Temperature and Humidity
Virtual Chart Recorder
Chart
http://192.168.1.200Address
iTHX-D3 Series
FCC-B
Optional
U Web Server
U Alarms by Email or Text Message
U Accurate Readings
U Data Logging
U No Special Software Required
View Temperature and Humidity
with a Web Browser
The OMEGA® iTHX transmitter lets you monitor and
record temperature, relative humidity, and dew point
over an Ethernet network or the Internet with no
special software except a web browser.
The iTHX serves active web pages to display realtime readings, charts of temperature and humidity,
or log data in standard data formats for use in a
spreadsheet or data acquisition program such as
Excel or Visual Basic.
The virtual chart viewed on the web page is a JAVA™
Applet that records a chart over the LAN or Internet
in real time. With the iTHX, there is no need to invest
time and money learning a proprietary software
program to log or chart the data.
Adjustable Charts
Chart scales are fully adjustable on the fly. For
example, the chart can display one minute, one
hour, one day, one week, one month or one year.
Temperature and humidity can be charted across the
full span [-40 to 124°C (-40 to 254°F)], and 0 to 100%
RH or within any narrow range such as 20 to 30°C
(68 to 86°F).
Award-Winning Technology
The OMEGA iTHX is simple to install and use, and
features OMEGA’s award-winning iServer technology
that requires no special software except a
web browser.
The iTHX connects to an Ethernet network with a
standard RJ45 connector and sends data in
standard TCP/IP packets. It is easily configured with
a simple menu using a web browser and can be
password protected.
From within an Ethernet LAN or over the Internet,
the user simply types the IP address or an easy-toremember name such as “Cleanroom5” or “Server
Room” in any web browser, and the iTHX serves a
web page with the current readings.
Email Alarms
All OMEGA iTHX models that are on a LAN that is
connected to the Internet can trigger an alarm that
can be sent by email to a user or a distribution list
anywhere in the world, including text messages to
cell phones and PDA’s.
Link to Web CAM or IP Camera
The web page includes a link to a “web cam” or “IP
camera” (not included). If you get a message about
an alarm condition, you can quickly click on the link to
view the actual scene over the Internet.
Readings Chart Web Link Setup
High 150.0 / Low 78.0 High 78.0 / Low 25.0
Max. 130.1 / Min. 95.9 Max. 69.9 / Min. 30.1
Temperature Humidity
200
20.0
C/Div
0
Data Source:
Alarm Relay Set Points:
Help[?]
108.2 F 50.8 %
Live
Bold
PROBE INCLUDED!
®
™
Recording: ON
1 Hour/Div
X-axis: 1 Day
Y-axis (left): TemperatureStyle: Bold
Y-axis (right):
Humidity Style: Bold
Thu Sep 17 12:05:07 PDT 2009Thu Sep 16 08:00:00 PDT 2009
Save Chart
Print Chart
RAIL-35-2
DIN Rail,
2 m (6.6'),
sold
separately.
iTHX-D3 iServer
MicroServer™
shown with industrial
temperature/humidity
probe (included).
10.0
%/Div
Sensor and Calibration
The iTHX transmitters come complete with a
temperature and humidity probe for measurement of
a single location.
OMEGA offers a choice of industrial probes in 2"
(with a 3' cable) or 5" length (with a 10' cable)
An optional Porous
Stainless Steel Cap
iP-SC is available
for industrial
probes.
iP-SC
Standard
100
0
To Order, Call or Shop Online at omega.com
SM
1
instead of the commercial wand style
probe which is designed for ambient
indoor applications.
It is not necessary to take the iTHX
out of service for routine calibration.
The temperature/humidity sensors
are interchangeable and can be
replaced for about the same cost as
a typical calibration.
OMEGA offers replacement probes
with optional 3-point NIST traceable
calibration certificates.
Typical Applications
The OMEGA iTHX is great for
monitoring temperature and humidity
in applications such as: clean rooms,
computer rooms, HVAC systems,
pharmaceutical and food processing
and storage, hospitals, laboratories,
semiconductor labs, electronic
assembly, warehousing, museums,
manufacturing, greenhouses and,
farm animal shelters.
