Omega Products PCL-130 Installation Manual

http://www.omega.com
e-mail: info@omega.com
®
User ’s Guide
PCL-130
Process Calibrator
CONTENTS
GENERAL OPERATING INSTRUCTIONS ...........................3
TURN ON ...................................................................5
CONNECTIONS .........................................................5
FIELD & BENCH USE (TILT STAND).........................6
CHANGING BATTERIES............................................6
CONFIGURING TEMPERATURE SCALES ...............7
ENABLING AUTO-OFF...............................................7
RESTORING DEFAULT SETTINGS...........................7
DISPLAY CONTRAST & BACKLIGHTING.................8
SELECTING FUNCTIONS & RANGES......................9
SOURCE MODE OPERATING INSTRUCTIONS...............10
STORING & RECALLING QUIK-CHEK®OUTPUTS10
SOURCE MODE CONNECTION INSTRUCTIONS
CALIBRATE MILLIAMP INPUTS ..............................11
SIMULATE 2-WIRE TRANSMITTERS......................12
CALIBRATE VOLTAGE INPUTS...............................13
CHECK 1-5 VOLT INPUTS WITHOUT
DISCONNECTING WIRES.......................................14
CALIBRATE THERMOCOUPLE INPUTS.................16
CALIBRATE RESISTANCE INPUTS........................17
CALIBRATE RTD INPUTS........................................18
CALIBRATE FREQUENCY INPUTS ........................20
READ MODE OPERATING INSTRUCTIONS ....................21
MIN/MAX ..................................................................21
OUT OF RANGE SIGNALS......................................21
READ MODE CONNECTION INSTRUCTIONS
READ MILLIAMP OUTPUTS....................................22
READ VOLTAGE OUTPUTS.....................................23
READ AC VOLTAGES...............................................24
MEASURE THERMOCOUPLE SENSORS..............25
READ RESISTANCE ................................................26
CHECK CONTINUITY..............................................27
MEASURE RTD SENSORS .....................................28
COUNT FREQUENCIES..........................................30
CALIBRATE 2 & 4 WIRE TRANSMITTERS .............32
PRESSURE (WITH COMPANION PRESSURE MODULES)
READ PRESSURE...................................................34
CALIBRATE PRESSURE TRANSMITTERS.............36
SPECIFICATIONS...............................................................40
WARRANTY........................................................................48
3
Your New PCL-130
Lighten your load…take the PCL-130 to ever y site. It’s like bringing a cartload of test equipment from the shop to the control room or the field. The PCL-130 sources and reads DC like a milliamp or voltage calibrator, simulates and measures T/Cs & R TDs lik e a temperature calibr ator, gener­ates and counts frequency and Counts-Per-Minute like a frequency calibrator and displays pressure like a precision test gauge. Troubleshooting? It checks continuity with a beeper and measures AC line voltage like a multimeter!
Calibrate Milliamp Inputs
Calibrate controllers, recorders and other devices in 4 to 20 or 0 to 20 mA loops. Source and read 0.00 to 24.00 mA, or Simulate a 2-Wire Transmitter.Use the optional AC adaptor for continuous operation.Display to 0.01 mA , 0.1 % of 4 to 20 and 0.1 % DP Flow.
Calibrate 2-Wire Transmitters
Easily calibrate 2-Wire Pressure and Electronic Transmitters by connecting the PCL-130 to both the input and output of the transmitter. The PCL-130 will simultane­ously indicate the input and output of the transmitter on the graphical display.
Calibrate Pressure Systems
Read pressure with extreme accuracy using an external PM130 pressure module in a companion PCL130-MH module holder. Attach the module directly to the pressure connection for the best accuracy or with optional tubing for tight spots. Display to five digits within 0.025% of reading in up to 20 engineering units including psi, pa, Kpa, Mpa, BAR, mBar, Atm, Kgf plus torr, inches and mm of mercury
GENERAL INFORMATION
54
or water at a variety of temperatures. Each module is fully characterized for temperature and linearity to give you labora­tory accuracy in the shop, control room or field.
