Omega Products HH506AE Installation Manual

INTRODUCTION
The HH506A is a portable digital thermometer that
measures external thermocouples of type K, J, R, S, T, E,
and N. The thermocouple types comply with the N.I.S.T. -
ITS 90 standard reference temperature/voltage tables. The
thermometer features a dual thermocouple input, an adjust-
able T/C offset.
To avoid electrical shock, do not use the device when
the working voltages exceed 24V AC or DC. To avoid
damages to the instrument and/or physical burns, do not
make measurements in microwave ovens. Avoid sharp
bending of the thermocouple leads to prolong its life,
especially near the connector.
WARNING
To avoid electrical shock, do not use this instrument
when working voltages at the measurement surface
are over 24V AC or DC.
WARNING
To avoid damage or burns, do not take temperature
measurement in microwave ovens.
CAUTION
Repeated sharp flexing can break the thermocouple
leads. To prolong lead life, avoid sharp bends in the
leads, especially near the connector.
GENERAL
Displays:
The Main and Second display panels are 4 1/2 digit liquid
crystal display (LCD) with maximum reading of 19999.
These panels are used for displaying the value of T1 or T2.
The Third panel displays the date, time, or the difference
value of T1 to T2. When the input measurement is
overloaded, the following is displayed “ ––––.–”.
Battery:
Standard 9V battery (NEDA 1604, IEC 6F22 006P)
Battery life “is“ about 100 hours when used with a carbon zinc
battery.
Dimensions:
192 mm (H) x 91 mm (W) x 52.5 mm (D)
Weight: 365 g
Supplied Wire:
4 feet Type “K” thermocouple beaded wire (teflon
tape insulated). Maximum insulation temperature
260C (500F). Wire accuracy 2.2C or 0.75% of
reading (whichever is greater) from 0C to 800C.
ENVIRONMENTAL
Ambient Operating Ranges:
0C to 50C (32F to 122F) <80% R.H.
Storage Temperature:
-20C to 60C (-4F to 140F) <70% R.H.
Input Connector:
Accepts standard miniature thermocouple connectors (fla
t
blades spaced 7.9 mm, center to center).
SPECIFICATIONS
ELECTRICAL
Temperature Scale:
Celsius or Fahrenheit
Measurement Range:
K-TYPE (0.1) -200C to 1372C,
-328F to 2501F
J-TYPE (0.1) -210C to 1200C,
-346F to 2192F
T-TYPE (0.1) -200C to 400C,
-328F to 752F
E-TYPE (0.1) -210C to 1000C,
-346F to 1832F
R-TYPE (1) 0C to 1767C,
32F to 3212F
S-TYPE (1) 0C to 1767C,
32F to 3212F
N-TYPE (0.1) -50C to 1300C,
-58F to 2372F
*Based on the ITS-90 temperature standard.
According to temperature standard ITS-90.
Accuracy:
K/J/T/E-TYPE
(0.05% rdg + 0.3C) on -50C to 1370C
(0.05% rdg + 0.7C) on -50C to -210C
(0.05% rdg + 0.6F) on -58F to 2501F
(0.05% rdg + 1.4F) on -58F to -346F
N-TYPE
(0.05% rdg + 0.8C) on -50C to 0C
(0.05% rdg + 0.4C) on 0C to 1300C
(0.05% rdg + 1.6F) on -58F to 32F
(0.05% rdg + 0.8F) on 32F to 2372F
R/S-TYPE
(0.05% rdg + 2C) on 0C to 1767C
(0.05% rdg + 4F) on 32F to 3212F
Temperature Coefficient:
0.1 times the applicable accuracy specification per C
from 0C to 18 and 28C to 50C (32F to 64F and 82F
to 122F)
Input Protecion:
24V dc or 24V ac rms maximum input voltage on any
combination of input pins.
Reading Rate: 2.5 time per second.
HH506A
K, J, T, E, R, S, N Thermometer
INSTRUCTION
SHEET
M4080/0704
Shop online at: omega.com e-mail: info@omega.com
For latest product manuals: omegamanual.info
HH506A
MULTILOGGER
THERMOMETER
Servicing North America:
U.S.A.:
One Omega Drive, Box 4047
Stamford, CT 06907-0047
ISO 9001
Certified
Tel: (203) 359-1660 FAX: (203) 359-7700
e-mail: info@omega.com
Canada:
976 Bergar
Laval (Quebec) H7L 5A1, Canada
Tel: (514) 856-6928 FAX: (514) 856-6886
e-mail: info@omega.ca
For immediate technical or application assistance:
U.S.A.
