Magnetic fields surrounding permanent magnets or electrical
conductors can be visualized as a collection of magnetic flux
lines; lines of force existing in the material that is being subjected
to a magnetizing influence. Unlike light, which travels away from
its source indefinitely, magnetic flux lines must eventually return
to the source. Thus all magnetic sources are said to have two
poles. Flux lines are said to emanate from the “north” pole and
return to the “south” pole, as depicted in Figure 1-1.
Figure 1-1
Flux Lines of a Permanent Magnet
One line of flux in the CGS measurement system is called a
maxwell (Mx), but the weber (Wb), which is 108 lines, is more
commonly used.
Flux density, also called magnetic induction, is the number of flux
lines passing through a given area. It is commonly assigned the
symbol “B” in scientific documents. In the CGS system a gauss
(G) is one line of flux passing through a 1 cm2 area. The more
1-1
INTRODUCTION
commonly used term is the tesla (T), which is 10,000 lines per
cm2 . Thus
1 tesla = 10,000 gauss
1 gauss = 0.0001 tesla
Magnetic field strength is a measure of force produced by an
electric current or a permanent magnet. It is the ability to induce
a magnetic field “B”. It is commonly assigned the symbol “H” in
scientific documents. The unit of “H” in the CGS system is an
oersted (Oe), but the ampere/meter (A/m) is more commonly
used. The relationship is
1 oersted = 79.6 ampere/meter
1 ampere/meter = 0.01256 oersted
It is important to know that magnetic field strength and magnetic
flux density are not the same. The only time the two are
considered equal is in free space (air). Only in free space is the
following relationship true:
1 G = 1 Oe = 0.0001 T = 79.6 A/m
MEASUREMENT OF FLUX DENSITY
A device commonly used to measure flux density is the Hall
generator. A Hall generator is a thin slice of a semiconductor
material to which four leads are attached at the midpoint of each
edge, as shown in Figure 1-2.
1-2
INTRODUCTION
Figure 1-2
Hall Generator
A constant current (Ic) is forced through the material. In a zero
magnetic field there is no voltage difference between the other
two edges. When flux lines pass through the material the path of
the current bends closer to one edge, creating a voltage
difference known as the Hall voltage (Vh). In an ideal Hall
generator there is a linear relationship between the number of
flux lines passing through the material (flux density) and the Hall
voltage.
The Hall voltage is also a function of the direction in which the
flux lines pass through the material, producing a positive voltage
in one direction and a negative voltage in the other. If the same
number of flux lines pass through the material in either direction,
the net result is zero volts. This sensitivity to flux direction makes
it possible to measure both static (dc) and alternating (ac)
magnetic fields.
The Hall voltage is also a function of the angle at which the flux
lines pass through the material. The greatest Hall voltage occurs
when the flux lines pass perpendicularly through the material.
Otherwise the output is related to the cosine of the difference
between 90° and the actual angle.
1-3
INTRODUCTION
The sensitive area of the Hall generator is generally defined as
the largest circular area within the actual slice of the material.
This active area can range in size from 0.2 mm (0.008”) to 19
mm (0.75”) in diameter. Often the Hall generator assembly is too
fragile to use by itself so it is often mounted in a protective tube
and terminated with a flexible cable and a connector. This
assembly, known as a Hall probe, is generally provided in two
configurations:
TRANSVERSE PROBE
AXIAL PROBE
Figure 1-3
Hall Probe Configurations
In “transverse” probes the Hall generator is mounted in a thin, flat
stem whereas in “axial” probes the Hall generator is mounted in a
cylindrical stem. The axis of sensitivity is the primary difference,
as shown by “B” in Figure 1-3. Generally transverse probes are
used to make measurements between two magnetic poles such
as those in audio speakers, electric motors and imaging
machines. Axial probes are often used to measure the magnetic
field along the axis of a coil, solenoid or traveling wave tube.
Either probe can be used where there are few physical space
limitations, such as in geomagnetic or electromagnetic
interference surveys.
1-4
INTRODUCTION
Handle the Hall probe with care. Do not bend the stem or
apply pressure to the probe tip as damage may result. Use
the protective cover when the probe is not in use.
