An OMEGA Technologies Company
OM-EL-2-12BIT
12 BIT PORTABLE DATALOGGER
INSTRUCTION
SHEET
M-3142
OM-EL-2-12BIT is an easy to usedisplay modulecapable of measuring, recording and displaying temperature(Pt 100),voltage and current. With12 bitA/D
resolution, a memory for 8000 readingsand abattery life of up to3 years,OM-EL-2-12BIT can operate asa 'standalone' logger or be permanently connected
to asystem. The OM-EL-2-12BIT serial linkis addressableand up to 8 loggers can be connected toone serialport. The PCsoftware operatesunder Windows
and doesnot requirespecialist skillto operate. Data outputis intext formatand canbe easilyintegrated intomost popularspreadsheets. Graphical outputis
possible under OM-EL-WIN. Consult theEasyLog softwaremanual for further details.
Battery Powered
Multi-function
Non-volatile Data Storage
°
C
Hand Held or Surface Mounting
Easy to Use
High Resolution Read-out
CONTROL SOFTWARE
Model No - OM-EL-WIN
Easy to install and use, the control software will run under
Windows 95 or 3.1 and enable the user to control one or more
EasyLogs, operating them as a complete system. Supplied on a
3½" diskwith a manualand seriallink.
ACCESSORIES - CABLES
Model No - OM-EL-EASYLINK
Extension cable to 'daisy chain' more than one EasyLog. One
extension will be needed for each extra OM-EL-2-12BIT module
that is attachedto thechain.
Ω
Ω
Ω
Input Impedance
200mV Range >1G
2V, 20V Range 0.5M
4-20mA Range 10
Data Logger OM-EL-2-12BIT
Specification Range Resolution Accuracy
Temperature - Low Range * -200 to +200°C 0.1°C ±0.5°C
-200 to +200 °F 0.1°F ±1°F
Temperature - High Range * -200 to +850°C 1°C ±1°C
-328°F to +1562°F 1°F ±2°F
Voltage - D.C. 0 to ±200mV 100 V
0 to ±2V 1mV
0 to ±20V 10mV
Current - D.C. 4 to 20mA 16 A
Battery 3.6V ½AA lithium (up to 3 years life) **
Serial link 8 Pin Mini DIN
Sensor connection 4 Pin Connector
Memory Up to 8000 samples**
Sample rate 1 sample per 5 seconds to 1 per 12 hours.
* Sensor dependent **Depending on sample rate
µ
µ
Model No
±0.05%
±1 Count
±0.1% ±1 Count
DIMENSIONS All dimensions in mm (inches)
70.0 (2.75)
46.5 (1.83)
54.0 (2.12)
17.5 (0.69)
15.0 (0.59)
26.0 (1.02)
54.0 (2.12)
8.00
(0.31)
37.5 (1.47)
LINK FUNCTIONS
Lk1: When daisy-chaining EasyLog Modules,
Lk2 & Lk3: Fitthese jumpers asshown in thetable (right).
remove the Lk1jumpers fromall modulesexcept one.
26.0 (1.02)
Screws M3 x 16
2.50
(0.09)
4.50
(0.17)
3.50
(0.13)
44.0 ((1.73)
60.0 (2.36)
65.5 (2.58)
V-
LA
HA
V+
AV+
LK3
LK2
RESET
LK1
+
LITHIUM
BATTERY
-
Measurement Range Jumper Link Setting
200mV L 2 open, L 3 Open
2V L 2 closed,L 3 open
20V L 2 closed, L 3 open
4-20mA L 2 open, L 3 closed
Temperature (Low Range) L 2 open,L 3 open
Temperature (High Range) L 2 open,L 3 open
kk
kk
kk
kk
kk
kk

PIN FUNCTIONS
SW: Switch input, normallypulled high. Connect momentarilyto V- to take a readingin One-Shot
mode or tostart loggingwhen configuredfor Push-to-Start inOM-EL-WIN.
AV+: Test pin. DoNot Use.
V-,V+: External power supplyconnections. Read ImportantNote belowprior touse.
External supply voltagerange 3 to 3.6Vdc.External power supplymust befloating with
respect to thesignal tobe measured.
HA, LA Normally at V-, these pinsgo highwhen theirrespective alarm levelshave beenreached
or exceeded.
Sensor Input
Vin- (R)
Vin+ (W)
Iin (R)
Iout (W)
IMPORTANT NOTE- Always removethe Lithium batteryfrom themodule
BEFORE
Internal Pin Header
connecting an external powersupply tothe module.
Failureto do somay causethe batteryto explode.
V+
LA
HA
AV+
V-
BATTERY REPLACEMENT
Only use ½AA 3.6V lithium. The listbelow is notexhaustive. Check with supplierthat the battery you areordering is 'pressfit' andis notfitted withsolder
tags. Take care toconnect correctly. DO NOT PRESSON LCDWHEN INSERTING BATTERY.
