INTRODUCTION
This instrument is a 4½ digit, compact-sized portable digital thermometer
designed to use external 100W Platinum RTD as temperature sensor. Temperature indication follows Reference Temperature/Resistance Tables (Pt385
for European Curve, Alpha=.00385 and Pt3926 for American Curve, Alpha=.003926).
SAFETY INFORMATION
It is recommended that you read the safety and operation instructions before
using the thermometer.
WARNING
To avoid electrical shock, do not use this instrument when working
voltages at the measurement surface over 24V AC or DC.
WARNING
To avoid damage or burns, do not make temperature
measurement in microwave ovens.
The symbol on the instrument indicates that the operator must refer to an
explanation in this manual.
1
Page 4
SPECIFICATIONS
ELECTRICAL
Temperature Scale: Celsius or Fahrenheit user-selectable.
Measurement Range:
Pt385(100W) -200°C to 800°C, (-328°F to 1472°F)
Pt3926(100W ) -200°C to 630°C, (-328°F to 1166°F)
Resolution: 0.1°C or 0.2°F
Accuracy: Accuracy is specified for operating temperatures over the range of
18°C to 28°C (64°F to 82°F), for 1 year, not including RTD probe error.
±(0.05% rdg + 0.2°C) on °C scale
±(0.05% rdg + 0.4°F) on °F scale
Temperature Coefficient: 0.1 times the applicable accuracy specification
per °C from 0°C to 18°C and 28°C to 50°C (32°F to 64°F and 82°F to 122°F).
Input Protection: 24V dc or 24V ac rms maximum input voltage on any
combination of input pins.
Maximum Differential Common Mode Voltage (Maximum Voltage between T1 and T2 during measurement): 1volt.
Reading Rate: one time per second.
Input Connector: MTP miniature RTD 3 prong connector.
2
Page 5
ENVIRONMENTAL
Ambient Operating Ranges: 0°C to 50°C (32°F to 122°F) <80% R.H.
Storage Temperature: -20°C to 60°C (-4°F to 140°F) <70% R.H.
GENERAL
Display: 4½ digit liquid crystal display (LCD) with maximum reading of 19999.
Overload: "----.-" is display.
Battery: Standard 9V battery.
Battery Life: 100 hours typical.
Auto power off: The meter key switch inactive for more than 30 minutes,
press power key to resume operation.
Dimensions: 192mm(H) x 91mm(W) x 52.5mm(D).
Weight: 365g.
3
Page 6
OPERATING INSTRUCTIONS
1
Power Switch
The key turns the thermometer on or off. In the data SET mode, can not
power off, must leave data SET mode then power off.
2
°C/°F Selecting the Temperature Scale (Main display)
Reading the main displayed in either degrees Celsius(°C) or degrees
Fahrenheit(°F). When the thermometer is turned on, it is set to the
temperature scale that was in use when the thermometer was last turned
off. To change the temperature scale, press the °C/°F key.
3
HOLD Mode (only Main display)
Pressing the HOLD key to enter the Data Hold mode, the
"HOLD" annunciator is displayed. When HOLD mode is selected, the
thermometer held the present readings and stops all further measurements.
Pressing the HOLD key again to cancel HOLD mode causing thermometer
to resume taking measurements.
In the MIN/MAX recording mode, press HOLD key to stop the recording.
Press HOLD key again to resume recording. (Previously recorded reading
are not erased).
4
Relative value Recall display mode.
Press REL RECALL key to display the Relative set value on second display.
Press REL RECALL key again to exit this mode.
4
Page 7
5
Pt385/Pt3926 T1 Input RTD Probe Select
The Pt385/Pt3926 key switch the T1 input selects the Pt385 or Pt3926 RTD
probe as input.
When the thermometer is turned on, it is set to the probe selected that was
in use when the thermometer was last turned off.