PRODUCT DESCRIPTION
The MODEL HHG-23 GAUSS / TESLAMETER is a portable
instrument that utilizes a Hall probe to measure magnetic flux
density in terms of gauss, tesla or ampere/meter. The
measurement range is from 0.01 mT (0.1 G or 0.01 kA/m) to
2.999T (29.99 kG or 2387 kA/m). The instrument is capable of
measuring static (dc) magnetic fields and alternating (ac) fields.
The MODEL HHG-23 consists of a palm-sized meter and various
detachable Hall probes. The meter operates on standard 9 volt
alkaline batteries or can be operated with an external ac-to-dc
power supply. A retractable bail allows the meter to stand upright
on a flat surface. A notch in the bail allows the meter to be wall
mounted when bench space is at a premium. The large display is
visible at considerable distances. The instrument is easily
configured using a single rotary selector and two pushbuttons.
Three measurement ranges can be selected or the meter can
automatically select the best range based on the present flux
density being measured. A “zero” function allows the user to
remove undesirable readings from nearby magnetic fields
(including earth’s) or false readings caused by initial electrical
offsets in the probe and meter. Included is a “zero flux chamber”
which allows the probe to be shielded from external magnetic
fields during this operation. Another feature called “relative
mode” allows large flux readings to be suppressed so that small
variations within the larger field can be observed directly. Both
the “zero” and “relative” adjustments can be made manually or
automatically.
1-5
INTRODUCTION
Other features include three “hold” modes, allowing either the
arithmetic maximum, minimum or true peak values to be held
indefinitely until reset by the user. An analog signal is available
from a standard BNC connector that is representative of the
magnetic flux density signal and is calibrated to ± 3 volts full scale
in dc mode or 3 Vrms in ac mode. This output can be connected
to a voltmeter, oscilloscope, recorder or external analog-to-digital
converter.
The meter can be fully configured and flux density readings
acquired from a remote computer or PLC using the RS-232
communications port. This is a standard 9-pin “D” connector
commonly used in personal computers. The commands follow
widely accepted protocols established by the IEEE-488.2 and
SCPI-1991 standards.
The meter, probes and accessories are protected when not in
use by a sturdy carrying case.
APPLICATIONS
• Sorting or performing incoming inspection on permanent
magnets, particularly multi-pole magnets.
• Testing audio speaker magnet assemblies, electric motor
armatures and stators, transformer lamination stacks,
cut toroidal cores, coils and solenoids.
• Determining the location of stray fields around medical
diagnostic equipment.
• Determining sources of electromagnetic interference.
• Locating flaws in welded joints.
• Inspection of ferrous materials.
• 3-dimensional field mapping.
• Inspection of magnetic recording heads.
1-6
Section 2
Specifications
INSTRUMENT
RANGE RESOLUTION
gauss tesla A/m gauss tesla A/m
300 G
3 kG 300 mT 238 kA/m
30 kG
ACCURACY (reading on display and from RS-232 port,
including probe)
AUXILIARY POWER: 6 to 12 Vdc, 300 mA minimum.
AUXILIARY POWER CONNECTOR: Standard 2.5 mm I.D. / 5.5 mm
O.D. connector. Center post is
positive (+) polarity.
ANALOG OUTPUT CONNECTOR: BNC
OPERATING TEMPERATURE: 0 to +50ºC (+32 to +122ºF)
2-2
SPECIFICATIONS
STORAGE TEMPERATURE: -25 to +70ºC (-13 to +158ºF)
METER DIMENSIONS:
Length: 13.2 cm (5.2 in)
Width:13.5 cm (5.3 in)
Height:3.8 cm (1.5 in)
WEIGHT: Meter w/batteries: 400 g (14 oz.)
Shipping: 1.59 kg (3 lb., 8 oz.)
REGULATORY INFORMATION:
Compliance was demonstrated to the following specifications as
listed in the official Journal of the European Communities:
EN 50082-1:1992 Generic Immunity
IEC 801-2:1991 Electrostatic Discharge
Second Edition Immunity
IEC 1000-4-2:1995
ENV 50140:1993 Radiated Electromagnetic
IEC 1000-4-3:1995 Field Immunity
EN 50081-1:1992 Generic Emissions
EN 55011:1991 Radiated and Conducted
Emissions
2-3
SPECIFICATIONS
COMMUNICATIONS PORT: Format: RS-232C
Lines supported: Transmit, receive, common.
