Omega Products HH506R Installation Manual

Page 1
OMEGA
HH506R
Digital Thermometer
Page 2
omega. com
OMEGAnetSM On-Line Service
http://www.omega.com
USA: ISO 9001 Certified Canada:
One Omega Drive, Box 4047 976 Bergar
Stamford, CT 06907-0047 Laval (Quebec) H7L5A1
Tel: (203) 359-1660 Tel: (514) 856-6928
FAX: (203)359-7700 FAX: (514) 856-6886
e-mail: info@omega.com e-mail: info@omega.com
USA and Canada: Mexico and Latin America:
Sales Service: 1-800-826-6342 / 1-800-TC-OMEGA
Customer Service: 1-800-622-2378 / 1-800-622-BEST
Engineering Service: 1-800-872-9436 / 1-800-USA-WHENSMEn Español: (203) 359-7803
TELEX: 996404 EASYLINK: 62968934 CABLE: OMEGA e-mail: espanol@omega.com
Benelux: Germany/Austria:
Postbus 8034, 1180 LA Amstelveen, Daimlerstrasse 26, D-75392
The Netherlands Deckenpfronn, Germany
Tel: (31) 20 6418405 FAX: (31) 20 6434643 Tel: 49 (07056) 3017 FAX: 49 (07056) 8540
Toll Free in Benelux: 06 0993344 Toll Free in Germany: 0800 82 66342
e-mail: nl@omega.com e-mail: germany@omega.com
Czech Republic: United Kingdom: ISO 9002 Certified
ul. Rude armady 1868, 733 01 Karvina- One Omega Drive
Hranice, Czech Repubic Riverbend Technology Centre Northbank, Irlam,
Tel: 420 (69) 6311627 FAX: 420 (69) 6311114 Manchester, M44 5EX, England
e-mail: czech@omega.com Tel: 44 (161) 777-6611 FAX: 44 (161) 777-6622
France:
9, rue Denis Papin, 78190 Trappes Toll Free in England: 0800-488-488
Tel: (33) 130-621-400 FAX: (33)130-699-120 e-mail: sales@omega.com.uk
Toll Free in France: 0800-4-06342
e-mail: france@omega.com
It is the policy of OMEGA to comply with all worldwide safety and EMC/EMI regulations that apply. OMEGA is constantly pursuing certification of its products to the European New Approach Directives. OMEGA will add the CE mark to every appropriate device upon certification. The information contained in this document is believed to be correct but OMEGA Engineering, Inc. accepts no liability for any errors it contains, and reserves the right to alter specifications without notice. WARNING: These products are not designed for use in, and should not be used for, patient connected application.
For immediate technical or application assistance:
Servicing North America:
SM
SM
Servicing Europe:
TM
Internet e-mail
Tel: (95) 800-TC-OMEGA
FAX: (95) 203-359-7807
SM
Page 3
INTRODUCTION
This instrument is a 4½ digit, compact-sized portable digital thermometer designed to use external K/J/T/E/R/S/N-type thermocouples as temperature sensor. Temperature indication follows Reference Temperature/Voltage Tables (N.I.S.T. Monograph 175 Revised to ITS-90) for K/J/T/E/R/S/N-type thermo­couples. Two K-type thermocouple are supplied with the thermometer.
SAFETY INFORMATION
It is recommended that you read the safety and operation instructions before using the thermometer.
WARNING
To avoid electrical shock, do not use this instrument when working
voltages at the measurement surface over 24V AC or DC.
WARNING
To avoid damage or burns, do not make temperature
measurement in microwave ovens.
CAUTION
Repeated sharp flexing can break the thermocouple leads. To prolong
lead life, avoid sharp bends in the leads, especially near the connector. The symbol on the instrument indicates that the operator must refer to an
explanation in this manual.
