Digital Anemometer HHF5000
Operating Instructions
Pressing this key once switches the instrument from OFF to ON or vice
versa. When switched on the HHF5000 will display for one second the
type of probe which is connected, i.e. 275 for the 2.75" (70mm) type or
100 for the 1" (25mm) size. If no probe is connected the instrument defaults to 275.
The values stored in the MAX and MIN memories are cleared each time the instrument
is switched off.
The MPS key converts the airspeed reading shown on the display to
metres per second, the FPM key converts the reading to feet per
minute. The measurement unit selected is shown by either MPS or
FPM on the right hand side of the display.
The 2S and 16S keys are used to select the period of time over
which the airspeed is measured, the default time period is two
seconds. Pressing the 2S key at any time starts a new two second
measurement cycle. The display will read 2S during that time, the
reading will then be displayed and subsequently updated every two seconds. The 16S
key sets the HHF5000 to measure the average airspeed over a sixteen second period.
A new sixteen second measurement is started each time the key is pressed. During
that time interval the display will show 16S and the MPS, FPM, MIN and MAX keys will
be disabled. At the end of the sixteen second period the averaged display reading will
be updated every two seconds. The 16S mode may be cancelled by pressing the 2S
key.
The MIN and MAX keys recall the lowest and highest airspeeds
measured by the HHF5000. Separate maximum and minimum values
are stored for the 2 second and 16 second measurement periods. In
this mode the display will alternate between the name of the
memory location (e.g. L2 for the minimum value measured using the 2 second
measurement period) and the numerical value stored in that location.
The airspeed continues to be measured while maximum or minimum values are being
displayed. Should a new maximum or minimum value be recorded the display will be
updated automatically. To cancel the display of the minimum or maximum values and
return the instrument to standard measurement mode the MIN or MAX key should be
pressed again. The maximum and minimum values are cleared by switching off the
HHF5000.
HHF5000 Specifications
RANGE: 2.75" (70mm) probe: 0.3 to 40 ms-1 (59 to 7875 ft min-1)
1.00" (25mm) probe: 0.3 to 35 ms-1 (59 to 6891 ft min-1)
ACCURACY: 2.75" (70mm) probe: ± 0.25% F.S plus ± 0.75% of reading
(SYSTEM) 1.00" (25mm) probe: ± 0.5% F.S plus ± 1.0% of reading
DISPLAY: 4-digit liquid crystal, 10mm character height
RESOLUTION: 0.01 ms-1 or 1 ft min-1
OPERATING Indicator: 0 to +50°C (32 to 122°F)
TEMPERATURE: Probes: -30 to +80°C (-22 to 176°F)
POWER SOURCE: 2 x 1 .5 volt alkaline LR6 (AA) cells
BATTERY LIFE: 250 hours approximately. Low battery warning after approximately
225 hours.
DIMENSIONS: 181 x 77 x 21.5 mm (indicator)
WEIGHT: 190g approx. with battery (indicator alone)
OTHER Automatic probe 'type identification
FEATURES: Automatic probe characteristic linearization
Switchable 2-second or 16-second measurement period
Maximum and minimum readings stored in memory