Portable Handheld Data Logger
OM-2041
U Input Signals: K, E, J, T, 0 to 5 V, and 0 to 20 mA
(Process Signals Need Input Adaptor)
U Simultaneous 4-Channel Measurement
and Display
U High/Low Alarms for Each Channel
U Computation: Maximum/Minimum/Average
U Logging Method: Log Mode and Tag Mode
U Data Processable Through a Personal Computer
U NEMA 3 Protection
U “AA” Battery Power Supply
The handy OM-2041 is a palm-size data logger capable of
storing up to 10,000 data records per channel. Powered
by either “AA” batteries or an AC adaptor, the OM-2041
works equally well as a portable or a fixed-location logger.
With the exclusive software, recording/recalling of data
and programming of logging parameters can be done
directly with the logger or through a personal computer.
OM-2041
Specifications
MEASURING RANGE
Thermocouple:
K: -200.0 to 1370.0°C (-328.0 to 2498.0°F)
E: -200.0 to 700.0°C (-328.0 to 1292.0°F)
J: -200 to 700.0°C (-328.0 to 1292.0°F)
T: -200.0 to 400.0°C (-328.0 to 752.0°F)
DC Voltage: 0 to 5 V (scale -9999 to 9999)
DC Current: 0 to 20 mA (scale -9999 to 9999)
ACCURACY RATING
Thermocouple:
100°C or More: (±0.1% of measuring value +0.3°C)
100°C or Less: (±0.1% of measuring value +0.6°C)
DC Voltage: (±0.1% of measuring value +0.2% of range)
Reference Junction Compensation Accuracy:
±0.4°C (at ambient temperature 15 to 35°C);
±0.7°C (at ambient temp. -10 to 15, 35 to 50°C)
Temperature Drift: ±0.01%/°C of measuring range
Allowable Signal Source Resistance:
100 Ω (burnout over upper limit)
Input Resistance: DC voltage inputs over 1 MΩ
Withstand Voltage Between Channels: 400 Vac
Measuring Cycle: 1 second/4 channels
Logging Interval:
1 second to 30 seconds (1 second steps);
1 minute to 24 hours (1 minute steps)
Logging Data:
Log Mode: 10,000 data records/channel
Tag Mode: 5500 data records/channel
Display: LCD with backlight
Dot-matrix 128 x 64 dot (view area 41 x 29 mm)
Operation Key: 18-key including numeric key
Logging Method:
Log Mode: Logging data per cycle (20 points) in log area
Tag Mode: Logging data (20 points), then press the
trigger in tag area
Display Contents: Measured value, alarm, maximum/
minimum/average value, computation value, etc.
Other Functions: Clock (year, month, day/hour, minute,
second), auto power-off, reading correction, backlight,
buzzer, residual battery, residual memory, alarm,
calculation, key lock, etc.
To Order, Call or Shop Online at omega.com
SM
1
Communications: RS232C or USB
Power Supply: 4 “AA” batteries (included)
or AC adaptor (100 to 240 Vac)
Battery Life: 500 hours or longer
(at 25°C, 1-minute interval)
Backup Battery: Backup log data, setting
and clock by lithium battery
Backup Battery Life: 5 years or longer (at 25°C)
Dimensions: 159 H x 70 W x 33 mm D
(6.2 x 2.7 x 1.3") excluding protrusions
Weight: Approx 300 g (10.5 oz) including batteries
Casing Material: Nylon ABS (antibacterial)
Ambient Temperature: -10 to 50°C (14 to 122°F)
Ambient Humidity: 10 to 80% RH (no condensation)
CE Approval: EN-61326, Class B (battery operation)
Protection: NEMA 3
Accessories: 4 “AA” batteries, wall-mount
holder (included)
SM
OMEGACARESM extended warranty program
is available for models shown on this page.
Extended Warranty
Program
Ask your sales representative for full details
when placing an order. OMEGACARE
covers parts, labor and equivalent loaners.
SM
To Order
Visit omega.com/om-2041 for Pricing and Details
Model No. Description
OM-2041-USB 4-channel handheld data logger
with USB port
OM-2041-C2 4-channel handheld data logger
OM-SPT-USB Software with USB cable
OM-SPN-C2 Software with RS232C cable
with C2 port
Accessories
Model No. Description
OM-9407 Current input adaptor
OM-9408 Voltage input adaptor
OM-AC AC adaptor
Comes complete with operator’s manual, beaded wire Type K
probe for each input, wall mount holder, and 4 “AA” batteries.
Ordering Example: OM-2041-USB, 4-channel handheld data
logger. OCW-3, OMEGACARESM extends standard 2-year warranty
to 5 years.
General Purpose Miniature
Probes with Miniature
Connectors
To order
ANSI
IEC color
color
code visit
code
omega.com
shown
U Works Great with Handheld Instruments
U 150 mm (6") Length Standard
†
U 3.2 mm (1⁄8") OD 304 SS Sheath
U Molded Junction Rated to 220°C (425°F)
U Color-Coded to Calibration
U 300 mm (12") Retractable Cable Expands
to 1.2 m (48")
U SMPW Male Connector Termination
The molded miniature probe works with most handheld
thermocouple meters. Its color-coded retractable
cable will stretch from 300 mm (12") to 1.2 m (48").
Grounded junction probes are suited for applications
involving benchtop meters or oscillating machines.
Calibration codes J, K, T, and E are available.
Must-Have
Accessory!
Shown smaller
than actual size.
Miniature connector
standard.
Thermocouple Model No.
Alloy ANSI Color Code
Iron Constantan
CHROMEGA
ALOMEGA
CHROMEGA® Constantan
Copper-
®
-
®
Constantan
Note: Hot end junction grounded.
† Other lengths are available, consult Sales department.
Ordering Example: KTSS-HH, Type K thermocouple probe,
150 mm (6") long, 3.2 mm (1⁄8") dia with 1' of RSC cable (which
stretches to 5') with a molded, miniature male thermocouple
connector.
JTSS-HH
KTSS-HH
ETSS-HH
TTSS-HH
2
To Order, Call or Shop Online at omega.com
SM