Voltage Calibration
Calibrate all your DC millivolt and voltage instrumenta­tion. Source from 0.00 to 110.00 mV and 0.00 to 10.25 V. Read up to 110.00 mV, 11.00 V and 200.0 VDC.
Temperature Calibration
Source and Read directly in °C and °F for T/C types J, K, T, E, R, S & N and four Pt 100 Ohm, Ni 120 Ohm and Cu 10 Ohm RTDs. Resolution to 0.1°. Cold junction compen­sation continuously tracks ambient temperature changes.
Frequency Calibration
Generate zero crossing square waves from 1 to 1000 Hz,
0.01 to 10.00 kHz and from 1 to 1000 CPM (Counts-Per­Minute).Built-in frequency counter measures Hz, kHz & CPM.
Measure AC Voltage
Check line voltage or mains from 0.0 to 250.0 volts AC. Great for troubleshooting power problems.
Check Continuity
Locate pairs of wires, open connections and shorts with the built-in beeper.
Rear Label - Condensed Operation Guide
GENERAL
TURN-ON
Each time you turn on the PCL-130 it runs through a self check then returns to the most recently selected settings.
SOURCE - The three QUIK-CHEK outputs will be the same as previously stored. Each time a different function is selected, the associated three QUIK-CHEK outputs will be recalled.
READ - The 130 is ready to measure the same signal as the last time it was used and automatically updates the MAX & MIN readings for recall at any time.
CONNECTIONS
PCL-130 has protected banana jacks compatible with standard and safety banana plugs. Included with your TechChek are: a pair of safety test leads with test probes, safety alligator clips, standard alligator clips and spade lugs for attachment to a wide variety of instruments. An additional test lead and spade lug are also included for 3-Wire RTD connections. A second pair of test leads with right angled safety banana plugs and alligator clips for mA Read and Power Transmitter functions. Thermocouple connections are made through a miniature thermo­couple socket.
CAUTION! To prevent accidently overloading the instrument being tested, correctly set up the outputs before connecting the PCL-130 to any instruments to be calibrated.
!
!
!
!
6
7
GENERAL
CONFIGURING TEMPERA TURE SCALES
The thermocouple and RTD ranges may be configured for full time use of °C, full time use of °F or selectable °C and °F operation. This configuration is part of the DEFAULT SETTINGS below.
AUTO-OFF
PCL-130 can be set up to turn itself off after 30 minutes of inactiv­ity. The inter nal timer is reset to 30 minutes each time the digital pot is turned or a pushbutton is pressed.
DEFAULT SETTINGS
PCL-130 may be restored to the factory default setting. This will reset the HI and LO “QUIK-CHEK”memories according to the tab le below and the SET memory to midrange between HI and LO.
OPERATING INSTRUCTIONS
1) Press and hold the STORE/RESET push-button while turning the 130 on.
2) Keep pressing the push-button until SETTING UP DEFAULT appears on the display then release the push-button.
3) °F/°C SETUP appears on the display.
4) Turn the digital pot to select °F & °C, °F Only or °C Only. °C & °F is the default if no selection is made.
5) Press STORE/RESET push-button to store your choice or wait five seconds for the 130 to automatically store your choice and AUTO-OFF SETUP appears on the display.
6) Turn the digital pot to select AUTO-OFF ENABLE. AUTO-OFF causes the 130 to turn itself off after 30 minutes of inactivity to preserve the batteries. Select AUTO-OFF DISABLE for contin­uous operation.
7) Press STORE/RESET push-button to store your choice or wait five seconds for the 130 to automatical­ly store your choice.
8) The firmware revision will be displayed, then the 130 will begin operation with the new settings.
OPERATING INSTRUCTIONS
GENERAL
FIELD & BENCH USE
PCL-130 comes with a carrying case and a built-in tilt stand/hanger. The 130 is held securely in the case by VELCRO® for use with the carrying case open. The carrying case also has a snap-on belt loop which can be looped around a pipe or rail. The tilt stand is easily raised by pulling the stand until it locks into place.The stand can also be reversed for use as a hanger to suspend the 130.