MANUFACTURER PARTNUMBER MANUFACTURER'SORDER CODE
MAXELL ER 3S TC n/a
SAFT LS3 n/a
SONNENSCHEIN SL-750/S 1107 501 100
TADIRAN ½AA/S 1551-02-210-000
WARNING: Handle lithiumbatteries carefully- observe warningson batterycasing. Dispose ofin accordance withlocal regulations.
NOTE -
APPLICATIONS
TEMPERATURE
Iin (R)
red
Pt100
sensor
Pt100
sensor
Pt100
sensor
Vin- (R)
Vin+ (W)
white
Iout (W)
Pt100: 2-wire connection
(Lk2 and Lk3 open)
Iin (R)
red
Vin- (R)
Vin+ (W)
white
white
Iout (W)
Pt100: 3-wire connection
(Lk2 and Lk3 open)
Iin (R)
red
Vin- (R)
red
Vin+ (W)
white
white
Iout (W)
Pt100: 4-wire connection
(Lk2 and Lk3 open)
It is possible to measure parameters other than those outlined below. Use an appropriate sensor and conditioning circuit to
convert the parameterto bemeasured into alinear voltageor current andapply thissignal to asuitably scaledOM-EL-2-12BITmodule.
Measurement signals must always be isolated from the communications signals.
EasyLog
socket
EasyLog
socket
EasyLog
socket
VOLTAGE
Vin- (R)
Vin+ (W)
200mV DC connection
(Lk2 and Lk3 open)
Vin- (R)
Vin+ (W)
2V DC connection
(Lk2 closed, Lk3 open)
Vin- (R)
Vin+ (W)
20V DC connection
(Lk2 closed, Lk3 open)
EasyLog
socket
EasyLog
socket
EasyLog
socket
CURRENT
Vin- (R)
Vin+ (W)
4-20mA Indication
Use 4-20mA indication range
(default calibration 0-1000)
(Lk2 open, Lk3 closed)
1R
Vin- (R)
Vin+ (W)
200mA DC connection
Use 200mV range with
annunciators set to mA.
Module will require calibration.
(Lk2 and Lk3 open)
Vin- (R)
R
Vin+ (W)
For higher ranges, use 200mV
range and R=
where = 2A or 20A fsd.
Module will require calibration.
(Lk2 and Lk3 open)
0.2
I
I
EasyLog
socket
EasyLog
socket
EasyLog
socket
OMEGAnet On-Line Service
http://www.omega.com
USA:
ISO 9001 Certified
Canada:
For immediate technical or application assistance:
USA and Canada:
Mexico and
Latin America:
Benelux:
Czech Republic:
France:
Germany/Austria:
United Kingdom:
ISO 9002 Certified
omega.com
SM
TM
®
Internet e-mail
@
info omega.com
Servicing North America:
One Omega Drive, Box 4047
Stamford, CT 06907-0047
Tel: (203) 359-1660 FAX: (203) 359-7700
@
e-mail: info omega.com
976 Bergar
Laval (Quebec) H7L 5A1
Tel: (514) 856-6928 FAX: (514) 856-6886
e-mail: canada omega.com
Sales Service: 1-800-826-6342 / 1-800-TC-OMEGA
Customer Service: 1-800-622-2378 / 1-800-622-BEST
Engineering Service: 1-800-872-9436 / 1-800-USA-WHEN
TELEX: 996404 EASYLINK: 62968934 CABLE: OMEGA
Tel: (95) 800-TC-OMEGA FAX: (95) 203-359-7807
En Español: (203) 359-1660 ext:2203 e-mail:
@
SM
SM
SM
SM
espanol omega.com
@
Servicing Europe:
Postbus 8034, 1180 LAAmstelveen, The Netherlands
Tel: (31) 20 6418405 FAX: (31) 20 6434643
Toll Free in Benelux: 06 0993344
@
e-mail: nl omega.com
Ostravska 767, 733 01 Karvina
Tel: 42 (69) 6311899 FAX: 42 (69) 6311114
e-mail: czech omega.com
9, rue Denis Papin, 78190 Trappes
Tel: (33) 130-621-400 FAX: (33) 130-699-120
Toll Free in France: 0800-4-06342
e-mail: france omega.com
Daimlerstrasse 26, D-75392 Deckenpfronn, Germany
Tel: 49 (07056) 3017 FAX: 49 (07056) 8540
Toll Free in Germany: 0130 1121 66
e-mail: germany omega.com
25 Swannington Road, PO Box 7, Omega Drive,
Broughton Astley, Leicestershire, Irlam, Manchester,
LE9 6TU, England M44 5EX, England
Tel: 44 (1455) 285520 Tel: 44 (161) 777-6611
FAX: 44 (1455) 283912 FAX: 44 (161) 777-6622
Toll Free in England: 0800-488-488
e-mail: uk omega.com
@
@
@
@
OMEGA ENGINEERING, INC. warrantsthis unit to be free of defects in materials and workmanship for a period of
from date of purchase. OMEGA Warranty adds an additional one (1) month grace period to the normal
to cover handling and shipping time. This ensures that OMEGA’s customers receive maximum coverage on each
warranty
product.