1
Page 4
SPECIFICATIONS
ELECTRICAL
Temperature Scale: Celsius or Fahrenheit user-selectable Measurement Range:
Thermocouple Range K-TYPE -200°C to 1372°C, -328°F to 2501°F J-TYPE -210°C to 1050°C, -346°F to 1922°F T-TYPE -200°C to 400°C, -328°F to 752°F E-TYPE -210°C to 790°C, -346°F to 1454°F R-TYPE 0°C to 1767°C, 32°F to 3212°F S-TYPE 0°C to 1767°C, 32°F to 3212°F N-TYPE -50°C to 1300°C, -58°F to 2372°F
Resolution: K-TYPE 0.1°C, 0.1°F (0°F to 700°F)
0.2°F on other range
J-TYPE 0.1°C, 0.1°F (0°F to 500°F)
0.2°F on other range
T-TYPE 0.1°C, 0.1°F (0°F to 600°F)
0.2°F on other range
E-TYPE 0.1°C, 0.1°F (0°F to 450°F)
0.2°F on other range R-TYPE 1°C , 1°F S-TYPE 1°C , 1°F N-TYPE 0.1°C , 0.2°F
2
Page 5
Accuracy: Accuracy is specified for operating temperatures over the range of
18°C to 28°C (64°F to 82°F), for 1 year, not including thermocouple error.
K/J/T/E-TYPE ±(0.05% rdg + 0.3°C) -50°C to 1370°C
±(0.05% rdg + 0.7°C) -50°C to -210°C ±(0.05% rdg + 0.6°F) -58°F to 2498°F ±(0.05% rdg + 1.4°F) -58°F to -346°F
N-TYPE ±(0.05% rdg + 0.7°C) -50°C to 0°C
±(0.05% rdg + 0.3°C) 0°C to 1300°C ±(0.05% rdg + 1.4°F) -58°F to 32°F ±(0.05% rdg + 0.6°F) 32°F to 2372°F
R/S-TYPE ±(0.05% rdg + 2°C) 0°C to 1767°C
±(0.05% rdg + 4°F) 32°F to 3212°F
Temperature Coefficient: 0.1 times the applicable accuracy specification
per °C from 0°C to 18°C and 28°C to 50°C (32°F to 64°F and 82°F to 122°F).
Input Protection: 24V dc or 24V ac rms maximum input voltage on any
combination of input pins.
Maximum Differential Common Mode Voltage (Maximum Voltage be-
tween T1 and T2 during measurement): 1volt. Reading Rate: one time per second. Input Connector: Accepts standard miniature thermocouple connectors (flat
blades spaced 7.9mm, center to center).
3
Page 6
ENVIRONMENTAL
Ambient Operating Ranges: 0°C to 50°C (32°F to 122°F) <80% R.H. Storage Temperature: -20°C to 60°C (-4°F to 140°F) <70% R.H.
GENERAL
Display: 4½ digit liquid crystal display (LCD) with maximum reading of 9999.9. Overload: "----.-" is display. Battery: Standard 9V battery. Battery Life: 100 hours typical. Auto power off: The meter key switch inactive for more than 30 minutes,
press power key to resume operation.
Dimensions: 192mm(H) x 91mm(W) x 52.5mm(D). Weight: 365g.
4
Page 7
OPERATING INSTRUCTIONS
1
Power Switch
The key turns the thermometer on or off. In the MIN MAX record mode can not power off, must leave MIN MAX record mode then power off.
2
°C/°F Selecting the Temperature Scale
Readings are dual displayed in either degrees Celsius(°C) or degrees Fahrenheit(°F). When the thermometer is turned on, it is set to the temperature scale that was in use when the thermometer was last turned off. To change the temperature scale, press the °C/°F key.
3
HOLD Mode (only Main display)
Pressing the HOLD key to enter the Data Hold mode, the "HOLD" annun­ciator is displayed. When HOLD mode is selected, the thermometer held the present readings and stops all further measurements. Pressing the HOLD key again to cancel HOLD mode causing thermometer to resume taking measurements. In the MIN/MAX recording mode, press HOLD key to stop the recording. Press HOLD key again to resume recording. (Previously recorded reading are not erased).
5
Page 8
T1 T2/T1-T2 Main display Input Selection
4
The input selection indicates which input is selected for main display; T1 thermocouple, T2 thermocouple or the difference between the two thermo­couples (T1-T2), when the thermometer is turned on, it is set to T1, When main display input selected T1, then T1 input can select alternate of thermocouple by pressing TYPE key switch.
