MULTIFUNCTION
CALIBRATOR
PORTABLE, HIGH ACCURACY
PCL1200, shown
smaller than
PCL1200 Series
Standard
actual size.
U Measures and Sources
T/C (13 Types), RTDs
(13 Types), Ohms,
Current, Voltage,
Frequency
U Sources Pulse Trains
U Isolated mA/V Readback
Circuit for Complete
Transmitter Calibration
U Pressure Module
Communication Port
Compatible with OMEGA
Pressure Modules
U Built-In 24V Supply Can
Drive 4 to 20 mA Loops
Up to 1000 Ω
U Direct Entry of Custom
RTD Coefficients
(R0, A, B, C)
U All Source Modes Can
be Programmed with
Dedicated Setpoints to
Speed Calibration and
Linearity Tests
U Highest Accuracy in
Class—to 0.015% of
Reading
U Meets CE Requirements
and is Designed to IEC
1010 Safety Standards
U Supplied with Full
Rubber Boot
Three sets
of leads
included!
The PCL1200 multifunction
calibrator has a feature set
unmatched by other high-accuracy
handheld calibrators in its price
range. It is portable, provides
the functions and accuracy of
a laboratory-grade instrument,
and has everything needed for
virtually any calibration task. It
can also measure and source
thermocouples, RTDs, current,
voltage, and frequency, and it
can source pulse trains. It has a
communications port for pressure
modules and an isolated mA/V
readback circuit. Arrow keys,
direct numeric keypad entry, and
3 software-driven function buttons
combine with a large, backlit, menudriven graphics display to provide
a highly intuitive yet powerful
interface. A built-in
250 Ω resistor for Hart®
compatibility, compatibility with
smart transmitters and PLCs,
full fuseless protection, and a
serial communications port for
control using ASCII commands
are additional features that make
the PCL1200 indispensable for
calibration tasks. It comes with
a tough rubber boot; an optional
carrying case is available.
Measuring Pressure
with the PCL1200
With a PCL-PMA pressure module
adaptor, the PCL1200 will work
with all OMEGA PCL-PM pressure
modules. See Specifications and
“To Order” chart.
Specifications
(23 ±5°C unless otherwise noted)
Operating Temperature: -10 to 50°C
(14 to 122°F)
Storage Temperature: -20 to 70°C
(-4 to 158°F)
Stability: ±0.005% rdg/°C outside of
23 ±5°C (73 ±9°F)
Cold Junction Compensation
(Thermocouple) Error: ±0.2°C (0.4°F)
Power Requirements: 6 Vdc
Batteries: 4 “AA”, alkaline (included)
Dimensions: 221 H x 107 W x 58 mm D
(8.7 x 4.2 x 2.3”)
Weight: 863 g (30.5 oz)
INSTRUMENTATION
D
D-120

MULTIFUNCTION CALIBRATOR
TYPE RANGE ACCURACY
VOLTAGE
Source 0.000 to 20.000 Vdc ±0.015% rdg + 2 mV
Read, Isolated 0.000 to 30.000 Vdc ±0.015% rdg + 2 mV
Read, Non-Isolated 0.000 to 20.000 Vdc ±0.015% rdg + 2 mV
THERMOCOUPLE mV
Read/Source -10.000 to 75.000 mV ±0.02% rdg + 10 µV
CURRENT
Read 0.000 to 24.000 mA ±0.015% rdg + 2 µA
FREQUENCY (1 TO 20V SELECTABLE AMPLITUDE)
CPM, Read 2.0 to 600.0 CPM ±0.05% rdg (+1 LSD read)
Hz, Read 1.0 to 1000.0 Hz ±0.05% rdg (+1 LSD read)
kHz, Read 1.00 to 10.00 kHz ±0.05% rdg +1 LSD
kHz, Source 1.00 to 10.00 kHz ±0.250% rdg
RESISTANCE (Ω)
Read 0.00 to 4000.0 Ω 400Ω range: ±0.025% rdg + 0.05Ω
Source 5.0 to 4000 Ω 4000Ω range: ±0.025% rdg + 0.5Ω
THERMOCOUPLE
J -200.0 to 1200.0°C -328.0 to 2192.0°F 0.2°C 0.4°F
K -200.0 to 1370.0°C -328.0 to 2498.0°F 0.3°C 0.5°F
T -200.0 to 400.0°C -328.0 to 752.0°F 0.2°C 0.4°F
E -200 to 950°C -328.0 to 1742.0°F 0.2°C 0.4°F
R 0.0 to 1750.0°C 32.0 to 3182.0°F 1.2°C 2.2°F
S 0.0 to 1750.0°C 32.0 to 3182.0°F 1.2°C 2.2°F
B 600.0 to 1800.0°C 1112.0 to 3272.0°F 1.2°C 2.2°F
C 0.0 to 2316.0°C 32.0 to 4200.0°F 0.6°C 1.1°F
XK -200.0 to 800.0°C -328.0 to 1472.0°F 0.2°C 0.4°F
BP 0.0 to 2500.0°C 32.0 to 4532.0°F 0.9°C 1.6°F
L -200.0 to 900.0°C -328.0 to 1652.0°F 0.2°C 0.4°F
U -200.0 to 400.0°C -328.0 to 752.0°F 0.25°C 0.45°F
N -200.0 to 1300.0°C -328.0 to 2372.0°F 0.4°C 0.7°F
RTD/THERMISTOR
Ni120 (672) -80.0 to 260.0°C -112.0 to 500.0°F 0.2°C 0.4°F
Pt100 (385) -200.0 to 800.0°C -328.0 to 1472.0°F 0.2°C 0.4°F
Pt100 (3926) -200.0 to 630.0°C -328.0 to 1166.0°F 0.2°C 0.4°F
Pt100 (3916) -200.0 to 630.0°C -328.0 to 1166.0°F 0.2°C 0.4°F
Pt200 (385) -200.0 to 630.0°C -328.0 to 1166.0°F 0.8°C 1.4°F
Pt500 (385) -200.0 to 630.0°C -328.0 to 1166.0°F 0.4°C 0.7°F
Pt1000 (385) -200.0 to 630.0°C -328.0 to 1166.0°F 0.2°C 0.4°F
Cu10 -100.0 to 260.0°C -148.0 to 500.0°F 1.4°C 2.5°F
Cu50 -180.0 to 200.0°C -292.0 to 392.0°F 0.4°C 0.7°F
Cu100 -180.0 to 200.0°C -292.0 to 392.0°F 0.3°C 0.5°F
Pt385-10 -200.0 to 800.0°C -328.0 to 1472.0°F 1.4°C 2.5°F
Pt385-50 -200.0 to 800.0°C -328.0 to 1472.0°F 0.4°C 0.7°C
400 Series Thermistor 15.00 to 50.00°C 15.00 to122.00°F 0.1°C 0.2°C
400Ω range: ±0.025% rdg + 0.05Ω
4000Ω range: ±0.025% rdg + 0.5Ω
D-121

READBACK DISPLAY
The top half of the display is dedicated to
readback from the device under test, or a
pressure module.
