Omega Products OM-EL-1 Installation Manual

®
An OMEGA Technologies Company
OM-EL-1
PANEL MOUNT DATALOGGER
INSTRUCTION
SHEET
M-3139
OM-EL-1 measures,displays, recordsand controls temperature, humidity,pH, voltage, current, etc., witha memoryfor 8064readings and abattery life of up to 3 years**. OM-EL-1 can operate as a 'stand alone' logger or be permanently connected to a system. The OM-EL-1 serial link is addressable and up to 8 loggers can be connected to one serial port. The PC software operates under Windows 3.1 and does notrequire specialist skill to operate. Data output is in text format and can be easily integrated into any popular spreadsheet. Graphical output is possible under OM-EL-WIN. Consult the EasyLog software manual for furtherdetails.
Battery Powered
Multi-function
°
Non-volatile Data Storage
Panel Mounting
Easy to Use
Cold Junction Compensation For Thermocouple Temperature Sensing
Windows 3.1 Control Software
Model No - OM-EL-WIN
Easy to install and use, the control software will enable the user to controlone or moreEasyLogs and operatethem asa complete system. Supplied on a 3½" disk with a manual and serial link. OM-EL-WINwill run underWindows95.
Accessories - Cables
Model No - OM-EL-EASYLINK
Extension connector to 'daisy chain' more than one EasyLog. One extension will beneededfor each extra logger.
* Sensor dependent. ** Battery life dependent on input mode and sample
rate. E.g. Rate Count,5 sec. sample rate. Battery lifeis approximately30 days.
DIMENSIONS
All dimensions in mm (inches)
58.0 (2.28)
55.0 (2.16)
46.5(1.76)
Data Logger OM-EL-1
Specification Range Resolution Accuracy
Temperature ('K' type thermocouple)
Temperature (internal)
Humidity 5 to 95% RH 1%RH * pH 0 to 14pH 0.1pH* ±0.1pH*
Voltage
Current 0 to 2A 0.01A ±1% Rate count 0 to 255 1 unit N/A Frequency 40 to 400Hz N/A ±2Hz Battery 3.6V ½AA lithium (up to 3 years life)** Serial link 8 Pin Mini DIN Sensor connection Screw terminal Memory capacity up to 8000 readings Sample rate 1 sample per 5 seconds to 1 per 12 hours.
-25 to +200°C 1°C ±1°C
-13 to +392°F 1°F ±2°F
-10 to +50°C 1°C ±1°C
+14 to +122°F 1°F ±2°F
0 to 2V D.C. 0.01V
0 to 20V D.C. 0.1V
1.6
(0.06)
16.0
(0.63)
NOTE - BEZEL SHAPE AND
DIMENSIONS ARE LIABLE
TO CHANGE WITHOUT
Model No
±1%
PRIOR NOTICE
a
44.0 (1.73)
34.5 (1.36)
30.0 (1.18)
18.0 (0.70)
25.0 (0.98)
9.0 (0.35)
64.5 (2.54)
47.0 (1.84)
Ø2.50
(0.10)
1.1 (0.044)
17.8 (0.70)
(0.05)
USING EXTERNAL SIGNALS
PIN FUNCTION
A1, A2, REF Output for test diagnostics. Do not use. HA High Alarm output =V+ whenalarmed. SW 'Pushto record event' input. Momentarily connect to V-. RS Reset input. Momentarily connectto V-. NOTE - logging will stopand restartat 0time. LA Low Alarm output =V+ whenalarmed. A+ Output= V+ when areading isin progress. V+ Batterypositive. V- Battery negative.
Only connect outputs to high impedance inputs, otherwise performance will be affected and battery life drastically reduced.
DO NOT APPLYAN EXTERNAL SUPPLY VOLTAGE ACROSS THESE PINS WITHBATTERY IN PLACE.
5.0 (0.20)
1.5
6.1
12.8
(0.24)
(0.50)
2.00 (0.08)
1.0
2.50
(0.09)
(0.04)
5.2 (0.20)
31.8 (1.25)
Panel cut-out 62 x 32 (2.44 x 1.26)
BATTERY REPLACEMENT
Only use ½AA 3.6V lithium. The list belowisnot exhaustive. Check with supplierthat the battery youare orderingis 'press fit'and isnot fitted withsolder tags or leads. Take careto connectcorrectly.
MANUFACTURER PARTNUMBER MANUFACTURER'SORDER CODE
MAXELL ER 3S TC n/a
SAFT LS3 n/a
SONNENSCHEIN SL-750/S 1107 501 100
TADIRAN ½AA/S 1551-02-210-000
WARNING: Handle lithiumbatteries carefully-observe warnings on batterycasing. Dispose ofin accordancewithlocal regulations.