MIN MAX with Time record Mode
6
Press MIN MAX key to enter the MIN MAX Recording mode, (displays the
Maximum reading with time, Minimum reading with time and Average
reading stored in record mode). In the this mode the automatic power-off
feature is disabled and key, °C/°F key, REL key, SET key, Hi/Lo LIMITS
key, Pt385/Pt3926 and all RECALL keys are disabled.
The beeper emits a tone when a new minimum or maximum value is
recorded.
Present temperature reading displayed on second display.
Push MIN MAX key to cycle through the MAX, MIN and AVG readings.
If an overload is recorded, the averaging function is stopped and average
value display"----.-".
The true average of all the reading taken over at least 22 hours preiod can
be displayed. If 22 hours is exceeded, new averages are no longer
calculated. The last calcalated value is retained as the average reading, but
the actual minimum and maximum reading will continue to be captured.
In the this mode, press HOLD key to stop the recording of readings, all
values are frozen, press again to restart recording.
To prevent accidental loss of MIN, MAX and AVG data, in this mode can only
be cancelled by pressing and hold down the MIN MAX key for 2 seconds to
exit and erased recorded readings.
5
Page 8
7
REL Relative mode
Pressing REL key to enter the Relative mode, zero the display, and store
the displayed Reading as a reference value and annunciator REL is
displayed. Present temperature reading displayed on second display.
Press REL key again to exit the relative mode.
The relative value can also be entered by the user. (See "SET mode" later
in this manual.)
When the desired Relative value has been entered, press REL key to enter
the Relative mode, press SET key use set Relative value as a reference
value. Press REL key again to exit the relative mode.
In the Relative mode, the value (can not >±3000.0 counts) shown on the
LCD is always the difference between the stored reference and the present
reading.
8
°C/°F Selecting the Temperature Scale (Second display)
Readings the second displayed in either degrees Celsius(°C) or degrees
Fahrenheit(°F). When the thermometer is turned on, it is set to the
temperature scale that was in use when the thermometer was last turned
off. To change the temperature scale, press the °C/°F key.
6
Page 9
9
SET mode (Relative value set, Time set and Hi/Lo
1. Press SET key to enter Relative valuse SET mode (Press ENTER key can
escape relative valuse set mode), REL set mode. = = = =.= is displayed in
main display.
Relative value is entered via overlay numbers, then press overlay ENTER
key, stored the relative value, enter Time set mode.
2. Time set mode, (Press ENTER key can escape Time set mode)
=.= = = : = = is displayed in second and third display. Time (hours, minutes,
seconds) value is entered via overlay numbers, then press overlay ENTER
key. Time start from set time value, enter Hi/Lo Limits value set mode.
Limits value set)
3. Hi Limit value set mode,
Limit value set mode), = = = =.= is displayed in main display, Hi Limit value
is entered via overlay numbers, then press overlay ENTER key, stored the
Hi Limit value, enter Lo Limit value set mode (Press ENTER key can escape
Lo Limit value set mode). = = = =.= is displayed in main display, Lo Limit
value is entered via overlay numbers, then press overlay ENTER key,
stored the Lo Limit value and exit SET mode.
4. When the thermometer is turned on. The Relative set value and Hi/Lo Limits
set value that was in use when thermometer was last turned off set values.
is displayed (Press ENTER key can escape Hi
7
Page 10
10
Hi Limit value Recall display mode
Press Hi RECALL key to display the Hi set Limit value on second display.
Press Hi RECALL key again to exit this mode.
Lo Limit value Recall display mode
11
Press Lo RECALL key to display the Hi set Limit value on second display.
Press Lo RECALL key again to exit this mode.
Hi/Lo Limits mode (only Main display)
12
Press Hi/Lo Limits key to enter the Hi/Lo Limits comparative mode, when
input temperature value exceed Hi or Lo Limits value. The beeper emits a
continuity pulse tone. Press Hi/Lo Limits key again to exit the Hi/Lo Limits
mode.
