer or sample. Rinse electrode. Place electrode in solution. Stir moderately.
Press Enter to continue.
Channel A: pH mode
Standardize Menu
1) Auto-enter a buffer
2) Manual buffer entry
3) Clear buffers
4) Options menu
5) Cal reminder menu
6) Select buffer set
Standardize
Channel A: pH mode
Manual Buffer Entry
Enter the actual pH of this
buffer, after correcting
for temperature:
pH
Press Enter to accept.
mV
4 7 10
pH buffers
samples
Electrode Potential,mV
Titrant Volume, mL
Redox Titration
Standardize
Millivolt measurements are used to measure ORP (oxi­dation-reduction potential) or redox potential,to check performance of pH or Ion Selective Electrodes, and for redox titrations.
The meter will measure millivolts (mV) by pressing Mode and selecting 2) mV.Relative mV can be measured by entering a mV offset or using the current mV value as the mV offset.
mV Standardization Menu
In mV mode,press Standardize and the mV mode Standardization Menu appears:
1.Auto-enter mV offset: use to set the relative mV off­set equal to the current mV reading. The current mV becomes 0.0 relative mV.
2.Set mV offset: use to manually enter a mV offset.
3.Clear mV offset: use to clear any offset that has been entered,returning the meter to absolute mV mode.
4.Options menu: a menu of additional settings specif­ic to the mV mode.See below.
mV Mode Options Menu
1.Select resolution: use to set mV readings to 1 or 0.1 millivolt resolution.
2.Select stability criteria: use to select stability criteria for slow, medium or fast response which use a tight, average or loose requirement to indicate a “stable” ( ) reading.
3.Select signal averaging: use to set the meter to average readings that are very slow (10 readings), slow (8),medium (6),fast (4) or very fast (2).
4.Set standardization delay: use to set a length of time for the meter to wait before entering a relative mV standardization.
5.Set mV offset: use to manually enter a mV offset (same as in the mV Standardize menu).
6.Standardization menu: retur ns to the mV Standardization Menu.
Clearing Relative mV Mode
Press Standardize,then press 3) Clear mV offset to clear offset and return the meter to absolute mV mode.
mV Mode
Channel A: mV mode
Standardize Menu
1) Auto-enter mV offset
2) Set mV offset
3) Clear mV offset
4) Options menu
Channel A: mV mode
Options Menu
1) Select resolution
2) Select stability criteria
3) Select signal averaging
4) Set standardization delay
5) Set mV offset
6) Standardize
13
S
14
Quick Start Guide for Ion Measurements
1.Connect the Ion Selective Electrode (ISE) and Reference Electrode,if required, to the meter.”Combination” ISE’s have a reference electrode built-in,and do not require a sepa­rate reference electrode or connection.
2.Prepare two or more ion standards at concen­trations bracketing typical sample solutions. Add the appropriate Ionic Strength Adjuster solution to each standard.
3.Set the meter to display the correct channel (the channel with the ISE attached,either A or B) using Channel.Set the meter to Ion mode: press Mode,then 3) Ion.
Note: The meter will display “— — —”(no valid data) until an ion standard has been entered.
4.Place the electrode(s) in the standard,provide stirring (a magnetic stirrer is recommended), and allow sufficient time (1 to 5 minutes depending on the ISE) for the electrode to reach a stable signal.
5.Press Standardize,1) Enter a standard and fol­low the prompts.Repeat these steps to enter up to seven ion standards.See the Standardizing and Measuring Ion section for more information.
6.Check the ISE response by pressing Cal Data to see the standards and the ISE slope between calibration points (standards).The meter will allow an ion electrode slope between 5.92 mV/decade (10% slope) and 70.99 mV/decade (120% slope).
Mode
Channel
+
ISA
Example: 1 mg/L 10 mg/L Ionic Strength
Adjuster
BNC
ref
measuring or Combination ISE
reference
Cal Data
Channel A:
Select Mode
1) pH
2) mV
3) Ion
Channel A: ion mode
Ion Cal Data
1.00 mg/l 08/01/1998 08:32 AM
57.15 mV/decade
10.0 mg/l 08/01/1998 08:35 AM
Channel A: ion mode
Standardize Menu
1) Enter a standard
2) Set ion slope
3) Clear standards
4) Options menu
5) Cal reminder menu
c
h
. A
re
f. A
c
h
T
e
m
p
. A
T
e
m
p
. B
R
S
2
3
2
C
p
o
w
e
r
E
Ion Mode Standardization Menu
Select channel A or B.Press Mode and then press 3) Ion for ion mode.Press Standardize and the Ion Mode Standardize Menu appears.
