and Humidity
Dimensions:About 250 (L) × 100 (W) × 45(H) mm
Weight:About 500g with battery
Printer:Interval print :
00 hour : 00 mi nute : 03 second 〜23 hour :
59 minute : 59 sec
Trigger print : Print three data
(date / time / Reading)
Accessories:Thermo carrying case, Test leads with alligator
clip, 2 rolls of thermo paper, 110V/220V AC to 9V
DC power adapter/power cord, RS-232 cable, P.C.
software disk, 6pcs 1.5V AA batteries, TL-A37
analog output cable, RS-232 convert header,
Operation manual.
4
2-2 Electrical Specification
Analog test range :DCV/ACV:2V,200mV
DCA: 20mA
Accuracy:(at 23℃±5℃,below 80%RH)
DCè(±1%rdg+3dgts)
ACè(±1%rdg+5dgts) at (50~450Hz)
Resolution:200mV/2V range is 0.1mV/1mV
20mA range is 0.01mA
A / D Sample Rate :2.5 times/sec
Power Source:6 pcs 1.5V,# 3 battery or 9V, DC adaptor
(inside negative, outside positive)
Battery Life:Approx. 100 hours (without printing)
Printer:Thermal printer (EPL1902 of Panasonic)
MEMORY:60000 records (up to 130000 records)
5
3. PARTS & CONTROLS
3-1 The Name of Each Parts and Positions:
200mV2V 200 mV 2V20mA
C¢XFcLuxdB
F
¢X
HiHiHiLoLoLo
DATE TIME START STOP
13456 789
20
,
3
SETTING
FEED
SETUP
LINEMIN
CLOCKALARM
PRINT
RECORD
,
1
TRIGGER
,
3 CONSECUTIVE
COM
mA
20mA
MAX
FUSED
600V
(Fig-2)
6
¢X
¢X
60V
24V
MAX
RH A A
UNITRANGE
MAX
V
TG
INTV
1.
V
A
2.
3.
4.
200mV
C / F
HiHi
DATA
0
2V
Lux/Fc
Hi
TIME
1/2/3
200mV
dB
LoLo
START
4/5/6
2
%RH
Lo
STOP
Toggle button for ( 200mV, ℃/℉, HiHi, DATA)
and digit “ o ”.
Toggle button for ( 2V, Lux/ Fc, Hi, TIME) and
digit “ 1/2/3 ”.
Toggle button for ( 200mV, dB, LoLo, START)
and digit “ 4/5/6 ”.
Toggle button for ( 2V, %RH, Lo, STOP) and
digit “ 7/8/9 ”.
5.
6.
7/8/9
20m
A/
G/T
INTV
3'SETTING
SETUP
Toggle button for ( 20mA, A / A,
G / T, INTV ) button and “ ” right shift
function.
Press and hold this button for 3 sec will into
setting mode, press again will end setting
mode.
7
7.
RANGE
FEED
LINE
Feed paper in normal mode, range function in
setting mode.
8.
9.
10.
11.
“ COM ” Measuring
12
13
14
“ V ” Measuring
“ mA ” Measuring
UNIT
MAX
MIN
CLOCK
PRINT
ALARM
=
RECORD
Socket
Socket
Socket
MAX/MIN function in normal mode “ UNIT ”
function in setting mode.
In normal mode, press 1 sec is trigger print
function, press and hold for 3 sec is start/stop
consecutive print function, but in setting mode
will be “ clock ” function.
In normal mode, press 1 sec is trigger record (
MAX 255 times), In setting mode is “ alarm ”
function, press and hold over 3 sec is start/stop
consecutive record function.
Power button.
To connect negative lead of voltage, current.
19 FEED ButtonFeed paper into printer
20 DC 9V SocketTo connect adaptor.
8
3-2 Power Status :
As battery power is not sufficient, the , will be shown on LCD
replace with six new 1.5V batteries is required.
3-3 Quick Manual for Easy Set-up
STEP1STEP2STEP3STEP4STEP5
Push this
button 3 second.
SETUP
Select one.