VARIOUS OPERATING MODES
Pluggable Links (Link 1)
SHORTED
“K" type thermocouple
YELLOW (+VE)
A B
RED (-VE)
TEMPERATURE (with thermocouple)
A 'K' type thermocouple can be connected directly to the OM-EL-1. The range should be selected inthe OM-EL-WINsoftware.
TEMPERATURE (internal)
No sensor needed. The OM-EL-1 uses its internal temperature
A
sensor. Ensure A to B is
B
SHORTED.
A B
A capacitive humidity sensor can be connected directly to the EasyLog.
NOTE
1. There is no RH annunciator on the LCD. Sensor range - 200 to 300pF
SENSOR
HUMIDITY
0.5 to 1.0pF per %RH
*NOTES
1. When using the serial link, ensure that all signals are isolated from the computer ground.
2. When using several loggers together, ensure all signals are isolated or use the same ground.
OPEN
A B
+ 0-20V D.C.
-
VOLTAGE*
D.C. voltage within the ranges of 0-2V or 0-20V can be connected directly to the OM-EL-1. Ensure the appropriate range has been selectedin advance within the OM-EL-WIN software. For higher voltage measurement an external potential divider network shouldbe added.
+
1
0-2A D.C.
A B
6watt
0V
CURRENT*
D.C. current can be measured up to 2A maximum by using an external 1 6watt
shunt across the input. Select the voltage '0-2V range' in OM-EL-WIN software and 'A' in thedisplay symbols.
On-Board Link (Link 2)
Normally SHORTED Cut to OPEN
A B
pH (with internal reference)
Connect a pH probe that has an internal reference cell directlyto the OM-EL-1. (refer to theprobe's instruction manual).
NOTES
1. There is no need to open Link 2 if the SENTEK P14 probe is used.
2. OPEN LINK 2
A B
OM-EL-1 counts the number of pulses on the input during the sample interval (e.g.: every 5s, 10s, 20s etc.). At the end of each sample interval, thecount number is stored and the counter is reset to zero. A pulsed input of 0-5V can be directly connected to the OM-EL-1. The counter is incremented on the fallingedge.
+
-
RATE COUNT*
Normally OPEN Solder to SHORT
pH PROBE
+
0-5V
0
OMEGAnet On-Line Service
http://www.omega.com
USA:
ISO 9001 Certified
Canada:
For immediate technical or application assistance:
USA and Canada:
Mexico and Latin America:
Benelux:
Czech Republic:
France:
Germany/Austria:
United Kingdom:
ISO 9002 Certified
omega.com
SM
Servicing North America:
One Omega Drive, Box 4047 Stamford, CT 06907-0047 Tel: (203) 359-1660 FAX: (203) 359-7700
@
e-mail: info omega.com 976 Bergar
Laval (Quebec) H7L 5A1 Tel: (514) 856-6928 FAX: (514) 856-6886 e-mail: canada omega.com
Sales Service: 1-800-826-6342 / 1-800-TC-OMEGA Customer Service: 1-800-622-2378 / 1-800-622-BEST Engineering Service: 1-800-872-9436 / 1-800-USA-WHEN TELEX: 996404 EASYLINK: 62968934 CABLE: OMEGA
Tel: (95) 800-TC-OMEGA FAX: (95) 203-359-7807 En Español: (203) 359-1660 ext:2203 e-mail:
Postbus 8034, 1180 LAAmstelveen, The Netherlands Tel: (31) 20 6418405 FAX: (31) 20 6434643 Toll Free in Benelux: 06 0993344 e-mail: nl omega.com
Ostravska 767, 733 01 Karvina Tel: 42 (69) 6311899 FAX: 42 (69) 6311114 e-mail: czech omega.com
9, rue Denis Papin, 78190 Trappes Tel: (33) 130-621-400 FAX: (33) 130-699-120 Toll Free in France: 0800-4-06342 e-mail: france omega.com
Daimlerstrasse 26, D-75392 Deckenpfronn, Germany Tel: 49 (07056) 3017 FAX: 49 (07056) 8540 Toll Free in Germany: 0130 11 21 66 e-mail: germany omega.com
25 Swannington Road, PO Box 7, Omega Drive, Broughton Astley, Leicestershire, Irlam, Manchester, LE9 6TU, England M44 5EX, England Tel: 44 (1455) 285520 Tel: 44 (161) 777-6611 FAX: 44 (1455) 283912 FAX: 44 (161) 777-6622 Toll Free in England: 0800-488-488 e-mail: uk omega.com
@
Servicing Europe:
@
@
@
@
@
TM
®
Internet e-mail
@
info omega.com
SM
SM
SM
SM
espanol omega.com
@
OMEGA ENGINEERING, INC.warrants this unit to be free of defects inmaterials and workmanship fora periodof from date of purchase. OMEGA Warranty adds an additional one (1) month grace period to the normal
to cover handling and shipping time.This ensures that OMEGA’s customers receive maximum coverage on each
warranty
product. Iftheunit should malfunction, it mustbereturnedtothe factory for evaluation.OMEGA’sCustomerServiceDepartment will issue anAuthorizedReturn (AR) number immediately upon phoneorwrittenrequest.UponexaminationbyOMEGA,if the unit is found to be defectiveit will be repaired or replaced at no charge.OMEGA’sWARRANTYdoes not apply to defects resulting from any actionofthepurchaser,includingbutnotlimitedto mishandling, improper interfacing, operation outside of design limits, improper repair,orunauthorizedmodification.This WARRANTY isVOIDifthe unit shows evidence of having been tamperedwithorshows evidence of being damaged as a result of excessive corrosion; or current, heat, moisture or vibration; improper specification; misapplication; misuse or other operating conditions outside of OMEGA’s control. Componentswhich wear are not warranted, includingbutnotlimitedtocontactpoints,fuses,andtriacs.
