Omega Products HH504 Installation Manual

OMEGA
HH504
Digital Thermometer
omega. com
OMEGAnetSM On-Line Service
http://www.omega.com
USA: ISO 9001 Certified Canada:
One Omega Drive, Box 4047 976 Bergar
Stamford, CT 06907-0047 Laval (Quebec) H7L5A1
Tel: (203) 359-1660 Tel: (514) 856-6928
FAX: (203)359-7700 FAX: (514) 856-6886
e-mail: info@omega.com e-mail: info@omega.com
USA and Canada: Mexico and Latin America:
Sales Service: 1-800-826-6342 / 1-800-TC-OMEGA
Customer Service: 1-800-622-2378 / 1-800-622-BEST
Engineering Service: 1-800-872-9436 / 1-800-USA-WHENSMEn Español: (203) 359-7803
TELEX: 996404 EASYLINK: 62968934 CABLE: OMEGA e-mail: espanol@omega.com
Benelux: Germany/Austria:
Postbus 8034, 1180 LA Amstelveen, Daimlerstrasse 26, D-75392
The Netherlands Deckenpfronn, Germany
Tel: (31) 20 6418405 FAX: (31) 20 6434643 Tel: 49 (07056) 3017 FAX: 49 (07056) 8540
Toll Free in Benelux: 06 0993344 Toll Free in Germany: 0800 82 66342
e-mail: nl@omega.com e-mail: germany@omega.com
Czech Republic: United Kingdom: ISO 9002 Certified
ul. Rude armady 1868, 733 01 Karvina- One Omega Drive
Hranice, Czech Repubic Riverbend Technology Centre Northbank, Irlam,
Tel: 420 (69) 6311627 FAX: 420 (69) 6311114 Manchester, M44 5EX, England
e-mail: czech@omega.com Tel: 44 (161) 777-6611 FAX: 44 (161) 777-6622
France:
9, rue Denis Papin, 78190 Trappes Toll Free in England: 0800-488-488
Tel: (33) 130-621-400 FAX: (33)130-699-120 e-mail: sales@omega.com.uk
Toll Free in France: 0800-4-06342
e-mail: france@omega.com
It is the policy of OMEGA to comply with all worldwide safety and EMC/EMI regulations that apply. OMEGA is constantly pursuing certification of its products to the European New Approach Directives. OMEGA will add the CE mark to every appropriate device upon certification. The information contained in this document is believed to be correct but OMEGA Engineering, Inc. accepts no liability for any errors it contains, and reserves the right to alter specifications without notice. WARNING: These products are not designed for use in, and should not be used for, patient connected application.
For immediate technical or application assistance:
Servicing North America:
SM
SM
Servicing Europe:
TM
Internet e-mail
Tel: (95) 800-TC-OMEGA
FAX: (95) 203-359-7807
SM
INTRODUCTION
This instrument is a 4½ digit, compact-sized portable digital thermometer designed to use external 100W Platinum RTD as temperature sensor. Tem­perature indication follows Reference Temperature/Resistance Tables (Pt385 for European Curve, Alpha=.00385 and Pt3926 for American Curve, Al­pha=.003926).
SAFETY INFORMATION
It is recommended that you read the safety and operation instructions before using the thermometer.
WARNING
To avoid electrical shock, do not use this instrument when working
voltages at the measurement surface over 24V AC or DC.
WARNING
To avoid damage or burns, do not make temperature measurement in
microwave ovens.
The symbol on the instrument indicates that the operator must refer to an explanation in this manual.
1
SPECIFICATIONS
ELECTRICAL
Temperature Scale: Celsius or Fahrenheit user-selectable. Measurement Range:
Pt385(100W) -200°C to 800°C, (-328°F to 1472°F) Pt3926(100W ) -200°C to 630°C, (-328°F to 1166°F)
Resolution: 0.1°C or 0.2°F Accuracy: Accuracy is specified for operating temperatures over the range of
18°C to 28°C (64°F to 82°F), for 1 year, not including RTD probe error.
±(0.05% rdg + 0.2°C) on °C scale ±(0.05% rdg + 0.4°F) on °F scale
Temperature Coefficient: 0.1 times the applicable accuracy specification
per °C from 0°C to 18°C and 28°C to 50°C (32°F to 64°F and 82°F to 122°F). Input Protection: 24V dc or 24V ac rms maximum input voltage on any
combination of input pins.
Maximum Differential Common Mode Voltage (Maximum Voltage be­tween T1 and T2 during measurement): 1volt. Reading Rate: one time per second. Input Connector: MTP miniature RTD 3 prong connector.
2
ENVIRONMENTAL
Ambient Operating Ranges: 0°C to 50°C (32°F to 122°F) <80% R.H. Storage Temperature: -20°C to 60°C (-4°F to 140°F) <70% R.H.
GENERAL
Display: 4½ digit liquid crystal display (LCD) with maximum reading of 19999. Overload: "----.-" is display. Battery: Standard 9V battery. Battery Life: 100 hours typical. Auto power off: The meter key switch inactive for more than 30 minutes,
press power key to resume operation.
Dimensions: 192mm(H) x 91mm(W) x 52.5mm(D). Weight: 365g.