Omega Products HH806AWE User Manual

HH806AW HH806AWE
WIRELESS MULTILOGGER THERMOMETER
INSTRUCTION
SHEET
M4553/0611
Shop online at: omega.com e-mail: info@omega.com For latest product manuals: omegamanual.info
MADE IN TAIWAN
OMEGAnetOn-Line Service
omega.com
Servicing North America:
U.S.A: One Omega Drive, Box 4047
ISO 9001 Tel: (203) 359-1660 FAX: (203) 359-7700 Certified
Canada: 976 Bergar
U.S.A Sales Service: 1-800-826-6342/1-800-TC-OMEGA and Customer Service: 1-800-622-2378/1-800-622-BEST Canada:
Mexico: En Espan ol: (001) 203-359-7803
Benelux: Postbus 8034, 1180 LA Amstelveen, The Netherlands
Czech Frystatska 184, 733 01 Karviná, Czech Republic Republic: Tel: +420 (0)59 6311899 FAX: +420 (0)59 6311114
France: 11, rue Jacques Cartier, 78280 Guyancourt, France
Germany/ Daimlerstrasse 26, D-75392 Deckenpfronn, Germany Austria: Tel: + 49 (0)7056 9398-0 FAX: +49 (0)7056 9398-29
United One Omega Drive, River Bend Technology Centre Kingdom: Northbank, Irlam, Manchester ISO 9002 M44 5BD United Kingdom Certified
It is the policy of OMEGA Engineering, Inc. to comply with all worldwide safety and EMC/EMI regulations that apply. OMEGA is constantly pursuing certification of its products to the European New Approach Directives. OMEGA will add the CE mark to every appropriate device upon certification. The information contained in this document is believed to be correct, but OMEGA accepts no liability for any errors it contains, and reserves the right to alter specifications without notice.
should not be used for, human applications.
Stamford, CT 06907-0047
e-mail: info@omega.com
Laval (Quebec) H7L 5A1, Canada Tel: (514) 856-6928 FAX: (514) 856-6886 e-mail: info@omega.ca
For immediate technical or application assistance:
e-mail: espanol@omega.com FAX: (001) 203-359-7807 info@omega.com.mx
Servicing Europe:
Tel: +31 (0)20 3472121 FAX: +31 (0)20 6434643 Toll Free in Benelux: 0800 0993344 e-mail: sales@omegaeng.nl
Toll Free: 0800-1-66342 e-mail: info@omegashop.cz
Tel: +33 (0)1 61 37 2900 FAX: +33 (0)1 30 57 5427 Toll Free in France: 0800 466 342 e-mail: sales@omega.fr
Toll Free in Germany: 0800 639 7678 e-mail: info@omega.de
Tel: +44 (0)161 777 6611 FAX: +44 (0)161 777 6622 Toll Free in United Kingdom: 0800-488-488 e-mail: sales@omega.co.uk
These products are not designed for use in, and
Internet e-mail
info@omega.com
®
®
INTRODUCTION
The instrument is a portable digital thermometer that measures external thermocouples of type K, J, R, S, T, E, N. The thermocouples types comply with the N.I.S.T. – ITS 90 standard reference temperature/voltage tables. The thermometer features a dual thermocouple input, an adjustable T/C offset and an USB interface for uploading data to a PC using optional software and cable.
SAFETY INFORMATION
To avoid electrical shock, do not use this instrument
when working voltages at the measurement surface
To avoid damage or burns, do not make temperature
measurement in microwave ovens.
Repeated sharp flexing can break the thermocouple
leads. To prolong lead life, avoid sharp bends in the
leads, especially near the connector.
FEDERAL COMMUNICATIONS
This device complies with Part 15 of the FCC Rules. Operation is subject to the following two condi­tions:(1) this device may not cause harmful interfer­ence, and (2) this device must accept any interference received, including interference that may cause unde­sired operation.
