Omega Products HHP8205 Installation Manual

Page 1
Page 2
Page 3
Congratulations on your purchase of the Manometer ! This instrument is portable, battery operated pressure measuring device. The Manometer is ideal for HVAC/R technicians measuring pressure level ,Medical equipment ,Computer peripherals , Pneumatic Controls.
INTRODUCTION
The meter will display all LCD s egments when it is f irst turned on for a pp r o x . 3 secon ds. Though you mi gh t have se en DATALOGGER,Y/M/D, REL, AVG ..these are not available for the meter.
The LCD is divided into two distinct sections : One large (Primary) top screen and one smaller right-bottom bottom screens (Relative Clock). The 2 display areas keep you constantly updated with the pressure measurements.
The Meter measures Gauge pressure-a measure of pressure in psi that is referenced to ambient pressure and Differential pressure - a measure of the difference of two pressures .
Meter has 11 selectable units of measure.: bar ,mmHg, ozin , kgcm , psi, inH2O, kPa, ftH2O,inHg ,mH2O,mbar.
2
2
Page 4
4 Measurement ranges are available: Pressure range :0~ 5 psi (HHP 8205) :0~ 15 psi ( 8215) :0~ 30 psi ( 8230) :0~ 100 psi ( 8210)
Please check the tubing is not leaking or damaged before using.
HHP HHP
HHP
CONVERSION & RESOLUTION
Quick Conversion Sheet
unit
model
HHP8205 HHP8215 HHP8230
HHP8210
Psi
5
15 30
100
inH2O
138 415
830
2768
mbar
345 0.35 1034 2068 6895
2
kg/cm
1.05
2.10
7.03
2
Page 5
CONTROLS AND INDICATORS
12
11
DATA LOGGER
1
8
9
7 6
14
2
5
13
15
HOLD REL DIF
4
3
REC AVG MAX MIN BAT
10
1. Primary Data Screen displays pressure value.
2. "-" . Minus pressure display.
3. MAX MIN presssure recorded.
4. REC starts recording mode and displays max./min. pressure recorded.
5. AVG. Average records (N/A).
6. DC power in Jack.
7. RS232 output port.
8. H/M/S 88:88:88 displays data for Hour / Minute / Second.
9. . Pressure unit indication.
10. BA T . Battery low indicator.
11. DIF. Differential pressure mode.
12. "+" Positive pressure connection. 13 . "-" Negative pressure connection.
14. HOLD. Freezes pressure reading.
3
Page 6
15. RE L. Establish a relative zero for the primary screen information.(N/A)
"-" Negative pressure
"+" Positive pressure
AUTO POWER OFF (SLEEP FUNCTION)
This instrument will shut off automatically in approx. 20 minutes for every power on. For recording or operating over longer
I
periods of time, you can disable the sleep mode by pressing
HOLD
and simultaneously before power on. An "n" will appear in the middle of the screen at which time you can release the button. (See Fig. A) The disable sleep
HOLD
mode will be invalid after power off.
Fig. A
MODE OPTIONS
Delete and replace with programmable user selectable startĀ­up mode.The display will default to the mode last used.
4
Page 7
HOLDHOLD
DIFDIF
RECREC
UNITUNIT
l l
++
==
==
==
DISABLE SLEEP MODE ZERO SETTING
SELECT UNIT / SKIP CAL.
HOLD
HOLD
Hold 2 sec.
The following table lists the modes of operation that can be invoked by pressing the button indicated.
I
Turns instrument on (Default set-ting) and off.
REC
Press momentarily and relative clock starts in the lower right screen. REC is displayed in the middle left of (Fig. B) other button functions are locked out except Power , Unit and Backlight. Press momentarily again and the unit cycles through MAX (Fig. C) and MIN (Fig.D) and back to current pressure ; the recordmode is displayed on the LCD. Press and hold for 3 seconds to turn off the record function
REC
and return to normal mode.
