Omega Products OM-CP-PR140 User Manual

High Temperature Pressure Data Logger
®
Part of the NOMAD
Family
OM-CP-PR140
Optional
U Operates up to 140°C (284°F) U Submersible (IP68) U User Replaceable Battery U Rugged U Programmable Start Time U Software Battery Life Indicator
OM-CP-IFC400, Windows®
The OM-CP-PR140 is a pressure data logger designed for use in autoclave validation and mapping. This rugged device can withstand temperatures up to 140°C (284°F) and is completely submersible (IP68).
The OM-CP-PR140 is built with a precision stainless steel pressure gauge. The device has an accuracy of ±0.03 Bar (±0.435 psi), which can be achieved over a wide temperature range, from 20 to 140°C (68 to 284°F).
The OM-CP-PR140 can be programmed to take readings as often as once per second (1 Hz), and has non-volatile memory that can store up to 32,700 measurements.
Simply place the OM-CP-PR140 in the OM-CP-IFC400 docking station, connect to an available USB port on your computer, and the device can be started, stopped or downloaded using our user friendly software. Graphical, tabular and summary data is provided for analysis and can be viewed in psia, mmHg, bar, Torr and kPa. The data can also be automatically exported to Excel® for further calculations. The storage medium is non-volatile solid state memory, providing maximum data security even if the battery becomes discharged.
Specifications
Pressure Sensor: Semiconductor
strain gauge Pressure Range: 0 to 5 bar
(0 to 72.5 psia) Pressure Resolution:
0.1 mBar (0.0015 psia) Calibrated Accuracy: ±0.03 bar
(±0.435 psi) (20 to 140ºC/68 to 284ºF)
software displays data in graphical
Pressure Response Time:
0.1 ms (10 to 90% FSR) Repeatability: ±0.5% FSR;
±0.2% typical Reading Rate: 1 second
to 1 every 24 hours Memory: 32,700 readings;
software configurable memory wrap Start Modes: Software
programmable immediate start or delay start up to six months
Real Time Recording: May be used with PC to monitor and record data in real time
Calibration: Digital calibration through software
Calibration Interval: Monthly calibration recommended
Data Format: psia, bar, mbar, atm, Torr, Pa, kPa, MPa, mmHg, inHg,
20, mH20, foot, meter
inH Battery Type: 3.6V lithium battery
included; user replaceable
Battery Life: 2 years typical Time Accuracy: ±1 minute/month
at 20ºC
To Order Visit omega.com/om-cp-pr140 for Pricing and Details
Model No. Description
OM-CP-PR140 High temperature pressure data logger OM-CP-PR140-CERT High temperature pressure data logger
and NIST calibration certificate OM-CP-IFC400 Docking station with USB cable, software
and manual OM-CP-BAT110 Replacement 3.6V high temperature lithium battery
Comes complete with 3.6V lithium battery. Operator’s manual and USB interface cable are included with the OM-CP-IFC400 software/cable package (required for data logger operation, sold separately).
Ordering Example: OM-CP-PR140-CERT, high temperature pressure data logger with NIST calibration certificate.
or tabular format.
OM-CP-PR140 data logger and OM-CP-IFC400 docking station
shown actual size.
Computer Interface: USB
(OM-CP-IFC400 interface cable required); 125,000 baud
Software: XP SP3/Vista/Windows 7 (32- and 64-bit)
Operating Environment:
-20 to 140ºC (-4 to 284ºF) 0 to 100% RH, may be used above 60°C (140°F) for up to 8 hrs per 24 hr period
Dimensions: 50.8 L x 25.4 mm dia. (2.0 x 1.0")
Weight: 85 g (3 oz) Enclosure: Stainless Steel
To Order, Call or Shop Online at omega.com
SM
1