This equipment has been tested and found to comply with the limits for a Class B digital device, pursuant to Par
t 15 of the FCC Rules. These limits are designed to provide reasonable protection. This equipment gener­ates, uses and can radiated radio frequency energy and, if not installed and used in accordance with the in­structions, may cause harmful interference to radio communications. However, there is no guarantee that interference will not occur in a particular installation If this equipment does cause harmful interference dio or television reception, which can be determined by turning the equipment off and on, the user is en­couraged to try to correct the interference by one or more of the following measures:
- Reorient or relocate the receiving antenna.
- Increase the separation between the equipment and
receiver.
- Connect the equipment into an outlet on a circuit dif-
ferent from that to which the receiver is connecte
- Consult the dealer or an experienced radio/TV tech-
nician for help. Shielded interface cables must be used in order to comply with emission limits.
over 24V AC or DC.
CAUTION
COMMISSION
NOTE
to ra-
d.
Changes or modifications not expressly approved by the party responsible for compliance could void the user‘s authority to operate the equipment. FCC (Domestic Use: USA) FCC ID: DONGLE 80TR FCC Identifer: VEA80TR
WIRELESS NOTE
Wireless receiver must keep a distance at least 40cm
from the meter and meter to meter must keep a dis-
tance at least 30cm.
SPECIFICATIONS
ELECTRICAL
Temperature Scale: Measurement Range:
K-TYPE (0.1°) -200°C to 1372°C (-328°F to 2501°F) J-TYPE (0.1°) -210°C to 1200°C (-346°F to 2192°F) T-TYPE (0.1°) -200°C to 400°C (-328°F to 752°F) E-TYPE (0.1°) -210°C to 1000°C (-346°F to 1832°F) R-TYPE (1°) 0°C to 1767°C (32°F to 3212°F) S-TYPE (1°) 0°C to 1767°C ( 32°F to 3212°F) N-TYPE (0.1°) -50°C to 1300°C ( -58°
*Based on the ITS-90 temperature standard.
According to temperature standard ITS-90.
Accuracy:
K/J/T/E-TYPE
±(0.05% rdg + 0.3°C) on -50°C to 1372°C ±(0.05% rdg + 0.7°C) on -50°C to -210°C ±(0.05% rdg + 0.6°F) on -58°F to 2501°F ±(0.05% rdg + 1.4°F) on -58°F to -346°F
N-TYPE
±(0.05% rdg + 0.8°C) on -50°C to 0°C ±(0.05% rdg + 0.4°C) on 0°C to 1300°C ±(0.05% rdg + 1.6°F) on -58°F to 32°F ±(0.05% rdg + 0.8°F)
R/S-TYPE
±(0.05% rdg + 2°C) on 0°C to 1767°C ±(0.05% rdg + 4°F) on 32°F to 3212°F
Temperature Coefficient:
0.1 times the applicable accuracy specification per °C from 0°C to 18°C and 28°C to 50°C (32°F to 64°F and 82°F to122°F).
Input Protection:
24V dc or 24V ac rms maximum input voltage on any combination of input pins.
Reading Rate: 2.5 time per second.
Wireless Features:
Frequency range: 910~920MHz (HH806AW)
Low current consumption less than 1mA. The transmitting distance can reach 25M without mag­netic interference.
External Connections:
1. USB Port
2. DC power JACK(12V)
Celsius or Fahrenheit user-selectable
on 32°F to 2372°F
868.1~868.5MHz (HH806AWE)
1
F to 2372°F)
2
ENVIRONMENTAL
Ambient Operating Ranges:
0°C to 50°C (32°F to 122°F) <80% R.H.
Storage Temperature:
-20°C to 60°C (-4°F to 140°F) <70% R.H.
GENERAL
Display:
There are three display areas on the HH806 series LCD (liquid crystal display). The Main and Second displays are 4 ½ digits with a maximum reading of 19999. These are used for displaying the value of T1 or T2. The Third display is for the date, time, or the differential value of T1 to T2.
Overload: “----.-” or “OL” is display. Battery: 1.5V x 4 PCS (SIZE AAA) UM-4 R03. Battery Life: 120 hours typical with carbon zinc battery. Dimensions: 160mm(H) x 83mm(W) x 38mm(D). Weight: Approx. 365g including batteries. Supplied Thermocouple:
1 meter (40”) type K insulated beaded wire thermocouple. Maximum insulation temperature is 482°C (900°F). Thermocouple accuracy is ±1.1°C or 0.4% of reading (whichever is greater) from 0°C to 1250°C.