Ifthe unitshouldmalfunction, itmust be returned to thefactory for evaluation.OMEGA’sCustomer ServiceDepartment willissue
anAuthorized Return(AR) numberimmediatelyupon phoneor written request.Uponexamination by OMEGA,if theunitis found
to be defective it will be repaired or replaced at no charge. OMEGA’s WARRANTYdoes not apply to defects resulting from any
actionof the purchaser,includingbut not limitedto mishandling, improperinterfacing,operation outside ofdesign limits,improper
repair,orunauthorized modification.ThisWARRANTY is VOID ifthe unit showsevidenceof havingbeen tampered withor shows
evidence of being damaged as a result of excessive corrosion; or current, heat, moisture or vibration; improper specification;
misapplication; misuse or other operating conditions outside of OMEGA’s control. Components which wear are not warranted,
includingbutnot limited tocontactpoints, fuses, andtriacs.
OMEGA is pleased to offer suggestions on the use of its various products. However, OMEGA neither assumes
responsibilityfor anyomissions or errorsnor assumes liability for anydamages that result from theuse ofits products
in accordance with information provided by OMEGA, either verbal or written. OMEGA warrants only that the parts
manufactured by it will be specified and free of defects. OMEGA MAKES NO OTHER WARRANTIES OR
REPRESENTATIONS OF ANY KIND WHATSOEVER, EXPRESSED OR IMPLIED, EXCEPT THAT OF TITLE, AND ALL
IMPLIED WARRANTIES INCLUDING ANY WARRANTY OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR
PURPOSE ARE HEREBY DISCLAIMED. LIMITATION OF LIABILITY: The remedies of purchaser set forth herein are
exclusive and the total liability of OMEGA with respect to this order,whether based on contract, warranty,negligence,
indemnification,strict liability or otherwise, shall not exceed the purchase price of the component upon whichliability
isbased.In no eventshallOMEGA be liableforconsequential,incidentalor special damages.
CONDITIONS: Equipment sold by OMEGA is not intended to be used, nor shall it be used:(1) as a “Basic Component” under
10CFR 21 (NRC),used in orwith any nuclearinstallationor activity;or (2)in medical applicationsor used onhumans.Shouldany
Product(s) be used in or with any nuclear installation or activity, medical application, used on humans, or misused in any way,
OMEGA assumes no responsibility as set forth in our basic WARRANTY/DISCLAIMER language, and additionally, purchaser
will indemnify OMEGA and hold OMEGA harmless from any liability or damage whatsoever arising out of the use of the
Product(s)insuch amanner.
Direct all warranty and repair requests/inquiries to the OMEGA Customer Service Department. BEFORE RETURNING ANY
PRODUCT(S) TO OMEGA, PURCHASER MUST OBTAIN AN AUTHORIZED RETURN (AR) NUMBER FROM OMEGA’S
CUSTOMERSERVICEDEPARTMENT(IN ORDERTOAVOIDPROCESSINGDELAYS).The assigned AR number shouldthen
bemarkedon theoutsideof the returnpackageand on anycorrespondence.
Thepurchaseris responsiblefor shippingcharges,freight,insurance and properpackagingto preventbreakage intransit.
WARRANTY NON-WARRANTY
FOR RETURNS, pleasehavethefollowing FOR REPAIRS,consultOMEGAfor
informationavailableBEFOREcontacting OMEGA: current repair charges.Havethe followinginformation
1.P.O.numberunder whichtheproduct was PURCHASED, availableBEFOREcontacting OMEGA:
2.Modeland serial numberofthe product underwarranty, 1. P.O.numberto cover the COSTofrepair,
and 2. Modeland serial numberofthe product, and
3.Repairinstructions and/or specificproblemsrelativeto 3.
theproduct. theproduct.
OMEGA’s policy is to make running changes, not model changes, whenever an improvement is possible. This affords our
customersthelatest intechnologyand engineering.
OMEGAisa registeredtrademarkof OMEGA ENGINEERING,INC.
© Copyright 1996 OMEGA ENGINEERING, INC. All rights reserved. This document may not be copied, photocopied,
reproduced, translated, or reducedto anyelectronic medium or machine-readable form, in wholeor in part, withoutprior written
consentofOMEGA ENGINEERING,INC.
Itis the policyofOMEGA to complywithall worldwide safetyandEMC/EMI regulations thatapply.OMEGA is constantly pursuingcertification of
itsproductsto the European NewApproach Directives.OMEGAwill add the CEmarktoeveryappropriate device upon certification.
The information containedin this document is believedtobe correctbut OMEGA Engineering, Inc. accepts noliability foranyerrors itcontains,
andreservesthe right to alterspecificationswithoutnotice.
Theseproductsare not designed foruse in, andshouldnotbe used for,patient connected applications.WARNING:
WARRANTY/DISCLAIMER
RETURNREQUESTS / INQUIRIES
Repairinstructions and/or specificproblemsrelativeto
13 months
one (1) year product