Temperature Monitoring and Alarming System
with Wireless Two-Way Communication
Part of the NOMAD
OM-CP-THERMALERT-GB
and
OM-CP-THERMALERT-P
Optional†
U Wireless Two-Way
Communication
U Precision RTD Sensing
Element
U View Data in Real-Time
U Glycol Bottle (Only for the
(OM-CP-THERMALERT-GB)
U 2 Year Battery Life
U Battery Life Indicator
U Field Upgradeable
®
Family
OM-CP-THERMALERT-P
The OM‑CP‑THERMALERT‑GB
and OM‑CP‑THERMALERT‑P
are temperature monitoring and
alarming systems designed for
use in hospitals, laboratories,
pharmaceutical facilities and other
critical temperature monitoring
areas. Also, these units can be used
to monitor a single temperature
location or expanded to include
hundreds of locations by adding
additional transmitters. These units
record both ambient temperature
and humidity.
The OM‑CP‑THERMALERT‑GB and
OM‑CP‑THERMALERT‑P provides
real time notification of temperature
deviations, and features a user
programmable alarm that can be
configured to send a message via
text message, screen alarm and/or
e‑mail if an alarm condition occurs.
The OM‑CP‑THERMALERT‑GB
system includes an RTD probe that
is mounted in a miniature bottle filled
with ethylene glycol. In this method,
a temperature “buffered” reading is
taken, preventing false alarms.
The OM‑CP‑THERMALERT‑P
system includes an RTD probe with
a 0.9 m (3’) lead wire and a 114 mm
(4.5”) probe sheath.
The OM‑CP‑THERMALERT‑GB and
OM‑CP‑THERMALERT‑P system
is equipped with wireless two‑way
communication. Full communication
can be performed directly from
a central PC, including starting,
stopping and downloading. In
addition to wirelessly transmitting
data, both units store each reading
OM-CP-THERMALERT-GB
OM-CP-RFC1000,
transceiver
to internal memory for backup.
This back up data may be retrieved
wirelessly at a later time. Customers
can set‑up automatic archiving
preferences, ensuring all data is
saved and retained to comply with
federal regulations.
†
Refer to next page for NIST calibration
ordering information.
1
To Order, Call or Shop Online at omega.com
SM
OM-CP-RFC1000,
Windows software displays
data in graphical or tabular
format (sold separately).
The RF-LogNet software
is used to configure the
OM-CP-THERMALERT-GB
and streams data to the
wireless transmitters
OM-CP software.
Specifications
TEMPERATURE
Probe Temperature Range:
OM-CP-THERMALERT-GB:
‑50 to 80°C (‑58 to 176°F)
OM-CP-THERMALERT-P:
‑200 to 440°C (‑328 to 824°F)
Probe Calibrated Accuracy:
OM-CP-THERMALERT-GB:
±0.1°C, ‑10 to 80°C (14 to 176°F)
±0.5°C (outside of that range)
OM-CP-THERMALERT-P:
±0.1°C, ‑10 to 150°C
(14 to 302°F) ±0.5°C (outside of
that range)
Resolution: 0.01°C
WIRELESS
RF Frequency: 2.45 GHz IEEE
802.15.4 ultra‑low power wireless
transceiver with fully bi‑directional
communication
Band: ISM band 2.405‑2.48 GHz
Output Power: +0dBm typical
Transmitter Range
(to Receiver or Repeater):
Outdoor Line-of-Sight
GENERAL
Reading Rate: One reading every
two seconds to one every 24 hours
Memory: 30,000 readings; software
configurable memory wrap
Wrap Around: Yes
Start Modes:
•Immediatestart
•Delaystartupto18months
Calibration: Digital calibration
through software
Calibration Date: Automatically
recorded within device
Battery Type: 3.6V lithium battery
(included); user replaceable
Battery Life: 2 years typical
Data Format: Date and time
stamped ºC, ºF, K, ºR
Time Accuracy: ±1 minute/month
(at 25°C)
Computer Interface:
OM‑CP‑RFC1000 required;
115,200 baud
Software: XP SP3/Vista/
Windows
®
7 (32‑ and 64‑bit)
Operating Environment:
‑20 to 80ºC (‑4 to 176ºF),
0 to 95% RH non‑condensing
Dimensions:
Data Logger:
55 W x 127 L x 33 mm D
(2.2 x 5.0 x 1.3")
Wire: 0.9 m (3')
Probe: 0.19 dia x 114 mm
3
⁄
16 x 4.5")
(
Enclosure Material: ABS plastic
OM-CP-RFC1000 Transceiver
(Acts as Receiver or Repeater)
Sensitivity: ‑95 dBm typical
Max Number of Transmitters:
Up to 20 transmitters sending data
directly to the transceiver or up to
300 transmitters using additional
transceivers (20 transmitters per
transceiver)
Max Distance to Transceiver
or from one Transceiver
to the Next:
Outdoor Line-of-Sight:
122 m (400')
Indoor: 46 m (150')
122 m (400’)
Indoor: 46 m (150’)
To Order
Model No. Description
OM-CP-THERMALERT-GB30 Wireless temperature transmitter with RTD probe and 30 ml glycol bottle
OM-CP-THERMALERT-GB30-CERT† Wireless temperature transmitter with RTD probe and 30 ml glycol bottle and
NIST calibration certificate
OM-CP-THERMALERT-GB60 Wireless temperature transmitter with RTD probe and 60 ml glycol bottle
OM-CP-THERMALERT-GB60-CERT† Wireless temperature transmitter with RTD probe and 60 ml glycol bottle and
NIST calibration certificate
OM-CP-THERMALERT-GB250 Wireless temperature transmitter with RTD probe and 250 ml glycol bottle
OM-CP-THERMALERT-GB250-CERT† Wireless temperature transmitter with RTD probe and 250 ml glycol bottle
and NIST calibration certificate
OM-CP-THERMALERT-P Wireless temperature transmitter with precision RTD probe
OM-CP-THERMALERT-P-CERT† Wireless temperature transmitter with precision RTD probe and
NIST calibration certificate
OM-CP-RFC1000
OM-CP-BAT109 Replacement 3.6V lithium battery
Comes complete with 3.6V lithium battery.
The OM-CP-RFC1000 transceiver package is required to operate the wireless transmitter and is sold separately.
OM-CP-THERMALERT-GB includes glycol bottle.
Ordering Example: OM-CP-THERMALERT-GB30, wireless temperature transmitter and OM-CP-RFC1000 wireless transceiver package.
Visit omega.com/om-cp-thermalert-gb_p for Pricing and Details
Wireless transceiver package includes RF transceiver, USB cable,
ac adaptor, OM‑CP Windows software, RF‑LogNet software
and operator’s manual
To Order, Call or Shop Online at omega.com
SM
2