Tel: (95) 800-826-6342
En Español: (95) 203-359-7803
e-mail: espanol@omega.com
SM
SM
SM
FAX: (95) 203-359-7807
Servicing Europe:
Benelux:
Czech Republic:
France:
Germany/Austria:
United Kingdom:
ISO 9002 Certified
It is the policy of OMEGA to comply with all worldwide safety and EMC/EMI regulations that apply. OMEGA is constantly
pursuing certification of its products to the European New Approach Directives. OMEGA will add the CE mark to every
appropriate device upon certification.
The information contained in this document is believed to be corrected but OMEGA Engineering Inc. accepts no liability for any errors it
WARNING: These products are not designed for use in, and should not be used for, patient connected applications.
Postbus 8034, 1180 LA Amstelveen, The Netherlands
Tel: (31) 20 6418405
Toll Free in Benelux: 0800 0993344
e-mail: nl@omega.com
One Omega Drive , River Bend Technology Centre
Northbank, Irlam, Manchester
M44 5EX, England
Tel: 44 (161) 777-6611
Toll Free in United Kingdom: 0800-488-488
e-mail: info@omega.co.uk
contains, and reserves the right to alter specifications without notice.
FAX: (31) 20 6434643
FAX: 420 (69) 6311114
FAX: (33) 130-699-120
FAX: 49 (07056) 8540
FAX: 44 (161) 777-6622
2
INTRODUCTORY NOTE
This publication contains operating instructions, as well as a description of the principles of operation, of
PCL12 pressure calibrator.
OMEGA has used the best care and efforts in preparing this book and believes the information in this
publication are accurate. The OMEGA products are subjected to continuous improvement, in order to pursue
the technological leadership; these improvements could require changes to the information of this book.
OMEGA reserves the right to change such information without notice.
No part of this document may be stored in a retrieval system, or transmitted in any form, electronic or
mechanical, without prior written permission of OMEGA Engineering Inc.
PCL12 units use sophisticated analogic and digital technologies. Any maintenance operation must be carried
out by qualified personnel
ONLY. We recommend to contact our technicians for any support requirements.
.
PCL12 is fully tested in conformity with the directive n°89/336/CEE Electromagnetic Compatibility. OMEGA
shall not be liable in any event, technical and publishing error or omissions, for any incidental and
consequential damages, in connection with, or arising out of the use of this book.
3
1 General
PCL12 is a microprocessor digital pressure gauge of new generation, its advanced analog section ensures a very high
long-term stability, and its converter A/D with 16 bits guarantee 65.000 internal divisions. This feature, together with the
0,05% accuracy class, make it an ideal laboratory instrument: the PCL12 is specifically proposed as First Line Sampling
for Metrologic Laboratories, and it's delivered together with the SIT certificate. The pressure gauge has internal batteries
with 1 year autonomy, which is also guaranteed by the AUTO-POWER-OFF function, that turns off the pressure gauge if
no pressure variation is detected within 30 minutes. The user can select different engineering units (mbar, bar, kPa,
MPa, psi), and can program different resolutions and digital filter according to the application. In order to increase the
integration level of the components, a combined traditional and SMT technology has been applied; which makes the
pressure gauge more resistant against mechanical stresses and vibrations and ensures the maximum reliability of the
electronic circuit. The display also show an analog indication bar of pressure always active also into a program-menu.
Main features:
• 1 YEAR AUTONOMY WITHOUT RECHARGE
• PROGRAMMABLE RESOLUTION
• DIGITAL FILTER
• PROGRAMMABLE MEASURE UNIT
• PEAK FUNCTION (positive and negative)
• RS232 OUTPUT (optional)
1.1 Specifications
♦ RELATIVE PRESSURE: 20 - 50 - 100 - 250 - 350 bar
♦ ACCURACY: < ±0.05% f.s. (Including linearity, hysteresis and temperature effect from 0 to +50°C)
♦ POWER SUPPLY: n° 4 Alkaline Battery size AA
♦ AUTONOMY: 1 YEAR
♦ MAX FULL SCALE: 60.000 div.