5
K/J/T/E/R/S/N T1 Input Thermocouple Type Select (Main display)
The TYPE key switch the T1 input circulating selects the K/J/T/E/R/S/N type thermocouple as input, when main display input selected T1. When the thermometer is turned on, it is set to the type selected that was in use when the thermometer was last turned off.
6
MIN MAX with Time record Mode (only Main display)
Press MIN MAX key to enter the MIN MAX Recording mode, (displays the Maximum reading with time, Minimum reading with time and Average reading stored in record mode). In the this mode the automatic power-off feature is disabled and key, °C/°F key, REL key, SET key, Hi/Lo Limits key and main display T1 T2 T1-T2 key, TYPE key are disabled. The beeper emits a tone when a new minimum or maximum value is recorded. Push MIN MAX key to cycle through the MAX, MIN and AVG readings. If an overload is recorded, the averaging function is stopped and average value display"----.-".
6
Page 9
The true average of all the reading taken over at least 22 hours preiod can be displayed. If 22 hours is exceeded, new averages are no longer calculated. The last calcalated value is retained as the average reading, but the actual minimum and maximum reading will continue to be captured. In the this mode, press HOLD key to stop the recording of readings, all values are frozen, press again to restart recording. To prevent accidental loss of MIN, MAX and AVG data, in this mode can only be cancelled by pressing and hold down the MIN MAX key for 2 seconds to exit and erased recorded readings.
REL Relative mode (only Main display)
7
Pressing REL key to enter the Relative mode, zero the display, and store the displayed Reading as a reference value and annunciator REL is displayed. Press REL key again to exit the relative mode. The relative value can also be entered by the user. (See "SET mode" later in this manual.) When the desired Relative value has been entered, press REL key to enter the Relative mode, press SET key use set Relative value as a reference value. Press REL key again to exit the relative mode. In the Relative mode, the value (can not >±3000.0 counts) shown on the LCD is always the difference between the stored reference and the present reading.
7
Page 10
Hi/Lo LIMITS mode (only Main display)
8
Press Hi/Lo LIMITS key to enter the Hi/Lo Limits comparative mode, "LIMIT" is displayed. In the this mode the automatic power-off feature is disabled and REL key, HOLD key, MAX/MIN key are disabled, when input temperature value exceed Hi. The beeper emits a continuity pulse tone and "Hi" is displayed, and when input temperature value exceed Lo value.The beeper emits a discontinuous pulse tone and "Lo" is displayed. Press Hi/Lo LIMIT key again to exit the Hi/Lo LIMIT mode.
9
SET mode (Relative value set, Time set and Hi/Lo Limits value set and Time set)
1. Press SET key to enter Relative valuse SET mode (Press ENTER key can escape relative valuse set mode), REL set mode. = = = =.= is displayed in main display.Relative value is entered via overlay numbers, when you want to get negative values push (- 0) key for end of numbers, then press overlay ENTER key, stored the relative value, enter Hi/Lo Limits value set mode.
2. Hi Limit value set mode, (Press ENTER key can escape Hi Limit value set mode), = = = =.= is displayed in main display, Hi Limit value is entered via overlay numbers, when you want to get negative values push (- 0) key for end of numbers, then press overlay ENTER key, stored the Hi Limit value, enter Lo Limit value set mode (Press ENTER key can escape Lo Limit value set mode), = = = =.= is displayed in main display, Lo Limit value is entered via overlay numbers, when you want to get negative values push (- 0) key for end of numbers, then press overlay ENTER key, stored the Lo Limit value, enter Time set mode.
8
Page 11
3. Time set mode, (Press ENTER key can escape Time set mode) = = : = = = =is displayed in third display. Time (hours, minutes, seconds) value is entered via overlay numbers, then press overlay ENTER key. Time start from set time value and exit set mode.
10
T1/T2 T1-T2 second display Input Selection
The input selection indicates which input is selected for second display; T1 thermocouple, T2 thermocouple or the difference between the two thermo­couples (T1-T2), when the thermometer is turn on, it is set to T2 input can select alternate of thermocouple by second display TYPE key switch.
11
K/J/T/E/R/S/N T2 Input Thermocouple Type Select (Second display)
The TYPE key switch the T2 input circulating selects the K/J/T/E/R/S/N type thermocouple as input, when second display input selected T2. When the thermometer is turned on, it is set to the type selected that was in use when the thermometer was last turned off.