Sales Service: 1-800-826-6342/1-800-TC-OMEGA
®
and
Customer Service: 1-800-622-2378/1-800-622-BEST
®
Canada:
Engineering Service: 1-800-872-9436/1-800-USA-WHEN®TELEX: 996404 EASYLINK: 62968934 CABLE: OMEGA
Mexico:
En Espan˜ ol: (001) 203-359-7803 e-mail: espanol@omega.com
FAX: (001) 203-359-7807 info@omega.com.mx
Servicing Europe:
Benelux:
Postbus 8034, 1180 LA Amstelveen, The Netherlands
Tel: +31 (0)20 3472121 FAX: +31 (0)20 6434643
Toll Free in Benelux: 0800 0993344
e-mail: sales@omegaeng.nl
Czech
Frystatska 184, 733 01 Karviná, Czech Republic
Republic:
Tel: +420 (0)59 6311899 FAX: +420 (0)59 6311114
Toll Free: 0800-1-66342 e-mail: info@omegashop.cz
France:
11, rue Jacques Cartier, 78280 Guyancourt, France
Tel: +33 (0)1 61 37 2900 FAX: +33 (0)1 30 57 5427
Toll Free in France: 0800 466 342
e-mail: sales@omega.fr
Germany/
Daimlerstrasse 26, D-75392 Deckenpfronn, Germany
Austria:
Tel:+ 49 (0)7056 9398-0 FAX: +49 (0)7056 9398-29
Toll Free in Germany: 0800 639 7678
e-mail: info@omega.de
United
One Omega Drive, River Bend Technology Centre
Kingdo
m: Northbank, Irlam, Manchester
ISO 9002
M44 5BD United Kingdom
Certified
Tel: +44 (0)161 777 6611 FAX: +44 (0)161 777 6622
Toll Free in United Kingdom: 0800-488-488
e-mail: sales@omega.co.uk
OMEGAnet
®
On-Line Service Internet e-mail
omega.com info@omega.com
It is the policy of OMEGA Engineering, Inc. to comply with all
worldwide safety and EMC/EMI regulations that apply. OMEGA
is constantly pursuing certification of its products to the
European New Approach Directives. OMEGA will add the CE
mark to every appropriate device upon certification.
The information contained in this document is believed to be
correct, but OMEGA accepts no liability for any errors it
contains, and reserves the right to alter specifications without
notice.
WARNING: These products are not designed for use in, and
should not be used for, human applications.
Top Side:
1. Sockets of thermocouples T1
2. Sockets of thermocouples T2
OPERATING INSTRUCTIONS
Operational Mode
There are three operation modes–Normal, Shift, and
Setup Mode.
Normal Mode:
This is the default mode; the operating functions for the
normal mode are printed on the top of each button in white.
Shift Mode:
The operating functions for the shift mode are printed
in yellow on the buttons. While in the normal mode,
push the SHIFT button to switch to shift mode. At the
lower righthand corner of the display panel, the word
“Shift” is displayed to indicate shift mode. To switch back
to normal mode, press the SHIFT button.
Setup Mode:
The operating functions for the setup mode are printed
in between two bracket signs “[ ]” on each button. Press
the SET[] button in normal mode to switch to setup mode,
the indicator “SET” is shown on the left side of the display
panel. To switch back to normal mode, press the SET[] button.
Normal Mode:
The following functions can only be used in the normal
mode.
(1) “
I
” Power Button
The “
I
” button turns the thermometer on or off. When
entering data set mode, the power off function is disabled.
(2) “Limits” Button (only main display)
The limits function will alert the user when a measure-
ment exceeds a specified limit. To set the limit values, refer
to limits function in the setup mode.
Press the [Limits] button to activate the limits function; the
word “LIMIT” should be displayed on the LCD.
When the value of the main display exceeds the Hi limit, the
word “Hi” will be displayed and the thermometer will beep
in an interval frequency. If the value of the main display is
lower then the Lo limit, the word “Lo” will be displayed
and the thermometer will beep in a continuous frequency.
(3) “ ” button
The backlight function is represented by this button “
Pressing the button will turn the backlight on or off in the
LCD.
(4) “SAVE/READ” button
The Read Data function works in conjunction with the
Save function in the Shift mode, it is used for reading saved
data. The Save function can be activated in Shift mode.
Press the SAVE/READ button to activate the Read Data
function; the word “READ” should be displayed on the
LCD. To navigate around the Saved Data table, press the
overlay “SECOND” button until the “#” sign is displayed on
the second display panel. The location of the read pointer
within the Saved Data table will be displayed. The arrow
buttons on the overlay are used for scrolling through the saved
data. Pressing the smaller arrow “
” or “ ” retrieves the
next saved data. Pressing the larger arrow “ ” or “
will retrieve the next ten saved data. Pressing the overlay
“ESC” button deactivates the read data function.
(5) “LOG/READ” button
The Log/Read function works in conjunction with the log
function, it is used for reading the log data. The log function
can be activated in the shift mode. Press the LOG/READ
button to activate the read log function; the word “READ”
should be displayed on the LCD. Press the overlay SEC-
OND button to rotate through following display menus:
T1, T2, GRP, and #.
T1 and T2: Displays the T1 or T2 saved value.
GRP: Displays the current group number.