OPERATING INSTRUCTIONS
1. OPERATIONAL MODE
There are three operation modes-Normal, Shift, and Setup Mode. NORMAL MODE:
This is the default mode, the operating functions for the normal mode are printed on the top of each button in white.
SHIFT MODE:
The operating functions for the shift mode are printed in gray on the buttons. While in the normal mode, press the SHIFT button to switch to shift mode. At the lower-right corner of the display, the word “Shift” is displayed to indicate shift mode. To switch back to normal mode, press the SHIFT button again.
SETUP MODE:
Press the set[ ] button in normal mode to switch to setup mode, the indicator “SET” is shown on the left side of the display. To switch back to normal mode, press SET[ ] button.
WIRELESS MODE:
Press the “ ” key for more than two seconds to start wireless function. Press the “ ” key again for another two seconds to stop wireless function. The wireless mode will shut down if there is no wireless signal for two minutes. To SET CH/ID to 00,00, by pressing “T1-T2” key and “ ” power key for more than 6 seconds with the me­ter powered down. The meter will set channel and ID to 00,00 status. The second display will show 00, which means that the channel and ID has been set to
00.
To check the channel and ID of the meter:
When the meter is off press “ ” key and “ ” for 5 seconds, LCD’s main display will show channel num­ber, the second display will show ID number.
2. NORMAL MODE
The following functions can only be used in the normal mode.
(1) “ ” Power Button
The “ ” button turns the thermometer on or off. When the meter is in MAX/MIN record mode, the power off function is disabled.
(2) “[LIMITS]” Button (only Main display)
The limits function will alert the user when a meas­urement exceeds a specified limit. To set the limit val­ues, refer to limits function in the setup mode. Press the [LIMITS] button to activate the limits function; the word “LIMIT” should be displayed on the LCD. When the value of the main display exceeds the Hi limit, the word “Hi” will be displayed and the ther­mometer will beep in a pulsed tone. If the value of the main display is lower then the Lo limit, the word “Lo” will be displayed and the thermometer will beep in a continuous tone. To exit the limits function, press the [LIMITS] button.
(3) “ ” Button
The backlight function is represented by this button “ ”. Pressing the button will turn on or off the LCD backlight. The backlight will turn off automatically af­ter.
(4) “SAVE/READ” Button
The read data function works in conjunction with the save function in the shift mode, it is used for reading saved data. The save function can be activated in shift mode. Press the SAVE/READ button to read saved data; the word “READ” should be displayed on the LCD. To navigate around the save data table, press the overlay “SECOND” button until the “#” sign is dis­played on the second display. The location of the read pointer within the saved data table will be displayed. The arrow buttons on the overlay are used for scrolling through the saved data. Press the smaller arrows “ ” or “ ” to step through the data one at a time. Press the larger arrows “ ” or “ ” to step through the data ten at a time. Pressing the overlay “ESC” button deac­tivates the read data function.
(5) “LOG/READ” Button
The read log function works in conjunction with the log function. It is used for reading logged data. The log function can be activated in the shift mode. Press the LOG/READ button to activate the log read function; the word “READ” is displayed on the LCD. Press the overlay SECOND button to rotate through following display menus: T1, T2, GRP, and #. T1 and T2: Dis­plays the T1 or T2 saved data. GRP: Displays the cur­rent group number. #: Displays the current location of the read pointer within a selected group. The arrow buttons on the overlay are used for scrolling through the data and groups. Press the smaller arrows “ ” or
“ ” to step through the logged data or groups one at a time. Press the larger arrows “ ” or “ ” to step through the data or groups ten at a time. To navigate the logged data and groups, press the overlay
SECOND button until GRP appears in the second dis­play panel. Then select the group using the arrows. Press the SECOND button again until the “#” sign is displayed. The location of the read pointer in the se­lected group will be displayed. Use the arrows to scroll through the data. Pressing the overlay “ESC” button deactivates the read data function.