CHANGING BATTERIES
Low battery is indicated by a battery symbol on the display. Approximately four hours of operation remain before the LCD blanks and PCL-130 shuts itself down. Turn the 130 off, loosen the captive screw securing the battery compartment and lift off the cover from the bottom of the case. The six “AA” batteries are easily removed and replaced (alkaline supplied and recom­mended).Replace the battery compartment cover by inserting the tabs and tightening the screw.
QUIK-CHEK DEFAULTS
RANGE LO SET HI
mA 4.00 12.00 20.00 mV 1.00 5.00 10.00
V 1.00 5.00 10.00 T/C J,T,E,K,N All points 0°C/32°F T/C R & S All points 538°C/1000°F
Ohms 100.0 200.0 400.0
RTD All points 0°C/32°F kHz 1.00 5.00 10.00
Hz 100 5001000
CPM 100 500 1000
GENERAL
SELECTING FUNCTIONS
Turn the selector knob to choose among mA, V, T/C, , RTD, FREQ and PRESSURE.
SELECTING RANGES
Press the RANGE/TYPE pushbutton to select the desired range and scale.
SOURCE RANGES
Milliamp:
Source: mA, %mA (% 4-20), DP% (% Differential Pressure) 2-Wire Simulator: mA, %mA (% 4-20), DP% (% Differential
Pressure)
VDC: mV, V T/C:Types J, T, E, K, N, R & S in °C &°F Ohms: Ohms RTD:Four Pt 100, Ni 120& Cu 10in °C &°F Frequency:KHz, Hz, CPM
READ RANGES
Milliamp: mA, %mA (% 4-20), DP% (% Differential Pressure) VDC: mV, 10V, 200V T/C:Types J, T, E, K, N, R & S in °C &°F Ohms: Ohms RTD:Four Pt 100, Ni 120& Cu 10in °C &°F Frequency:KHz, Hz, CPM AC V olts:V A C
TRANSMITTER RANGES
Pwr Xmtr: Supplies nominal 24 VDC to power the transmitter while
measuring the transmitter milliamp signal . Can be simul­taneously displayed with any Source Range or Pressure Measurement
Read mA:Measures transmitter output
98
OPERATING INSTRUCTIONS
GENERAL
DISPLAY CONTRAST
The contrast of the Liquid Crystal Display may be adjusted for best readability.
1) Press and hold the DISPLAY/SOURCE/READ push-button while turning the 130 on to adjust the contrast of the display.
2) Turn the digital pot (knob) until the display is most legible.
3) Press the STORE/RESET push-button or wait five seconds for your selection to be stored.
DISPLAY BACKLIGHTING
The LCD may have too little contrast where the lighting is dim or the 130 is in shadow.Turn on the LCD backlighting to make the display easy to read.
1) If the PCL-130 is off, turn it on and wait for the display to come up in normal mode.
2) Press and hold the DISPLAY/SOURCE/READ push-button for three seconds and the backlight will turn on.
To extend the life of the batteries, it is recommended that the backlighting be turned off when the LCD is in normal light. Press and hold the DISPLAY/SOURCE/READ push-button for three seconds and the backlight will turn off.
TURN OFF
Press the POWER push-button to turn the 130 off.If AUT O-OFF is enabled, the 130 will turn itself off after 30 minutes of inactivity.
OPERATING INSTRUCTIONS
POWER
mA
V
T/C
RTD
FREQ
PRESSURE
TYPE
ENG UNITS
11
CALIBRATE MILLIAMP INPUTS
mA, mA % (Percent of 4 to 20 mA), DP% (DP Flow)
Choose this function to provide an output from 0.00 to 24.00 milliamps.The compliance voltage is a nominal 24 VDC to pro vide the driving power to your milliamp receivers.