12
" " Button
Press " " button to toggle on and off of backlight. The backlight will switch­off automatically after 60 seconds.
9
Page 12
OPERATOR MAINTENANCE
WARNING
To avoid possible electrical shock, disconnect the thermocouple
connectors from the thermometer before removing the cover.
Battery Replacement
Power is supplied by a 9 volt "transistor" battery. The " " appears on the LCD display when replacement is needed. To replace the battery, remove the two screws from the back of the meter and lift off the battery cover. Remove the battery from battery contacts.
10
Page 13
RS-232 Operation Using Quick BASIC
The following example shows how to send command instruction and receive data responses over the RS-232 interface using Quick BASIC. The program example is set up on "COM1".
CLS E$ = "A" ST1: CLOSE #1 OPEN "COM1:1200,E,7,1,DS,RS" FOR OUTPUT AS #1 PRINT #1, E$; CLOSE #1 OPEN "COM1:1200,E,7,1,DS,RS" FOR RANDOM AS #1 ST2: INPUT #1, A$ B$ = MID$(A$, 2, 6) DEC = 0 F$ = "0123456789ABCDEF" N = 6 FOR I = 0 TO 5 C$ = MID$(B$, N, 1) N = N - 1 D = INSTR(F$, C$) - 1 DEC = DEC + (16 ^ I) * D NEXT DEC = DEC / 1000 C$ = MID$(A$, 1, 1)
11
Page 14
IF C$ = "-" THEN DEC = DEC * -1 END IF B$ = MID$(A$, 10, 6) DEC1 = 0 N = 6 FOR I = 0 TO 5 C$ = MID$(B$, N, 1) N = N - 1 D = INSTR(F$, C$) - 1 DEC1 = DEC1 + (16 ^ I) * D NEXT DEC1 = DEC1 / 1000 C$ = MID$(A$, 9, 1) IF C$ = "-" THEN DEC1 = DEC1 * -1 END IF CLS LOCATE 12, 33 PRINT " T1 T2" LOCATE 13, 30 PRINT USING "######.# ######.#"; DEC; DEC1 E$ = INKEY$ IF E$ = "" THEN GOTO ST2 END IF GOTO ST1
12
Page 15
RS-232 Transmit and Receive Comammand Summary
1. Control letter should use Capital Letter.
2. If you write you own program, the host computer must send the "A" letter to activate data transmission.
3. If you should stop data transmission the host computer must send the "B" letter to extend battery life.
4. Communication Parameters Baud rate: 1200 Parity check: EVEN Data bits: 7 Stop bits: 1 "A" Activate data transmission "B" Stop data transmission "C" °C/°F Key "D" HOLD Key "E" T1/T2/T1-T2 Key (main display) "F" TYPE Key (main display) "G" MIN/MAX Key "H" Exit record mode " I " REL Key "J" LIMITS Key "K" T1/T2/T1-T2 Key (second display) "L" TYPE Key (second display) "M" " " Key "P" Turn off power of meter
13
Page 16
OUTPUT DATA FORMATS
The data format consists of 32 bytes. byte
0000000001111111 1234567890123456 SHHHHHHASHHHHHHA
T1 present status T2 present status
S: Negative sign or positive sign H: Hexadecimal digits D: Numeric digits A: Capital Letter or "-" sign CR: Carriage return character nl: newline character 01 T1 polarity "+" Positive 08 T2 polarity "-" Negative 02 — MSD T1 Values=[H07x160+H06x161+H05x16 07 — LSD +H04x163+H03x164+H02x165] / 1000 10 — MSD T2 Values=[H15x160+H14x161+H13x16 15 — LSD +H12x163+H11x164+H10x165] / 1000 08 — T1 Thermocouple Type "K" ® K type 16 — T2 Thermocouple Type "J" ® J type
1112222222222333 7890123456789012 DDDDDDAAAAAAAA
3 rd display reading Meter status
"T" ® T type
"E" ® E type "R" ® R type "S" ® S type "N" ® N type
14
CR nl
2
2
Page 17
17 — MSD 18 — LSD 19 — MSD 20 — LSD 21 — MSD 22 — LSD 23 — Record mode "R" ® Record mode
24 — REL mode "R" ® Relative mode 25 — HOLD mode "H" ® HOLD mode 26 — LIMITS mode "L" ® Limits mode 27 — Hi LIMITS "H" ® Hi Limits 28 — Lo LIMITS " L" ® Lo Limits 29 — none "-"
30 — Battery status "B" ® Low battery
Hours
Minutes
Seconds
"M" ® MAX mode " I " ® MIN mode
"A" ® AVG mode "-" ® normal mode
"-" ® normal mode "-" ® normal mode "-" ® normal mode "-" ® normal mode "-" ® normal mode
