The Omega ISE-8882 Surfactant Electrode indicates the potentiometric endpoint when titratinganionic or
cationic
surfactants in solution. Titration procedures for manual titrations are discussed
in this manual, though adaptation to automatic titration techniques is quite simple. The electrode
0.05M
comes packaged with one 50 ml bottle of
O.OlM
one 50 ml bottle of
sodium lauryl sulfate (sodium dodecyl sulfate) titrant, and one 50 ml
Hyamine 1622 (benzethonium chloride) titrant,
bottle of sample additive, diluted Triton X- 100.
Required Equipment
1.
2.
3.
A
pH/mV meter, either line operated or portable.
A hand controlled delivery system, such as a 10 ml
ISE-8882 Surfactant Electrode (glass)
Required Solutions
1.
2.
Distilled or de-ionized water to prepare all solutions and standards.
Titrant for the titration of anionic surfactants is Hyamine 1622,
To prepare this titrant from your own laboratory stock, add 22.405 grams ofHyamine 1622 and 5 ml of 1 M de-ionized water.Swirl the flask to dissolve the solid and fill to the mark withdistilled water. Cap the flask and invert several times to mix the solution.
3.
Titrant for the titration of (SLS) To prepare this titrant from your own laboratory stock, add 2.883 gramssodium lauryl sulfate (SLS) to a one liter volumetric flask about half full of distilledwater.Swirl the flask to dissolve the solid and fill to the mark with distilled water.Cap the flask and invert several times to mix the solution.
pipet
or burette.
0.05M
NaOH to a 1 liter volumetric flask about half full of
O.OlM
cationic
surfactants is
Sodium Lauryl Sulfate
3.
Sample Additive, diluted Triton X- 100, to keep electrodes clean when added to all
Page 2
samples. To prepare, add 10 ml of reagent-grade Triton X-100 to a one liter volumetricflask about half full of distilled water.Cap the flask and invert several times to mix thesolution.
4.
Electrode Filling Solution, 4M KC1 (with
Ag’), for filling the reference chamber of
the electrode.
5.
6.
(O.OlM
(O.lMHCl) and polyacrylates
NaOH).
Electrode Rinse Solutions consisting of about 50 ml 0. for acidic rinse (anionic or
cationic
surfactant analysis) and 50 ml 0.
cationic
HCl diluted to 1000 ml
1M
1M
surfactantspH of both anionic and pH Adjuster Solutions for adjusting the
NaOH
diluted to 1000 ml for alkaline rinse (polyacrylate analysis).
GENERAL PREPARATION
Electrode Preparation
Remove the rubber cap(s) covering the electrode tip(s) and the rubber insert covering the fillinghole of the reference electrode. Fill the combination electrode or the reference electrode with the
filling solution shipped with the electrode to a level just below the fill hole. No preparation isrequired with a sealed reference electrode. Gently shake the electrode downward in the samemanner as a clinical thermometer to remove any air bubbles which might be trapped behind the
surfactant membrane.
O.OlM
Prepare
O.OOOlM
SLS by diluting 1 ml of the
first usage, or after long-term storage, soak the tip of the surfactant electrode in
SLS to 100 ml with distilled water. Prior to
O.OOOlM
SLS for
10 minutes before using the electrode each day. Use fresh solution daily. The electrode is now
ready for use.
Connect the electrode(s) to the proper terminal(s) as recommended by the meter manufacturer.
If the stock solution becomes cloudy or contaminated in any way, discard it.
Titrant Preparation
Based on the recommendations found in
Required Solutions,
select an appropriate titrant.
Determine the concentration of titrant needed for the analysis from Table 1.
Page 3
TABLE 1: Recommended Titrant Concentrations
Recommended Titrant
Concentration
(M)
Expected SampleConcentration
(M)
0.050.050 to 0.001
0.0050.001 to 0.0001
0.001
0.0001 to 0.00001
The titrant concentration may need to be adjusted depending on the concentration of the sample andthe method of titration in use. Use the formula:
~&_.Ys_
Vt
Ct
=
Ct
where:
= concentration of titrant
C,
= concentration of sample
V, = volume of sample
Vt
= volume of titrant
For example, for the titration of anionic surfactants, dilute the
0.05M Hyamine 1622 solution
provided to the appropriate concentration as calculated above.
Titrate against a known concentration of SLS to standardize the titrant.Calculate the exactconcentration of the titrant using the same formula given above.
Units of Measurement
Any convenient unit of measurement may be used for the result.Units may be chosen for samplesmeasured in volume or weight.
If doing the titrations with an automatic titrator, note whether specific units must be entered for thetitrant and/or specific units are required for the result.
Measuring Hints
The sensing membrane is normally subject to water uptake and might appear milky.This has no
effect on performance.
All samples and standard should be at ambient temperature for precise measurement.