1) Disconnect one or both input wires from the device to be calibrated.
2) Turn the Selector Knob to mA
3) Press the DISPLAY/SOURCE/READ push-button until SOURCE or 2-WIRE appear on the display
4) If 2-WIRE is on the display, press the TYPE/ENG UNITS push­button once to indicate SOURCE on the display
5) Press the mA/%/% DP FLO W push-b utton to display mA, % 4-20 or % DP Flow.
6) Connect the red SOURCE lead of the calibrator to the plus (+) input of the device and the black SOURCE lead to the minus (-).
Output current is continuously adjustable with the "QUIK-CHEK" switch in the SET position. Zero & Span (or any other values) are available by using the LO and HI "QUIK-CHEKs".
OPERATING INSTRUCTIONS
SOURCE MODE
Select source by pressing the DISPLAY/SOURCE/READ pushbutton until the word SOURCE appears on the LCD display. To change the output value, turn the speed sensitive digital pot. Turning the knob slowly will cause a g radual change in the output. A faster rate of change will occur when the knob is turned faster. This function operates in all three output positions (HI, SET & LO).
STORING QUIK-CHEK OUTPUTS
1) Switch to HI or LO
2) Turn the knob to desired value
3) Press the STORE push-button The LCD will flash once to show that the value was saved
If a value is in the SET position and you want that value stored in HI or LO, press and hold the STORE push-button while moving the switch to HI or LO. The display will flash once to indicate the value has been stored. Then release the STORE
push-button.
RECALLING QUIK-CHEK OUTPUTS
When you need a stored value just flip the QUIK-CHEK switch. Any value for the selected range may be stored in HI & LO. The PCL-130 remembers the HI, LO and SET values for each function with the power on or off. Each time a different function is selected, the last three QUIK­CHEK values for that function will be recalled.
OPERATING INSTRUCTIONS
10
HI SET LO
SOURCE
STORE RESET
HI SET LO
SOURCE
RED
BLACK
Milliamp Receiver Input
Controller
Transmitter
Computer
Logger
I/P
DCS
13
CALIBRA TE V OL TAGE INPUTS
V, mV
Choose this function to provide an output from 0.00 mV to 110.00 mV and from 0.00 to 10.25 VDC.Current compliance up to 20 mA to provide the driving power to your voltage receivers.
1) Disconnect one or both input wires from the device to be calibrated
2) Turn the Selector Knob to V
3) Press the DISPLAY/SOURCE/READ push-button until SOURCE and V or mV appear on the display
4) Press the TYPE/ENG UNITS push-button once to switch between V and mV on the display
5) Connect the red SOURCE lead of the calibrator to the plus (+) input of the device and the black SOURCE lead to the minus (-).
Output voltage is continuously adjustable with the "QUIK-CHEK" switch in the SET position. Zero & Span (or any other values) are available by using the LO and HI "QUIK-CHEKs".
OPERATING INSTRUCTIONS
12
OPERATING INSTRUCTIONS
SIMULATE 2-WIRE TRANSMITTERS
2-WIRE mA, 2-WIRE % (Percent of 4 to 20 mA) 2-WIRE DP %
Choose this function to simulate a 2-Wire Transmitter output from
1.00 to 24.00 milliamps. Operates in loops with power supply voltages from 3 to 45 VDC.
1) Disconnect existing 2-Wire Transmitter from the loop
2) Turn the Selector Knob to mA
3) Press the DISPLAY/SOURCE/READ push-button until SOURCE or 2-WIRE appear on the display
4) If SOURCE is on the display, press the TYPE/ENG UNITS push­button once to indicate 2-WIRE on the display
5) Press the mA/%/% DP FLO W push-b utton to display mA, % 4-20 or % DP Flow.
6) Connect the red SOURCE lead of the calibrator to the plus (+) input of the device and the black SOURCE lead to the minus (-).