"-" ® normal mode
15
Page 18
COMMUNICATION
4612-IR232
RS232C CABLE
506R
RS-232C "COM1" PORT
1. Install the RSR232C cable between computer "COM1" and Thermometer RS232C port.
2. Turn on Thermometer.
3. Utilize the Quick BASIC for inputting programs.
4. Use the capital control letter and input "A" letter from the keyboard to start sending messages to computer.
5. Input "C" letter from the keyboard and then you can change the temperature scale.
6. Input "F" letter from the keyboards and then you can circulating select the T1 input type.
7. Input "G" letter from the keyboard and then the meter enter record mode, and then you can push "G" again to circulate through Maximum/Minimum/Average/ Record mode.
8. When you want to stop transmission mode you should push "B" letter from the keyboard in order to conserve battery life.
16
Page 19
HH 506(R)/HH 507(R) CALIBRATION PROCEDURE
Note: The following calibration procedure should perform only by qualified technicians
who have access to the items as following.
Equipment: The class of calibrator had better 10 times greater than the measured meter.
1. Set the jumper position on J1 position.
2. Input DCV 17mV to the T1 after the display is stabilized (forward 4 digits), then press “ HOLD “ key and "0.0" is displayed in main display.
3. Input DCV 60mV after the display is stabilized (first 4 digits), then press “ HOLD “ key and "0.1" is displayed in main display.
4. Input 0°C (K-type) after the disp lay is stabilized (first 4 digits), then press “ HOLD “ key and "0.2" is displayed in main display.
5. Don't push any key and the meter will turn off automatically after 20 seconds.
6. After adjustment set the jumper back to J3 position.
7. Turn on the meter then input 0°C (K-type). The display reading reads 0°C if the calibration procedure is right.
1
Page 20
OMEGA ENGINEERING, INC. warrants this unit to be free of defects in materials and workmanship for a period of 13 months from date of purchase. OMEGA Warranty adds an additional one (1) month grace period to the normal one (1) year product warranty to cover handling and shipping time. This ensures that OMEGA's customers receive maximum cov erage on each product.If the unit should malfunction, it must be returned to the factory for evaluation. OMEGA's Customer Service Department will issue an Authorized Return (AR) number immediately upon phone or written request. Upon examination by OMEGA, if the unit is found to be defective it will be repaired or replaced at no charge. OMEGA's W ARRANTY does not apply to defects resulting from any action of the purchaser, including but not limited to mishandling, improper interfacing, operation outside of design limits, improper repair, or unauthorized modification. This WARRANTY is V OID if the unit shows e vidence of having been tampered with or shows evidence of being damaged as a result of excessive corrosion; or current, heat moisture or vibration; improper specification; misapplication; misuse or other operating conditions outside of OMEGA's control. Components which wear are not warranted, including but not limited to contact points, fuses, and triacs.
OMEGA is pleased to offer suggestions on the use of its various products. However , OMEGA neither assumes responsibility for any omissions or errors nor assumes liability for an y damages that result fr om the use of its products in accordance with information provided by OMEGA, either verbal or written. OMEGA warrants only that the parts manufactured by it will be as specified and free of defects. OMEGA MAKES NO OTHER WARRANTIES OR REPRESENTATIONS OF ANY KIND WHATSOEVER, EXPRESSED OR IMPLIED, EXCEPT THAT OF TITLE AND ALL IMPLIED WARRANTIES INCLUDING ANY WARRANTY OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE ARE HEREBY DISCLAIMED. LIMITATION OF LIABILITY: The remedies of purchaser set forth herein are exclusive and the total liability of OMEGA with respect to this order, whether based on contract, warranty, negligence, indemnification, strict liability or otherwise, shall not exceed the purchase price of the component upon which liability is based. In no event shall OMEGA be liable for consequential, incidental or special damages.