(7) HOLD Mode (only Main display)
When HOLD mode is selected, the thermometer holds the present readings and stops all further measure­ments. To activate the data hold mode, press the HOLD button, and “HOLD” is displayed on the LCD. Pressing the HOLD button again cancels the function, and the instrument will automatically resume meas­urements.
(8) MIN/MAX with Time record Mode
(only Main display)
The MIN/MAX function records the highest and low­est value recorded, and it calculates the average read­ing, and the differences of MAX to MIN. Press MIN/MAX button to enter the MIN/MAX recording mode. The beeper emits a tone when a new minimum or maximum measurement is recorded. Press the MIN/MAX button again to rotate through the current readings: MAX: The highest measurement recorded. MIN: The lowest measurement recorded. MAX-MIN: The difference of the highest and the lowest measure­ment. AVG: The average values of the measurements. This mode works in conjunction with the hold function, pressing the HOLD button will stop the recording and measurements (Previously recorded readings are not erased). Press HOLD button again to resume recording and measurements. To prevent accidental loss of MIN, MAX and AVG data, the MIN/MAX function can only be cancelled by pressing and holding down the MIN MAX key for more then 2 seconds. The automatic power-off feature, and the power, °C/°F, REL, SET, Hi/Lo Limits, TYPE, T1/T2 buttons are also disabled.
(10) “T1/T2” Button (Main display)
The input selection button [T1/T2] selects the input for the main display, T1 thermocouple or T2 thermocou­ple. Press the T1/T2 button to switch between the two inputs. When meter is turned on, it is set to the display that was last in use.
(11) “T1/T2” Button (Second display)
The input selection button [T1/T2] selects the input for the second display, T1 thermocouple or T2 thermo­couple. Press the T1/T2 button to switch between the two inputs. When meter is turned on, it is set to the display that was last in use.
(12) “T1-T2/TIME” Button (Third display)
The input selection button [T1-T2] selects the system time and date, or the differential between the two ther­mocouples (T1-T2) for the third display. Press the T1-T2 button to switch the display options. When me-
ter is turned on, it is set to the display that was last in use.
3. SHIFT MODE
The following functions can only be used in the shift mode.
(2) “°C/°F” Button
Press the °C/°F button to select the temperature scale, readings can be displayed in Celsius (°C) or Fahrenheit (°F). When the thermometer is turned on, it is set to the temperature scale that was last in use.
(4) “SAVE” Button
The save function stores the T1, T2 data in a nonvola­tile memory. Press the SAVE button to save the cur­rent data, the word SAVE is displayed to indicate the data has been saved. The built in memory can store up to 256 data. The data can be read using the read func­tion in the normal mode.
(5) “LOG” Button
The data log function continuously records the data according to a specified time interval. The time inter­val can be set using the interval setup function [INVT] in the set up mode. Press the LOG button to activate the log function; the indicatosr “LOG” and “MEM” will be displayed on the LCD. There are 16 groups that are used for storing the log data. A maximum of 16,000 data point can be saved in one log session. Press the LOG button again to exit the data log func­tion.
(6) “CLR ?” Button
The CLR function clears all the saved and logged data in memory. When the CLR button is pressed, indicator “MEM” is displayed and the “CLR” on upper-right of LCD will blink. Pressing the “ENTER” button printed on the overlay in white will clear all saved and logged data. Press “ESC” button to exit this function without clearing data.
(7) “REL” Button (Main display)
The relative value function can be used for comparing the saved reference value with other measurements. Press the “REL” button to store the current measure­ment as the reference value, and “REL” should be dis­played on the right part of the LCD. The next meas­urement will display the current value compared to the reference value. Press “REL” button again to clear the reference value and deactivate the relative value meas­urement function.
(8) “APO” Button
Press the APO button to turn the “Auto power off” function on or off. When this function is enabled, the indicator “APO” is shown at the upper left part of the LCD. When APO (Auto power off) is enabled, it will automatically turn the thermometer off no button is pressed for a period longer than the set time interval (the default time for APO is 5 minutes). Press power button to resume operation.
(10) “TYPE” Button (Main display)
Press this button to change the type of thermocouple in the main display (K/J/T/E/R/S/N). If the inputs of the
Loading...
+ 1 hidden pages