MAIN DISPLAY
The lower half of the display is for all input
and output combinations.
FUNCTION KEYS
Three software-controlled function keys;
functions displayed over each button at
bottom of display.
POWER ON/OFF
Tu rns power on/off. Auto shut-off button.
HOME KEY
HOME key displays main operating screen.
CLEAR ENTRY KEY
Allows clearing of entry.
NUMERIC KEYPAD
Rapid numeric entry of data values.
ENTER KEY
Accepts entries into memory and
updates outputs.
CURRENT (mA)
Input/output for current.
PRESSURE MODULE
COMMUNICATIONS PORT
Pressure modules connector; compatible
with pressure modules (not shown).
Pressure measurements from 1 inH
10,000 psi, module dependent.
2
O to
ARROW KEYS
Arrow keys allow rapid movement of
cursor and setting of output values.
COMPUTER
COMMUNICATION CONNECTION
Mini-jack (stereo) serial communications
connection. Allows full remote control.
VOLTAGE/Ω/Hz
Input/output for voltage/Ω/Hz.
THERMOCOUPLE
Input/output for thermocouple.
ISOLATED READBACK JACKS
Volts, mA, and mA with loop power.
To Order Visit omega.com/pcl1200 for Pricing and Details
MODEL NO. DESCRIPTION
PCL1200 High-accuracy multifunction calibrator
MODULES – MOST POPULAR RANGES*
MODEL NO. PARAMETER RANGE GAUGE psig ACCURACY OVERPRESSURE
PCL-PM030G 0 to 30 (0 to 2 bar) ±0.025% 300%
PCL-PM050G 0 to 50 (0 to 3.5 bar) ±0.03% 300%
PCL-PM100G 0 to 100 (0 to 7 bar) ±0.025% 300%
PCL-PM300G 0 to 300 (0 to 20 bar) ±0.025% 200%
PCL-PM1KG 0 to 1000 (0 to 70 bar) ±0.05% 200%
PCL-PM3KG 0 to 3000 (0 to 200 bar) ±0.1% 200%
PCL-PM5KG 0 to 5000 (0 to 340 bar) ±0.1% 200%
PCL-PM015VAC 0 to -15 (0 to -1 mbar) ±0.025%, ±0.0025 psi 300%
PCL-PM015A 0 to 15 (0 to 1 bar) absolute ±0.025%, ±0.0025 psi 300%
PCL-PM030A 0 to 30 (0 to 2 bar) absolute ±0.025% 300%
PCL-PM015C -15 to 15 (-1 to 1 bar) ±0.025%, ±0.0025 psi 300%
PCL-PM030C -15 to 30 (-1 to 2 bar) ±0.025%, ±0.0025 psi 300%
PCL-PM030D 0 to 30 (0 to 2 bar) differential ±0.025% 300%
For all ranges consult sales.
*
One PCL-PMA adaptor required for use of pressure modules.
Note: Gauge, absolute and compound types are isolated and accept any media compatible with 316 SS. Vacuum and differential types are
compatible with pressure media that are clean, dry, non-corrosive air or gas.
Comes complete with test leads, protective rubber boot, 1 Type K thermocouple with cable, 4 “AA” alkaline batteries, NIST certificate and
operator’s manual.
Ordering Example: PCL1200, high-accuracy multifunction calibrator. For pressure applications, order one PCL-PMA (required) plus the
modules required. Example: PCL-PMA, pressure module adaptor, PCL-PM030G, 0 to 30 psig pressure module.
PCL-ACA-EU, 230 vac 50 Hz European adaptor/charger.
PCL-ACA-J, 120 vac 60 Hz Japan adaptor/charger
PCL-ACA-UK, 230 vac 50 Hz UK adaptor/charger
INSTRUMENTATION
D
ACCESSORIES
MODEL NO. DESCRIPTION
PCL-PMA Pressure module adaptor (one required for pressure)
PCL-ACA 120 Vac adaptor
D-122