1.Enter a standard: use to add a new standard or update (re­enter) an existing standard.Follow the prompts.With the first standard you select the ion name and units.
2.Set ion slope: use to manually enter a slope for the selected ion electrode.Used with a one-point ion calibration.Useful if the ISE has a known,stable slope,so that measurements can be made after entering a single ion standard.
Note: when two or more standards are entered,the meter uses the actual determined slope(s).
3.Clear standards: use to clear standards for the electrode standardization selected.
4.Cal reminder menu: use to set a timer reminding you to recalibrate.A icon will a ppear on the main screen and an exclamation mark will appear beside the standards which need to be re-entered.
5.Options Menu: use to set various additional parameters to the ion mode.See below.
Ion Mode Options Menu
1.Resolution: use to set the readings to 1,2, or 3 significant dig­its.
2. Select stability criteria: use to set the sta bility cr iteria to slow,
medium or fast to match the electrode's speed and stability of response,providing tight,medium and loose requirements for a stable ( ) indication.
3.Select signal averaging: use to set filtering of the electrode signal to very slow (10 readings),slow (8),medium (6),fast (4) or very fast (2). Slower settings give more stable readings, although may require longer times to reach stability.
4.Set standardization delay: use to set a reading delay time for the meter to wait before accepting an electrode signal dur­ing standardization.Programming a standardization reading delay helps slow responding electrodes reach equilibrium before the electrode signal is accepted.Delays of one minute for fast ISE’s and five to ten minutes for slow ISE’s are appropriate.
5.Set ion slope: use to enter a known ion electrode slope for a one point standardization.
6.Standardize menu: returns to the ion standardize menu.
7.Manual temperature menu: use to set a manual temperature for use in the absence of an ATC probe or when manually overriding the ATC.
8.Data alarm menu: use to enter ion limits to be entered. If the limits are exceeded an alarm indication (“
*
”) is displayed.
9.Set isopotential point: use to enter an isopotential point. See page 20.
0.Enable incremental: use to turn on the known addition/sub­traction type ion methods.See page 17.
Ion Mode
Channel A: ion mode
Standardize Menu
1) Enter a standard
2) Set ion slope
3) Clear standards
4) Options menu
5) Cal reminder menu
Channel A: ion mode
Options Menu
1) Select resolution
2) Select stability criteria
3) Select signal averaging
4) Set standardization delay
5) Set ion slope
6) Standardize menu
7) Manual temperature menu
8) Data alarm menu
9) Set isopotential point
0) Enable incremental modes
15
CAL!
S
16
Standardizing and Measuring Ion
Standardizing and Measuring Ion
1.Set the meter to ion mode (use Mode) and turn ON the channel (use Channel) with the Ion Selective Electrode (either Channel A or B).
The meter displays “— — —”,indicating no valid
measurement,until at least one ion standard has been entered.
2.Prepare a standard,and add the appropriate Ionic Strength Adjuster (ISA) solution to the stan­dard.
3.Immerse the electrode(s) in the solution and stir continuously.
4.Press Standardize,select the correct channel if prompted to do so,and select 1) Enter a stan- dard to add a standard. If this is the first standard to be entered,select the ion name and units.
Follow the prompts.Be sure to allow enough time for the electrode to reach a stable signal.
Note: The default standardization delay for ion mode is 30 seconds.This can be set by the user. See Ion Mode Options Menu,page 15.
5.The meter waits for a stable signal and enters the standard.The entered standard appears in the display (in single channel mode).Alternatively, if the signal is not stable,you can press Enter when the reading stabilizes according to your toler­ance criteria.The meter then enters the stan­dard.
6.Repeat steps 2 through 5 to enter additional standards.Up to seven standards can be entered.With more than one standard,the meter performs a diagnostic check on the electrode.
Helpful Hints:
• Provide stirring.
• Allow the electrode time to reach a stable read-
ing before entering the standard into the meter.
• To achieve better accuracy, standardize using at
least two standards,bracketing the expected range of your samples.
• Standardize from low to high concentrations.
• Always use fresh standards.
• Use standards and samples near the same tem-
perature.
• Remember to add Ionic Strength Adjuster to
each standard and sample.
mV
ion standards
log [ion]
Channel B: ion mode
A 25.0°C
No standards.
08/01/1998 08:40 PM
Channel B: ion mode
Ion Checklist
Prepare standard or sample. Add Ionic strength Adjuster. Rinse electrode. Place electrode in solution. Stir moderately. Press Enter to continue.
