Omega Products OM-EL-1-12BIT Installation Manual

®
An OMEGA Technologies Company
OM-EL-1-12BIT
12 BIT PANEL MOUNT DATALOGGER
INSTRUCTION
SHEET
M-3141
OM-EL-1-12BIT isan easy to usedisplay modulecapable of measuring, recording,displaying and controllingtemperature (Pt100), voltage andcurrent. With 12 bit A/D resolution, a memory for 8000 readings and a battery life of up to 3 years, OM-EL-1-12BIT can operate as a 'stand alone' logger or bepermanently connected to a system. The OM-EL-1-12BIT serial link is addressableand upto 8loggers canbe connected to one serial port. The PC software operatesunder Windowsand doesnot require specialist skill tooperate. Data output is intext formatand can be easily integratedinto mostpopular spreadsheets. Graphical output is possibleunder OM-EL-WIN. Consult theEasyLog softwaremanual forfurther details.
Battery Powered
Multi-function
°
Non-volatile Data Storage
C
Panel Mounting
Easy to Use
High Resolution Read-out
CONTROL SOFTWARE
Model No - OM-EL-WIN
Easy to install and use, the control software will run under Windows 95 or 3.1 and enable the user to control one or more EasyLogs, operating them as a complete system. Supplied on a 3½" diskwith a manualand seriallink.
ACCESSORIES - CABLES
Model No - OM-EL-EASYLINK
Extension cable to 'daisy chain' more than one EasyLog. One extension will be needed for each extra OM-EL-1-12BIT module that is attachedto thechain.
Ω Ω
Input Impedance
200mV Range >1G 2V, 20V Range 0.5M 4-20mA Range 10
Specification Range Resolution Accuracy
Temperature - Low Range * -200 to +200°C 0.1°C ±0.5°C
-200 to +200 °F 0.1°F ±1°F
Temperature - High Range * -200 to +850°C 1°C ±1°C
-328°F to +1562°F 1°F ±2°F
Voltage - D.C. 0 to ±200mV 100 V
0 to ±2V 1mV
0 to ±20V 10mV Current - D.C. 4 to 20mA 16 A Battery 3.6V ½AA lithium (up to 3 years life) ** Serial link 8 Pin Mini DIN Sensor connection Screw terminal Memory Up to 8000 samples** Sample rate 1 sample per 5 seconds to 1 per 12 hours.
* Sensor dependent **Depending on sample rate
µ
µ
Model No
±0.05%
±1 Count
±0.1% ±1 Count
DIMENSIONS All dimensions in mm (inches)
58.0 (2.28)
55.0 (2.16)
46.5(1.76)
30.0 (1.18)
25.0 (0.98)
44.0 (1.73)
9.0 (0.35)
64.5 (2.54)
46.5 (1.83)
34.5 (1.36)
18.0 (0.69)
LINK FUNCTIONS
Lk1: When daisy-chaining EasyLog Modules,
Lk2 & Lk3: Fitthese jumpers asshown in thetable (right).
remove the Lk1jumpers fromall modulesexcept one.
a
Ø2.50
(0.10)
1.1 (0.044)
17.8 (0.70)
2.50
(0.09)
(0.04)
1.0
1.5
(0.05)
(0.24)
6.1
1.6
(0.06)
2.00 (0.08)
5.2 (0.20)
12.8
(0.50)
31.8 (1.25)
16.0
(0.63)
Panel cut-out 62 x 32 (2.44 x 1.26)
Measurement Range Jumper Link Setting
200mV L 2 open, L 3 Open 2V L 2 closed,L 3 open 20V L 2 closed, L 3 open 4-20mA L 2 open, L 3 closed Temperature (Low Range) L 2 open, L 3 open Temperature (High Range) L 2 open,L 3 open
kk kk kk kk kk kk
PIN FUNCTIONS
TxD, RxD: No pins fitted.Input andoutput connectionsfor Infra-Red communications,e.g.: connectto aPANEL-IR module. SW: Switch input, normallypulled high. Connect momentarilyto V- to take a readingin One-Shot
AV+: Testpin. DoNot Use. V-,V+: External power supplyconnections. Read ImportantNote belowprior touse.
HA, LA Normally at V-, these pinsgo highwhen theirrespective alarm levelshave beenreached
IN- IN+
V , V : MeasurementInputs, extension ofscrew terminalblock connections. Iin, Iout: CurrentLoop connection, extension of screwterminal block connections.
mode or tostart loggingwhen configuredfor Push-to-Start inOM-EL-WIN.
External supply voltagerange 3 to 3.6Vdc.External power supplymust befloating with respect to thesignal tobe measured.
or exceeded.
Rear Pin Header
SW
HA
V-
IN-
Iin
RxD AV+ V+ LA V
IN+
Iout
TxD
V
RS
IMPORTANT NOTE- BEFORE
Always remove theLithium batteryfrom the module connecting an external powersupply tothe module. Failureto do somay causethe batteryto explode.