#: Displays the current location of the read pointer within
a selected group. The arrow buttons on the overlay are used
for scrolling through the data and groups. Press the smaller
arrows “
” or “ ” to retrieve the next log data or
group. Press the larger arrows “ ” or “ ” to retrieve
the next ten log data or groups. To navigate around the log
data and groups, press the overlay SECOND button till
GRP appears in the second display panel, and then select
the group using the arrows, then press the SECOND button
again till the “#” sign is displayed. The location of the read
pointer in the selected group will be displayed, and then use
the arrows to scroll through the data. Pressing the overlay
“ESC” button deactivates the read data function.
(7) HOLD mode (only main display)
When HOLD mode is selected, the thermometer holds
the present readings and stops all further measurements. To
activate the data hold mode, press the HOLD button, and
“HOLD” is displayed on the LCD. Pressing the HOLD
button again cancels the function, and the instrument will
automatically resume measurements.
(8) MIN/MAX with time record mode (only main
display)
The MIN/MAX function records the highest and lowest
value recorded, and it calculates the average reading, and
the differences of MAX to MIN. Press MIN/MAX button
to enter the MIN/MAX recording mode. The beeper emits
a tone when a new minimum or maximum measurement is
recorded. Press the MIN/MAX button again to rotate
through the current readings:
MAX: The highest measurement recorded.
MIN: The lowest measurement recorded.
MAX-MIN: The difference of the highest and the lowest
measurement.
AVG: The average values of the measurements. This mode
works in conjunction with the hold function, pressing the
HOLD button will stop the recording and measurements
(Previously recorded readings are not erased). Press HOLD
button again to resume recording and measurements. To
prevent accidental loss of MIN, MAX and AVG data, the
MIN/MAX function can only be cancelled by pressing and
holding down the MIN MAX key for more than 2 seconds.
The Automatic Power Off function, and the power, C/F,
REL, SET, Hi/Lo Limits, TYPE, T1/T2 buttons are also
disabled.
(10) “T1/T2” button (main display)
The input selection button [T1/T2] selects the input
for the main display panel, T1 thermocouple or T2
thermocouple. Press the T1/T2 button to switch between
the two inputs. When meter is turned on, it is set to the
display that was last in use.
(11) “T1/T2” button (SECOND display)
The input selection button [T1/T2] selects the input for
the second display panel, T1 thermocouple or T2 thermo-
couple. Press the T1/T2 button to switch between the two
inputs. When meter is turned on, it is set to the display that
was last in use.
(12) “T1-T2/TIME” button (THIRD display)
The input selection button [T1/T2] selects the system
time and date, or the differential between the two thermo-
couples (T1-T2) for the third display panel. Press the T1/
T2 button to switch the display options. When meter is
turned on, it is set to the display that was last in use.
SHIFT MODE
The following functions can only be used in the shift mode.
(2) “C/F” button
Press the C/F button to select the temperature scale,
readings can be displayed in Celsius (C) or Fahrenheit
(F). When the thermometer is turned on, it is set to the
temperature scale that was last in use.
(4) “SAVE” button
The save function stores the T1, T2 data in nonvolatile
memory. Press the SAVE button to save the current data.
The word SAVE is displayed to indicate the data are saved.
The build in memory can store up to 128 data. The data can
be read using the read function in the normal mode.
(5) “LOG” button
The data log function continuously records the data
according to a specified time interval. The time interval can
be set using the interval time setup function [INVT] in the
set up mode. Press the LOG button to activate the log
function; the indicator “LOG” and “MEM” will be dis-
played on the LCD. There are 16 groups that are used for
storing the log data, and each group uses 64 data slots. If
the current log session exceeds 64 data, the log function
will automatically use the next group to store the following
data. A maximum of 1024 data can be saved in one log
session. Pressing the LOG button again to exit the data log
function.
(6) “CLR ?” button
The CLR function clears all the saved and logged data in
memory. When the CLR button is pressed, indicator
“MEM” is displayed and the “CLR” word on upper-right
of LCD will blink. Pressing “ENTER” button printed on
the overlay in white word to clear all saved and logged data
or “ESC” button to exit this function.
(7) “REL” button (only main display)
The relative value function can be used for comparing the
saved reference value with other measurements. Press the
“REL” button to store the current measurement as the
reference value, and the “REL” should be displayed on the
right part of the LCD. The next measurement will display
the relative value compared to the reference value.
Press “REL” button again to clear the reference value and
deactivate the relative value measurement function.
(8) “[APO]” button
Press the [APO] button to trigger “Auto power off”
function on or off. In this function, the indicator “APO” is
shown at the upper lefthand panel of the LCD. When APO
(Auto power off) is enabled, it will automatically turn the
thermometer off if the key switch is inactivated according to
the set time (the default time for APO is 5 minutes). Press
the Power button to resume operation.
”.
In Limits function, on K or J Type, and with the scale of
temperature at °F, the reading counts should be times 10.
(ex: 2100°F means 21000 counts)
To exit the Limits function, press the [Limits] button.