Standardize
Channel A: ion mode
Standardize Menu
1) Enter a standard
2) Set ion slope
3) Clear standards
4) Options menu
5) Cal reminder menu
Measuring Ion using Known Addition type (Incremental Ion) Methods
The meter provides known ("standard") addition/sub­traction and analate ("sample") addition/subtraction incremental methods for measuring ion concentra­tions. These advanced ion measurement techniques are useful in overcoming certain problems in ion analysis.
In known addition/subtraction a volume of sample is obtained,Ionic Strength Adjuster is added,and the ion electrode potential is obtained. Then a small vol­ume of standard is added to the sample,and a sec­ond electrode potential is obtained. From the change in electrode potential,the ion concentration in the sample can be calculated.Interference from complexation and other ions can often be overcome by the known (standard) addition method.
In analate addition/subtraction the ion electrode is placed in a volume of standard and the potential obtained. Then a small volume of sample is added and a second electrode potential is obtained. This method helps overcome problems from widely differ­ing sample ionic strengths or temperatures.
Enable Incremental Ion Modes Press Standardize, select the channel (if necessary),
4) Options menu, 0) Enable incremental modes, then
1) Yes.
This will "turn on" a special softkey in the main mea­sure screen which is a direct access softkey to start a known addition type measurement.
Using a known addition type incremental ion mea­surement
Press the incremental method softkey, select 1) Known (standard) addition or 2) Analate (sample) addition. Follow the prompts to place the electrode
in the first solution and obtain a reading,add an aliquot (a known volume) of standard or sample, obtain a second electrode reading,and enter the sample volume and standard volume and concen­tration. The meter then displays the calculated ion concentration in the original sample. Press Enter to leave the result screen and return to the measure screen to use direct reading ion measurements or start another known addition type measurement.
Channel A: ion mode
Enable Incremental Modes
Display ion soft key?
1) Yes
2) No If incremental modes are enabled, a ‘+/- ion’ softkey will appear on the measurement screen.
Select Measurement Technique
1) Known (standard) addition
2) Analate (sample) addition Select ‘known addition’ to add a small volume of known standard to the sample.
Select ‘analate addition’ to add a volume of sample to a volume of known standard.
The meter will automatically detect subtraction techniques.
Channel A: Known addition mode Incremental Ion Measurement
S 58.7 mV
Prepare sample and note volume. Add Ionic Strength Adjuster. Rinse electrode. Place electrode in solution. Stir moderately.
Press Enter to accept.
17
Channel A: Known addition mode
Incremental Ion Measurement
S 95.2 mV
Delta = 36.5 mV Add aliquot of known standard. Stir moderately. For best accuracy, add volume until Delta > 4.0 mV. Note total volume added.
Press Enter to accept.
+
18
Quick Start Guide for Conductivity/Resistivity/Salinity/TDS Measurements
1.Connect the conductivity/ATC cell to the meter.
2.Prepare one or more conductivity/resistivity stan­dard solutions at values near typical sample solu­tions.
3.Set the meter to display channel C using Channel. Set the meter to the correct mode (Conductivity,Resistivity,NaCl salinity,Practical salinity or Total dissolved solids) using Mode.
4.Place the conductivity cell in the standard, immerse the cell past the fill vent hole,then lift and allow the solution to drain out. Immerse and drain at least three times to fully flush the inner chamber of the cell. Gently tap the cell to dis­lodge any air bubbles.
5.Press Standardize,then 1) Enter a standard and follow the prompts to enter the value of the stan­dard. Repeat these steps to enter up to five con­ductivity/resistivity standards. Each standard is displayed in the main measuring screen when in single channel display.
6.Check the cell performance by pressing Cal Data to display the standards and the cell constants between standards.
Channel C: Conductivity mode
Select Mode
1) Conductivity
2) Practical salinity
3) NaCl salinity
4) Resistivity
5) Total dissolved solids
Channel C: Conductivity mode
Standardize Menu
1) Enter a standard
2) Set cell constant
3) Clear standards
4) Options menu
5) Cal reminder menu
Channel C: Conductivity mode
Cal Data
1.0051 /cm
100.1 µS/cm 08/01/98 08:45 AM
1.0092 /cm
1000 µS/cm 08/01/98 08:47 AM
Conductivity/ATC cell
c
h
. A
c
h
re
f. A
c
h
. C
C
o
T
n
e
m
d
u
p
c
. A
tiv
ity
T
e
m
p
c
. B
h
. F
F
E
T
R
S
2
3
2
C
p
o
w
e
r
Conductivity/Resistivity/Salinity/TDS Menus
Conductivity/Resistivity/Salinity/TDS Standardize Menu
Turn ON Channel C using the Channel key. Press Standardize and the Standardize Menu is displayed.