BATTERY REPLACEMENT
Only use ½AA 3.6V lithium. The listbelow is notexhaustive. Check with supplierthat the battery you areordering is 'pressfit' andis not fitted withsolder tags. Take care toconnect correctly. DO NOT PRESSON LCD WHENINSERTING BATTERY.
MANUFACTURER PARTNUMBER MANUFACTURER'SORDER CODE
MAXELL ER 3S TC n/a
SAFT LS3 n/a
SONNENSCHEIN SL-750/S 1107 501100
TADIRAN ½AA/S 1551-02-210-000
WARNING: Handle lithiumbatteries carefully- observe warningson batterycasing. Dispose ofin accordance withlocal regulations.
APPLICATIONS
TEMPERATURE
red
2-wire Pt100 connection (Lk2 and Lk3 open)
NOTE -
It is possible to measure parameters other than those outlined below. Use an appropriate sensor and conditioning circuit to
convert the parameterto bemeasured into alinear voltageor current andapply thissignal to asuitably scaledOM-EL-1-12BITmodule.
Measurement signals must always be isolated from the communications signals.
white
Vin- (R)
Vin+ (W)
Iout (W) Iin (R)
VOLTAGE
200mV DC connection (Lk2 and Lk3 open)
Vin- (R) Vin+ (W)
Iout (W) Iin (R)
CURRENT
Vin- (R) Vin+ (W)
Iout (W)
4-20mA Indication Use 4-20mA indication range (default calibration 0-1000) (Lk2 open, Lk3 closed)
Iin (R)
OMEGAnet On-Line Service
http://www.omega.com
USA:
ISO 9001 Certified
Canada:
For immediate technical or application assistance:
USA and Canada:
Mexico and Latin America:
Benelux:
Czech Republic:
France:
Germany/Austria:
United Kingdom:
ISO 9002 Certified
white
white
white
Vin- (R) Vin+ (W)
Iout (W) Iin (R)
Vin- (R)
Vin+ (W)
Iout (W) Iin (R)
TM
®
Internet e-mail
info omega.com
@
red
white
3-wire Pt100 connection (Lk2 and Lk3 open)
red
red
4-wire Pt100 connection (Lk2 and Lk3 open)
omega.com
SM
Servicing North America:
One Omega Drive, Box 4047 Stamford, CT 06907-0047 Tel: (203) 359-1660 FAX: (203) 359-7700
@
e-mail: info omega.com 976 Bergar
Laval (Quebec) H7L 5A1 Tel: (514) 856-6928 FAX: (514) 856-6886 e-mail: canada omega.com
Sales Service: 1-800-826-6342 / 1-800-TC-OMEGA Customer Service: 1-800-622-2378 / 1-800-622-BEST Engineering Service: 1-800-872-9436 / 1-800-USA-WHEN TELEX: 996404 EASYLINK: 62968934 CABLE: OMEGA
Tel: (95) 800-TC-OMEGA FAX: (95) 203-359-7807 En Español: (203) 359-1660 ext:2203 e-mail:
@
SM
SM
SM
SM
espanol omega.com
@
Servicing Europe:
Postbus 8034, 1180 LAAmstelveen, The Netherlands Tel: (31) 20 6418405 FAX: (31) 20 6434643 Toll Free in Benelux: 06 0993344
@
e-mail: nl omega.com Ostravska 767, 733 01 Karvina
Tel: 42 (69) 6311899 FAX: 42 (69) 6311114 e-mail: czech omega.com
9, rue Denis Papin, 78190 Trappes Tel: (33) 130-621-400 FAX: (33) 130-699-120 Toll Free in France: 0800-4-06342 e-mail: france omega.com
Daimlerstrasse 26, D-75392 Deckenpfronn, Germany Tel: 49 (07056) 3017 FAX: 49 (07056) 8540 Toll Free in Germany: 0130 1121 66 e-mail: germany omega.com
25 Swannington Road, PO Box 7, Omega Drive, Broughton Astley, Leicestershire, Irlam, Manchester, LE9 6TU, England M44 5EX, England Tel: 44 (1455) 285520 Tel: 44 (161) 777-6611 FAX: 44 (1455) 283912 FAX: 44 (161) 777-6622 Toll Free in England: 0800-488-488 e-mail: uk omega.com
@
@
@
@
0.2
Vin- (R) Vin+ (W)
Iout (W) Iin (R)
Vin- (R)
Vin+ (W)
Iout (W) Iin (R)
I
13 months
one (1) year product
Vin- (R) Vin+ (W)
Iout (W)
2V DC connection (Lk2 closed, Lk3 open)
20V DC connection (Lk2 closed, Lk3 open)
OMEGA ENGINEERING, INC. warrantsthis unit to be free of defects in materials and workmanship for a period of from date of purchase. OMEGA Warranty adds an additional one (1) month grace period to the normal
to cover handling and shipping time. This ensures that OMEGA’s customers receive maximum coverage on each
warranty
product. Ifthe unitshouldmalfunction, itmust be returned to thefactory for evaluation.OMEGA’sCustomer ServiceDepartment willissue anAuthorized Return(AR) numberimmediatelyupon phoneor written request.Uponexamination by OMEGA,if theunitis found to be defective it will be repaired or replaced at no charge. OMEGA’s WARRANTYdoes not apply to defects resulting from any actionof the purchaser,includingbut not limitedto mishandling, improperinterfacing,operation outside ofdesign limits,improper repair,orunauthorized modification.ThisWARRANTY is VOID ifthe unit showsevidenceof havingbeen tampered withor shows evidence of being damaged as a result of excessive corrosion; or current, heat, moisture or vibration; improper specification; misapplication; misuse or other operating conditions outside of OMEGA’s control. Components which wear are not warranted, includingbutnot limited tocontactpoints, fuses, andtriacs.
