Omega Products PCL-2A-E User Manual

BENCHTOP PRESSURE STANDARDS
WITH INTERCHANGEABLE RANGE MODULES
PCL-2A Series
PCL-2A, $2430, with 2 pressure modules, PCL-MB-010G, $1110, and PCL-MB-1KG, $1265.
Data Logging and
Hi/Lo Programmable
Relays Standard
Starts at
$
2430
Standard
Interchangable
Pressure Modules Available for Ranges from 0.25 inH2O to 7500 psi
Large Display Indicates
Dual Readings Simultaneously
Accurate to 0.06% FS
The PCL-2A is the perfect companion product to OMEGAs PCL-1B handheld calibrator. For the ultimate in measurement flexibility, this benchtop or panel mount unit uses the same PCL-M pressure modules as the PCL-1B (for details, see page A-42).
An intuitive menu-driven user interface puts all of the PCL-2A power at the simple press of a key. It uses PCL-M plug-in modules to provide the ultimate in measurement flexibility.
A-39
Two plug-in pressure modules allow range changes without calibration, shown smaller than actual size. Modules sold separately, see page A-42.
Now you can purchase a calibration system to serve your field and bench mounting requirements while enjoying the significant cost savings gained through the availability of commom pressure modules and software.
BENCHTOP PRESSURE CALIBRATORS
Standard Features
• Measurement of pressure, temperature, voltage and current
• Two-channel—simultaneous measurement and display of pressure, temperature, voltage or current in any combination
• Accuracy ratings of to 0.05% FS (pressure)
• Pressure ranges from
0.25
inH2O to 7500 psi
• Multiple engineering units— 12 factory-programmed plus 1 user-programmable engineering unit
• Programmable damping
• Tare capability
• Display hold
• RS232 two-way communications
• Standard NIST-traceable certificate of calibration
• Data logging—automatic, manual and delayed actuation (user programmable)
• Relays—Hi/lo programmable configurations—N/O and N/C
Optional Features
• Supplies 24 Vdc to power transmitters, transducers, switches
• High-static differential pressure measurement with dual sensor
• Battery power—5 “AA” Ni-Cads with built-in charger (included)
Application Capabilities
• Switch testing, including trip, reset and deadband
• Leak detection in either leak rate or pressure decay modes
To Order (Specify Model Number)
MODEL NO. PRICE DESCRIPTION
PCL-2A $2430 Base unit, including data logging, high and low alarm
relays and backlit display (requires 1 or more plug-in pressure modules, see page A-42)
ORDER ADD’L SUFFIX PRICE DESCRIPTION
-B $155 Rechargeable Ni-Cad batteries
-E 365 Built-in 24 Vdc loop power supply
Comes complete with calibration certificate and operator’s manual.
Ordering Example: PCL-2A-B, benchtop pressure calibrator with Ni-Cad battery option, PCL-MA-100BWC, ±100 inH2O, and PCL-MB-030G, 0 to 30 psig plug-in pressure module, $2430 + 155 + 1295 + 1110 = $4990. Modules shown on page A-42.
• Flow velocity and volume measurement using standard pitot tubes and anubars
• Automatic % error calculation simplifies transmitter or transducer calibration
• Valve test and set using the min/max tracking feature
AVAILABLE FOR FAST DELIVERY!
A
PRESSURE INSTRUMENTS
mA and V display
Menu-driven setup
Push to zero
Flow velocity and volume
Data log up to 714 records
Control backlight
Min/max memory
12 engineering units at the press of a key
Trip detector for switch calibration
A-40
Tare to remove pressure preload
14.2
(0.56)
234.7
(9.24)
27.94 (1.10)
160.8 (6.33)
84
(3.31)
38.3
(1.51)
31.75
(1.25)
171
(6.74)
101.6 (4.00)
88
(3.47)
36.5
(1.44)
23.6
(0.93)
24.6
(0.97)
BENCHTOP PRESSURE CALIBRATORS
SPECIFICATIONS
Physical
Dimensions: 102 H x 171 W x 277 mm D
(4.00 x 6.74 x 10.9")
Panel Cutout: 90 H x 167 mm W (3.53 x 6.56") Weight: 9.0 kg (4.08 lb) with 2 modules max Case: High impact ABS Sensor Module Capacity: 2 bays for PCL-M series
plug-in modules Display: Backlit 2 line LCD, 0.94 mm (0.037")
height per line; display of two plug-in modules simultaneously
Electrical Connection: Standard banana jacks Calibration: Both plug-in modules and electronics
base unit may be field calibrated
Operating
Operating Ambient: 0 to 49°C (32 to 120°F) Storage Ambient: -20 to 70°C (-4 to 158°F) Update Rate: 130 ms nominal, with one sensor installed Resolution: ±0.002% span, 60,000 count max Warm-Up: 5 m for rated accuracy Serial Interface: RS232, baud rates of 300/1200/2400/9600 Power: 110 Vac; optional NiCad batteries with built-in
charger (or 5 “AA” alkaline batteries, non-rechargeable) Battery Life: Up to 20 hr with no backlight, 2 hr with
backlight on Data Logging: Manual or automatic; store up to 715
measured values for upload to PC (utility software included) High-Low Relay: Programmable setpoints for activation of
alarms or control valves
Voltage/Current Input
Input Range Accuracy
0 to 10 Vdc ±0.025% FS 0 to 30 Vdc ±0.10% FS
0 to 20 mA ±0.03% FS 0 to 50 mA ±0.05% FS
Temperature Effect: ±0.0005% span per °C (0.001% span per °F) over compensated range
PCL-M PLUG-IN MODULES
See page A-42 for ranges and specifications and ordering information
Gage/Differential/Compound Pressure Modules Input Ranges: 0.25 to 200 inH2O gage; ±0.125 to
±100 inH2O differential; non-isolated, for clean, dry, non-conductive, non-corrosive gases
Accuracy: ±0.07% span (ranges below 2 inH ±0.06% span (2 to 200 inH2O ranges)
Media: Clean, dry, non-conductive, non-corrosive gas Under/Overpressure: -15 to 50 psi Max Static (Line) Pressure: 100 psi Gage/Absolute/Vacuum/Differential Pressure Modules
Input Ranges: 5 to 7500 psig or psia; 5 to 15 psi vacuum;
±5 to ±15, -15 to 30 or -15 to 60 psi compound
Accuracy: 0.05% span (0.1% for 10k psi range) Media: Any medium compatible with 316 SS isolation; 5 psi
modules are non-isolated, for clean, dry, non-conductive, non-corrosive gases only
2
O);
PCL-2A, $2430.