1.Enter a standard: use to enter or re-enter a conduc­tivity standard. Follow the prompts.
2.Set cell constant: use to manually enter a known conductivity cell constant for use with no standards. If the cell constant is known and stable,then this allows standardizing the cell without using standard solutions. If a standard is entered,the actual cell constant is calculated and used.
3.Clear standards: use to clear all existing standards. This is useful if new standards are to be entered.
4.Options menu: accesses additional settings used with each conductivity-type mode.See below.
5.Cal reminder menu: use to set a timer reminding you when to recalibrate. A icon appears on the main screen and an exclamation mark appears beside those standards which need to be re­entered.
6.[TDS mode only] Calculate solids factor: use this to allow the meter to calculate a solids factor. *
7.[TDS mode only] Set solids factor: use to manually enter a known solids factor * for a particular sample type.
* The "solids factor" is used to correlate the conductivi-
ty measurement with the weight based TDS mea­surement for a sample type.
Conductivity Mode Options Menu
1.Select resolution: use to set readings to 1 through 4 significant digits resolution.
2.Select stability criteria: sets stability criteria used to determine when the meter indicates Stable .
3.Select signal averaging: use to set filtering of the cell signal to very slow (average 10 readings),slow (8),medium (6), fast (4) or very fast (2 readings). Slower settings give more stable readings, and are recommended with conductivity measurements.
4.Set standardization delay: use to set a reading delay time used by the meter when entering conductivity standards.Programming a reading delay helps by ensuring sufficient time for the cell signal to become stable before being entered into the meter.
5.Set cell constant: use to manually enter a known conductivity cell constant. (Same as in Standardize Menu).
6.Standardize Menu: use to return to the Standardize Menu.
Channel C: Conductivity Mode
Standardize Menu
1) Enter a standard
2) Set cell constant
3) Clear standards
4) Options menu
5) Cal reminder menu
Options Menu
1) Select resolution
2) Select stability criteria
3) Select signal averaging
4) Set standardization delay
5) Set cell constant
6) Standardize menu
7) Manual temperature menu
8) Data alarm menu
9) Select display units
0) Set temperature coeff.
19
CAL!
S
20
Conductivity/Resistivity/Salinity/TDS Modes
7.Manual temperature menu: Use to set a manual temperature for use in the absence of an ATC probe or when manually overriding the ATC.
8.Data alarm menu: use to set ion limits to be entered.If the limits are exceeded an alarm indica­tion (“*”) is displayed.
9.Select display units: the meter automatically switch­es between uS/cm and mS/cm in conductivity, or between ohm-cm,Kiloohm-cm and Megohm-cm in resistivity. If it is better to display a fixed unit,the "Fixed" settings allow that.
0.Set temperature coefficient: use to set the reference temperature to correct all conductivity and TDS measurements to,and set the temperature coeffi­cient for the temperature correction. Salinity mea­surements by definition are corrected to 20°C. Resistivity measurements are not temperature cor­rected.
Standardizing and Measuring Conductivity/Resistivity/Salinity/TDS
1.Set the meter to display channel C (use Channel). Set the meter to the correct mode (Conductivity, Resistivity,NaCl salinity,Practical salinity or Total dis­solved solids) using Mode.
2.Place the conductivity cell in the standard,immerse the cell past the fill vent hole,then lift and allow the solution to drain out. Immerse and drain at least three times to fully flush the inner chamber of the cell. Gently ta p the cell to dislodge any air bub­bles.
3.Press Standardize, select the channel if necessar y, then 1) Enter a standard and follow the prompts to enter the value of the standard. Repeat these steps to enter up to five conductivity/resistivity standards. Each standard is displayed in the main measuring screen when in single channel display. Use multiple standards that cover the range of values expected in samples. Generally, standards should be a factor of ten apart in conductivity.
Channel C: Conductivity mode
Standardize Menu
1) Enter a standard
2) Set cell constant
3) Clear standards
4) Options menu
5) Cal reminder menu
Channel C: Conductivity mode
Set T
emperature Coefficient
Enter the value for this cell:
%/°C
Press Enter to accept.
Channel C: Conductivity mode Select Reference T
emperature
1) 15°C
2) 20°C
3) 25°C
Channel C: Conductivity mode
Select display units
1) Automatic
2) Fixed µS/cm
3) Fixed mS/cm
1.90
The meter will store up to 500 data points in an internal data log. Press Print when in the main mea­sure screen to store the current result with units, temperature,time,date,channel, sample number and stability in the data log. Pr int also outputs this data through the RS232 serial port. All channels dis­played are printed and data logged.