♦ PROGRAMMABLE MEASURING UNITS: *mbar- bar- MPa- Kpa- psi (other units on request)
♦ PROGRAMMABLE RESOLUTION: 1-2-5-10
♦ PROGRAMMABLE BAUD RATE: 9600, 4800,2400
♦ ZERO FUNCTION: 100%
♦ PEAK FUNCTION: positive and negative
♦ READING PER SEC.: 3
♦ DISPLAY: 5 digit custom LCD
♦ MECHANICAL LIMIT VALUES RELATED TO NOMINAL PRESSURE:
service pressure 100%
max. permissible pressure 150%
breaking pressure >300%
high dynamic pressure 75%
♦ SERVICE TEMPERATURE RANGE: from 0 to 50 °C
♦ STORAGE TEMPERATURE: from -10 to +60°C
♦ STANDARD PROCESS COUPLING:
Pressure gauge 1/2” BSP male
Pump 1/4” NPT female
♦ GASKET ADVISED: USIT A 63-18
♦ TIGHTENING WRENCH: 27mm
♦ TIGHTENING TORQUE: 28Nm
♦ PROTECTION CLASS (DIN 40050): IP60
♦ MATERIAL OF THE SENSOR: INOX 17-4 pH
♦ MATERIAL OF THE CONTAINER: ALUMINIUM
♦ ELECTRICAL CONNECTION (RS232C): tank SUB D 9 pole FEMALE
4
2 Operating instructions
The PCL12 LCD can be divided into 3 different areas as follow:
♦ The central area display the measured pressure value using the selected unit.
♦ The lower area display the measured value as bargraph referred to the full scale value (this is active in program
mode also).
♦ The upper area show the active function (Zero, Peak+, Peak- e Low-Bat).
2.1 Keyboard
3 keys are available to program and operate with the calibrator.
ON
Set
ZERO
ON Switch the PCL12 on. The instrument power off is automatic with a
programmable delay (see par. 2.2).
Set Press this key for 3 seconds to go in programming mode.
ZERO Zeroing the pressure displayed. It is possible to run the function only if the
displayed value is less than 10% of full scale.
ª In programming mode allows to decrease the values of the selected parameter.
Peak
Peak Activating of peak+ (press it for 1 sec.) or peak- (press it for 5 sec.). This
function display the max or min measured pressure.
It is possible, pressing the “SET” key for 3 seconds to activate the programming mode.
To select the next parameter and confirm the previous value, you have to press the “SET” key.
After you have run all parameters, PCL12 returns in measuring mode.
Set the auto power off time (from 1 to 99 minutes). The 0 value, disable the function
and the instrument will be on all time. The function begin the time counting if the
PCL12 doesn’t measure pressure changes higher than 10% f.s.
Change the RS232 serial interface communication speed (if installed).
Possible values are: 1=4800baud, 2=9600baud, 3=19200baud, 0=disable.
5
2.3 Power Supply
PCL12 calibrator is powered by 4 type AA 1.5V batteries. These batteries allows ≈1 Year autonomy operative life. The
“LOW BAT” message will be displayed if the batteries needs to be changed. Measured pressure will be incorrect if the
LOW BAT message is displayed.
6
WARRANTY/DISCLAIMER
OMEGA ENGINEERING, INC. warrants this unit to be free of defects in materials and workmanship for a period of 13
months from date of purchase. OMEGA Warranty adds an additional one (1) month grace period to the normal one (1)
year product warranty to cover handling and shipping time. This ensures that OMEGA’s customers receive maximum
coverage on each product.