Fig. B
2
bar ozin psi kPa inHg mbar
mmHg kgcm inH2O ftH2O cmH2O
2
5
Page 8
Fig. C
2
bar ozin psi kPa inHg mbar
mmHg kgcm inH2O ftH2O cmH2O
2
Fig. D
2
bar ozin psi kPa inHg mbar
mmHg kgcm inH2O ftH2O cmH2O
HOLD
Press momentarily to freeze the pressure reading . (Fig.E)
2
Fig. E
2
bar ozin psi kPa inHg mbar
mmHg kgcm inH2O ftH2O cmH2O
DIFDIF
Press momentarily ,DIF appears on top of the LCD and the
2
display indicates the relative zero (Relative zero causes the value of the display to show as "0.0")-only the amount of pressure change will be indicated. Press momentarily again and the unit returns to the normal mode of pressure differential (see Fig.F).
Fig. F
DIF
2
bar ozin psi kPa inHg mbar
mmHg kgcm inH2O ftH2O cmH2O
2
6
Page 9
Differential Pressure : A measure of the difference between two pressures , i.e. use differential pressure sensor to measure gauge pressure by leaving one process connection open to atmosphere and connecting the second sensor port to your system.
UNIT
Press momentarily and the units will cycle through "bar" , "mmHg","ozin2","kgcm2" ,"psi","inH2O","kPa", "ftH2O" ," inHg" ,"cmH2O" ,"mbar" ,which are
indicated on the bottom of the display (See Fig.G &H).
Fig. G
2
bar ozin psi kPa inHg mbar
mmHg kgcm inH2O ftH2O cmH2O
2
Fig. H
2
bar ozin psi kPa inHg mbar
mmHg kgcm inH2O ftH2O cmH2O
2
Press momentarily and the back-light illuminates for approx. 30 seconds then turns off automatically. Or press momentarily to decrease the figure when calibration is
being performed.
7
Page 10
MAINTENANCE
The meter is calibrated in house before shipping.
When properly maintained , the meter will maintain its accuracy specification, to ensure your meter is performing at its peak, send it to the factory or a qualified instrument calibration facility for annual calibration.
Always set zero before measurement .Refer to the zero setting procedure in page 10.
Cleaning: Use a d am p cloth and mild soap to clean the case of the Manometer, do not use harsh detergents or abrasives as these may mar the finish or damage the unit's case with an adverse chemical reaction.
CALIBRATION MODE
Calibration mode is only applicable for a standard Manometer calibrator or any qualified meter calibration facility for annual calibration.
1. First , manually set the display to zero ( no pressure applied to the connector ), refer to the Manual zero procedure.
8
Page 11
2. Turn the meter off.
3. Press & simultaneously ,"CA" appears on the
RECREC
display, (See Fig.I )the meter enters to the calibration mode ,make sure the units setting on "PSI" to start positive (+) pressure calibration
Fig. I
psi
4. The meter has defaulted as 80 psi calibration point , the adjustable pressure range is from 78.0 to 82.0 if calibration pressure source is not 80psi ,increase by pressing key ,
RELREL
or decrease by pressing key to set calibration point as required.
5. Save the calibration point by pressing key , "SA" and
RECREC
small "CA"appears on the display (See Fig. J ) in 2 seconds , the meter auto-skips to the negative pressure (-) point for next calibration mode.
Fig. J
psi
9
Page 12
6. Follow the same procedure as step 4 for the negative pressure calibration point by pressing key , the LCD now displays
UNITUNIT
" -80.0 " and small"CA" (See Fig.K ) , do the necessary calibration figure refer to your pressure standard if needed.
Fig. K
psi
7. Again save the calibration point by pressing key, "SA"
RECREC
and "CA"appears in 2 seconds and then "End"and "CA" appears in another 2 seconds, the meter turns back to the normal mode (See Fig. L ).
Fig. L
psi
If you can't save by pressing key, i.e. no "SA" appeared ,
RECREC
please check: (a) The cabliration pressure source is between
75.0 and 85.0, or check (b) if you entered the right positive pressure (+) or negative pressure (-).
If you want to skip positive (+) calibration when in the Calibration mode , press to skip to negative (-) calibration point.
UNITUNIT
10
Page 13
Above calibration is an example for model HHP8210 , i.e. the pressure range is from 0 to +100psi (Positive pressure ) or from 0 to - 100 psi (Negat iv e pressure).