8
Page 11
OPERATOR MAINTENANCE
WARNING
To avoid possible electrical shock, disconnect the thermocouple
connectors from the rmometer before removing the cover.
Battery Replacement
Power is supplied by a 9 volt "transistor" battery. The " " appears on the LCD
display when replacement is needed. To replace the battery, remove the two
screws from the back of the meter and lift off the battery cover. Remove the
battery from battery contacts.
OMEGA ENGINEERING, INC. warrants this unit to be free of defects in materials and workmanship for a period of 13 months
from date of purchase. OMEGA Warranty adds an additional one (1) month grace period to the normal one (1) year productwarranty to cover handling and shipping time. This ensures that OMEGA's customers receive maximum cov erage on each product.If
the unit should malfunction, it must be returned to the factory for evaluation. OMEGA's Customer Service Department will issue an
Authorized Return (AR) number immediately upon phone or written request. Upon examination by OMEGA, if the unit is found to
be defective it will be repaired or replaced at no charge. OMEGA's W ARRANTY does not apply to defects resulting from any action
of the purchaser, including but not limited to mishandling, improper interfacing, operation outside of design limits, improper repair,
or unauthorized modification. This WARRANTY is V OID if the unit shows e vidence of having been tampered with or shows evidence
of being damaged as a result of excessive corrosion; or current, heat moisture or vibration; improper specification; misapplication;
misuse or other operating conditions outside of OMEGA's control. Components which wear are not warranted, including but not
limited to contact points, fuses, and triacs.
OMEGA is pleased to offer suggestions on the use of its various products. However , OMEGA neither assumes responsibility
for any omissions or errors nor assumes liability for an y damages that result fr om the use of its products in accordance
with information provided by OMEGA, either verbal or written. OMEGA warrants only that the parts manufactured by it
will be as specified and free of defects. OMEGA MAKES NO OTHER WARRANTIES OR REPRESENTATIONS OF ANY KIND
WHATSOEVER, EXPRESSED OR IMPLIED, EXCEPT THAT OF TITLE AND ALL IMPLIED WARRANTIES INCLUDING ANY
WARRANTY OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE ARE HEREBY DISCLAIMED.
LIMITATION OF LIABILITY: The remedies of purchaser set forth herein are exclusive and the total liability of OMEGA with
respect to this order, whether based on contract, warranty, negligence, indemnification, strict liability or otherwise, shall
not exceed the purchase price of the component upon which liability is based. In no event shall OMEGA be liable for
consequential, incidental or special damages.
CONDITIONS: Equipment sold by OMEGA is not intended to be used, nor shall it be used: (1) as a "Basic Component" under 10
CFR 21 (NRC), used in or with any nuclear installation or activity; or (2) in medical applications or used on humans. Should any
Product(s) be used in or with any nuclear installation or activity, medical application, used on humans, or misused in any way,
OMEGA assumes no responsibility as set forth in our basic WARRANTY / DISCLAIMER language, and additionally, purchaser will
indemnify OMEGA and hold OMEGA harmless from any liability or damage whatsoever arising out of the use of the Product(s) in
such a manner.
Direct all warranty and repair requests/inquiries to the OMEGA Customer Service Department. BEFORE RETURNING ANY
PRODUCT(S) TO OMEGA, PURCHASER MUST OBTAIN AN AUTHORIZED RETURN (AR) NUMBER FROM OMEGA'S
CUSTOMER SERVICE DEPARTMENT (IN ORDER TO AVOID PROCESSING DELAYS). The assigned AR number should then
be marked on the outside of the return package and on any correspondence.
The purchaser is responsible for shipping charges, freight, insurance and proper packaging to prevent breakage in transit.
FOR WARRANTY RETURNS, please have the f ollowing
information available BEFORE contacting OMEGA:
1. P.O. number under which the product was PURCHASED.
2. Model and serial number of the product under warranty, and
3. Repair instructions and/or specific problems relative to the
product.