High Temperature Data Logger with 607 mm (24") Stainless Steel
Flexible Probe with Optional Thermal Shield
®
Part of the NOMAD
Family
OM-CP-HITEMP140-PT Series
Optional†
U 607 mm (24") Flexible Probe
U Operating Temperature Range:
-40 to 140°C (-40 to 284°F)
U Optional Thermal Shield for
Operation up to 250°C (482°F)
U Submersible NEMA 4X (IP68)
U Programmable Start Time
U User Calibration Through Software
U Real-Time Operation
U Autoclave Verification
U Pharmaceutical
U Implement HACCP Programs
U Food Preparation and Processing
The OM-CP-HITEMP140-PT is
a submersible, temperature data
logger that can operate up to 140°C
(284°F) and has an accuracy of
±0.1°C (0.18°F). The device features
a 607 mm (24") flexible stainless
steel probe for measuring extended
temperatures up to 260°C (500°F).
The probe is durable and can be
spiraled, bent or angled in any
direction, making it easy to log
temperatures within bottles, vials or
other hard to reach places.
The OM-CP-HITEMP140-PT
records date and time stamped
readings, and has non-volatile
solid state memory that will retain
data even if the battery becomes
discharged. Data retrieval is simple.
Plug the device into an available
USB port and our easy-to-use
software does the rest. The software
converts a PC into a real-time strip
chart recorder.
Using the software, starting,
stopping and downloading the
OM-CP-HITEMP140-PT is simple
and easy. Graphical, tabular and
summary data is provided for
analysis and data can be viewed in
°C, °F, K or °R. The data can also be
automatically exported to Excel
®
for
further calculations.
OM-CP-HITEMP140-PT,
shown smaller than
actual size.
Docking station
sold separately.
The OM-CP-HITEMP140-TSK
is a kit that includes a
OM-CP-HITEMP140-PT data logger
housed in a thermal shield.
The combined features of
the ±0.1°C accuracy of the
OM-CP-HITEMP140-PT and the
properties of the durable thermal
shield allow the device to be used
for a wide range of validation
applications.
Using the software, the data
logger is fast and easy to setup.
Remove the thermal shield and
place the OM-CP-HITEMP140-PT
into the OM-CP-IFC400 docking
station (sold separately). Using the
software, an immediate or delay
start can be chosen, as well as the
reading rate. Select Start to program
the settings and start the data
logger.
OM-CP-HITEMP140-PT-TSK,
shown smaller than
actual size.
OM-CP-IFC400, Windows software
displays data in graphical
or tabular format.
Place the thermal shield around the
OM-CP-HITEMP140-PT and screw
it back together. The device is ready
to be deployed.
The OM-CP-HITEMP140-TSK can
be completely submerged and is
built for applications that require
extreme temperature monitoring.
These two devices can be started
and stopped directly from a
computer and their compact design
allows them to fit almost anywhere.
† Refer to ordering chart on next page for
NIST calibration ordering information.
To Order, Call or Shop Online at omega.com
SM
1
Specifications
OM-CP-HITEMP140-PT-TSK
(Thermal Shield)
Operating Environment:
-200 to 250ºC (-328 to 482ºF)
(time limited) 0 to 100% RH
Enclosure Material: PTFE
Dimensions: 110 H x 51 mm dia.
(4.3 x 2.0")
Weight: 274 g (9.7 oz)
not including data logger
OM-CP-HITEMP140-PT
(Without Thermal Shield)
Temperature Sensor: 100Ω
Platinum RTD
Probe Measurement Range:
-200 to 260ºC (-328 to 500ºF)
Temperature Resolution:
0.01ºC (0.02ºF)
Calibrated Accuracy:
±0.1ºC (±0.18ºF);
20 to 140ºC (68 to 284ºF)
Start Modes: Software
programmable immediate start
or delay start up to eighteen
months in advance.
Stop Modes: Manual through
software timed (specific data
and time)
Real Time Recording: May be
used with PC to monitor and
record data in real time
Password Protection: An optional
password may be programmed
into the device to restrict access to
configuration options, Data may be
read out with the password,
Memory: 32,700 readings
Wrap Around: Yes
Reading Rate: One second up
to once every 24 hours
Battery Type: 3.6V high-
temperature lithium battery
(included); user replaceable
Battery Life: 1 year typical
[1 minute reading rate at
25ºC (77ºF)]
Calibration: Digital calibration
through software
Calibration Date: Automatically
recorded within device
Data Format: Date and time
stamped ºC, ºF, K, ºR
Time Accuracy: ±1 minute/month
at 20 to 30ºC (68 to 86ºF)
Computer Interface:
Software: Windows
VISTA/7 (32- and 64-bit)
Operating Environment:
-20 to 140ºC (-4 to 284°F),
0 to 100% RH
Weight: 120 g (4.2 oz)
Material: 316 Stainless Steel
Dimensions:
Body: 67 x 25 mm dia.
(2.7 x 0.97")
Probe Tip: 3.2 OD x 38 mm L
(0.125 x 1.5")
Flexible Probe Portion:
607 x 1.6 mm OD dia.
(24 x 0.062")
®
XP SP3/
OM-CP-IFC400 docking station
required; 125,000 baud
OM-CP-HITEMP150-PT-TSK (Data Logger with Thermal Shield)
Time in Air to Max Internal Time in Liquid to Max Internal
Ambient Temperature Temp (140°C / 284°F) Temp (140°C / 284°F)
-200°C (-328°F) 18 min N/A
-180°C (-292°F) 19 min N/A
-160°C (-256°F) 21 min N/A
-140°C (-220°F) 24 min N/A
-120°C (-184°F) 27 min N/A
-100°C (-148°F) 32 min N/A
-80°C (-112°F) 40 min N/A
-60°C (-76°F) 55 min 25 min
-40°C (-40°F) 70 min 32 min
-20 to 140°C (-4 to 284°F) Indefinitely Indefinitely
150°C (302°F) 88 min 40 min
160°C (320°F) 75 min 34 min
170°C (338°F) 63 min 29 min
180°C (356°F) 55 min 26 min
190°C (374°F) 50 min 23 min
200°C (392°F) 45 min 21 min
210°C (410°F) 42 min 19 min
220°C (428°F) 39 min 18 min
230°C (446°F) 36 min 17 min
240°C (464°F) 34 min 16 min
250°C (482°F) 32 min 15 min
To Order Visit omega.com/om-cp-hitemp140-pt for Pricing and Details
Model No. Description
OM-CP-HITEMP140-PT High temperature data logger with 607 mm (24") stainless steel flexible probe
OM-CP-HITEMP140-PT-CERT† High temperature data logger with 607 mm (24") stainless steel flexible probe
and NIST certificate
OM-CP-HITEMP140-PT-TSK OM-CP-HITEMP140-PT high temperature data logger with 607 mm (24") SS flexible
probe and thermal shield
OM-CP-HITEMP140-PT-TSK-CERT† OM-CP-HITEMP140-PT high temperature data logger with 607 mm (24") SS flexible
probe and thermal shield and NIST certificate
OM-CP-IFC400 Docking station with software, USB cable and manual
OM-CP-BAT110 Replacement 3.6V lithium battery
Comes complete with 3.6V lithium battery. Operator’s manual and USB interface cable are included with the OM-CP-IFC400 software/cable
package (required for data logger operation, sold separately).
Ordering Example: OM-CP-HITEMP140-PT data logger with 607 mm (24") stainless steel flexible probe and OM-CP-IFC400 docking station
with software.
2
To Order, Call or Shop Online at omega.com