RANGE
UNIT
CLOCK
ALARM
Choose one from specified
function Bar.
2V
LUX / FC
TIME START
Hi
200mV
dB%RH
STOP
LoLo
Lo
200mV2V
F
c
DATE
HiHi
20mA
A /
INTV
G / T
A
Reconfirm
input setting.
RANGE
UNIT
CLOCK
ALARM
a. You could only set up one of the functions at each time.
b. To change position of point decimal, follow step1 & step2, then press button.
c. To input digit, follow step1 & step3 to set the number.
Press to
return
normal.
SETUP
3-4 ACV、DCV and DCA Test Setting :
1. ACV 2V Test setting sequence:Press and hold
à press
3'SETTING
RANGE
à press
, to end setting mode.
2. ACV 200mV Test setting sequence:Press and hold
sec à press
3'SETTING
press
RANGE
à press
, to end setting mode.
3. DCV 2V Test setting sequence:Press and hold
à press
3'SETTING
RANGE
à press
, to end setting mode.
2V
à press
200mV
2V
à press
9
3'SETTING
RANGE
à press
3'SETTING
RANGE
3'SETTING
for 3 sec
à press
for 3
RANGE
for 3 sec
à press
à
4. DCV 200mV Test setting sequence:Press and hold
sec à press
3'SETTING
press
RANGE
à press
, to end setting mode.
200mV
à press
3'SETTING
RANGE
for 3
à
5. DCA 20mA Test setting sequence:Press and hold
sec à press
3'SETTING
press
RANGE
à press
, to end setting mode.
20mA
3-5 UNIT Setting Step:
1. ℃/℉ Setting step:press and hold
UNIT
press
à press
3'SETTING
, to end setting mode.
C/ F
change℃/℉ à press
2. Lux/Fc Setting step:press and hold
UNIT
press
à press
3'SETTING
Lux/Fc
change Luc/Fc à press
, to end setting mode.
3. dB Setting step:press and hold
UNIT
à press
dB
à press
to end setting mode.
4. %RH Setting step:press and hold
UNIT
à press
%RH
à press
to end setting mode.
5. / A Setting step:press and hold
UNIT
à press change A/A à press
3'SETTING
, to end setting mode.
3'SETTING
3'SETTING
3'SETTING
UNIT
3'SETTING
UNIT
3'SETTING
à press
3'SETTING
for 3
RANGE
for 3 sec à press
UNIT
for 3 sec à press
UNIT
for 3 sec à press
à press
3'SETTING
for 3 sec à press
à press
3'SETTING
for 3 sec à press
UNIT
à press
à
à
à
,
,
3-6 ALARM Setting Step:
1. HIHI ALARM Setting step:press and hold
press
ALARM
à press
shift button setting “ALARM" à press
to end setting mode.
2. HI ALARM Setting step:press and hold
mode à press
ALARM
HiHi
à press
10
3'SETTING
for 3 sec à
à assort 0~9 number è right
ALARM
3'SETTING
Hi
à press
fo r 3 sec into setting
à assort 0~9 number
3'SETTING
,
è right shift button setting “ALARM" à press
3'SETTING
, to end setting mode.
ALARM
à press
3. LOLO ALARM Setting step:press and hold
setting mode à press
ALARM
à press
number è right shift button setting “ALARM" à press
3'SETTING
press
, to end setting mode.
4. LO ALARM Setting step:press and hold
setting mode à press
ALARM
à press
number è right shift button setting “ALARM" à press
3'SETTING
press
, to end setting mode.
3'SETTING
3'SETTING
LoLo
Lo
for 3 sec into
à assort 0~9
ALARM
for 3 sec into
à assort 0~9
ALARM
3-7 Real Time Clock , Time/ Date Setting Step:
1. Real time clock , date setting step:press and hold
sec into setting mode à press
CLOCK
à press
assort 0~9number è right shift button setting date à press
CLOCK
à press
3'SETTING
, to end setting mode.
2. Real time clock, time setting step:press and hold
sec into setting mode à press
CLOCK
à press
assort 0~9number è right shift button setting time à press
CLOCK
à press
3'SETTING
, to end setting mode.
