GENERAL
The OMEGA®LCGD Series bonded foil strain gage load cells are compression only units that are highly cost effective. To ensure excellent long term stability and reliability in severe environments, the LCGD Series utilizes high quality strain gages, precision gaging techniques and all stainless steel construction. These units have a machined load button as an integral part of the basic load cell. The LCGD Load Cell is designed to operate by mounting on a flat surface. The LCGD must rest on a flat surface at least the diameter D1dimension for proper operation.
The part number indicates the load cell type (LCGD) and load range (eg., 500 for 500 lb). Models are available with load ranges of 100, 250, 500, 1000, 2000, 5000, 10,000 and 50,000 lb. LCGD series load cells are compatible with the DP41-S and DP25B-S digital panel meters. These and other compatible instruments can be found in Section D of the OMEGA Pressure, Strain and Force Handbook™.
SPECIFICATIONS
Signal Output: 2mV/V nominal
see cal. sheet for exact output
Linearity: 0.25% FS, BFSL
Hystersis: 0.2%
Zero Balance: +2%
Repeatability: ±0.1% FS
Compensated Temperature Range: 60°F to 160°F
Shop online at: omega.com e-mail: info@omega.com For latest product manuals: omegamanual.info
3 MOUNTING HOLES, "A" THREAD X "B" DEEP, EQUALLY SPACED AS SHOWN ON "C" DIAMETER BOLT CIRCLE
L
H
D
1
D
2
S.S. SPRING STRAIN RELIEF
SPHERICAL LOAD BUTTON
1.5m (5 ft.) Teflon
®
Cable
Temperature Effect:
Zero: ±0.009% FSO/°C Span: ±0.018% reading/°C
Bridge Resistance: 350 ohm foil strain gage
Excitation Voltage: 10 V
Safe Overload: 150% of full scale
Electrical Connection: 5 feet of shielded cable with pigtail
end
Operating Temperature Range: -65° to 250°F
CABLE COLOR CODE
+EXC RED +SIG WHITE
- SIG GREEN
- EXC BLACK