Data Log Menu
Press Data Log and the Data logging menu will appear.
1.View data log: Shows the stored data,one screen at a time.Press the arrow soft keys to page up and down through the stored data. Press Clear or Enter to return to the menu.
2.Data logging on/off: Turns the data logging on or off for all displayed channels.
3.Set logging frequency: Allows you to enter the time interval for automatic data logging.
4.Clear data log: Clear s all stored data points from memory.
5.Set sample number: Allows a number to be assigned to the first sample.This number will increment for each consecutive sample logged.
6.Print data log: Use to send all data points in the Data Log to the RS232 serial port.
Warning The Data log is kept in instrument RAM (Random Access Memory), which is powered by the external power supply. Removing power from the meter will lose the stored data in the Data log.
Note:
Meter settings and electrode standardization data are kept in separate non-volatile memory. Unplugging the meter has no effect on these stored items.
Data logging
!
Data Log Menu
1) View data log
2) Turn logging On/Off: ON
3) Set logging interval
4) Clear data log
5) Set sample number
6) Print data log
Number of points in log:
350
Data Log
A S 25.3C 4.176 pH
00102 06/26/98 9:32AM
B S 25.4C 1.03 mg/L F-
00101 06/26/98 9:31AM
A S 25.3C 6.713 pH
00100 06/26/98 9:31AM
21
Set Logging Interval
Enter the time between
data samples.
hr:min:sec
Please Enter to accept
01 : 00 : 00
Printing Log
Points remaining:
321
Press any to abort printing.
22
Testing the Electrode and Meter
To test the meter for correct operation with a pH, ORP or Ion Selective electrode,shor t the BNC input connector (either Channel A or B) using the BNC Shorting Cap that was supplied with the meter on the BNC connector(s).Select the correct channel using Channel. Select mV mode by pressing Mode and selecting 2) mV. Verify meter is in absolute mV mode (display shows “mV”, not “rel mV”). If the meter reads 0 ± 0.1 mV*, and is sta ble, the meter is measuring correctly.
To test the pH electrode,place it in a fresh pH 7 buffer.Select the correct channel for the electrode using Channel.Press Mode and select 2) mV.Verify that the meter is in absolute mV mode (display shows "mV",not "rel mV") and note the mV reading. Repeat for either a pH 4 or pH 10 buffer. If the elec­trode potential is within the limits shown,it is mea­suring correctly.
pH 7 0 ± 30 mV pH 4 159 to 186 mV higher than pH 7 reading pH 10 159 to 186 mV lower than pH 7 reading
* Note: Meter accuracy is ±0.1 mV at calibration temperature,not including long term drift and a temperature error. The zero and slope temperature coefficients of the meter over the range of 15 to 40°C specify 85ppm/°C,or ±4 mV at full scale (worst case).The long ter m dr ift will not exceed 0.1 mV per month.
Troubleshooting
PHB-215, PHB-220, PHB-225 and PHB-250
Modes pH mV Temperature Range -2.000 to 20.000 ±1800.0 -5.0 to 105.0˚C Resolution 0.001/ 0.01 / 0.1 0.1 / 1 0.1 Accuracy ±0.002 ±0.1 ± 0.3 Temperature
Automatic & manual: -5 to 105˚C
Compensation Slope Control Automatic,90 to 105%
Manual,80 to 120%
Environmental
15 to 40°C,humidity from 0 to 90% (noncondensing)
Conditions Power Requirements 115V 50/60Hz (Additional voltages available) Ordering Information
Meter with kit includes: Meter,power supply,high performance glass-body pH/ATC “3-in-1”
electrode,electrode ar m and operation manual.
Meter only kit includes: Meter,power supply, electrode arm and operation manual.
Models 225 and 250
Mode Ion Range 1.00E-9 to 9.99E9 Resolution 1,2,or 3 significant figures Accuracy ± 0.17%n (n = ion charge) Slope Control Automatic or manual,5.9 mV/decade (10% slope)
to 71 mV/decade (120%)
PHB-220 and PHB-250
Practical NaCl
Mode Conductivity Resistivity Salinity Salinity TDS Range 0.01 – 300,000 µS/cm• 30 - 20M½•cm* 0.01 - 42 ppt* 0.01 - 70ppt* 0.005 - 150,000* Resolution 1,2, 3, or 4 significant figures Accuracy ± 0.5% ±0.01µS/cm Cell Constant Automatic or manual,0.01 - 100 /cm Temperature Coeff Off or On (0 - 4% /°C) Number of Standards 5
*depending on cell constant (from 0.1 to 10 /cm)
Meter Specifications
23
24
pH Theory
The measurement of pH plays an important role in water quality, industry and research.pH is a measure of acidity or alkalinity of a solution,and is usually written:
pH = -log [H+]
Where [H+] is the concentration of hydrogen ions. pH levels generally range from 0 to 14,with a pH value
of 7 being the neutral point.pH values greater than 7 are alkaline,and pH values less than 7 indicate acidic solutions.