3'SETTING
3'SETTING
for 3
DATE
for 3
TIME
3-8 Record/Print Start Time, Stop Time, Interval Time Setting Step:
à
à
à
à
1.
Start time setting step
setting mode
à press
:press and hold
CLOCK
à press
3'SETTING
START
number è right shift button setting start time à press
3'SETTING
press
2.Stop
time setting step
setting mode
0~9number è right shift button
3'SETTING
press
3. Interval time setting step:press and hold
setting mode à press
number è right shift button setting in terval time à press
to end setting mode.
:press and hold
à press
to end setting mode.
CLOCK
setting stop time à press
CLOCK
à press
11
3'SETTING
à press
3'SETTING
INTV
for 3 sec into
à assort
CLOCK
for 3 sec into
STOP
à
CLOCK
for 3 sec into
à assort 0~9
CLOCK
0~9
à
assort
à
à
3'SETTING
press
to end setting mode.
3-9 Print Mode Setting Step:
Printer work mode setting sequence:press and hold
sec into setting mode à press
G/T MODE à press
3'SETTING
ALARM
, to end setting mode.
à press
3-10 Clear Record Step:
3'SETTING
G/T
for 3
change
Press
=
RECORD
+
Power ON in the same time
The LCD should display “dEL”
3-11 Disable Start、Stop、Intv and Setting Step:
1.Disable START、STOP、INTV sequence:press and hold
3'SETTING
START
for 3 sec into
3'SETTING
start, stop, intv à press
for 3 sec into setting mode à press
3'SETTING
to end setting mode.
2.Disable ALARM sequence:press and hold
setting mode à press
STOP
disable alarm à press
to end setting mode.
3-12 Decimal Point Setting Step:
Shift point setting sequence:press and hold
enter setting mode à press
position à press
3'SETTING
to end setting mode.
INTV
3'SETTING
to change decimal point
3-13 Consecutive Record and Print Setting Step:
Setting sequence:press and hold
mode à press
TIME
setting / = à press
end setting mode.
※ If already in setting MODE, press
3'SETTING
TIME
for 3 sec to enter setting
3'SETTING
button only.
3-14 Reset System Setting Step:
disable
3'SETTING
for 3 sec to
, to
Press
3'SETTING
+ Power ON
12
3-15 Record / Print Start and Step Operation Method:
way)
b). Into setting mode, setting Intv.
c). Back to general mode.
d). Press
=
RECORD
PRINT
button 3 second, the LCD will be show
= or symbol, into consecutive Record / Print of hand
setting mode.
PRINT
or
(2). End Record/Print :
1. Automatic mode : The function will be stopped automatically while
the STOP time is up.
2. Hand setting mode :To stop consecutive Record/Print by user press
=
and hold
PRINT
RECORD
symbol will be disappear.
or
PRINT
button for 3 seconds. The LCD
3-16 Memory Full Warning :
The = symbol will be flash.
3-17 Alarm Print in Graphic Mode :
When printing in G mode, if you have setting any alarm points
(HiHi, Hi, LoLo, Lo) when the read value over the set points, the
printer the will print alarm value on the paper.
13
=
RECORD
or
3-18 Scale Change in Graphic Mode
HiHi
ALARMHiALARMLoALARM
üNo change
üNo change
üNo change
üüChange scale by HiHi Hi ALARM
üüChange scale by HiHi Lo ALARM
üüChange scale by HiHi LoLo ALARM
üüChange scale by Hi Lo ALARM
üüChange scale by Hi LoLo ALARM
üüChange scale by Lo LoLo ALARM
üüüChange scale by HiHi Lo ALARM
üüüChange scale by HiHi LoLo ALARM
üüüChange scale by HiHi LoLo ALARM
üüüChange scale byHi LoLo ALARM
LoLo
ALARM
üNo change
Scale change by alarm setting
üüüüChange scale by HiHi Lo ALARM
No setting ALARM, no change
ü : had setting
3-19 How to ConnectsTL-(A37) Analog Output Cable
14
Loading...
+ 43 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.