HIGH-ACCURACY PRESSURE CALIBRATORS
SINGLE AND DUAL RANGE MODELS
PCL341 Series
PCL342-30 pressure
calibrator complete
with 4 “AA” batteries,
test leads, soft case,
and NIST calibration
certificate, $1895.
Starts at
$
1495
Standard
A
PRESSURE INSTRUMENTS
⻬ Triple Display: Pressure,
Temperature, Process
Voltage and Current
⻬ Isolated, Stainless Steel
Pressure Sensors with
0.025% FS Accuracy
⻬ Temperature
Compensation
Ensures Accuracy in
Field Applications
⻬ External Connection
Supports All OMEGA
PCL-PM Series Pressure
Modules (25 Ranges) for
Greater Measurement
Capability
⻬ Pt100 RTD Input
for Temperature
Measurement—
Accurate to 0.1°C (0.2°F)
⻬ Measures 4 to 20 mA Input
⻬ Internal 24 V Loop Power
Supply Can Power a
Transmitter Under Test
⻬ Measures Up to 30 Vdc
⻬ Large, Graphic-Style
LCD with Backlight
®
⻬ Can Display Up to
3 Inputs Simultaneously
(e.g., Pressure 1,
Pressure 2, and RTD)
⻬ Ideal for Gas Flow
Calibration (Custody
Transfer) Applications
⻬ Switch Test on
Pressure 1, Pressure 2,
or External Pressure
Module
⻬ Operating-Mode
Layers Support Both
A-28
1
⁄4 FNPT
ports for
pressure
use.
See page A-32 or visit
omega.com/pcl_pm for
available pressure modules.
Simple and Complex
Applications
⻬ Up to 5-Setups Can be
Stored; Last Setup
Automatically Recalled
on Power-Up
⻬ % Error and Damping
Functions
⻬ Small, Rugged,
Compact Design
⻬ Operates on 4 Standard
“AA” Batteries (Included)
HIGH-ACCURACY
PRESSURE CALIBRATORS
OMEGA’s PCL341 and PCL342 set a new
standard for handheld pressure calibrators.
Both models have internal, isolated, stainless
steel pressure sensors with 0.025% FS
accuracy. For greater measurement capability,
an external connection supports all OMEGA
PCL-PM Series pressure modules. A Pt100 RTD
input allows temperature readings accurate to
0.1°C (0.2°F) (One Pt100 RTD probe comes with
the “-KIT” models). The PCL341 and PCL342
measure 4 to 20 mA loop current and up to
30 Vdc. An internal 24V loop power supply can
power a transmitter under test.
PCL340-030, $1099,
see page A-27 for
more information.
The PCL342 can display up to 3 measurements
simultaneously—for example, 3 pressure
measurements. The ability to display 2 pressure
readings along with an RTD temp makes the
PCL342 ideal for gas flow calibration (custody
transfer) applications.
The PCL341 is available in 9 ranges from
5 to 10,000 psig. The PCL342 comes in
2 standard ranges—15/1500 psi or
30/3000 psi—or the user can select 2 ranges
from 5, 15, 100, 300, 500, 1000, 3000, 5000,
or 10,000 psig. It also comes in a kit, which
includes the PCL342 pressure calibrator,
Pt100 RTD probe, 4 “AA” batteries, operator’s
manual, NIST-traceable certificate,
test leads, and hard-sided carrying case.
SPECIFICATIONS
(18 to 28°C Unless Otherwise Noted)
General
Can store up to 5 setups; recalls last used on power-up
Operating Temperature: -10 to 50°C (14 to 122°F)
Storage Temperature: -20 to 60°C (-4 to 140°F)
Power Requirement: 6 Vdc
Batteries: 4 standard “AA” cells (included)
Battery Life: >35 hours, typical usage
Dimensions: 211 H x 99 W x 46 mm D
(8.3 x 3.9 x 1.8")
Weight: 0.57 kg (1.25 lb)
Connectors/Ports: Pressure one (PCL341) or two
(PCL342),
1
⁄8 NPT OMEGA-P pressure module and RTD
See page
A-32 or visit
omega.com/
pcl_pm for
available
pressure
modules.
Ranges
Available Pressure (Select Any Two):
5, 15, 30, 100, 300, 500, 1000, 3000, 5000, 10,000 psi
mA: 0 to 24 mA
V: 0 to 30 Vdc
RTD: -40 to 105°C (-40 to 220°F)
Engineering Units (Keypad Selectable):
psi, bar, mbar, kPa, kg/cm
4 or 20°C, inH2O @ 4 or 20°C, inH2O @ 60°F, mmHg, inHg,
ftH2O @ 4 or 20°C, or 60°F
2
, cmH2O @ 4 or 20°C, mH2O @
Accuracy
Pressure: 15 to 3000 psi ±0.025% FS; for all ranges 5, 5000,
10,000 psi ±0.035% FS
mA: ±0.015% of rdg ±0.002 mA
V: ±0.015% of rdg ±0.002 V
RTD: ±0.015% of rdg ±0.02 Ω or ±0.1°C @ 0°C for Pt100
Temperature Effect: No effect on accuracy from
15 to 35°C. Add ±0.002% FS/°C for temps falling outside
the 15 to 35°C range.
INTERNAL PRESSURE MODULE RANGES
psig 5.0000 15.000 30.000 100.00 300.00 500.00 1000.00 3000.0 5000.0 10,000.0
bar 0.7000 1.0342 2.0684 6.8947 20.684 34.474 68.9475 206.84 344.74 689.475
kPa 70.000 103.42 206.84 689.48 2068.4 3447.4 6894.75 20684 34474 68947.6
A-29