Sensor Specifications
RELATIVE HUMIDITY (RH)
Accuracy/Range:
10 to 90%: ±2%
5 to 10% and 90 to 95%: ±3%
0 to 5% and 95 to 100%: ±4%
Non-Linearity: ±3%
Hysteresis: ±1% RH
Response Time: 8 seconds, tau 63%
Repeatability: ±0.1%
Resolution: 0.1%, 12 bit
TEMPERATURE
Accuracy/Range**:
Industrial Probe: (see chart in
manual) ±0.5°C (±1°F) for 5 to 45°C
(41 to 113°F); up to ±1.5°C (up to
±2.7°F) for -40 to 5°C and 45 to 124°C
(-40 to 41°F and 113 to 255°F)
** Note: Extended temperature range is for
industrial probe only, the iServer’s operating
temp is -40 to 85°C (-40 to 185°F).
Response Time: 5 to 30 seconds, tau 63%
Repeatability: ±0.1°C
Resolution: 0.1°C, 14 bit
Probe Specifications
iTHP-5, iTHP-2: 137 or 51 L x 16 mm Dia
(5 or 2 x 0.63")
Housing Material: 316 SS
Cable: 3 or 0.9 m L (10 or 3')
Cable Operating Temperature:
-40 to 125°C (-40 to 257°F)
NIST-Traceable Calibration
Certificate available.
In compliance with ISO9001:2008,
ISO10012-1.1992(E),
ANSI/NCSL Z540-1.1994 and
MIL-STD-45662A.
iServer
Mechanical Specifications
Specifications
INTERFACES
Ethernet: (RJ45):
Fixed or auto-negotiating
10/100BASE-T, auto
MDI/MDIX
Sensor: Removable 4 pos
screw terminal
Supported Protocols:
TCP, UDP, SNMP, SMTP,
NTP, ARP, ICMP, DHCP,
DNS, HTTP, and Telnet
Embedded Web Server:
Serves web pages
containing real-time data
and live updated charts
within
definable time intervals
Software: OPC Server;
macro for data logging in
Excel program; compatible
with Windows
systems
Power Input: 10 to 32 Vdc
switching power supply
(sold separately)
ENVIRONMENTAL
Operating Temperature:
-40 to 85°C (-40 to 185°F)
Storage Temp: -40 to 125°C
(-40 to 257°F)
Packaging Material:
Polycarbonate case with
DIN rail mount
®
operating
90.2
(3.54)
25.1
(0.99)
iDRN-PS-1000
U 24 Vdc Supply
U Switching Power Supply
Powers Up to 7 Units
Industrial Probes
shown smaller
than actual size.
iTHX-D3: 113 g (0.25 lb)
iTHP-5.
Dimensions: mm (in)
115 (4.53)
iTHP-2.
To Order Visit omega.com/ithx-d3 for Pricing and Details
Model No. Description
iTHX-D3 DIN rail case iServer MicroServer™ for temperature + humidity
iTHX-D3-2 DIN rail case iServer MicroServer™ for temperature + humidity
Accessories
iTHP-2 Replacement industrial probe 51 mm (2"), 0.9 m (3') cable
iTHP-5 Replacement industrial probe 137 mm (5"), 3 m (10') cable
iP-PCI-10P Porous polyethylene probe cap, for wet environments, 10 caps
iP-SC Porous stainless steel probe cap, 5 µm porosity, for dusty,
iDRN-PS-1000 Power supply (switching), 95 to 240 Vac input, 24 Vdc
CAL-3-HU
CT485B-CAL-KIT Calibration kit, 33 and 75% RH standards
* -CAL-3-HU
RAIL-35-2 DIN rail, 2 m (6.6')
Volume discounts available.
Ordering Examples: iTHX-D3, iServer MicroServer™ with temp and humidity and dew point,
127 mm (5") industrial probe.
iTHP-5-CAL-3-HU*, calibrated replacement probe, with NIST certificate.
and dew point, with industrial 137 mm (5") probe on 3 m (10')
cable with stripped wire leads
and dew point, with industrial 51 mm (2") probe on 0.9 m (3')
cable with stripped wire leads
per pack
pressurized (<35 psi) environments
output @ 850 mA (powers up to 7 units
NIST-traceable calibration certificate. 3 points of 25%, 50%, 75% RH,
temperature 25°C (77°F), for new units
Calibrated replacement probe and NIST calibration certificate
2
To Order, Call or Shop Online at omega.com