Omega Products LVU104 User Manual

NON-CONTACT ULTRASONIC LEVEL TRANSMITTER
LVU100 Series
U Simple Compact Sensor
with Target Calibration and LED Status Indicators
U 4 Measurement Ranges
U 5 cm (2") Minimum Beam
Width for Applications with Restricted Space
U 5 cm (2") Deadband
Optimizes the Filling Capacity of Small Vessels
U PC/ABS Enclosure Rated
NEMA 6 (IP67) with Optional NEMA 4X (IP65)
U Fail-Safe Intelligence with
Diagnostic Feedback for Easy Troubleshooting
The LVU100 Series is a general purpose 2-wire ultrasonic transmitter. It provides non-contact level measurement up to 8 m (26'), and is well suited for handling ultrapure, corrosive, slurry or waste liquids. The LVU100 Series transmitter is often used in atmospheric bulk storage, day tank, waste sump, process vessel, IBC tote, 55-gallon drum and waste sump applications. Media examples include acetic acid and resin.
SPECIFICATIONS
Range: LVU104: 5 cm to 1.2 m (2" to 4')
LVU109: 10 cm to 3 m (4" to 9.8') LVU116: 10 cm to 5 m (4" to 16.4') LVU126: 20 cm to 8 m (8" to 26.2')
Accuracy:
LVU104: 3 mm (0.125") LVU109/116/126: ±0.2% of span in air
Resolution: LVU104/109: 0.5 mm (0.019") LVU116/126: 1 mm (0.039")
Beam Width:
LVU104/109: 5 cm (2") dia. LVU116/126: 7.6 cm (3") dia.
Deadband:
LVU104: 5 cm (2") LVU109/116: 10 cm (4") LVU126: 20 cm (8") LED Indication: Power, calibration and diagnostics
Memory: Non-volatile Supply Voltage: 12 to 28 Vdc Loop Resistance: 500 Ω @ 24 Vdc Signal Output: 4 to 20 mA, 2-wire Signal Invert: 4 to 20 mA or 20 to 4 mA Calibration: Target, calibration wire Fail-Safety: Reverts to 22 mA Process Temperature: -20 to 60°C
(-4 to 140°F)
Temperature Compensation: Automatic Electronics Temperature: -40 to 71°C
(-40 to 160°F)
To Order Visit omega.com/lvu100 for Pricing and Details
Model No. Description
LVU104 2-wire transmitter—1.2 m (4') range, 1 NPT, NEMA 6 LVU104G 2-wire transmitter—1.2 m (4') range, 1 straight thread, NEMA 6 LVU109 2-wire transmitter—3 m (9.8') range, 1 NPT, NEMA 6 LVU109G 2-wire transmitter—3 m (9.8') range, 1 straight thread, NEMA 6 LVU116 2-wire transmitter—5 m (16.4') range, 2 NPT, NEMA 6 LVU116G 2-wire transmitter—5 m (16.4') range, 2 straight threads, NEMA 6 LVU126 2-wire transmitter—8 m (26.2') range, 2 NPT, NEMA 6 LVU126G 2-wire transmitter—8 m (26.2') range, 2 straight threads, NEMA 6 LVM-30 Side mount bracket 2 NPT
Comes complete with operator’s manual and 3 m (10') cable. For NEMA 4X (IP65) rated enclosure add “-NEMA4” to the model number for additional cost.
(cable not included with this model). Ordering Examples: LVU116, 2-wire transmitter, 5 m (16.4') range, 2 NPT mounting, NEMA 6
(IP67) enclosure with 3 m (10') cable. LVU104-NEMA4, 2-wire transmitter, 1.2 m (4') range, 1 NPT mounting, optional NEMA 4X
(IP65) enclosure.
K-9
LVU104, shown smaller than actual size.
Pressure: 30 psi (2 bar) @ 25°C,
derated @ 1.667 psi (0.115 bar) per °C above 25°C Enclosure Rating: NEMA 6 (IP67), NEMA 4X (IP65) optional
Enclosure Material: PC/ABS FR Transducer Material: PVDF Cable Jacket Material: PP Cable: 3-conductor, shielded 3 m (10') Conduit Entrance: Dual,
on NEMA 4 (IP65) version
Process Mount:
LVU104/109: 1 NPT or 1" G LVU116/126: 2 NPT or 2" G
Mounting Gasket: FKM Classification: General purpose
CE compliance