It is the policy of OMEGA to comply with all worldwide safety and EMC/EMI regulations that apply. OMEGA is constantly pursuing certification
of its products to the European New Approach Directives. OMEGA will add the CE mark to every appropriate device upon certification.
The information contained in this document is believed to be correct but OMEGA Engineering, Inc. accepts no liability for any errors it contains,
and reserves the right to alter specifications without notice.
WARNING: These products are not designed for use in, and should not be used for, patient connected application.
For immediate technical or application assistance:
Servicing North America:
SM
SM
Servicing Europe:
TM
Internet e-mail
info@omega.com
Tel: (95) 800-TC-OMEGA
FAX: (95) 203-359-7807
SM
INTRODUCTION
This instrument is a 4½ digit, compact-sized portable digital thermometer
designed to use external K/J/T/E/R/S/N-type thermocouples as temperature
sensor. Temperature indication follows Reference Temperature/Voltage Tables
(N.I.S.T. Monograph 175 Revised to ITS-90) for K/J/T/E/R/S/N-type thermocouples. Two K-type thermocouple are supplied with the thermometer.
SAFETY INFORMATION
It is recommended that you read the safety and operation instructions before
using the thermometer.
WARNING
To avoid electrical shock, do not use this instrument when working
voltages at the measurement surface over 24V AC or DC.
WARNING
To avoid damage or burns, do not make temperature
measurement in microwave ovens.
CAUTION
Repeated sharp flexing can break the thermocouple leads. To prolong
lead life, avoid sharp bends in the leads, especially near the connector.
The symbol on the instrument indicates that the operator must refer to an
explanation in this manual.
1
SPECIFICATIONS
ELECTRICAL
Temperature Scale: Celsius or Fahrenheit user-selectable
Measurement Range:
ThermocoupleRange
K-TYPE-200°C to 1372°C, -328°F to 2501°F
J-TYPE-210°C to 1050°C, -346°F to 1922°F
T-TYPE-200°C to 400°C, -328°F to 752°F
E-TYPE-210°C to 790°C, -346°F to 1454°F
R-TYPE0°C to 1767°C,32°F to 3212°F
S-TYPE0°C to 1767°C,32°F to 3212°F
N-TYPE-50°C to 1300°C, -58°F to 2372°F
Resolution:
K-TYPE0.1°C, 0.1°F (0°F to 700°F)
0.2°F on other range
J-TYPE0.1°C, 0.1°F (0°F to 500°F)
0.2°F on other range
T-TYPE0.1°C, 0.1°F (0°F to 600°F)
0.2°F on other range
E-TYPE0.1°C, 0.1°F (0°F to 450°F)
0.2°F on other range
R-TYPE1°C , 1°F
S-TYPE1°C , 1°F
N-TYPE0.1°C , 0.2°F
2
Accuracy: Accuracy is specified for operating temperatures over the range of
18°C to 28°C (64°F to 82°F), for 1 year, not including thermocouple error.
K/J/T/E-TYPE±(0.05% rdg + 0.3°C) -50°C to 1370°C
±(0.05% rdg + 0.7°C) -50°C to -210°C
±(0.05% rdg + 0.6°F) -58°F to 2498°F
±(0.05% rdg + 1.4°F) -58°F to -346°F
N-TYPE±(0.05% rdg + 0.7°C) -50°C to 0°C
±(0.05% rdg + 0.3°C) 0°C to 1300°C
±(0.05% rdg + 1.4°F) -58°F to 32°F
±(0.05% rdg + 0.6°F) 32°F to 2372°F
R/S-TYPE±(0.05% rdg + 2°C) 0°C to 1767°C
±(0.05% rdg + 4°F) 32°F to 3212°F
Temperature Coefficient: 0.1 times the applicable accuracy specification
per °C from 0°C to 18°C and 28°C to 50°C (32°F to 64°F and 82°F to 122°F).
Input Protection: 24V dc or 24V ac rms maximum input voltage on any
combination of input pins.
Maximum Differential Common Mode Voltage (Maximum Voltage be-
tween T1 and T2 during measurement): 1volt.
Reading Rate: one time per second.
Input Connector: Accepts standard miniature thermocouple connectors (flat
blades spaced 7.9mm, center to center).
3
ENVIRONMENTAL
Ambient Operating Ranges: 0°C to 50°C (32°F to 122°F) <80% R.H.
Storage Temperature: -20°C to 60°C (-4°F to 140°F) <70% R.H.
GENERAL
Display: 4½ digit liquid crystal display (LCD) with maximum reading of 9999.9.
Overload: "----.-" is display.
Battery: Standard 9V battery.
Battery Life: 100 hours typical.
Auto power off: The meter key switch inactive for more than 30 minutes,
press power key to resume operation.
Dimensions: 192mm(H) x 91mm(W) x 52.5mm(D).
Weight: 365g.
4
OPERATING INSTRUCTIONS
1
Power Switch
The key turns the thermometer on or off. In the MIN MAX record mode
can not power off, must leave MIN MAX record mode then power off.
2
°C/°F Selecting the Temperature Scale
Readings are dual displayed in either degrees Celsius(°C) or degrees
Fahrenheit(°F). When the thermometer is turned on, it is set to the
temperature scale that was in use when the thermometer was last turned
off. To change the temperature scale, press the °C/°F key.
3
HOLD Mode (only Main display)
Pressing the HOLD key to enter the Data Hold mode, the "HOLD" annunciator is displayed. When HOLD mode is selected, the thermometer held
the present readings and stops all further measurements.
Pressing the HOLD key again to cancel HOLD mode causing thermometer
to resume taking measurements.
In the MIN/MAX recording mode, press HOLD key to stop the recording.
Press HOLD key again to resume recording. (Previously recorded reading
are not erased).
5