Conventional pH meters use a combination glass pH electrode,which includes a reference electrode.The reference electrode provides a stable reference point and completes the electrical circuit.The pH meter reads the voltage of the two electrodes,converts it to pH units, and displays the result.
These meters can also use a Field Effect Transistor (FET) pH/ATC electrode for measuring pH.The FET uses an ion-sensing solid state membrane attached to a transis­tor to measure the hydrogen ion concentration of a solution.These non-glass pH electrodes offer durability, dry storage and no glass construction.
The electrode signal varies with the pH,according to the Nernst Equation:
E = E° + S • log [H+]
Where:
E = measured electrode potential E° = standard potential of the system (constant) S = slope
pH 0 Strong Acid 1 2 Lemon Juice 3 4 Tomato Juice 5 Coffee 6
Neutral 7 Pure water
8 Baking Soda 9 10 11 12 Ammonia 13 14 Strong Base
More Acidic
More Basic
pH scale showing the relative acidity or basicity of some common substances
Ion Selective Electrode Theory
The measurement of ions plays an important role in water quality, industry,research and environmental monitoring. Ion-selective Electrodes (ISE’s) respond, more or less exclusively, to a specific type of ion in solution.The particular ion to which an ISE responds depends on the chemical makeup of its sensing membrane.ISE’s operate according to a form of the Nernst equation:
E = Eo+ S • log [ion]
The Isopotential point is the potential of an electrode system which does not change with temperature.Typical pH electrodes have isopotential points near zero mV (which is the default setting for the meter).For high accuracy pH measurements,or for ion measure­ments where the sample temperature may widely vary,the isopotential of the pH or ion electrode may be experimentally determined and entered into the meter.
• Prepare a set of buffers or ion standards spanning the linear range of the electrode. Place the buffers or standards in a temperature bath at known temperature.
• Place the meter into mV mode.
• Measure and record mV readings of each pH or concentration,and repeat at sev­eral temperatures.
• Plot the log of concentration or pH value versus mV reading.
• Connect the points for each temperature.
Where the lines intersect is the Isopotential point.
Determining the Isopotential Point
200
100
0
10
-3
10
-4
Log of Ion Concentration, moles/L
Electrode
Output
(mV)
10°C
60°C
Isopotential Point:
3.07 x 10-4moles/L, 162 mV
Ion Electrode Isopotential Point
50
150
-180
0
180
4 7 10
pH
Electrode
Output
(mV)
25°C
10°C
Isopotential Point:
7.04pH 9 mV
pH Electrode Isopotential Point
25
26
Omega meters have a bi-directional RS-232 serial port,which can be used to send commands to the meter and output data from the meter. Special char­acters (W, µ, é) are coded using ASCII (not ANSI); use an ASCII font like “Terminal”.Also use a ter minal emulation like TTY or ANSI,not VT100. Serial commands follow either "keystroke" mode or high level command mode consisting of "SET","GET" and "DO" instructions. (Note: “GET”and “DO”are optional).
Keystroke instructions
KEYS M Mode
Z Standardize C Cal Data H Channel S Setup L Data Log R Clear N Enter/Print [0 to 9] Equivalent to pressing a numeric key
- +/- key . Decimal key E Used within a number to enter values in exponential form ! Press softkey #1 (at top,usually Help) @ Press softkey #2 (usually return to measuring screen) # Press softkey #3 (usually up arrow) $ Press softkey #4 (at bottom,usually down arrow)
Notes: Key commands are acknowledged by the meter with a reply
Keys = COMMAND_RECEIVED.Multiple keys can be concatenated together into a single command,for example,Keys Z413@ (Standardize,options, resolution, set to 3,main),or Keys Z4721.2-N@ (Set manual temperature to -1.2).
High level instructions
Use commands SET,GET,DO. Follow command by a keyword like MODE,STDZPH,STDZCONDO, CALDATA,STDZCLEAR, CHANNEL,DATETIME,TIMESTAMP,DISPLAY, READ, INFO Typical Syntax: [command] [keyword] [channel] [variable(s)]
Error conditions are replied to with an Error response; for example:
"Error: Need channel","Error: Need mode","Error: Unspecified" Accepted commands have a response; indicated below for each com­mand.