If the unit should malfunction, it must be returned to the factory for evaluation. OMEGA’s Customer Service Department
will issue an Authorized Return (AR) number immediately upon phone or written request. Upon examination by OMEGA,
if the unit is found to be defective it will be repaired or replaced at no charge. OMEGA’s WARRANTY does not apply to
defects resulting from any action of the purchaser, including but not limited to mishandling, improper interfacing,
operation outside of design limits, improper repair, or unauthorized modification. This WARRANTY is VOID if the unit
shows evidence of having been tampered with or shows evidence of being damaged as a result of excessive corrosion;
or current, heat, moisture or vibration; improper specification; misapplication; misuse or other operating conditions
outside of OMEGA’s control. Components which wear are not warranted, including but not limited to contact points,
fuses, and triacs.
OMEGA is pleased to offer suggestions on the use of its various products However, OMEGA neither assumes
responsibility for any omissions or errors nor assumes liability for any damages that result from the use of its
products in accordance with information provided by OMEGA, either verbal or written. OMEGA warrants only
that the parts manufactured by it will be as specified and free of defects. OMEGA MAKES NO OTHER WARRANTIES OR REPRESENTATIONS OF ANY KIND WHATSOEVER, EXPRESSEO OR IMPUED, EXCEPT THAT OF
TITLE, AND ALL IMPLIED WARRANTlES INCLUDING ANY WARRANTY OF MERCHANTABIUTY AND RTNESS
FOR A PARTlCULAR PURPOSE ARE HEREBY DISCLAIMED. LIMITATlON OF LIABILITY: The remedies of
purchaser set forth herein ate exclusive and the total liability of OMEGA with respect to this order, whether
based on contract, warranty, negligence. Indemnification, strict liability or otherwise, shall not exceed the
purchase price of the component upon which liability is based. In no event shall OMEGA be liable for
consequential, incidental or special damages.
CONDITIONS: Equipment sold by OMEGA is not intended to be used, nor shall it be used: (1) as a ”Basic Component”
under 10 CFR 21 (NRC), used in or with any nuclear installation or activity; or (2) in medical applications or used on
humans. Should any Product(s) be used in or with any nuclear installation or activity, medical application, used on
humans, or misused in any way, OMEGA assumes no responsibility as set forth in our basic WARRANTY/DISCLAIMER
language, and additionally, purchaser will indemnify OMEGA and hold OMEGA harmless from any liability or damage
whatsoever arising out of the use of the Product(s) in such a manner.
RETURN REQUESTS / INQUIRIES
Direct all warranty and repair requests/inquiries to the OMEGA Customer Service Department. BEFORE RETURNING
ANY PRODUCT(S) TO OMEGA, PURCHASER MUST OBTAIN AN AUTHORIZED RETURN (AR) NUMBER FROM
OMEGA’S CUSTOMER SERVICE DEPARTMENT (IN ORDER TO AVOID PROCESSING DELAYS). The assigned AR
number should then be marked on the outside of the return package and on any correspondence.
The purchaser is responsible for shipping charges, freight, insurance and proper packaging to prevent breakage in
transit.
FOR WARRANTY
following information available BEFORE contacting
OMEGA:
1. P.O. number under which the product was
PURCHASED,
2. Model and serial number of the product under
warranty, and
3. Repair instructions and/or specific problems
relative to the product.
OMEGA’s policy is to make running changes, not model changes, whenever an improvement is possible. This affords our customers the
latest in technology and engineering.
OMEGA is a registered trademark of OMEGA ENGINEERING, INC.
(C) Copyright 1999 OMEGA ENGINEERING, INC. All rights reserved. This document may not be copied, photocopied, reproduced,
translated, or reduced to any electronic medium or machine-readable form, in whole or in part, without prior written consent of OMEGA
ENGINEERING, INC.
RETURNS, please has the
FOR NON-WARRANTY
current repair charges. Have the following information
available BEFORE contacting OMEGA:
1. P.O. number to cover the COST of the repair,
2. Model and serial number of product, and
3. Repair instructions and/or specific problems relative to
the product.