Overpressure: 200% for ranges up to 1000 psi; 150% for
ranges over 1000 psi
MODULE COMMON SPECIFICATIONS
Engineering Units: psi, inHg, inWC, ftSW, bar, mbar, kPa,
mPa, mmHg, cmWC, mmWC, kg/cm2and any single user­defined, field-programmable engineering unit
Damping (Measurement Averaging): Programmable, averaging from zero through 16 consecutive readings
Compensated Temperature Range: -6 to 49°C (20 to 120°F)
Reference Temperature: 21±2°C (70±3°F) Temperature Effect: ±0.002% per °C (0.004% per °F) over
compensated range Repeatability: ±0.01% span (ranges above 0.5 inH
±0.02% span (ranges below 1 inH2O)
Sensitivity: ±0.002% span, typical Process Connection:
Dimensions: mm (in)
1
8 FNPT
2
O);
NEW! Thermocouple and RTD Modules Available for Temperature Calibration.
Visit omega.com/pcl-m for details.
A-41
One Omega Drive | Stamford, CT 06907 | 1-888-TC-OMEGA (1-888-826-6342) | info@omega.com
EPG05
www.omega.com
UNITED KINGDOM
www. omega.co.uk
Manchester, England
0800-488-488
UNITED STATES
www.omega.com
1-800-TC-OMEGA
Stamford, CT.
CANADA
www.omega.ca
Laval(Quebec)
1-800-TC-OMEGA
GERMANY
www.omega.de
Deckenpfronn, Germany
0800-8266342
Karviná, Czech Republic
FRANCE
www.omega.fr
Guyancourt, France
088-466-342
CZECH REPUBLIC
www.omegaeng.cz
596-311-899
BENELUX
www.omega.nl
Amstelveen, NL 0800-099-33-44
More than 100,000 Products Available!
Temperature
Calibrators, Connectors, General Test and Measurement Instruments, Glass Bulb Thermometers, Handheld Instruments for Temperature Measurement, Ice Point References, Indicating Labels, Crayons, Cements and Lacquers, Infrared Temperature Measurement Instruments, Recorders Relative Humidity Measurement Instruments, RTD Probes, Elements and Assemblies, Temperature & Process Meters, Timers and Counters, Temperature and Process Controllers and Power Switching Devices, Thermistor Elements, Probes and Assemblies,Thermocouples Thermowells and Head and Well Assemblies, Transmitters, Wire
Flow and Level
Air Velocity Indicators, Doppler Flowmeters, Level Measurement, Magnetic Flowmeters, Mass Flowmeters, Pitot Tubes, Pumps, Rotameters, Turbine and Paddle Wheel Flowmeters, Ultrasonic Flowmeters, Valves, Variable Area Flowmeters, Vortex Shedding Flowmeters
pH and Conductivity
Conductivity Instrumentation, Dissolved Oxygen Instrumentation, Environmental Instrumentation, pH Electrodes and Instruments, Water and Soil Analysis Instrumentation
Data Acquisition
Auto-Dialers and Alarm Monitoring Systems, Communication Products and Converters, Data Acquisition and Analysis Software, Data Loggers Plug-in Cards, Signal Conditioners, USB, RS232, RS485 and Parallel Port Data Acquisition Systems, Wireless Transmitters and Receivers
Pressure, Strain and Force
Displacement Transducers, Dynamic Measurement Force Sensors, Instrumentation for Pressure and Strain Measurements, Load Cells, Pressure Gauges, Pressure Reference Section, Pressure Switches, Pressure Transducers, Proximity Transducers, Regulators, Strain Gages, Torque Transducers, Valves
Heaters
Band Heaters, Cartridge Heaters, Circulation Heaters, Comfort Heaters, Controllers, Meters and Switching Devices, Flexible Heaters, General Test and Measurement Instruments, Heater Hook-up Wire, Heating Cable Systems, Immersion Heaters, Process Air and Duct, Heaters, Radiant Heaters, Strip Heaters, Tubular Heaters
click here to go to the omega.com home page