CONDITIONS: Equipment sold by OMEGA is not intended to be used, nor shall it be used: (1) as a "Basic Component" under 10 CFR 21 (NRC), used in or with any nuclear installation or activity; or (2) in medical applications or used on humans. Should any Product(s) be used in or with any nuclear installation or activity, medical application, used on humans, or misused in any way, OMEGA assumes no responsibility as set forth in our basic WARRANTY / DISCLAIMER language, and additionally, purchaser will indemnify OMEGA and hold OMEGA harmless from any liability or damage whatsoever arising out of the use of the Product(s) in such a manner.
Direct all warranty and repair requests/inquiries to the OMEGA Customer Service Department. BEFORE RETURNING ANY PRODUCT(S) TO OMEGA, PURCHASER MUST OBTAIN AN AUTHORIZED RETURN (AR) NUMBER FROM OMEGA'S CUSTOMER SERVICE DEPARTMENT (IN ORDER TO AVOID PROCESSING DELAYS). The assigned AR number should then be marked on the outside of the return package and on any correspondence. The purchaser is responsible for shipping charges, freight, insurance and proper packaging to prevent breakage in transit.
FOR WARRANTY RETURNS, please have the f ollowing
information available BEFORE contacting OMEGA:
1. P.O. number under which the product was PURCHASED.
2. Model and serial number of the product under warranty, and
3. Repair instructions and/or specific problems relative to the product.
OMEGA's policy is to make running changes, not model changes, whenever an improvement is possible. This affords our customers the latest in technology and engineering. OMEGA is a registered trademark of OMEGA ENGINEERING, INC. © Copyright 1999 OMEGA ENGINEERING, INC. All rights reserved. This document may not be copied, photocopied, reproduced, translated, or reduced to any electronic medium or machine-readable from, in whole or in part, without prior written consent of OMEGA ENGINEERING, INC.
WARRANTY / DISCLAIMER
RETURN REQUESTS / INQUIRIES
FOR NON-WARRANTY REPAIRS, consult OMEGA for
current repair charges. Have the following information available BEFORE contacting OMEGA:
1. P.O. number to cover the COST of the repair.
2. Model and serial number of product , and
3. Repair instructions and/or specific problems relative to the product.
Page 21
Where Do I Find Everything I Need for
Process Measurement and Control?
OMEGA...Of Course!
TEMPERATURE
þ Thermocouple, RTD & Thermistor
Probes, Connectors, Panels & Assemblies
þ Wire: Thermocouple, RTD & Thermistor
þ Calibrators & Ice Point References
þ Recorders, Controllers & Process Monitors
þ Infrared Pyrometers
PRESSURE/STRAIN AND FORCE
þ Transducers & Strain Gauges
þ Load Cells & Pressure Gauges
þ Displacement Transducers
þ Instrumentation & Accessories
FLOW/LEVEL
þ Rotameters, Gas Mass Flowmeters
& Flow Computers
þ Air Velocity Indicators
þ Turbine/Paddlewheel Systems
þ Totalizers & Batch Controllers
pH/CONDUCTIVITY
þ pH Electrodes, Testers & Accessories
þ Benchtop/Laboratory Meters
þ Controllers, Calibrators, Simulators
& Pumps
þ Industrial pH & Conductivity Equipment
M-2812/0799
DATA ACQUISITION
þ Data Acquisition &
Engineering Software
þ Communications-Based
Acquisition Systems
þ Plug-in Cards for Apple, IBM
& Compatibles
þ Datalogging Systems
þ Recorders, Printers & Plotters
HEATERS
þ Heating Cable
þ Cartridge & Strip Heaters
þ Immersion & Band Heaters
þ Flexible Heaters
þ Laboratory Heaters
ENVIRONMENTAL MONITORING AND CONTROL
þ Metering & Control Instrumentation
þ Refractometers
þ Pumps & Tubing
þ Air, Soil & Water Monitors
þ Industrial Water & Wastewater
Treatment
þ pH, Conductivity & Dissolved Oxygen
Instruments
Loading...