The simulated output of the 2-Wire Transmitter is continuously adjustable from 1.00 to 24.00 mA with the "QUIK-CHEK" switch in the SET position. Zero & Span (or any other values) are available by using the LO and HI "QUIK-CHEKs".
+IN-
REF +OUT-
Power Supply
2 to 45 VDC
Receiver
(Powers external
2-Wire Transmitter)
To
Sensor
Typical
2-Wire
Transmitter
(Disconnected)
RED
BLACK
Voltage Receiver Input
RED
BLACK
Controller
Transmitter
Computer
Logger
DCS
1514
CHECK 1-5 VOLT INPUTS
WITHOUT DISCONNECTING WIRES
Most 1-5 Volt receivers in 4-20mA loops have a 250 Ohm resis-
tor across the input of the receiver.This resistor may be mounted internally or externally. PCL-130 is connected directly across the input of the 1-5 Volt receiver without disconnecting any field wiring. This saves a great deal of time when a large number of voltage receivers, such as chart recorders or computer systems, require calibration.
Make certain that changing the signal input will not disturb the
process or cause unexpected alarms when checking on-line instruments. It is impor tant to remember the 130 drives only the device to which it is connected. It has no effect on other devices in the 4 to 20 mA loop.PCL-130 will clamp the selected value in the mV and V Ranges to the maximum source or sink current of >16 mA.
1) Turn the Selector Knob to V
2) Press the DISPLAY/SOURCE/READ push-button until SOURCE and V or mV appear on the display
3) Press the TYPE/ENG UNITS push-button once to switch between V and mV on the display
4) Connect the red SOURCE lead of the calibrator to the plus (+) input of the device and the black SOURCE lead to the minus (-). Any associated 250 Ohm resistor must not be disconnected.
OPERATING INSTRUCTIONS
CHECK 1-5 VOLT INPUTS
WITHOUT DISCONNECTING WIRES
OPERATING INSTRUCTIONS
CONTROLLER
(TYPICAL)
RED
BLACK
250
RECORDER COMPUTER
LOGGER
ETC,
1-5V DC
ADDITIONAL
4-20 mA OR 1-5V
INSTRUMENTS
Typical 2-Wire
REF +OUT-
Transmitter
+IN-
1716
CALIBRATE RESISTANCE INPUTS
Choose this function to simulate a resistance into a variety of instruments.
1) Disconnect one or both input wires from the device to be calibrated
2) Turn the Selector Knob to
3) Press the DISPLAY/SOURCE/READ push-button until SOURCE appears on the display
4) Connect the red SOURCE lead of the calibrator to the plus (+) input of the device and the black SOURCE lead to the minus (-).
OPERATING INSTRUCTIONS
CALIBRA TE THERMOCOUPLE INPUTS
Choose this function to simulate a thermocouple signal into any instrument requiring a thermocouple input. The output of the 130 is automatically cold junction compensated.
1) Disconnect the thermocouple from the instrument being calibrated.
2) Turn the Selector Knob to T/C
3) Press the DISPLAY/SOURCE/READ push-button until SOURCE and any T/C TYPE appear on the display
4) Use the proper thermocouple wire and corresponding miniature T/C connector to connect PCL-130 to the thermocouple.
Output temperature is continuously adjustable with the "QUIK­CHEK" switch in the SET position. Zero & Span (or any other values) are available by using the LO and HI "QUIK-CHEKs
To Change the Thermocouple type
1) Press the TYPE/ENG UNITS push-button.The words SELECT
T/C TYPE and the current selection appear on the display.
until the desired T/C type and temperature scale appear.
2) Turn the digital pot (knob) until the required type appears on the display.
3) Press the STORE/RESET push-button or wait 5 seconds to store the selection
OPERATING INSTRUCTIONS
-
+
Instrument with
Thermocouple Input
Controller
Temperature Indicator
Temperature Trip or Alarm
Temperature T ransmitter
RED
BLACK
Receiver Input
Ohms or 2-Wire RTD
Controller
Transmitter
Computer
Loading...
+ 18 hidden pages