Standard
U 0.05% FS Accuracy U Gage and Absolute
Pressure Ranges
U Portable and Lightweight U NIST Certificate Provided U Hoses, Fittings,
Connectors Included
U Voltage/Current Display
Mode for Transducer Testing
U Overpressure Protection
For rapid on-site calibration of transducers, transmitters, gages, and many other pressure-sensitive devices, OMEGA® PCL-5000 Series portable pressure calibrators provide high accuracy, reliability, and pushbutton operation.
The PCL-5000 covers the higher pressure ranges (20 to 2000 psi) with a self-contained 2000 psi nitrogen cylinder. Its pressure modes—gage or absolute—are switch selectable. An extremely accurate instrument (±0.05% FS), the PCL-5000 comes with quick­disconnect fittings and hoses for input and output pressure.
PCL-5000A shown smaller than actual size.
The PCL-5100, a low-pressure unit with ranges from 30 inHg to 100 psi, offers 5 independent selectable ranges. Designed for field use, this rugged calibrator weighs less than 25 lb and has its own built-in pressure and vacuum source. For immediate on-site use, it comes with with hose and couplings. Its internal microcircuitry automatically maintains zero and span calibration data to ensure long-term stability and accuracy. A pressure limit control provides overpressure protection.
K-79-2 pocket thermometer is included at no extra charge. To order extra pocket thermometers, visit omega.com/k-79
A-45
0.05% ACCURACY, 30 inH20 to 2000 psi RECHARGEABLE BATTERIES—STANDARD
A
PRESSURE INSTRUMENTS
Hoses and couplings included.
SPECIFICATIONS
Overall Accuracy: ±0.05% FS max;
accuracy statement includes all effects of linearity, hysteresis, repeatability and ambient temperature
Operating Temperature: 5 to 43°C (41 to 109°F)
Internal Pressure Cylinder: Capacity: 9.0 standard cubic feet
Volume: 80 cubic inches Rating: 2015 psi
Overpressure Rupture Disk Rating: 3000 psig, nominal
Pressure Media Filter: Rating: 7 microns, nominal
Type: Field-replaceable filter cartridge
Control Valves: Micrometering with
positive shut-off, adjustable stop and replaceable seat; non-destructible.
Fill and Test Ports: Rating: 3000 psig, min (connected or disconnected) Type: Quick-disconnect
Pressure Hoses: 2 (1 input, 1 output)
1.5 m (5'), nominal each Built-In Carrying Case: Rugged
aluminum case with sealed cover and carrying handle
PCL-5100 shown smaller than actual size.
Weight: 11 kg (24 lb)
(including all hoses and cables)
Dimensions: 250 W x 410 L x 290 mm D (10 x 16 x 11.5")
Metal cover included.
To Order Visit omega.com/pcl-5000 for Pricing and Details
MODEL NO. TYPE RANGES RESOLUTION
PCL-5000A psig and psia 2000/1000/400 0.5/0.2/0.05 PCL-5000B psig and psia 1000/500/200 0.2/0.1/0.05 PCL-5000C psig and psia 500/250/100 0.1/0.05/0.002 PCL-5000D psig 100/50/20 0.02/0.01/0.005 PCL-5000DB psia 100/50/20 0.02/0.01/0.005
) PCL-5100 5 ranges: ±30 inHg to 100 psi 0.5/0.1/0.01
OPTION
SUFFIX DESCRIPTION
-2 Second scale, inH2O, on PCL-5000D and PCL-5000DB
ACCESSORY
MODEL NO. DESCRIPTION
K-79-2 Extra pocket thermometer, -40 to 160°F range
Comes complete with hoses, couplings, K-79-2 pocket thermometer, NIST traceable calibration certificate, and operator’s manual.
Ordering Examples: PCL-5000D-2, 100/50/20 psig range calibrator with second scale option.
PCL-5100, 5 range, ±30 inHg to 100 psi calibrator.
Easy carry handles
A-46