Connector type: 9-pin “D” female
Cable length: 3 m (9.8 ft.) maximum
Receive input resistance: 3 kΩ minimum
Receive voltage limit: ± 30 V maximum
Transmit output voltage: ± 5 V min, ± 8 V typical
Baud rate: 2400
Stop bits: 1
Character length: 8
Parity: None
Standards supported: IEEE-1987.2, SCPI-1991
EMC APPLICATION NOTE
Use only high quality, double shielded cables for RS-232
connection. Keep the length of the cables less than
3 meters (9.8 ft.). Long cables (>3m) with insufficient EMI
shielding can cause excessive emissions or may be
susceptible to external interference.
2-4
SPECIFICATIONS
STANDARD TRANSVERSE PROBE
MODEL NUMBER: STD58-0404
FLUX DENSITY RANGE: 0 to ± 3 T (0 to ± 30 kG)
FREQUENCY BANDWIDTH: 0 - 20 kHz
OFFSET CHANGE WITH
TEMPERATURE: - 0.05% / ºC typical
OPERATING TEMPERATURE RANGE: 0 to +75 ºC (+32 to +167ºF)
STORAGE TEMPERATURE RANGE: -25 to +75 ºC (-13 to +167ºF)
2-6
Figure 2-2
Standard Axial Probe
SPECIFICATIONS
OPTIONAL PROBE EXTENSION CABLE
MODEL NUMBER: X5000-0006
OPERATING TEMPERATURE RANGE: 0 to +75 ºC (+32 to +167ºF)
STORAGE TEMPERATURE RANGE: -25 to +75 ºC (-13 to +167ºF)
Figure 2-3
Optional Probe Extension Cable
2-7
SPECIFICATIONS
ZERO FLUX CHAMBER
MODEL NUMBER: YA-111
CAVITY DIMENSIONS:
Length: 50.8 mm (2”)
Diameter: 8.7 mm (0.343”)
ATTENUATION: 80 dB to 30 mT (300 G)
PURPOSE: To shield the probe from
external magnetic fields during
the ZERO or RELATIVE
operations.
2-8
Figure 2-4
Zero Flux Chamber
Section 3
Operating Instructions
OPERATOR SAFETY
Do not connect the auxiliary power connector to an ac power
source. Do not exceed 15 Vdc. Do not reverse polarity. Use
only an ac-to-dc power supply certified for country of use.
Figure 3-1
Auxiliary Power Connector Warnings
3-1
OPERATING INSTRUCTIONS
Do not allow the probe to come in contact with any voltage
source greater than 30 Vrms or 60 Vdc.
Figure 3-2
Probe Electrical Warning
This instrument may contain ferrous components which will
exhibit attraction to a magnetic field. Care should be utilized
when operating the instrument near large magnetic fields, as
pull-in may occur. Extension cables are available to increase
the probe cable length, so that the instrument can remain in a
safe position with respect to the field being measured with
the probe.
3-2
OPERATING INSTRUCTIONS
OPERATING FEATURES
Figure 3-3
Operating Features
1 Display. Liquid crystal display (LCD).
2 Manual ZERO / RELATIVE Control. In the ZERO and
RELATIVE modes of operation the user can manually
adjust the zero or relative point using this control.
3 Function Selector. This control allows the operator to
change the meter’s range, units of measure, ac or dc
measurement, hold modes and operation of the analog
output. It also engages the ZERO, RELATIVE and
MEASURE modes of operation.
5 Battery Compartment Cover. This cover slides open to
allow one or two 9 volt batteries to be installed.
3-3
OPERATING INSTRUCTIONS
6 Power Switch. Push-on / push-off type switch to apply
power to the meter.
7 SELECT Switch. Momentary pushbutton used in
conjunction with the Function Selector 3 to configure
the meter’s range, units of measure, ac or dc
measurement, hold modes and operation of the analog
output.
8 AUTO/HOLD RESET Switch. Momentary pushbutton
used to reset the held reading when one of the HOLD
modes is being used, or to start an automatic ZERO or
RELATIVE operation when in the ZERO or RELATIVE
modes.
9 Auxiliary Power Connector. This is an industry standard
2.5 mm I.D. / 5.5 mm O.D. dc power connector. The
meter will accept a dc voltage in the range of 6 - 15 Vdc at
300 mA minimum current. The center pin is positive (+).
The internal batteries are disconnected when using this
connector.
Do not connect the auxiliary power connector to an ac power
source. Do not exceed 15 Vdc. Do not reverse polarity. Use
only an ac-to-dc power supply certified for country of use.