3
Page 4
Constant, but not violent, stirring is necessary for accurate measurement.Slow stirring isrecommended to avoid foaming.
Always rinse the electrode tip(s) with the slightly acidic (or alkaline) rinse solution described in
Required Solutions
and blot dry with a fresh tissue between titrations to prevent solution
carryover.
Check the electrode for air bubbles adhering to the membrane surface after immersion in solution.Agitate the electrode gently to remove any air bubbles.
A slow or sluggish electrode response may indicate surface contamination of the electrodemembrane.Soak the electrode tip in distilled water for about 5 minutes to clean the membrane.Rinse the membrane and soak in
O.OOOlM
SLS for about 5 minutes to restore performance.
The electrode should be reconditioned daily before storage as described in
Cleaning,
Reconditioning, and Storage.
Sample Requirements
To help keep the electrode clean and working properly, add sample additive, diluted Triton X- 100,to all samples. For every 50 ml of sample, use 1 ml of sample additive.
Samples should be diluted to approximately
10m5
to
lOA
M to help preserve electrode life, help
avoid foaming during the titration, and help improve long term results.
pH
Adjust the
of the sample depending on the method being used.
Anionic surfactants, as well as sulfated and sulfonated surfactants, may be titrated with Hyamine
1622. Adjustment to
Polyacrylates should be adjusted to
Cationic
acidification to
surfactants should be titrated with an anionic reagent, such as sodium lauryl sulfate,
O.OlM
HCl.
pH 3 with
pH
1 O-l 1 with 0.
O.OlM
1M
NaOH before analysis.
HCl.
pH 2.5-4.5 should be done by addition of
after
Page 5
ANALYTICAL PROCEDURES
Sample Analysis
For potentiometric endpoint determination, the surfactant electrode is used as an endpoint indicator.
An example of the titration procedure is illustrated using the analysis of an anionic surfactant as an
example.
2.
3.
1.
Using the acid rinse solution, rinse the surfactant electrode and blot dry with a soft,
lint-free tissue before the titration. Fill the single junction reference electrode, or thereference chamber of the combination electrode with fresh filling solution to a leveljust below the fill hole.
Assure that the electrodes are plugged into the
mV mode. To prevent air entrapment, mount the electrode at a 20” angle from
the
pipet,
the vertical. Using a
O.OlM
Add 3 ml of
HCl
add 50 ml of the unknown sample to a 150 ml beaker.
and 1 ml of the sample additive, diluted Triton X-100.
pH/mV meter and that the meter is in
Place the beaker on a magnetic stirrer, and start stirring at a constant, but moderate,
rate.
Lower the electrodes into the solution so that the tips are completely covered
+l
and wait until the
to 2 drift is mV reading is stable,
mV/minute, before adding
any titrant. Remove any bubbles by re-dipping electrode.
Add
0.05M Hyamine 1622 titrant to a 10 ml buret until filled. Once t&rant
has been reached, add the
in 0.5-l .O ml increments at the beginning of the
titration, and in increments of 0.1-0.25 ml in the region of the endpoint.
mV stability
Theendpoint is at that volume of titrant where the potential changes dramatically withthe slightest addition of titrant.The electrode potential should be recorded aftereach addition of titrant. Continue titrating until 1 or 2 ml past the endpoint. Onstandard coordinate graph paper, plot milliliters of titrant added versus
mV reading.
The endpoint is the point of greatest inflection.Calculate the unknown surfactant
concentration:
Vtitrant
X
unknown
Of
theunknown
where:
C
unknown
=
&horn
Ctibant
VtibantVunknown
=COnCenkitiOn
=
concentration of the titrant
=
volume of the titrant in milliliters
= volume of the unknown in milliliters
Ctitmt
V
Depending on sample concentration and method used, this basic procedure may need modification
Page 6
ELECTRODE CHARACTERISTICS
Electrode Response
The time for the analysis may vary, depending on the sample, the titrant, the method, and theequipment used.The average time for manual titration of anionic surfactants is 2-5 minutes.
Temperature
The surfactant electrode should be used in the operating range of
0-4O’C.
The membrane may be
permanently destroyed at other temperatures.
Reproducibility
The reproducibility of the surfactant electrode will depend heavily on the good laboratory practicesof the technician, but will usually be less than 1% with manual techniques and less than 0.5% with
automatic techniques.
Limit of Detection
For anionic surfactants, the lower limit of detection is
-10W5M.
Good laboratory practice and
selection of titrant may allow lower levels of detection for some sample types.
pH Effects
The surfactant electrode has an operating
pH
range of 2-12.Use at other
pH values can adversely
affect the membrane.
pH
For anionic, sulfated and sulfonated surfactants, the analysis should take place at a
between 2.5
and 4.5.
For other samples, the
pH
range may need to be adjusted.Polyacrylates require adjustment to
pH
10, for example.
