Before attempting to interface the PCL9001 with any other device, carefully read the
following instructions.
II.. GGEENNEERRAALL DDEESSCCRRIIPPTTIIOONN
The PCL9001 is designed as a full function pressure calibration system combining highly accurate
pressure and electrical range in a simple to use hand-held device. The PCL9001 also incorporates a
built in 24 volt loop power supply to simultaneously power the loop while reading the resultant current. Arugged manifold allows quick connections for the device under test and or connection of the
optional hand pump.
IIII.. IINNIITTIIAALL PPRROOCCEEDDUURREESS
UUNNPPAACCKKIINNGG
Remove the Packing List and verify that all equipment has been received. If there are any questions
about the shipment, please call OMEGA Engineering at 1-800-826-6342.
Upon receipt of shipment, inspect the container and equipment for any signs of damage. Take particular note of any evidence of rough handling in transit. Immediately report any damage to the shipping
agent.
NNOOTTEE
The carrier will not honor any claims unless all shipping material is saved for their
examination. After examining and removing contents, save packing material and carton in the
event reshipment is necessary.
Check to see if your calibrator is complete - it should include:
Calibrator only includes: Calibrator Kit includes:
• Carrying Case • Carrying Case
• T est Leads • T est Leads
• Batteries already installed • Batteries already installed
• Allen Wrench • Allen Wrench
• T eflon Tape • T eflon Tape
• Instructions & Specifications • Instructions & Specifications
• Battery Eliminator/Charger
• Pump
Remove sliding battery cover door located on the bottom of the instrument and verify battery installation.
Become familiar with the designation and polarities of the five jacks located below the keypad of the
calibrator. Refer to the back panel for polarity information.
On the right hand side of the calibrator is the connection for the RS-232 port. This connection is only
used when data communication is required between the PCL9001 and another device. Refer to section
V for more detailed information.