LOAD WASHERS
FOR MEASURING BOLT COMPRESSIONAL FORCES
STANDARD AND METRIC MODELS
Compression
0-200 lb to 0-200,000 lb
0-1000 to 0-200 kN
1 Newton = 0.2248 lb
1 daNewton = 10 Newtons
1 lb = 454 g
1 t = 1000 kgf = 2204 lb
LCWD/LCMWD Series
Standard
U 200 kN/200,000 lb Capacity
U Compact Design
U Spherical Washers
Included
U Load Buttons Included
U Spherical Washers
Recommended
When Used with Bolts
Load Button Recommended
U
When Used as
Compression Load Cell
SPECIFICATIONS
Excitation: 10 Vdc (15V max)
Output: 2 mV/V
Accuracy (Linearity, Hysteresis and
Repeatability Combined):
≤10K lb/50 kN: ±0.5% BFSL
>10K lb/50 kN: ±0.75% BFSL
Zero Balance: ±4.0% FSO
Operating Temp Range:
-40 to 127°C (-40 to 260°F)
Compensated Temp Range:
16 to 71°C (60 to 160°F)
Thermal Effects:
Zero: ±0.018% FSO/°C
Span: ±0.018% FSO/°C
Safe Overload: 120% of capacity
Ultimate Overload: 200% of capacity
Bridge Resistance:
<50K lb/200 kN: 350 Ω min
≥50K lb/200 kN: 700 Ω
Full Scale Deflection: 0.001" typical
Construction: Stainless steel
Electrical:
1.5 m (5') insulated shielded cable
STANDARD
LCWD-500, with spherical
washers (included), shown
actual size.
METRIC
with spherical washers
and load button (included),
WIRING CODE
RED +EXC
BLACK -EXC
GREEN +OUT
WHITE -OUT
O.D.
LCMWD-2KN,
shown actual size.
I.D.
Dimensions: mm (inch)
CAPACITY ID OD H
≤10K lb (50 kN) 10.2 (0.40) 25 (1.0) 9.0 (0.35)
20K lb (100 kN) 12.7 (0.50) 30 (1.2) 10 (0.40)
50K lb (200 kN) 19.3 (0.76) 44 (1.75) 13 (0.50)
100K lb 25.7 (1.01) 57 (2.25) 18 (0.70)
200K lb 38.4 (1.51) 77 (3.00) 32 (1.25)
Caution
This unit is not field serviceable.
H
CABLE
STANDARD MODELS
To Order
lb N NO. METERS
200 890 LCWD-200
500 2224 LCWD-500
1000 4448 LCWD-1K
2000 8897 LCWD-2K
5000 22,242 LCWD-5K
10,000 44,484 LCWD-10K
20,000 88,968 LCWD-20K
50,000 222,420 LCWD-50K
100,000 444,840 LCWD-100K
200,000 889,680 LCWD-200K
Comes complete with 4-point NIST-traceable calibration and 59 kΩ shunt data.
4-digit meter.
*
† One set of spherical washers and one set of load buttons supplied with each load cell.
Ordering Examples: LCWD-500, 500 lb capacity load washer supplied with 2 load buttons
and a set of spherical washers.
LCWD-100K, 100,000 lb capacity load washer supplied with 2 load buttons and set of
spherical washers.
Visit omega.com/lcwd for Pricing and Details
CAPACITY MODEL COMPATIBLE
Modifying and/or tampering will forfeit
the warranty and may be dangerous.
Use appropriate protective equipment
and/or garments when testing or
installing, including safety glasses.
