Infrared Thermocouple Data Logger
Part of the NOMAD® Family
OM-CP-IRTC101A
OM-CP-WATERBOX101A,
enclosure assembly shown
smaller than actual size.
OM-CP-IRTC101A
data logger with infrared
thermocouple, shown
actual size.
OS36-01 infrared
thermocouple, sold
separately.
Optional†
U 10-Year Battery Life
U Wide Temperature Range
U Uses Type K
Infrared Thermocouple
U High Speed Download
U NIST Traceable Calibration
U Real-Time Operation
U Programmable Start/Stop Time
U Miniature Size
The OM-CP-IRTC101A is a battery
powered infrared thermocouple
based temperature data logger.
It provides nearly instantaneous,
non-contact temperature
measurements, making it a perfect
solution for applications such as
surface temperature recording and
monitoring moving objects.
The OM-CP-IRTC101A features
a 10 year battery life, 1 second
reading rate, a multiple start/stop
function, ultra-high speed download
capability, 500,000 reading storage
capacity, optional memory wrap,
battery life indicator, optional
password protection, programmable
high-low alarms and more.
The OM-CP-IRTC101A was
designed with our customers in
mind. OMEGA offers free firmware
upgrades for the life of the product
so that data loggers already
deployed in the field can grow with
new technological developments.
Units do not need to be returned to
the factory for upgrades. The user
can do this from any PC.
This is an all-in-one compact,
portable, easy to use device that will
measure and record up to 500,000
measurements per channel. The
storage medium is non-volatile solid
state memory, providing maximum
data security even if the battery
becomes discharged. The device
can be started and stopped directly
from your computer and its small
size allows it to fit almost anywhere.
The OM-CP-IRTC101A makes data
retrieval quick and easy. Simply plug
it into an empty COM or USB port
and our user-friendly software does
the rest.
Specifications
DATA LOGGER INTERNAL
TEMPERATURE CHANNEL
Range: -40 to 80°C (-40 to 176°F)
Temperature Resolution: 0.1°C
(0.18°F)
Logger Calibrated Accuracy:
±0.25°C (±0.45°F)
REMOTE CHANNEL
Thermocouple Type: K (infrared,
contact sales for other types)
Thermocouple Connection:
Female subminiature (SMP)
Cold Junction
Compensation: Automatic, based
on internal channel
Maximum Thermocouple
Resistance: 3000Ω
Thermocouple Type: K
Range (°C): See To Order chart
on next page
Sensor: ±2.0°C
Field of View: 60° (1:1)
Minimum Spot Size: 8 mm (0.3")
Spectral Response: 6.5 to 14 µ
Reading Rate: 1 reading every
2 seconds to 1 every 12 hours
Memory: 500,000 readings per
channel (1,000,000 readings total)
software configurable memory wrap
250,000 readings in multiple start/
stop mode
Wrap Around: Yes
Start Modes:
• Immediate start
• Delay start up to 18 months
• Multiple pushbutton start/stop
Stop Modes:
• Manual through software
• Timed (specific date and time)
Multiple Start/Stop Mode:
Start and stop the device multiple
times without having to download
data or communicate with a PC
Multiple Start/Stop Mode
Activation:
To Start the Device:
Press and hold the pushbutton
for 5 seconds, the green LED will
flash during this time. The device
has started logging
To Stop the Device:
Press and hold the pushbutton for
5 seconds, the red LED will flash
during this time. The device has
stopped logging.
† Refer to ordering chart on next page for
NIST calibration ordering information
To Order, Call or Shop Online at omega.com
SM
1
Average battery life vs. reading rate of
OM-CP-IRTC101A recording in a 25°C environment.
OM-CP-IFC200, Windows® software displays
data in graphical or tabular format.
Real Time Recording: May be
used with PC to monitor and record
data in real time
Alarm: Programmable high and
low limits; alarm is activated when
temperature reaches or exceeds
set limits
LED Functionality:
Green LED Blinks: 10 second
rate to indicate logging; 15 second
rate to indicate delay start
Red LED Blinks: 10 second rate
to indicate low battery and/or full
memory; 1 second rate to indicate
an alarm condition
Password Protection: An optional
password may be programmed
into the device to restrict access to
configuration options. Data may be
Battery Type: 3.6V lithium battery
included; user replaceable
Battery Life: 10 years typical at a
15 minute reading rate
Data Format: Date and time
stamped °C, °F, K, °R, µV, mV, V
Time Accuracy: ±1 minute/month
(at 20°C, stand alone data logging)
Computer Interface:
USB (interface cable required);
115,200 baud
Software: XP SP3/Vista and
Windows
®
7 (32 and 64-bit)
Operating Environment:
-18 to 70°C (-40 to 158°F)
Dimensions: 36 x 56 x 16 mm
(1.4 x 2.2 x 0.6")
Weight: 24 g (0.9 oz)
Infrared Thermocouple
Dimensions:
3.25 L x 1.80 cm Dia.
(1.28 x 0.71"); 0.9 m (36") PFA
coated unshielded stranded wire
OM-CP-WATERBOX101A
Operating Temperature:
-20 to 80ºC (-4 to 176ºF)
IP Rating: IP65
Dimensions:
62 H x 93 W x 25 mm D
(2.45 x 3.66 x 0.93")
Material: Anodized aluminum with
polycarbonate cover
Weight: 140 g (5 oz)
read out without the password.
To Order Visit omega.com/om-cp-irtc101a for Pricing and Details
Model No. Description
OM-CP-IRTC101A Infrared thermocouple data logger
OM-CP-IRTC101A-CERT† Infrared thermocouple data logger with NIST calibration certificate
OM-CP-IFC200 Windows software and 1.8 m (6') USB interface cable
OM-CP-BAT105 Replacement 3.6V lithium battery
OM-CP-WATERBOX101A Water resistant enclosure for data logger
Comes complete with 3.6V lithium battery. Operator’s manual and USB interface cable are included with the OM-CP-IFC200 software/cable
package (required for data logger operation, sold separately).
Infrared Thermocouples
Model No. Description
OS36-01-K-50F-SMPW Type K infrared thermocouple, -18 to 27°C (0 to 80°F)
OS36-01-K-80F-SMPW Type K infrared thermocouple, 10 to 49°C (50 to 120°F)
OS36-01-K-98.6F-SMPW Type K infrared thermocouple, 16 to 39°C (60 to 102°F)
OS36-01-K-140F-SMPW Type K infrared thermocouple, 25 to 80°C (80 to 180°F)
OS36-01-K-180F-SMPW Type K infrared thermocouple, 70 to 104°C (160 to 220°F)
OS36-01-K-240F-SMPW Type K infrared thermocouple, 95 to 130°C (200 to 270°F)
OS36-01-K-280F-SMPW Type K infrared thermocouple, 125 to 155°C (260 to 310°F)
OS36-01-K-340F-SMPW Type K infrared thermocouple, 145 to 175°C (290 to 350°F)
OS36-01-K-440F-SMPW Type K infrared thermocouple, 170 to 250°C (340 to 480°F)
Ordering Example: OM-CP-IRTC101A-CERT, infrared thermocouple data logger with NIST calibration certificate, OS36-01-K-440F-SMPW,
infrared thermocouple with 170 to 250°C (340 to 480°F) temperature range, and OM-CP-IFC200 Windows software and USB cable.
2
To Order, Call or Shop Online at omega.com
SM