Mode operations set mode "channel character" "mode id"
Examples: SET MODE A PH
SET MODE B MV SET MODE B ION SET MODE C CONDUCTIVITY
Returns confirmation; for example "SET MODE A PH" returns "A mode = PH". Valid modes are (depending on the meter model) PH,MV,CONDUCTIVITY, RESISTIVITY, PRAC_SALINITY, NACL_SALINITY, DISSOLVED_SOLIDS.
RS-232 Serial Interface Meter Command Set
27
RS-232 Serial Interface Meter Command Set
[get] mode "channel character" "mode id"
Examples: [GET] MODE A
Returns mode information on selected channel; "A Mode = MV".
Channel operations set channel "channel character" "on|off"
Examples: SET CHANNEL A ON
SET CHANNEL C OFF
Returns confirmation; for example "A Channel = ON". [get] channel "channel character" Example: [GET] CHANNEL B
Returns channel information; "A Channel = OFF". (DO) READ “channel character” (Take reading w/temperature with-
out sending to data log).
Standardization operations [DO] STDZPH "channel character" [DO] STDZCLEAR "channel character" [DO] STDZCONDO "conductivity standard value" [DO] CALDATA "channel character"
[DO] STDZPH A [Auto-enter a buffer] Returns "Stdz pH = COMMAND_RECEIVED",followed by the Calibration Data printout.
[DO] STDZCLEAR A [Clear all buffers/standards.] Returns "Stdz Clear = COMMAND_RECEIVED".
[DO] STDZCONDO 1000 [Enter a standard of 1000 uS/cm.] Returns "C Stdz Condo = COMMAND_RECEIVED",followed by the Calibration Data printout.
[DO] CALDATA A Returns "A Cal Data = COMMAND RECEIVED" followed by the Calibration Data printout.
General meter setup operations
SET DATETIME MM/DD/YYYY HH:MM:SS [Leading 0's required, 24
hour time]
[GET] DATETIME Returns: "MM/DD/YYYY
HH:MM:SS"
[GET] INFO Returns Model,Ver sion,
Serial#. (DO) DISPLAY display_text_string (at 0,0) [x = pixel from left,0 - 319] (DO) DISPXY x y display_text_string [y = pixel from top,0 - 239]
Example: DO DISPXY 15 0 Device ready, press any key. SET TIMESTAMP # (Set date/time using ‘unix’
(GET) TIMESTAMP seconds).
28
Maintenance
This product contains no user serviceable parts . All replacement parts should be obtained from the manufacturer or an authorized distributor.
Cleaning
The exterior surfaces of this product may be cleaned with a damp cloth or with mild detergent.
CAUTION
Changes or modifications not expressly approved by the manu­facturer will void the user’s warranty for this equipment.
!
WARRANTY/DISCLAIMER
OMEGA ENGINEERING, INC. warrants this unit to be free of defects in materials and workmanship for a period of 37 months from date of purchase. OMEGA Warranty adds an additional one (1) month grace period to the normal three (3) years product warranty to cover handling and shipping time. This ensures that OMEGA’s customers receive maximum coverage on each product.
If the unit should malfunction, it must be returned to the factory for evaluation. OMEGA’s Customer Service Department will issue an Authorized Return (AR) number immediately upon phone or written request. Upon examination by OMEGA, if the unit is found to be defective it will be repaired or replaced at no charge. OMEGA’s WARRANTY does not apply to defects resulting from any action of the purchaser, including but not limited to mishandling, improper interfacing, operation outside of design limits, improper repair, or unauthorized modification. This WARRANTY is VOID if the unit shows evidence of having been tampered with or shows evidence of being damaged as a result of excessive corrosion; or current, heat, moisture or vibration; improper specification; misapplication; misuse or other operating conditions outside of OMEGA’s control. Components which wear are not warranted, including but not limited to contact points, fuses, and triacs.
OMEGA is pleased to offer suggestions on the use of its various products. However, OMEGA neither assumes responsibility for any omissions or errors nor assumes liability for any damages that result from the use of its products in accordance with information pro­vided by OMEGA, either verbal or written. OMEGA warrants only that the parts manufactured by it will be as specified and free of defects. OMEGA MAKES NO OTHER WARRANTIES OR REPRESENTATIONS OF ANY KIND WHATSOEVER, EXPRESSED OR IMPLIED, EXCEPT THAT OF TITLE, AND ALL IMPLIED W ARRANTIES INCLUDING ANY WAR­RANTY OF MERCHANTABILITY AND FITNESS FOR A P AR TICULAR PURPOSE ARE HEREBY DISCLAIMED. LIMITATION OF LIABILITY: The remedies of purchaser set forth herein are exclusive and the total liability of OMEGA with respect to this order, whether based on con­tract, warranty, negligence, indemnification, strict liability or otherwise, shall not exceed the purchase price of the component upon which liability is based. In no event shall OMEGA be liable for consequential, incidental or special damages.