Calibration point reference
Model:
psi range
HHP8205 HHP8215 HHP8230 HHP8210
0~ 5 0~ 15 0~ 30 0~ 100
Calibration point
4.000
12.00
24.00
Reommend
( )( )
3.900~4.100
11.70 ~12.30
23.40~24.60
80.00 78.00~82.00
MANUAL ZERO SETTING
When you set the display to zero( no pressure applied to the connector), press button for 2 seconds , now the meter displays "0 .0.0.0" from right to left (See Fig.M) and then disappears , the LCD display shows a normal mode .(See Fig.N)
HOLDHOLD
Fig. M
Fig. N
2
bar ozin psi kPa inHg mbar
mmHg kgcm inH2O ftH2O cmH2O
2
11
Page 14
TROUBLESHOOTING
Power on but no display. Check the battery connections
?
Replace with new battery or attach optional AC adaptor.
BAT indication. Replace with a new battery when LCD
?
displays BAT at the middle bottom.
No Display. Make sure battery is not empty , if the display
?
disappear, check sleep mode is active . Refer to the Disable sleep mode function for a long time using the measurement. Or check the tubing is connected to the meter tightly.
Err.1. For the pressure value exceeding the maximum range ,
?
"Err.1" appears on the display ( See Fig. O ).Do not exceed
rated over pressure range of manometer. Sensor will bedamaged.
Fig.O
Err. 2. For the measurement pressure less than minimum
?
range ,"Err. 2 " will appear (See Fig.P ).
12
Page 15
Fig. P
Err.3.For a differential pressure value larger than maximum
?
display , Err .3 appears on the display (See Fig.Q ).
Fig. Q
Err.4. When you set zero ,make sure you have disconnected
?
the tubing. If you see an Err.4 appear on the display, it means the manometer is damaged (See Fig.R ).
P.S. Err.4 will be also appear if the tubing is connected during zero set.
Fig.R
E1OL or E2UL. When you see the errors while operating
?
Rs232 software, it means pressure source is less or over than the range of the instrument.
13
Page 16
REPLACING THE BATTERY
Replace your 9-volt battery when:
The BAT icon appears on the right of the screen. The meter will not power on. Use of the back-light causes the BAT icon to appear.
Even if the battery was recently replaced, check its voltage level if you get no response from your instrument.To replace the battery:
1. Remove the tubing of the instrument.
2. Lay the instrument face-down on a clean, flat surface.
3. Remove battery cover.
Remove battery from instruments that you do not plan to use for a month or more. Do not leave battery in instrument.
OPERATING CONDITIONS
Compensated temperature range:0~50 C.
Operating temperature 0 ~50 C (32~122 F)
oo o
Storage temperature range: -20~60 C
14
o
o
Page 17
Operating Humidity Max. 80% RH Power : One 9.0 volt battery
Exceeding Maximum pressure will cause permanent sensor damage.
Model Max.Pressure HHP 8205 HHP 8215 HHP 8230 HHP 8210
20 psi 30 psi 60 psi
150 psi
SPECIFICA TION
Pressure
Range Resolution
Accuracy Dimension
Unit Weight Response
time Format
0~ (5 or 15 or 30 or 100 psi) See Page 8 data sheet
0.3% of full scale at 25 C 72 x 182 x 30 mm (meter)
Approx. 220 gram (with battery)
0.5 seconds Baud Rate : 2400 bit/sec
Data Bit : 8 , Stop Bit : 1 P XXXXX , P - XXXXX (unit)
o
15
Page 18
MATERIAL SUPPLIED
This package contains:
The meter Battery (9.0 volt) Operation manual Hard carrying case Connection hose
OPTIONAL ACCESSORY:
RS232 software disk or CD with D-sub connector DC Adaptor
RS232 OUTPUT:
The meter can link with a personal computer to capture on-line data ,display pressure records with real- time output, you can retrieve files , save the data for operating data analysis, record
statistics ,multi-files display in the screen,....versatile functions
for your choice.
Connection procedures:
1. P lug the optional accessory Rs232 cable onto the DC jack port ( at the right side of the meter)
2. Insert the D-sub 9P type connector onto computer's Com.1 or 2 port or..
3. S t art to set up RS232 software by inserting the CD-ROM or Floppy diskette.
16
Page 19
4. When installing the RS232 software,please follow the operation manual procedure in the software package.
17
Page 20
Page 21
Loading...