SPHERICAL WASHERS
MODEL ID
DP41-S, DP25B-S, DPiS
DP41-S, DP25B-S, DPiS
DP41-S, DP25B-S, DPiS
DP41-S, DP25B-S, DPiS
DP41-S, DP25B-S, DPiS
DP41-S, DP25B-S*, DPiS*
DP41-S, DP25B-S*, DPiS*
DP41-S, DP25B-S*, DPiS*
DP41-S, DP25B-S*, DPiS*
DP41-S, DP25B-S*, DPiS*
SPW-2 8.7 (0.34)
SPW-2 8.7 (0.34)
SPW-2 8.7 (0.34)
SPW-2 8.7 (0.34)
SPW-2 8.7 (0.34)
SPW-2 8.7 (0.34)
SPW-3 10 (0.41)
SPW-5 17 (0.66)
SPW-7 24 (0.94)
SPW-10 33 (1.31)
mm (inch)
†
F-83
LOAD WASHERS
FOR MEASURING BOLT COMPRESSIONAL FORCES
STANDARD AND METRIC MODELS
METRIC MODELS
To Order Visit omega.com/lcwd for Pricing and Details
MODEL COMPATIBLE
LCMWD-1KN
LCMWD-2KN
LCMWD-5KN
LCMWD-10KN
LCMWD-20KN
LCMWD-50KN
LCMWD-100KN
LCMWD-200KN
DPiS, DP41-S, DP25B-S
DPiS, DP41-S, DP25B-S
DPiS, DP41-S, DP25B-S
DPiS*, DP41-S, DP25B-S*
DPiS*, DP41-S, DP25B-S*
DPiS*, DP41-S, DP25B-S*
DPiS*, DP41-S, DP25B-S*
DPiS*, DP41-S, DP25B-S*
METRIC
LCMWD-1KN, with spherical
washers (included), shown
actual size.
CAPACITY
N lb NO. METERS MODEL ID
1000 225
2000 450
5000 1124
10,000 2248
20,000 4496
50,000 11,240
100,000 22,481
200,000 44,962
Comes complete with 5-point NIST-traceable calibration and 59 kΩ shunt data.
4-digit meter.
*
One set of spherical washers and one set of load buttons supplied with each load cell.
†
Ordering Examples: LCMWD-10KN, 10 kN capacity load washer supplied with 2 load buttons
and a set of spherical washers.
LCMWD-200KN, 200 kN capacity load washer supplied with 2 load buttons and a set of
spherical washers.
SPHERICAL WASHERS
SPW-2 8.7 (0.34)
SPW-2 8.7 (0.34)
SPW-2 8.7 (0.34)
SPW-2 8.7 (0.34)
SPW-2 8.7 (0.34)
SPW-2 8.7 (0.34)
SPW-3 10 (0.41)
SPW-5 17 (0.66)
†
STRAIN GAGE/BRIDGE
TRANSDUCER-SIGNAL
CONDITIONERS
iDRX-ST
U Unipolar/Bipolar
30 mV to 100 mV
U 13-Bit Resolution
U 10 Vdc Excitation
U 0.2% FS Accuracy
U 1800 Vdc Isolation
U 250 Vac/1 Min Input
Overvoltage Protection
U Free Setup and
Configuration Software
The iDRN-ST and iDRX-ST
signal conditioners provide
highly accurate, stable,
isolated measurement for
strain gage transducers
such as load cells,
torque transducers,
non-amplified pressure
transducers, and other bridge
based transducers. Both
models can accept signals
from 30 to 100 mV full scale
and provide 10 Vdc
reference voltage which
may be used for
transducer excitation.
Shown smaller
than actual size.
To Order Visit omega.com/drn-drx for Pricing and Details
MODEL NO. DESCRIPTION
iDRX-ST Digital signal conditioner for strain gages and bridge transducers
with RS485 output
iDRN-ST Signal conditioner for strain gages and bridge transducers with
analog output
-FS Factory setup and scaling
Comes complete with operator’s manual.
Ordering Example: iDRX-ST, digital signal conditioner for strain gages and bridge
transducers with RS485 output, CAT-285, bi-directional RS232-RS485 converter for
iDRX series.
For iDRN/iDRX accessories and power supplies, please visit omega.com/drn-drx
F-84
LOAD CELLS