RH/Temp Humidity Transmitters
Wall, Duct Mount, and Remote Probe Models
With Optional Dual Built-In Displays
HX93AC
(without displays)
†
Optional
U 4 to 20 mA or
0 to 1 Vdc Output
for Both RH and
Temperature
U Splashproof Enclosure
U Accurate to 2.5% RH
and 0.6°C
U New Models with
Built-In Display to
Indicate Humidity and
Temperature
U Duct Mount Kit Included
with Duct Mount and
Remote Probe Models
Relative Humidity Control System
The HX93A and HX93DA Series interface
directly with OMEGA’s iSeries meters and
controllers to create a complete relative
humidity control system.
iSeries meters and controllers
feature totally programmable color
displays and embedded internet
connectivity. Visit omega.com
HX93AC wall
mount.
The HX93DA and HX93A Series
transmitters provide remote and
on-site monitoring of relative
humidity and temperature. Both
series output a linearized current
or voltage signal proportional
to the measured humidity or
temperature. RH outputs are
temperature compensated. Current
output models enable placing of
the transmitter at a remote location
virtually any distance away from
the measuring device. The HX93DA
models feature built-in dual displays
that indicate both humidity and
temperature simultaneously. All
use a thin-film polymer capacitor
to sense relative humidity and a
Platinum 100 Ω RTD to sense
temperature. A stainless steel
mesh type filter, easily removable
for cleaning, protects the sensors.
Mounting screws are easily
accessible inside the rugged,
polycarbonate enclosure which
houses and protects the electronics
to NEMA 13 specifications.
†
Refer to ordering chart on next page for
NIST calibration ordering information.
U Rugged Industrial
Splashproof Design
U “Humidity Only” Models
Also Available, see page
Hu-17 for HX92A Series
Hu-19
1
⁄32 DIN temperature/process
DPi32
meter, shown actual size.
HX93AC-D duct mount,
shown smaller than
actual size.
HX93A(*)-RPI remote
probe humidity
transmitter, shown
smaller than actual
size, supplied with a
1 m (3') cable attached
to a 124.5 mm (4.9")
stainless steel probe.
HX93DAC,
shown smaller
than actual size.
Displays
Humidity and
Temperature
Simultaneously!
Specifications
Input Voltage Range: 24 Vdc nominal
HX93A Series: 6 to 30 Vdc
HX93DA Series: 12 to 30 Vdc
Measuring Range
RH: 3 to 95%
Temperature: -20 to 75°C (-4 to 167°F)
Accuracy
RH: ±2.5% RH from 20 to 80%RH;
±3.1% RH below 20 and above 80%
RH@ 25°C (76°F) with temp coefficient
of ±0.1% RH/°F
Temperature: 0.6°C (±1°F)
Output
HX93AC or HX93DAC:
4 to 20 mA for 0 to 100% RH
and -20 to 75°C (-4 to 167°F)
HX93AV: 0 to 1 Vdc for 0 to 100% RH
and -20 to 75°C (-4 to 167°F)
RH Temperature Compensation:
-20 to 75°C (-4 to 167°F)
RH Time Constant (90% Response
at 25°C, in Moving Air @ 1m/s):
>10 seconds: 10 to 90% RH;
>15 seconds: 90 to 10% RH
Repeatability: ±1% RH, 0.5°F
Housing: Gray polycarbonate,
(IP65, DIN 40050) NEMA rated up
to 13, UL listed
Connections: Liquid-tight nylon with
neoprene gland, for 0.09 to 0.265" diameter
cable; internal 4-pin terminal block accepts
14 to 22 gauge wire
Dimensions:
HX93A(*) and HX93A(*)-D Enclosure:
79.8 x 81.8 x 54.9 mm
(3.14 x 3.22 x 2.16")
HX93A(*)-D Probe: 15.7 D x 124.5 mm L
(0.620 x 4.9")
Weight:
HX93A(*): 170 g (6 oz)
HX93A(*)-D: 187 g (6.6 oz)
Displays (HX93DA Series Only)
Dual Miniature Loop-Powered
Indicators: 3
Reading Rate: 2.5 times per second
Display Accuracy:
Relative Humidity: ±2.5% RH
Temperature: ±0.6°C (1°F)
Display Resolution:
Relative Humidity: 0.1% RH
Temperature: 0.1°C (0.1°F)
Display Accuracy in Relation
to Output:
Relative Humidity: ±0.3% RH
Temperature: ±0.3°C (0.5°F)
1
⁄2 digit LCD
To Order Visit omega.com/hx93a for Pricing and Details
Model No. Description
HX93A(*) Wall mount RH and temperature transmitter
HX93A(*)-D Duct mount RH and temperature transmitter
HX93A(*)-RP1 Remote probe relative humidity/temp transmitter
HX93DAC-(**) Wall mount, dual display RH/temperature transmitter/indicator
HX93DAC-D-(**) Duct mount, dual display RH/temperature transmitter/indicator
HX93DAC-RP1-(**) Remote probe, dual display RH/temperature transmitter/indicator
HX92-CAL Calibration kit, 11 and 75% RH standards
PSR-24S
PSR-24L
PSR-24L-230
PSU-93 Unregulated power supply, 16 to 23 Vdc, 300 mA max, screw terminal
TX4-100 4-conductor shielded transmitter cable 30 m (100')
CAL-3-HU† NIST-traceable calibration
HX90DM-KIT
For 4 to 20 mA output, specify “C” or for 0 to 1 Vdc output specify “V”.
*
For temperature display in °C insert “C” or for temperature display in °F insert “F”.
**
Ordering Example: HX93DAC-F, wall mount, dual display RH/temp transmitter indicator with °F display with 4 to 20 mA output,
with HX92-CAL calibration kit.
Regulated power supply, US plug, 90 to 264 Vac input, 24 Vdc output, 400 mA, screw terminals, UL
Regulated power supply, US plug, 90 to 264 Vac input, 24 Vdc output, 400 mA, stripped leads, UL
Regulated power supply, European plug, 230 Vac input, 24 Vdc output, 400 mA, stripped leads, CE
Spare duct mount kit [for HX93DAC-(**)-D, HX93DAC-(**)-RP1, HX93A(*)-D & HX93A(*)-RP1 models only]
H
u
Hu-20