OMEGA is pleased to offer suggestions on the use of its various products. However, OMEGA neither assumes responsibilityfor any omissions or errorsnor assumesliabilityfor any damages that result fromtheuse of its products in accordance with information provided by OMEGA, either verbal or written. OMEGA warrants only that the parts manufactured by it will be specified and free of defects. OMEGA MAKES NO OTHER WARRANTIES OR REPRESENTATIONS OF ANY KIND WHATSOEVER, EXPRESSED OR IMPLIED, EXCEPT THAT OF TITLE, AND ALL IMPLIED WARRANTIES INCLUDING ANY WARRANTY OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE ARE HEREBY DISCLAIMED. LIMITATION OF LIABILITY: The remedies of purchaser set forth herein are exclusive and thetotal liability of OMEGA with respect to this order, whether based on contract, warranty,negligence, indemnification,strict liability orotherwise, shall notexceed the purchase priceof the componentupon which liability isbased.InnoeventshallOMEGAbeliablefor consequential, incidental or special damages.
CONDITIONS: Equipment sold by OMEGA is not intended to be used, nor shall it be used: (1) as a “Basic Component”under 10CFR21(NRC), used in or with any nuclear installation or activity;or (2) in medicalapplicationsorusedonhumans.Should any Product(s) be used in or with any nuclear installation or activity, medical application, used on humans, or misused in any way, OMEGA assumes no responsibility as set forth in our basicWARRANTY/DISCLAIMER language, and additionally, purchaser will indemnify OMEGA and hold OMEGA harmless from any liability or damage whatsoever arising out of the use of the Product(s)insuchamanner.
Direct all warranty and repair requests/inquiries to the OMEGA Customer Service Department. BEFORE RETURNING ANY PRODUCT(S) TO OMEGA, PURCHASER MUST OBTAIN AN AUTHORIZED RETURN (AR) NUMBER FROM OMEGA’S CUSTOMERSERVICEDEPARTMENT(IN ORDERTOAVOIDPROCESSINGDELAYS).TheassignedAR number should then bemarkedontheoutsideofthereturnpackageandonanycorrespondence. Thepurchaserisresponsibleforshippingcharges,freight, insurance and proper packaging to preventbreakageintransit.
WARRANTY NON-WARRANTY
FOR RETURNS,pleasehavethefollowing FOR REPAIRS, consult OMEGA for informationavailableBEFORE contacting OMEGA: current repair charges.Havethefollowinginformation
1.P.O.numberunderwhichtheproductwasPURCHASED, availableBEFORE contacting OMEGA:
2.Modelandserialnumberoftheproductunderwarranty, 1.P.O.number to cover the COST of repair, and 2. Modelandserialnumberoftheproduct,and
3.Repairinstructionsand/orspecificproblemsrelativeto 3. theproduct. theproduct.
OMEGA’s policy is to make running changes, not model changes, whenever an improvement is possible. This affords our customersthelatestintechnologyandengineering. OMEGAisaregisteredtrademarkofOMEGAENGINEERING,INC. © Copyright 1996 OMEGA ENGINEERING, INC. All rights reserved. This document may not be copied, photocopied, reproduced, translated, or reducedto any electronicmedium or machine-readableform,in whole orin part, without prior written consentofOMEGAENGINEERING,INC.
ItisthepolicyofOMEGAtocomplywithallworldwidesafetyandEMC/EMIregulationsthatapply.OMEGAisconstantly pursuing certification of itsproductstotheEuropean New Approach Directives.OMEGA will add the CEmarktoeveryappropriatedeviceupon certification.
The information contained in this documentisbelievedtobe correct but OMEGA Engineering, Inc.accepts no liabilityforanyerrors it contains, andreservestherighttoalter specifications without notice.
Theseproductsarenotdesigned for use in,andshouldnotbeused for,patientconnectedapplications.WARNING:
WARRANTY/DISCLAIMER
RETURNREQUESTS/INQUIRIES
Repairinstructionsand/orspecificproblemsrelativeto
13 months
one (1) year product