3-4
OPERATING INSTRUCTIONS
10 Analog Output Connector. A voltage signal
representative of the magnetic flux density being
measured is available at this BNC connector. Calibration
is set to ± 3.0 V full scale dc or 3.0 Vrms ac, depending
upon the mode of operation . Minimum load is 10 kΩ. The analog output will be active when the ANALOG
ON/OFF function has been turned ON using the Function
Selector 3 .
11 Probe Connector. The Hall probe or probe extension
cable plugs into this connector and locks in place. To
disconnect, pull on the body of the plug, not the cable !
12 Meter Stand. Retractable stand that allows the meter to
stand upright when placed on a flat surface. A notch in
the stand allows the meter to be mounted to a vertical
surface.
3-5
OPERATING INSTRUCTIONS
INSTRUMENT PREPARATION
1) With the power switch turned off (POWER pushbutton in the
full up position) apply pressure to the battery compartment cover
at the two points shown in Figure 3-4. Slide the cover open and
remove.
2) Install one or two 9 volt alkaline batteries (two batteries will
provide longer operating life). The battery compartment is
designed so that the battery polarity cannot be reversed.
Reinstall the battery compartment cover.
3-6
Figure 3-4
Battery Installation
OPERATING INSTRUCTIONS
3) If using an ac-to-dc power supply review Figure 3-1 for safety
notes and the SPECIFICATIONS section for voltage and current
ratings. When using a power supply the batteries are
automatically disconnected.
4) Install the probe or probe extension cable by matching the key
way in the connector to that in the mating socket in the meter.
The connector will lock in place. To disconnect, pull on the body
of the plug, not the cable!
Figure 3-5
Probe Connection
3-7
OPERATING INSTRUCTIONS
POWER-UP
Depress the POWER switch. There will be a momentary audible
beep and all display segments will appear on the display.
Figure 3-6
Power-Up Display
The instrument will conduct a self test before measurements
begin. If a problem is detected the phrase “Err” will appear on the
display followed by a 3-digit code. The circuitry that failed will be
retested and the error code will appear after each failure. This
process will continue indefinitely or until the circuitry passes the
test. A condition in which a circuit fails and then passes should
not be ignored because it indicates an intermittent problem that
should be corrected.
If the self test is successful the meter will perform a self
calibration. During this phase the meter will display the software
revision number, such as “r 1.0”. Calibration will halt if there is no
Hall probe connected. Until the probe is connected the phrase
“Err” will appear accompanied by a flashing “PROBE” annunciator
as shown in Figure 3-7.
3-8
OPERATING INSTRUCTIONS
Figure 3-7
Missing Probe Indication
After power-up the position of the FUNCTION selector switch will
determine what happens next. For instance if the selector is in
the RANGE position the meter will wait for the user to change the
present range. If in the MEASURE position flux density
measurements will begin.
Allow adequate time for the meter and probe to reach a stable
temperature. See the SPECIFICATIONS section for specific
information.
POWER-UP SETTINGS
The meter permanently saves certain aspects of the instrument’s
setup and restores them the next time the meter is turned on.
The conditions that are saved are:
RANGE setting (including AUTO range)
MODE (ac or dc)
UNITS of measure (gauss, tesla or ampere/meter)
HOLD mode (min, max or peak)
3-9
OPERATING INSTRUCTIONS
Other aspects are not saved and default to these conditions:
RELATIVE mode (turned OFF)
RELATIVE value (set to 0)
ZERO mode (inactive)
OUTPUT function (turned OFF and output set to 0 Vdc)
NOTE: The present setup of the instrument is saved only when
the FUNCTION selector is returned to the MEASURE position.
For example assume the meter is in the MEASURE mode on the
30 mT range. The FUNCTION selector is now turned to the
RANGE position and the 300 mT range is selected. The meter is
turned off and on again. The meter will be restored to the 30 mT
range because the FUNCTION selector was never returned to the
MEASURE mode prior to turning it off.
LOW BATTERY CONDITION
The meter is designed to use one or two standard 9V alkaline
batteries (two batteries will provide longer operating life). When
the battery voltage becomes too low the battery symbol on the
display will flash, as shown in Figure 3-8. Replace the batteries
or use an external ac-to-dc power supply.
Instrument specifications are not guaranteed when a low
battery condition exists !
3-10
Loading...
+ 69 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.