CONDITIONS: Equipment sold by OMEGA is not intended to be used, nor shall it be used: (1) as a “Basic Component” under 10 CFR 21 (NRC), used in or with any nuclear installation or activity; or (2) in medical applications or used on humans. Should any Product(s) be used in or with any nuclear installation or activity, medical application, used on humans, or misused in any way, OMEGA assumes no responsibility as set forth in our basic WARRANTY/DISCLAIMER language, and additionally, purchaser will indemnify OMEGA and hold OMEGA harmless from any liability or damage whatsoever arising out of the use of the Product(s) in such a manner .
RETURN REQUESTS / INQUIRIES
Direct all warranty and repair requests/inquiries to the OMEGA Customer Service Department. BEFORE RETURNING ANY PRODUCT(S) TO OMEGA, PURCHASER MUST OBTAIN AN AUTHORIZED RETURN (AR) NUMBER FROM OMEGA’S CUSTOMER SERVICE DEPARTMENT (IN ORDER TO AVOID PROCESSING DELAYS). The assigned AR number should then be marked on the outside of the return package and on any correspondence.
The purchaser is responsible for shipping charges, freight, insurance and proper packaging to prevent breakage in transit.
FOR W
ARRANTY RETURNS, please have the following information available BEFORE contacting OMEGA:
1. P.O. number under which the product was
PURCHASED,
2. Model and serial number of the product
under warranty, and
3. Repair instructions and/or specific
problems relative to the product.
FOR NON-WARRANTY REPAIRS,
consult OMEGA for current repair charges. Have the following information available BEFORE contacting OMEGA:
1. P.O. number to cover the COST
of the repair,
2. Model and serial number of product, and
3. Repair instructions and/or specific problems
relative to the product.
OMEGA’s policy is to make running changes, not model changes, whenever an improvement is possible. This affords our customers the latest in technology and engineering.
OMEGA is a registered trademark of OMEGA ENGINEERING, INC. © Copyright 1996 OMEGA ENGINEERING, INC. All rights reserved. This document may not be copied, photocopied,
reproduced, translated, or reduced to any electronic medium or machine-readable form, in whole or in part, without prior written consent of OMEGA ENGINEERING, INC.
UUSASA
MADE
ININ
Where Do I Find Everything I Need for
Process Measurement and Control?
OMEGA…Of Course!
TEMPERATURE
MU
Thermocouple, RTD & Thermistor Probes, Connectors, Panels & Assemblies
MU
Wire: Thermocouple, RTD & Thermistor
MU
Calibrators & Ice Point References
MU
Recorders, Controllers & Process Monitors
MU
Infrared Pyrometers
PRESSURE, STRAIN AND FORCE
MU
Transducers & Strain Gauges
MU
Load Cells & Pressure Gauges
MU
Displacement Transducers
MU
Instrumentation & Accessories
FLOW/LEVEL
MU
Rotameters, Gas Mass Flowmeters & Flow Computers
MU
Air Velocity Indicators
MU
Turbine/Paddlewheel Systems
MU
Totalizers & Batch Controllers
pH/CONDUCTIVITY
MU
pH Electrodes, Testers & Accessories
MU
Benchtop/Laboratory Meters
MU
Controllers, Calibrators, Simulators & Pumps
MU
Industrial pH & Conductivity Equipment
DATA ACQUISITION
MU
Data Acquisition & Engineering Software
MU
Communications-Based Acquisition Systems
MU
Plug-in Cards for Apple, IBM & Compatibles
MU
Datalogging Systems
MU
Recorders, Printers & Plotters
HEATERS
MU
Heating Cable
MU
Cartridge & Strip Heaters
MU
Immersion & Band Heaters
MU
Flexible Heaters
MU
Laboratory Heaters
ENVIRONMENTAL MONITORING AND CONTROL
MU
Metering & Control Instrumentation
MU
Refractometers
MU
Pumps & Tubing
MU
Air, Soil & Water Monitors
MU
Industrial Water & Wastewater Treatment
MU
pH, Conductivity & Dissolved Oxygen Instruments
M2834/0298
301121.1
Loading...