OMEGA is pleased to offer suggestions on the use of its various products. However, OMEGA neither assumes responsibilityfor anyomissions or errorsnor assumes liability for anydamages that result from theuse ofits products in accordance with information provided by OMEGA, either verbal or written. OMEGA warrants only that the parts manufactured by it will be specified and free of defects. OMEGA MAKES NO OTHER WARRANTIES OR REPRESENTATIONS OF ANY KIND WHATSOEVER, EXPRESSED OR IMPLIED, EXCEPT THAT OF TITLE, AND ALL IMPLIED WARRANTIES INCLUDING ANY WARRANTY OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE ARE HEREBY DISCLAIMED. LIMITATION OF LIABILITY: The remedies of purchaser set forth herein are exclusive and the total liability of OMEGA with respect to this order,whether based on contract, warranty,negligence, indemnification,strict liability or otherwise, shall not exceed the purchase price of the component upon whichliability isbased.In no eventshallOMEGA be liableforconsequential,incidentalor special damages.
CONDITIONS: Equipment sold by OMEGA is not intended to be used, nor shall it be used:(1) as a “Basic Component” under 10CFR 21 (NRC),used in orwith any nuclearinstallationor activity;or (2)in medical applicationsor used onhumans.Shouldany Product(s) be used in or with any nuclear installation or activity, medical application, used on humans, or misused in any way, OMEGA assumes no responsibility as set forth in our basic WARRANTY/DISCLAIMER language, and additionally, purchaser will indemnify OMEGA and hold OMEGA harmless from any liability or damage whatsoever arising out of the use of the Product(s)insuch amanner.
Direct all warranty and repair requests/inquiries to the OMEGA Customer Service Department. BEFORE RETURNING ANY PRODUCT(S) TO OMEGA, PURCHASER MUST OBTAIN AN AUTHORIZED RETURN (AR) NUMBER FROM OMEGA’S CUSTOMERSERVICEDEPARTMENT(IN ORDERTOAVOIDPROCESSINGDELAYS).The assigned AR number shouldthen bemarkedon theoutsideof the returnpackageand on anycorrespondence. Thepurchaseris responsiblefor shippingcharges,freight,insurance and properpackagingto preventbreakage intransit.
WARRANTY NON-WARRANTY
FOR RETURNS, pleasehavethefollowing FOR REPAIRS,consultOMEGAfor informationavailableBEFOREcontacting OMEGA: current repair charges.Havethe followinginformation
1.P.O.number under whichtheproduct was PURCHASED, availableBEFOREcontacting OMEGA:
2.Modeland serial numberofthe product underwarranty, 1. P.O.numberto cover the COSTofrepair, and 2. Modeland serial numberofthe product, and
3.Repairinstructions and/or specificproblemsrelativeto 3. theproduct. theproduct.
OMEGA’s policy is to make running changes, not model changes, whenever an improvement is possible. This affords our customersthelatest intechnologyand engineering. OMEGAisa registeredtrademarkof OMEGA ENGINEERING,INC. © Copyright 1996 OMEGA ENGINEERING, INC. All rights reserved. This document may not be copied, photocopied, reproduced, translated, or reducedto anyelectronic medium or machine-readable form, in wholeor in part, withoutprior written consentofOMEGA ENGINEERING,INC.
Itis the policyofOMEGA to complywithall worldwide safetyandEMC/EMI regulations thatapply.OMEGA is constantly pursuingcertification of itsproductsto the European NewApproach Directives.OMEGAwill add the CEmarktoeveryappropriate device upon certification.
The information containedin this document is believedtobe correctbut OMEGA Engineering, Inc. accepts noliability foranyerrors itcontains, andreservesthe right to alterspecificationswithoutnotice.
Theseproductsare not designed foruse in, andshouldnotbe used for,patient connected applications.WARNING:
Iin (R)
Vin- (R)
Vin+ (W)
Iout (W) Iin (R)
WARRANTY/DISCLAIMER
RETURNREQUESTS / INQUIRIES
1R
200mA DC connection Use 200mV range with annunciators set to mA. Module will require calibration. (Lk2 and Lk3 open)
R
For higher ranges, use 200mV range and R=
where = 2A or 20A fsd.
I
Module will require calibration. (Lk2 and Lk3 open)
Repairinstructions and/or specificproblemsrelativeto