8/16-Channel Thermocouple/Voltage Input USB Data Acquisition Module
OM-DAQ-USB-2401
FREE Software for
Configuration, Logging,
Charting and Real-Time
Display of Data
U 8 Differential or 16 Single-
Ended Analog Inputs
U 24 Bit Resolution with Up to
1000 Samples/Sec
U User Programmable for Type J,
K, T, E, R, S, B, N Thermocouple
or Voltage Input
U Built-In 4 Cold Junction
Compensation Temperature
Sensors and Open
Thermocouple Detection
U Powered Directly by USB Port
or External DC Power Supply
U 500 V Isolation Between
Input and PC For Safe and
Noise-Free Measurements
U FREE.NET API/Driver for
Visual Basic, C#, and Visual
C++ for Windows XP, Vista
and Windows 7
U Provides +12 Vdc Output for
Sensor Excitation
U Includes Hardware for
Benchtop, DIN Rail or
Wall Mounting
TMQSS-125U-12.
KMQSS-125U-12.
The OM-DAQ-USB-2401 is a USB
2.0 full speed thermocouple/voltage
input data acquisition module (fully
compatible with both USB 1.1 and
USB 2.0 ports). This stand-alone
module draws power from the USB
port to operate. An external power
supply (optional) can be used. All
configurable options (including
individual channel input type and
range) are software programmable.
The OM-DAQ-USB-2401 has
user programmable voltage inputs
that range from ±30 mV to ±10V,
full scale. The compact, modular
packaging ensures ease of use
The compact OM-DAQ-USB-2401
is ideal for portable data
acquisition applications, with
accessory probes (shown L to R),
KMQSS-125U-12, TMQSS-125U-12,
JMQSS-125U-12.
OM-DAQ-USB-2401,
shown smaller than
actual size.
Terminal Block
(4 included).
in a variety of applications. Units
can be DIN rail or wall mounted
with the included hardware or easily
operated on a bench. All analog
input channels can be measured
sequentially at about 1 ms per
channel. A total of 1000 samples
per second can be taken, divided
across all active channels.*
*Note: At highest scan rate, 1000 samples/sec
±1% with one channel on, ±5% with all
channels on.
SM
Extended Warranty
Program
OMEGACARESM extended warranty program
is available for models shown on this page.
Ask your sales respresentative for full details
when placing an order. OMEGACARESM
covers parts, labor and equivalent loaners.
To Order, Call or Shop Online at omega.com
Laptop not included.
SM
1
Specifications
GENERAL
Isolation: 500V from PC
External Excitation Output
Voltage: 12 Vdc regulated, max
total current output 67 mA
Power Requirements: Powered
direct from USB port, max 500 mA,
or from external 7.5 to 12 Vdc
Environmental: 0 to 50°C
(0 to 122°F) 95% RH (non-condensing)
Operating Temperature:
0 to 50°C (32 to 122°F),
0 to 95% RH non-condensing
Storage Temperature:
-40 to 85°C (-40 to 185°F)
Weight: 0.23 kg (0.5 lb)
Dimensions
107 W x 128 L x 39 mm H
(4.2 x 5.1 x 1.5")
Input Voltage Range: Software
programmable on a per-channel
basis differential/single-ended
-10 to 10V -500 to 500 mV
-5 to 5V -250 to 250 mV
-2.5 to 2.5V -125 to 125 mV
-2 to 2V -75 to 75 mV
-1 to 1V -30 to 30 mV
TC Input Range
Type J: -18 to 1200°C
(0 to 2192°F)
Type K: -129 to 1372°C
(-200 to 2502°F)
Type T: -101 to 400°C
(-150 to 752°F)
Type E: -184 to 1000°C
(-300 to 1832°F)
Type R: 204 to 1768°C
(400 to 3214°F)
Type S: 204 to 1768°C
(400 to 3214°F)
Type B: 538 to 1820°C
(1000 to 3308°F)
Type N: -129 to 1300°C
(-200 to 2372°F)
:
TC Input
Thermocouple Accuracy: Typical, in
very slow mode, 24 bit resolution
J = ±1.1°C
K = ±1.2°C
T = ±1.1°C
E = ±1.0°C
R = ±2.5°C
S = ±2.6°C
B = ±3.3°C
N = ±1.5°C
Cold Junction Compensation
Accuracy: ±1.0°C
Analog Input Accuracy:
Differential Input: Typical, in very
slow mode, 0.015% of reading
+0.004%
(exclusive of noise)
Single-End Input: Typical, in very
slow mode, 0.05% of reading
+0.01%
of range +10uV
of range +50uV (exclusive
The OM-DAQ-USB-2401 comes
complete with hardware for
both DIN rail and wall mounting,
both shown here.
of noise)
USB Device Type: USB 2.0
(full-speed)
Device Compatibility: USB 1.1,
USB 2.0
Power Supply: From USB or
9 Vdc universal adaptor (included)
DIN Rail Mounted for Rack
Application: Optional
Open Thermocouple Detect:
Automatically enable when a
channel is configured for a
thermocouple sensor
OMEGA® DAQ Central
Software
Each OM-DAQ-USB-2401 comes
complete with an impressive
FREE, easy-to-use software suite
that allows charting, data logging,
file storage and virtual instrument
viewing of data in real time. Users
can use the provided .NET driver
for C#, Visual Basic and Visual
C++ to design and develop their
own application software.
OMEGA® DAQ Central Software Device Controls
A demo version of DAQ central can be downloaded from omega.com/software
OMEGA® DAQ Central Software Configuration Screen OMEGA® DAQ Central Software Data Displays
2
To Order, Call or Shop Online at omega.com
SM