Interferences
Interferences may be caused by any organic anion or cation which chemically resembles the speciesof interest.
Page 7
Cleaning, Reconditioning, and Storage
Acidic (or alkaline) rinse solution should be used to rinse the electrode between measurements.
To recondition an electrode when the response had become noisy, sluggish, or
10m4M
in slightly acidic (or alkaline) distilled water for one hour, followed by
SLS solution for 10
&producible,
soak
minutes.
The ISE-8882 Surfactant Ion Selective Electrode may be stored in
O.OOOlM
SLS for short periodsof time.For storage over 3 weeks, rinse and dry the membrane element and cover the tip with anyprotective cap shipped with the electrode(s). The reference portion of the combination electrode (orthe reference chamber of the reference electrode) should be drained of filling solution, if refillable,
and the rubber sleeve placed over the filling hole.
Electrode Life
The surfactant electrode will last six months in normal laboratory use.
Continuous titrations on anautomatic sample changer might shorten operational lifetime to several months. In time, theresponse time will increase and the titration endpoint breaks will not be as sharp. At this point,titration is impossible and electrode replacement is required.
ELECTRODE THEORY
Electrode Operation
The surfactant electrode is an endpoint indicator for the potentiometric determination of anionic
surfactants in solution.
Cationic
surfactants may also be determined with this electrode.
The reaction that occurs when a sulfated or sulfonated anionic surfactant is titrated with Hyamine
1622 is as follows:
R
-
&N+Cl-
RS03NR4
6 S03Na+ +
+
NaCl
where: R = surfactant carbon chain
R4Nf
= Hyamine ion
Page 8
TROUBLESHOOTING GUIDE
The goal of troubleshooting is the isolation of the problem through checking each of the systemcomponents in turn: the instrumentation, the electrodes, the reagents, the sample, and the technique.
Instrumentation
For manual titration, assure that the
mV meter is operating correctly and that the glassware is clean.
Most meters are provided with an instrument check-out procedure in the instruction manual and a
shorting strap for ease of troubleshooting.
Consult the manual for complete instructions and verify
the instrument operates as indicated.
when rinsed with distilled or de-ionized water, the water does
Clean glassware will drain clean..
.
not bead on the inside walls of the glassware.
If using automatic titration instrumentation, check the instrument instruction manuals/operators’handbook for the correct check-out procedure or call the instrument manufacturer for the check-outprocedure.
Electrodes
1.
2.
3.
Using distilled or de-ionized water, rinse the electrodes thoroughly.
Titrate a known standard to check the electrode’s operation.
If the electrode fails to respond as expected, see the section
Measuring Hints.
Repeat Step 2.
4.
5.
6.
If the electrode still fails to respond as expected, substitute another surfactant
electrode that is known to be in good working order for the questionable electrode.If the problem persists, try the same routine with a working reference electrode.
If the problem persists the reagent may be of poor quality, interferences in thesample may be present or the technique may be faulty.See
Technique
sections below.
Reagents, Sample, and
If another electrode is not available for test purposes, or if the electrode in use issuspect, review the instruction manual and be sure to:
-
Clean and rinse the electrode(s) thoroughly.
-
Prepare the electrode(s) properly.
-
Use the proper filling solution, titrant, and sample additives.
pH
-
Adjust the
of the solution according to the method being used for the analysis.
Page 9
-
Measure correctly and accurately.
-
Review
TROUBLESHOOTING HINTS.
Reagents
Whenever problems arise with the measuring procedure that has been used successfully in the past,be sure to check the reagent solutions.prepare them again.
Errors may result from contamination of the titrant, incorrect dilution, poor
If in doubt about the credibility of any of the reagents,
quality distilled/ de-ionized water or additive, or a simple mathematical miscalculation.
Sample
Look for possible interferences,
complexing agents, or substances which could affect the response
or physically damage the sensing electrode or the reference electrode if the electrodes work
perfectly in the standard, but not in the sample.
Try to determine the composition of the samples prior to testing to eliminate a problem before itstarts. See
Sample Requirements,
Effects,
and
Interferences.
pH
Technique
Be sure that the electrodes’ limit of detection has not be exceeded.Be sure that the analysis method
is clearly understood and is compatible with the sample.
Refer to the instruction manual again. Reread
and
PROCEDURES,
ELECTRODE CHARACTERISTICS.
GENERAL PREPARATION, ANALYTICAL
If trouble still persists, call Omega Engineering Customer Service Department.
Page 10
Symptom
Possible Causes
Next Step
Out of Range
Reading
defectiveinstrumentation
electrode(s) notplugged inproperly
no referenceelectrode
reference electrodenot filled
air bubble on membrane
electrode(s) not insolution
check instrumentby using instrumentcheck-out procedure
unplug electrode(s)and reseat
use referenceelectrode described
Required Equipment
in
add fillingsolution to theouterchamber of thereference electrode