TRIPLE-RANGE HIGH-ACCURACY
PRESSURE STANDARDS
DUAL DISPLAY—DIGITAL AND BAR GRAPH
Used to Test and
Calibrate Pressure
PCL-3000 Series
Standard
U Absolute and Gage
Pressure Measurement
U Accuracy of ±0.05% FS
U Ranges from 10 to
10,000 psi
U Automatic Zeroing (No
Potentiometer Needed)
U Self-Calibrating
U RS232 Output Optional
U Integral Stainless Steel
Transducer
PCL-3000 Series digital pressure
standards are designed for use in the
manufacture, testing, or calibration
of pressure-sensitive devices. Using
a patented, bonded-foil, strain-gage
sensor and advanced microcircuitry,
these rugged, compact instruments
provide simultaneous digital and
analog readouts. The digital display
shows precise pressure information,
while the analog display gives
direction and pressure level.
Features include auto-zeroing
and self-calibration. A front-panel
knob allows the user to select
pressure ranges.
SPECIFICATIONS
Accuracy: ±0.05% FS max, each range.
Accuracy statement includes linearity,
hysteresis and repeatability.
Operating Temperature: 10 to 43°C
(50 to 109°F)
Pressure Media: Any media compatible
with 17-4 PH stainless steel
Overpressure Capability: 750% FS on
lowest range; 300% FS on middle range;
150% FS on highest range
Display Resolution: Nominally
0.02% FS for each pressure range
Transducers
PCL-3000B shown smaller than actual size.
Reading Rate: 10/s
Input Voltage: 117 or 220 Vac
±10%, 50/60 Hz; voltage selected via
transformer taps
Pressure Connection:
7/16-20 UNFM
To Order
MODEL NO. RANGE TYPE
PCL-3000A 5000/2500/1000 psig and psia
PCL-3000B 1000/500/200 psig and psia
PCL-3000C 500/250/100 psig and psia
PCL-3000D 100/50/20 psig with conversion to inH
PCL-3000E 50/25/10 psig with conversion to inH
PCL-3000G 10000/5000/2000 psig and psia
OPTIONS
SUFFIX DESCRIPTION
-1
-4 RS232 simplex output
-5 Peak hold*
-7 Battery operation**
-9 Panel mounting kit
-10 Freeze mode for testing pressure switches*
The peak hold and freeze mode options require a setup charge when ordered in one unit.
*
Please list as separate line item.
Excludes bar graph display.
**
Comes complete with pocket thermometer, NIST traceable certificate and operator’s manual.
Ordering Examples: PCL-3000C-4, 500/250/100 psig and psia range unit, with RS232
simplex output.
PCL-3000B, 1000/500/200 range unit, psig and psia.
Visit omega.com/pcl-3000 for Pricing and Details
Analog output, selectable 0 to 10 Vdc or 4 to 20 mA, isolated
Power Consumption: 8 W, typical
Nominal Size: 27 W x 12 H x 25 cm D
(10.8 x 4.9 x 9.9")
Panel Cutout: 26.9 x 12 cm
(10.62 x 4.75")
Weight: 2.3 kg (5 lb)
O
2
O
2
A-47