Omega Products LCKD Installation Manual

omega.com
e-mail: info@omega.com
For latest product manuals:
omegamanual.info
User’s Guide
LCKD SERIES
Subminiature Compression
Load Cells
Shop online at
LCKD SERIES Subminiature Compression Load Cells
1
Each LCKD Series Load Cell incorporates a small printed circuit board into the load cell’s lead wire. DO NOT remove this board or cut the cable between the compensation board and load cell. Removal of this board voids the calibration and warranty of the load cell.
General
The OMEGA®LCKD Series subminiature compression load cells are compression-only units that are highly cost-effective. To ensure excellent long-term stability and reliability in severe environments, the LCKD Series utilize high quality strain gages, precision gaging techniques and all stainless steel construction. These units have a load button machined as an integral part of the basic load cell. The load cell is designed to operate by mounting on a flat surface. The LCKD must rest on a flat surface the same diameter as the D
1
dimension for proper
operation.
Shunt Calibration
The LCKD Series are highly accurate millivolt output type load cells with shunt calibrator for quick calibration checks. Shunt calibration allows the user to install and calibrate the instrument in the field without the use of a dead weight tester. A 59 kilohm resistor is shorted across negative excitation and negative signal output at the factory, which produces a simulated millivolt signal out of the transducer. The shunt calibration signal is equivalent to a simulated pressure of:
Shunt Cal mV/V
Simulated Load = x Full Scale Load
Calibration Factor mV/V
Example: Model LCKD
Where:
Calibration Factor - 2.0315 mV/V Shunt Cal - 1.4962 mV/V
Simulated Load = 1.4962
x 50 = 36.85
2.0315
To set up the transducer in the field, follow these steps:
1. Connect transducer excitation terminals to dc power supply.
2. Connect transducer signal output terminals to readout instrument (DVM, Analog meter, etc.)
3. Turn power on.
4. Null transducer signal output with zero adjust potentiometer on meter.
5. Short a 59 kilohm resistor across negative excitation and negative signal output.